Part 11. CAPA. Batch release. Each question answered, with its source.
229 answers on regulations, validation, quality, batch records, the laboratory, suppliers, choosing systems and AI.
Find your question.
Regulations and guidance43 answers
Regulations and guidance
- What is GxP?
GxP means good “x” practice, a collective term for the quality rules regulators set for medicines, where the x names the field: manufacturing (GMP), laboratory safety studies (GLP), clinical trials (GCP), distribution (GDP) and pharmacovigilance (GVP).
- What is the difference between GMP and cGMP?
There is no difference in kind.
- What does 21 CFR Part 11 require?
Part 11 requires validated systems, secure time-stamped audit trails and access limited to authorized people.
- What is an FDA Form 483?
An FDA Form 483, Inspectional Observations, is the signed list of significant objectionable conditions an FDA investigator gives a site’s management at the close of an inspection.
- What is the difference between an FDA 483 and a warning letter?
A Form 483 lists conditions an FDA investigator observed, handed to management as the inspection closes.
- What is the FDA QMSR, and what changed from the Quality System Regulation (QSR)?
The Quality Management System Regulation (QMSR) is 21 CFR Part 820, FDA’s good manufacturing practice rule for medical devices, effective 2 February 2026.
- What is 21 CFR Part 211, and how does it differ from Part 210?
Part 211 is FDA’s detailed CGMP rule for finished drug products: 60 sections in 11 subparts, from the quality unit to returned goods, the part inspectors cite.
- What is the difference between ISO 9001 and ISO 13485?
ISO 9001 is the quality management standard for any organization; ISO 13485 is the one for medical devices, written for regulatory use.
- What is the difference between a 503A pharmacy and a 503B outsourcing facility?
A 503A pharmacy compounds for identified patients on valid prescriptions, is overseen mainly by state boards of pharmacy and is exempt from federal CGMP if its drugs meet every 503A condition.
- What is the difference between GLP, GMP and GCP?
GLP (good laboratory practice) governs nonclinical safety studies, such as the toxicology behind an IND; GMP (good manufacturing practice) governs making and testing medicines, trial supplies included; GCP (good clinical practice) governs trials in people.
- What is the difference between 21 CFR Part 11 and EU GMP Annex 11?
21 CFR Part 11 is a binding US regulation on electronic records and signatures; EU GMP Annex 11 is guidance on computerized systems across their whole life.
- What does the draft revision of EU GMP Annex 11 change, and when does it take effect?
Not yet: the revised EU GMP Annex 11 is still a draft with no effective date, so the 2011 annex applies.
- What is EU GMP Annex 1, and what does it require?
EU GMP Annex 1 is the European Commission’s guideline for manufacturing sterile medicines, part of EudraLex Volume 4 and shared with PIC/S.
- How do you respond to an FDA Form 483, and how many days do you have?
You don’t have to respond to an FDA Form 483, but FDA recommends one response within 15 business days of the date it was issued, and reviews a timely response in detail before deciding on a warning letter.
- What do NAI, VAI and OAI mean after an FDA inspection?
NAI, VAI and OAI are FDA’s three inspection outcomes.
- Where can I look up FDA 483s, warning letters and inspection results for a company or supplier?
Search the firm’s name or FEI number on FDA’s Data Dashboard: its Firm Profile shows inspections with their NAI, VAI or OAI result, 483 citations, recalls, import refusals, current import alerts and the last five years’ warning letters.
- What is the difference between ISO 13485 and 21 CFR Part 820 (the QMSR)?
Since 2 February 2026, 21 CFR Part 820, the Quality Management System Regulation, incorporates ISO 13485:2016, so the core requirements are the same.
- What is ICH Q7 (GMP for active pharmaceutical ingredients)?
ICH Q7, the Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, is the International Council for Harmonisation’s GMP guideline for APIs.
- What is quality by design (QbD) in ICH Q8, and what are its steps?
Quality by design (QbD) is ICH Q8(R2)’s way of developing a product: decide up front what quality it needs, learn how materials and process affect it, and control them.
- What is a Qualified Person (QP) in pharma, and what do they certify?
QP stands for Qualified Person: the legally qualified individual every EU manufacturing or import authorization holder must have.
- What are the main differences between EU GMP and FDA cGMP?
Both require written procedures, batch records, investigated deviations and quality control.
- What is EudraLex Volume 4, and which EU GMP chapters and annexes are in force?
EudraLex Volume 4 is the EU GMP guide: the European Commission’s good manufacturing practice guidelines for medicines, one PDF per chapter and annex.
- What is PIC/S, and how does the PIC/S GMP Guide relate to EU GMP?
PIC/S, the Pharmaceutical Inspection Co-operation Scheme, is a non-binding arrangement between 57 medicines regulators, including the FDA, MHRA, Health Canada, TGA and most EU inspectorates, that harmonizes GMP standards, trains inspectors and assesses inspectorates.
- What is a GMP certificate, and how do you check one?
A GMP certificate is the inspecting regulator’s statement that one site complied with GMP at its inspection.
- What are the basic principles of GMP?
GMP’s core principle is that medicines are consistently made and controlled to the quality their use and authorisation require: quality is built in, not tested in.
- What is an HCT/P, and what is the difference between a 361 and a 351 product?
An HCT/P is human cells or tissue intended for implantation, transplantation, infusion or transfer into a person, such as bone, skin or cord blood.
- What is the difference between an NDA, an ANDA and a BLA?
A New Drug Application (NDA), under FD&C Act section 505(b), proves a new drug safe and effective with the applicant’s own studies, or partly others’ in a 505(b)(2).
- What is a drug master file (DMF)?
A drug master file (DMF) is a confidential submission to FDA describing facilities, processes or articles used to make, package or store human drugs.
- What is the Common Technical Document (CTD), and what goes in each module?
The Common Technical Document (CTD) is the ICH M4 format for the dossier that registers a human medicine.
- Does FDA regulate laboratory-developed tests (LDTs) now that the LDT rule has been vacated?
Not as medical devices.
- Can FDA inspectors now see management review, internal audit and supplier audit reports?
Yes, for device manufacturers.
- What is an FDA pre-approval inspection (PAI), and how do you prepare for one?
A pre-approval inspection (PAI) is FDA’s on-site check of a facility named in a pending application; for an original BLA it is called a pre-license inspection.
- What should you expect during an FDA inspection of a drug manufacturer?
FDA usually arrives unannounced for domestic drug surveillance inspections and calls ahead for pre-approval and most foreign ones.
- What must a site master file contain, and is there a template?
A site master file describes one manufacturing site’s GMP activities in, as far as possible, no more than 25 to 30 pages plus appendices.
- Is a post-approval change a prior approval supplement, a CBE-30, a CBE-0 or an annual report change?
It depends on how much the change could harm the product’s quality.
- What is the difference between a Field Alert Report (FAR) and a Biological Product Deviation Report (BPDR), and which one do you file?
The approving application decides.
- What is the difference between Type IA, IB and II variations in the EU, and what changed in 2025 and 2026?
Type IA has minimal or no impact: make it, then notify, usually in one annual update 9 to 12 months after the first.
- Which drug and device CGMP requirements apply to a combination product under 21 CFR Part 4 now that the QMSR is in force?
Each constituent part keeps its own CGMP rules; a site making both can meet every set in full or run one base system plus the other’s named provisions.
- When is a medical device correction or removal reportable to FDA under 21 CFR 806?
A device manufacturer or importer must report a correction or removal to FDA within 10 working days of starting it if it was started to reduce a risk to health, or to remedy a violation that may present one.
- What happened to the DHF, DMR and DHR under FDA’s QMSR?
The design history file records how a device’s design was developed; the device master record holds the specifications and procedures for building it; the device history record shows how each lot or unit was built.
- How do you write a comparability protocol (post-approval change management protocol) for a CMC change?
A comparability protocol is a plan the regulator approves before you make a CMC change, so the change files in a lower category.
- What goes in an NDA or ANDA annual report under 21 CFR 314.81(b)(2), and when is it due?
An NDA or ANDA annual report is due each year within 60 days of the anniversary of US approval, covering the year that ends on that date, with Form FDA 2252, in eCTD.
- How do you respond to an FDA records request under section 704(a)(4) or a remote regulatory assessment?
Send every record listed on Form FDA 4003 by its date: a section 704(a)(4) request is mandatory, and FDA’s example is 15 days, or 30 with translation.
Validation and data integrity39 answers
Validation and data integrity
- What is computer system validation (CSV), and what are its steps and deliverables?
Computer system validation (CSV) is documented evidence that a GxP computerized system does what you need it to, and keeps doing so until it is retired.
- What is computer software assurance (CSA), and how does it differ from CSV?
CSV and CSA both show software is fit for its intended use.
- What are the ALCOA+ and ALCOA++ principles of data integrity?
ALCOA+ stands for attributable, legible, contemporaneous, original and accurate (ALCOA), plus complete, consistent, enduring and available: the attributes regulators expect of GxP data, on paper or electronic.
- What are IQ, OQ and PQ in equipment qualification?
IQ, OQ and PQ are installation, operational and performance qualification, the three stages that prove equipment is fit for use: IQ that it is installed as designed, OQ that it works across its operating ranges, worst cases included, and PQ that it performs reproducibly in normal use.
- What is data integrity in pharma, and why does it matter?
Data integrity means your GMP data are complete, consistent and accurate, and stay that way from creation to disposal.
- How do you perform an audit trail review, and how often should audit trails be reviewed?
An audit trail review is the documented check that the creations, changes and deletions behind GxP data are complete, attributable and justified.
- What are the GAMP 5 software categories, and which category is a LIMS, eQMS or MES?
GAMP 5 Second Edition groups software components into four categories: 1, infrastructure software, tools and IT services; 3, standard system components; 4, configured components; and 5, custom applications and components.
- What are the three stages of process validation?
FDA’s 2011 guidance divides process validation into three stages: process design, process qualification and continued process verification.
- What is cleaning validation, and what do FDA and EU GMP require?
Cleaning validation is documented evidence that an approved cleaning procedure reproducibly removes the previous product, cleaning agents and microbial contamination from product-contact equipment to below a scientifically justified limit, in the EU one based on toxicology.
- What do the FDA, MHRA, PIC/S and WHO data integrity guidances require, and how do they differ?
Four guidances set the reference: FDA’s Data Integrity and Compliance With Drug CGMP Q&A (December 2018), MHRA’s GxP Data Integrity Guidance and Definitions (Revision 1, March 2018), PIC/S PI 041-1 (July 2021) and WHO TRS 1033 Annex 4 (2021).
- What are the audit trail requirements in Part 11 and EU GMP Annex 11?
An audit trail is a secure, computer-generated, time-stamped record of who created, changed or deleted a GxP record, what changed and when, without obscuring the earlier value.
- What does EU GMP Annex 11 require for computerized systems?
Annex 11 requires you to validate each GMP computerized system to a depth set by risk assessment, and to qualify the IT infrastructure it runs on.
- What is GAMP 5, and what changed in the Second Edition?
GAMP 5 is ISPE’s guide to risk-based validation of GxP computerized systems; GAMP stands for Good Automated Manufacturing Practice.
- What is a user requirements specification (URS), and how do you write one for a GxP system?
A user requirements specification (URS) is the approved, controlled document stating what equipment, a utility or a computerised system must do for its intended use.
- What is a requirements traceability matrix in validation, and is it required?
A requirements traceability matrix links each user requirement to the design or configuration that meets it, the test or other verification that proves it, and the result.
- What is risk-based validation, and how do you decide how much testing a system needs?
Risk-based validation sets how much you test each function of a computerised system from a documented assessment of what its failure could do to patients, product quality and data integrity.
- How do you validate a SaaS or cloud system for GxP use, and what is the vendor responsible for?
You validate a SaaS system for your own intended use, and you stay accountable even though the vendor builds, hosts, tests and runs it.
- Can you use Excel in GMP, and how do you validate a spreadsheet?
Yes.
- What should a validation master plan (VMP) template include?
No regulation sets a format for a validation master plan (VMP): it is your plan for what you will validate, why, how and who approves it.
- What is the difference between validation, verification, qualification and calibration?
Calibration shows an instrument’s readings agree with a reference standard within set limits.
- What is continued process verification (CPV), and how does it differ from continuous and ongoing process verification?
Continued process verification (CPV) is Stage 3 of FDA’s process validation lifecycle: an ongoing programme that collects and statistically trends data from commercial batches to show the process stays in a state of control.
- How many PPQ batches are required, and is it still three?
No regulation fixes the number.
- What is the difference between prospective, concurrent and retrospective validation?
For medicines, they differ in when evidence is gathered relative to sale.
- How do you calculate MACO and set HBEL- or PDE-based cleaning validation limits?
MACO, the maximum allowable carryover, is the most of the previous product that can safely carry over into the next: its PDE or other HBEL (µg/day) × the next product’s smallest batch (kg) ÷ the next product’s maximum daily dose (mg/day), in grams.
- What are common data integrity violations in pharma?
FDA’s 2026 warning letters cite familiar violations: shared logins and analysts with administrator rights; audit trails off, or changes made outside them; trial injections of real samples; failing results dropped while a passing retest is reported; deleted data with no backup; original printouts thrown away; backdated or fabricated records; and uncontrolled blank forms.
- What are the acceptance criteria for blend and content uniformity now that FDA has withdrawn its 2003 draft guidance?
FDA no longer gives a numeric blend uniformity limit: the RSD of 5.0% or less in its 2003 draft went with the draft.
- What does EU GMP Annex 15 require, section by section, and what will the revision change?
EU GMP Annex 15 requires you to qualify your facilities, utilities and equipment and validate your processes, cleaning and test methods, and to keep them in a validated state for their whole life.
- How do you run an airflow visualization (smoke) study, and what are the acceptance criteria?
In a smoke study, release a visible tracer at each critical location and film where the air goes, at rest and in operation, with operators simulating their interventions as in production.
- When is manual integration of chromatography peaks allowed, and what must the integration SOP say?
Only by exception: when the validated method’s automatic integration cannot integrate a peak correctly and an approved SOP covers that case.
- How long does each phase of pharmaceutical water system validation take, and what is tested in phases 1, 2 and 3?
Under FDA’s guide, phase 1 is two to four weeks of daily sampling after each treatment step and point of use.
- What does USP <1058> require for analytical instrument qualification, and what does the proposed revision change?
USP <1058> is an informational chapter: it recommends a risk-based way to show analytical instruments are fit for intended use but prescribes no exact tests or limits, which come from technique chapters and your methods.
- What does 21 CFR 211.68 require for automated and computerized equipment, and when can a system replace the second-person check?
21 CFR 211.68 allows automatic, mechanical and electronic equipment, including computers, if it is calibrated, inspected or checked under a written program, with records.
- How do you run a dirty and clean equipment hold time study, and how long can a cleaning campaign run?
No regulation sets the hours or batches: each site proves its own dirty hold (end of production to start of cleaning), clean hold (end of cleaning to next use) and campaign length, in time and batches.
- How do you run a swab recovery study for cleaning validation, and how do you correct results for recovery?
Spike coupons of every product-contact material you sample at levels spanning the limit, swab and extract them exactly as in production, and report recovery as found ÷ spiked × 100.
- How do you validate a depyrogenation tunnel or dry heat oven, and what endotoxin reduction must it show?
At least a 3 log10 (99.9%) reduction of endotoxin spiked onto worst-case containers: EU GMP Annex 1 and FDA both set that minimum for the process, and protocols apply it to every indicator.
- How do you validate a sterilizing-grade filter, and what does the bacterial retention test have to show?
A bacterial retention test must show no passage of at least 10⁷ CFU of Brevundimonas diminuta, or another justified worst-case organism, per cm² of filter area, under your worst-case conditions.
- How do you choose the worst-case product for cleaning validation, and how do you score a product matrix?
Choose it on two counts, not one: the product whose residue has the lowest health-based carryover limit, and the product your procedure finds hardest to remove.
- How often does GMP equipment need requalification, and what should a periodic review check?
No general GMP rule sets one interval: EU GMP, PIC/S and WHO expect a periodic review at a risk-justified interval, and requalification when the review, a change or a repair shows the need.
- What should factory and site acceptance tests cover, and can FAT and SAT results replace IQ and OQ testing?
A factory acceptance test (FAT) checks equipment at the vendor, before shipment, against your URS and functional specification: documents, build, functions, alarms, interlocks and software.
Quality management31 answers
Quality management
- What is CAPA (corrective and preventive action) in pharma?
CAPA, corrective and preventive action, is the pharmaceutical quality system process that finds why a problem happened or could happen and removes that cause.
- What is the difference between QA and QC in pharma?
Quality control (QC) samples and tests materials and products against their specifications and judges each batch’s results.
- What is a deviation in GMP, and does every deviation need to be investigated?
A deviation is a departure from an approved instruction or established standard, such as an exceeded hold time, a skipped check or an instrument used past calibration.
- What is root cause analysis, and how do you perform one for a deviation or CAPA?
Root cause analysis finds why a deviation or failure happened, so the CAPA removes the cause, not the symptom.
- What is change control in pharma, and what are the steps of the change control process?
Change control is the GMP system for proposing, assessing, approving, implementing and reviewing planned changes that could affect product quality, a validated state or the marketing authorization.
- What is an annual product review (APR or PQR), and what must it include?
An annual product review is a documented evaluation, normally once a year, of a product’s batches, complaints, recalls, returns and investigations, to decide whether its specifications, process or controls need to change.
- How do you perform an FMEA in pharma and calculate the risk priority number (RPN)?
In an FMEA, the risk priority number is RPN = severity × occurrence × detection: rate each process step’s failure modes on all three and multiply, so 5 × 3 × 4 = 60 on 1 to 5 scales, where 125 is the maximum.
- What is quality risk management (ICH Q9(R1)), and what changed in the revision?
Quality risk management is a systematic process for assessing, controlling, communicating and reviewing risks to a medicine’s quality across its lifecycle, set out in ICH Q9.
- How do you get ISO 13485 certification, how long does it take, and what does it cost?
ISO 13485 certification comes from a certification body, not ISO: a Stage 1 readiness audit, a Stage 2 audit of the whole system, then a three-year certificate with yearly surveillance.
- What is ICH Q10, and what are the elements of a pharmaceutical quality system?
ICH Q10 is the ICH guideline, final since 4 June 2008, that describes a pharmaceutical quality system for a product’s whole lifecycle, run under senior management’s responsibility.
- What is the difference between a correction, a corrective action and a preventive action?
A correction fixes the nonconformity itself, for example by reworking, regrading or rejecting a batch.
- How do you perform a CAPA effectiveness check, and how long should it run?
A CAPA effectiveness check, or verification of effectiveness, tests whether the problem stopped, not whether tasks were done.
- How do you write a deviation investigation report?
Write a deviation investigation report so a reviewer can follow the event to the batch decision: what should have happened and what did, immediate actions, affected batches, the evidence, the root or most likely cause, product impact, the batch decision and CAPA, signed off by the quality unit.
- What is the difference between a deviation and a nonconformance?
A deviation is a departure from an approved instruction or established standard, such as a skipped step or an exceeded hold time; it may or may not affect quality.
- What is a nonconformance report (NCR), and when does an NCR become a CAPA?
A nonconformance report (NCR) records an item that fails a requirement: what failed, against which specification, how it was contained and evaluated, and the authorised disposition.
- What is a planned deviation, and should it go through change control instead?
A planned deviation is a one-off departure from an approved instruction, approved in writing by QA before the work and limited to a set scope such as one batch.
- Is human error an acceptable root cause for a deviation?
Only when the investigation proves it.
- How should pharmaceutical product complaints be handled and investigated?
Log every written and oral complaint under a written procedure, then sort it: possible quality defect, adverse event or suspected falsification, and class any defect as critical, major or minor.
- What is a management review, what are its inputs and outputs, and how often must it happen?
A management review is senior management’s formal, periodic review of whether the quality system remains suitable, adequate and effective.
- What does document control require under GMP and ISO 13485?
Document control means every GMP instruction, from SOPs and specifications to master batch records and blank forms, is uniquely identified, approved by authorised people, given a defined effective date and available in its current version where the work happens.
- How do you write a GMP SOP, and how is an SOP different from a work instruction?
Write a GMP SOP for your facility: a header (number, revision, effective date); purpose, scope and responsibilities; numbered imperative steps, each check with its limit, each decision with its criteria and what to do when a step fails; then records, change history and approvals.
- What are the GMP training requirements, and what must training records show?
Everyone whose work can affect quality must be trained in their operations and in GMP, continually; no general refresher interval is set.
- What is a GMP internal audit (self-inspection), and how often should one be done?
A GMP internal audit, which EU GMP calls a self-inspection, is a planned, recorded check by competent people independent of the area that a site follows GMP and its own procedures, with corrective actions tracked to completion.
- What are pharmaceutical quality metrics, and what is FDA’s Quality Metrics Reporting Program?
Pharmaceutical quality metrics are objective measures of how well manufacturing and the quality system perform, such as lot acceptance, right first time, invalidated out-of-specification results, complaints and CAPA effectiveness.
- What is a quality manual, and is one required for pharma and medical devices?
A quality manual is the top-level document describing a company’s quality system: its policy, scope, processes and how they link, responsibilities and documentation structure.
- What is clinical quality assurance, and what does a GCP QA function do?
Clinical quality assurance is the sponsor’s system of documented procedures and risk-based checks, applied throughout a trial, that finds the causes of serious noncompliance with the protocol, GCP and the law so they can be corrected and prevented.
- How do you run a mock recall for medicines, and how often is one required?
A mock recall is an unannounced drill: pick a released batch, trace it to every customer, contact each one and reconcile the stock, timing each step against your targets.
- What are FDA recall classes I, II and III, and how does a recall differ from a market withdrawal, stock recovery or correction?
FDA assigns the class after its own health hazard evaluation.
- What are the good documentation practice rules for handwritten GMP entries: date format, late entries, blank fields and ink?
Handwritten GMP entries are made in indelible ink, legibly, at the time, by the person who did the task, under their own initials.
- When is a medical device complaint MDR reportable, and when are FDA’s 30-day, 5-day and supplemental reports due?
A complaint is MDR reportable when information reasonably suggests your device may have caused or contributed to a death or serious injury, or malfunctioned in a way likely to cause or contribute to one if it recurred; FDA presumes recurrence.
- How often must SOPs be reviewed, and what does a periodic review of an SOP check?
No regulation sets a frequency.
Manufacturing and batch records31 answers
Manufacturing and batch records
- What is an electronic batch record (eBR), and does FDA require one?
An electronic batch record (eBR) is a batch’s production record, issued from the approved master and completed in a computer system as the work happens: entries saved when made, steps and signers enforced, changes kept in an audit trail, signatures linked to the record.
- What is the difference between a master batch record and a batch production record?
A master batch record is the approved set of instructions for making one product at one batch size: its formula, equipment, steps, limits and expected yields.
- What does EU GMP Annex 1 require for environmental monitoring?
EU GMP Annex 1 asks for a documented, risk-based environmental monitoring programme within the contamination control strategy, covering total particles, viable particles in air, on surfaces and on people, and temperature and humidity.
- What is batch record review by exception, and do FDA and EU inspectors accept it?
Review by exception means QA reviews a batch record through a validated report of predetermined exceptions, such as out-of-limit values, changed data, alarms and deviations, instead of reading every entry.
- How do you review a batch record before release, and who has to do it?
In the US, before release, the quality unit reviews and approves every production and control record, packaging and labelling included, and has any unexplained discrepancy, out-of-limit yield or failed specification investigated (21 CFR 211.192); under contract manufacture, the product owner’s quality unit approves or rejects the product.
- What must a batch production record include under 21 CFR 211.188?
Under 21 CFR 211.188, every batch needs a production and control record with complete information on how it was made and controlled.
- How do you correct a mistake in a batch record under good documentation practice?
Draw a single line through the wrong entry so it stays readable, write the correct entry beside it, and initial or sign and date the change.
- What is line clearance in pharma, and what do GMP regulations require for it?
Line clearance is the documented check, made immediately before a packaging or processing operation starts, that the line, room and equipment hold nothing from the previous job, whether product, printed materials, documents or printer settings, and that equipment is clean and set up for the next one.
- What are cleanroom grades A, B, C and D, and how do they compare with ISO 14644 classes?
EU GMP Annex 1 defines four cleanroom grades for sterile products.
- What is a contamination control strategy (CCS) under EU GMP Annex 1, and what goes in a CCS template?
A contamination control strategy (CCS) is a documented, site-wide plan showing how all your controls, from design and people to cleaning and monitoring, keep microbial, endotoxin and particle contamination out of sterile products.
- How often must a media fill (APS) be run, and what are the acceptance criteria?
Qualify a line with at least three consecutive successful media fills covering every shift, then run one about every six months for each line and shift, with every operator taking part at least once a year.
- What is PUPSIT, and does EU GMP Annex 1 require it?
PUPSIT, the pre-use post-sterilization integrity test, checks a sterilizing grade filter after sterilization and before product passes through it, so damage from preparation is found before product can mask it.
- What is chain of identity, and how does it differ from chain of custody in cell and gene therapy?
Chain of identity is the permanent link between a patient or donor and their cells, kept from collection to treatment by unique identifiers checked at every handoff.
- What is technology transfer in pharmaceutical manufacturing, and what should a tech transfer protocol cover?
Technology transfer is the planned, documented move of a product, process or analytical method, with the knowledge behind it, from a sending unit to a receiving unit: from development to manufacturing, or between sites.
- What is lot genealogy, and what forward and backward traceability do GMP rules require?
Lot genealogy is the record of which material lots went into which batches, and where each batch went.
- What GMP requirements apply to cell and gene therapy manufacturing, from Phase 1 to commercial?
In the US, cell and gene therapies need CGMP at every phase.
- How long should batch records be kept?
In the US, batch production, control and distribution records are kept at least one year after the batch’s expiry (21 CFR 211.180(a)), or three years after distribution for expiry-exempt OTC products.
- How do you calculate batch yield and reconcile materials and labels, and when must a discrepancy be investigated?
Yield is actual ÷ theoretical × 100, worked out at each phase and compared with its limits.
- What must GMP equipment use, cleaning and maintenance logbooks record?
A logbook records, in time order, every use of major equipment with the date, time, product and batch, and every cleaning and maintenance with who did it.
- What are the hold time study guidelines, and how long can material be held?
No regulator sets a maximum hold time: you set each one from your own data.
- How do you calculate F0 and validate a steam sterilization cycle?
F0 is a steam cycle’s lethality in equivalent minutes at 121 °C, assuming a z-value of 10 °C.
- How do you adjust an API quantity for potency (as-is, anhydrous or dried basis) in a batch formula?
Put the lot’s assay on the as-is basis, then scale the label-claim quantity by it.
- What pressure differential, air change rate and clean-up time does a GMP cleanroom need?
Keep a pressure differential of at least 10 Pa between cleanroom grades and clean up within 20 minutes: those are EU GMP Annex 1’s guidance values, and FDA suggests at least 10 to 15 Pa.
- How do you qualify cleanroom operators for aseptic gowning, and how often must they be requalified?
Gowning qualification proves a person can don a sterile gown without contaminating it: training, an observed gowning against a checklist, and samples from gloved fingers, forearms, chest and hood over enough sessions to show consistency.
- What is the difference between a RABS and an isolator, and what background grade does each need under Annex 1?
An isolator keeps its interior continuously isolated from the room and is bio-decontaminated automatically; a RABS is a rigid, glove-fitted barrier that is not fully sealed.
- What is the difference between reprocessing and reworking in pharma, and what approvals does each need?
In ICH Q7’s terms, reprocessing repeats a step that is already part of the established process, such as a second crystallization; reworking puts nonconforming material through different steps, such as recrystallizing from a different solvent.
- What are the requirements for visual inspection of injectables, including AQL and inspector qualification?
Every filled unit of an injectable must be visually inspected, by qualified inspectors or a validated machine at least as good, so the batch is essentially free of visible particles.
- What does a permit to work and LOTO system require in pharma?
Lockout makes the machine safe; the permit controls the job.
- What bioburden limit applies before sterile filtration, and where should the sample be taken?
In the EU, normally not more than 10 CFU per 100 mL (TAMC), tested on 100 mL of the bulk immediately before sterile filtration, and before any precautionary prefilter.
- How do you leak-test installed HEPA filters in a cleanroom, how often, and what are the acceptance criteria?
For FDA, one probe reading of 0.01% of the upstream concentration is a leak: replace the filter or repair a limited area, then retest.
- What in-process quality control tests does 21 CFR 211.110 require for tablets, and how do you set the sampling plan?
21 CFR 211.110 requires written in-process controls for every batch and names, where appropriate, tablet weight variation, disintegration, adequacy of mixing, dissolution and bioburden.
Laboratory and quality control30 answers
Laboratory and quality control
- What is the difference between OOS, OOT and OOE results?
An OOS result fails its specification, an OOT result breaks the expected trend, and an OOE result is unexpected but neither.
- How do you investigate an out-of-specification (OOS) result?
Investigate an out-of-specification (OOS) result in phases: the laboratory first, then the process.
- What are the ICH stability storage conditions for long-term, intermediate and accelerated studies?
ICH Q1A(R2) sets long term at 25°C/60% RH or 30°C/65% RH, intermediate at 30°C/65% RH and accelerated at 40°C/75% RH, each ±2°C and ±5% RH, for climatic zones I and II.
- What is the difference between alert and action limits, and how do you set them?
An alert level warns you that results are drifting, so you look into it; an action limit means something has gone wrong, so you investigate the root cause, assess product impact and raise CAPA.
- How often must GMP equipment and instruments be calibrated, and what must a calibration program include?
GMP sets no calibration interval: you set each one from the instrument’s use, its risk and its as-found results.
- What must a certificate of analysis (CoA) contain?
A certificate of analysis identifies the material, grade, batch number, release date and expiry or retest date.
- What does ICH Q2(R2) require for analytical method validation, and what changed from Q2(R1)?
ICH Q2(R2) asks you to validate a method against a protocol with acceptance criteria set in advance, testing what its use needs: specificity alone for identity, and specificity, response, accuracy and precision for an assay.
- What is the difference between analytical method validation, verification and transfer?
Validation proves a procedure is fit for its intended purpose by studying characteristics such as specificity, accuracy, precision and range.
- What are reserve (retention) samples, how much must be kept, and for how long?
Reserve samples, called reference and retention samples in the EU, let a batch be retested or its pack examined later.
- What changes in the new ICH Q1 stability guideline, and when does it apply?
ICH adopted the new ICH Q1 as final in November 2026 but has not yet published the text, so Q1A(R2) to Q1E and Q5C apply until each regulator implements it.
- What is sample management in a GxP lab, and what must a sample record show?
Sample management is the control of each sample from the plan that calls for it through sampling, labeling, receipt, storage, testing, retention and disposal, so every result traces to a known, representative sample.
- How do you design an in-use stability study for a multidose or reconstituted product?
Simulate use after opening, reconstitution or dilution: withdraw as instructed, store as labeled, and test physical, chemical and microbial quality at intermediate points and on the last dose.
- What are the acceptance criteria for growth promotion testing of culture media, and how is the factor of 2 calculated?
Each batch of medium for a pharmacopeial test must grow its organisms from not more than 100 CFU.
- What is the difference between a retest date and an expiry date, and how do you assign a retest period?
A retest date is when material should be re-examined against its specification before further use; an expiry date is when it must no longer be used.
- How many batches must go into the ongoing stability program each year?
At least one batch a year of every strength and every primary packaging type made that year, unless otherwise justified, under EU GMP 6.32.
- How do you qualify a working (secondary) reference standard against a pharmacopoeial standard?
Start from a high-purity batch that meets the monograph.
- How do you calculate the f2 similarity factor for dissolution profiles, and when is an f2 comparison valid?
f2 = 50 × log10{100 ÷ √[1 + (1/n)Σ(Rt − Tt)²]}, from mean percent dissolved of 12 or more units per product at shared time points; 50 to 100 means similar.
- How do you calculate shelf life from stability data using ICH Q1E regression?
Shelf life is the earliest time the 95% confidence limit for the fitted average line reaches the acceptance criterion.
- What are the TAMC and TYMC limits for non-sterile products, and how is a limit of 10² CFU read?
The harmonized table recommends a TAMC of 10³ and a TYMC of 10² CFU/g or mL for non-aqueous oral and rectal products, and 10² and 10¹ for aqueous oral, mouth, skin, nasal, ear, inhaled and vaginal products and patches, most with named organisms absent.
- How do you calculate the reporting, identification and qualification thresholds for impurities under ICH Q3A and Q3B?
Identification and qualification thresholds are the lower of the table percentage and the daily intake limit divided by the maximum daily dose.
- How do you calculate residual solvent limits with ICH Q3C Option 1 and Option 2?
A Class 2 solvent’s ICH Q3C limit in ppm is 1000 × PDE (mg a day) ÷ dose (g of drug product a day).
- How do you calculate an endotoxin limit and the maximum valid dilution (MVD)?
Endotoxin limit = K ÷ M.
- What are the dissolution acceptance criteria at S1, S2 and S3, and for extended- and delayed-release products?
At S1, each of 6 units must dissolve at least Q + 5%.
- How many containers does a sterility test need, and how much of each must be tested?
For parenterals, test at least 20 containers from batches over 500, or 2% if fewer; 10 from 101 to 500; and 10% or 4, whichever is more, up to 100.
- How do you run an ICH Q1B photostability study, and how do you know the light dose was reached?
Expose samples to at least 1.2 million lux hours and 200 Wh/m² of near-UV energy, from a D65/ID65-type source or cool white fluorescent plus near-UV lamps, with dark controls.
- What system suitability criteria does an HPLC method need under USP <621>, and how is the maximum RSD worked out?
HPLC tests meet their monograph’s criteria plus two USP <621> defaults: symmetry (tailing) factor 0.8 to 1.8 for the quantified peak in organic impurities, related substances and assays, waived in impurity tests that prescribe an RSD; and S/N of at least 10 at a reporting threshold.
- When is container closure integrity testing required, and which test method should you use?
For sterile products: to validate the package and its sealing, in routine production, and on stability if it replaces the sterility test.
- How do you design a forced degradation study, and how much degradation is enough?
No ICH or FDA text sets a target; WHO asks for about 10 to 30% loss of assay, forming primary degradants without secondary ones.
- What are the subvisible particle limits for injections under USP <788> and Ph. Eur. 2.9.19, and how is the test run?
By light obscuration, an injection of 100 mL or less may contain at most 6,000 particles of 10 µm or larger and 600 of 25 µm or larger per container; above 100 mL, and at 100 mL in JP, 25 and 3 per mL.
- How do you run a tablet friability test, how is friability calculated, and what is the limit?
Friability is the percentage of mass a dedusted tablet sample loses in 100 drum rotations at 24 to 26 rpm: (initial mass − final mass) ÷ initial mass × 100.
Suppliers and materials13 answers
Suppliers and materials
- How do you qualify a supplier in the pharmaceutical industry?
Rank the material’s risk, trace its supply chain to the original manufacturer, assess the supplier, audit where required (EU GMP requires it for active substance manufacturers and distributors), agree specifications and responsibilities in writing, then have the quality unit approve a named manufacturer, site and material.
- What is a quality agreement, and what should a quality agreement template include?
A quality agreement is the written split of GMP duties between a product owner and a contract manufacturer, laboratory or other contractor: for each activity, who performs it, who approves it and who reports it.
- How do you choose a GMP or validation consultant?
Write the brief and the deliverables before you call anyone.
- How should a temperature excursion during storage or shipment be handled?
Treat it as a deviation.
- How should incoming materials be received, quarantined, sampled and released for use?
Check each delivery against the order and approved supplier list, inspect every container for labeling, damage, broken seals and contamination, and code each lot in each shipment.
- How many containers of an incoming raw material must be sampled, and is the √n+1 rule required?
None of FDA, EU GMP or Health Canada requires √n+1.
- Can you rely on a supplier’s certificate of analysis instead of testing raw materials yourself?
Partly.
- What is good distribution practice (GDP) for medicines?
Good distribution practice (GDP) is the part of quality assurance that keeps a medicine’s quality and integrity while it is bought, stored, transported and supplied, and keeps falsified products out of the supply chain.
- What does the ICH Q3E draft require for extractables and leachables, and when will it apply?
ICH Q3E is still a draft.
- How do you run a temperature mapping study, and how many data loggers does it need?
Temperature mapping uses calibrated loggers to find hot and cold spots that decide where product and monitoring probes go.
- What is controlled room temperature, and is USP changing it to 15–25 °C?
Controlled room temperature (CRT) is USP’s “temperature maintained thermostatically that encompasses the usual and customary working environment of 20°–25°”, with mean kinetic temperature (MKT) not above 25 °C.
- Is there an FDA requirement for shipping validation, and how do you qualify a temperature-controlled shipping lane?
No FDA rule is titled shipping validation.
- How do pharma sites manage chemical segregation, GHS labels and SDSs?
Sort storage in two passes.
Choosing systems29 answers
Choosing systems
- What is the difference between a LIMS and an ELN?
A LIMS is organized around the sample: it registers each sample, assigns tests, methods and specifications, records results and routes them for review and disposition.
- What is a LIMS, and what is it used for in a pharma QC lab?
A LIMS (laboratory information management system) is software that runs a laboratory’s testing and keeps its records.
- What is the difference between MES and ERP in pharmaceutical manufacturing?
ERP (enterprise resource planning) runs the business side: orders, purchasing, inventory and distribution.
- What makes software 21 CFR Part 11 and GMP compliant?
No software is Part 11 or GMP compliant on its own, and FDA certifies none.
- How much does a LIMS cost?
Published subscriptions start at $550 a month billed yearly by testing volume (Confident LIMS), $310 a user a month for three users (CloudLIMS) and $16,500 a year for five users (QBench).
- What is an eQMS (electronic quality management system), and what does it do?
An eQMS (electronic quality management system) is software that runs a company’s quality processes and keeps their records: document control and training, deviations and nonconformances, investigations and CAPA, change control, complaints, audits, suppliers, risk and management review.
- How much does eQMS software cost?
Few vendors publish eQMS prices.
- What is an MES (manufacturing execution system) in pharma, and how does it differ from an eBR?
A manufacturing execution system (MES) is the software that runs and records production on the shop floor: it issues the approved master record as steps, checks materials and equipment, captures weights and process data, and reports exceptions.
- How do you choose an eQMS, and what should its requirements include?
Choose an eQMS from the quality processes you run and the rules behind them: 21 CFR 211, EU GMP Chapter 1 and ICH Q10 for medicines, the QMSR and ISO 13485 for devices.
- How do you choose a LIMS, and what should LIMS requirements include?
Choose a LIMS by testing suppliers on your requirements, not their feature lists.
- How long does a LIMS implementation take?
No regulation sets a duration.
- What is the difference between an eQMS and a document management system?
A document management system (DMS or EDMS) controls instructions: review, approval, versions, distribution and retention.
- What is an electronic lab notebook (ELN), and what should it do in a GxP lab?
An electronic lab notebook (ELN) is software that replaces the paper notebook: scientists record methods, samples, materials, observations, calculations and results as they work.
- When should you move from a paper QMS to an eQMS, and how do you migrate?
Neither FDA nor EU GMP requires an eQMS; controlled paper is acceptable.
- How do you move from paper batch records to electronic batch records?
Treat it as redesigning the batch record, not scanning it.
- How do you choose an MES or electronic batch record (eBR) system?
Choose an MES or eBR by having each shortlisted supplier run your master batch record, failure paths included, and judging it against 21 CFR Parts 211 and 11, EU GMP Chapter 4 and Annex 11.
- How do LIMS, MES and QMS fit together, and do you need separate systems?
A LIMS holds samples, methods, specifications and results; an MES runs each batch from the approved master record and creates the batch record; a QMS holds procedures, training, deviations, CAPA and change control.
- How long does it take to implement and validate an eQMS?
No regulation sets a timetable.
- How do you integrate lab instruments with a LIMS?
Connect each instrument so results reach the LIMS without retyping, carrying sample identity, units, method and times, and link each result to its original electronic record, metadata and audit trail.
- Can you sign GMP records with DocuSign, Adobe Acrobat Sign or a scanned signature under 21 CFR Part 11?
Yes, with DocuSign’s Part 11 module or Acrobat Sign’s enterprise service, configured and validated by you; not with a pasted image.
- How do you validate a data migration from one GxP system to another?
Treat the migration as a validated process with its own plan.
- How do you validate a LIMS, and what should LIMS validation test scripts cover?
Validate a LIMS against your intended use: write requirements, assess the supplier, record the configuration, rate each function’s risk, then test with approved scripts and report before GMP use.
- How much does an eBR or MES cost, and how long does implementation take?
An eBR or MES programme’s cost and timeline are driven less by software licences than by recipe authoring, ERP and equipment interfaces, and validation for intended use.
- How do you choose validation management software (VLMS) for GxP?
Choose a VLMS from how often your regulated systems change, not only how you run initial qualification projects.
- How do you choose a GxP CMMS and calibration management system?
Choose a CMMS by what it can prove after something fails.
- How do you choose an electronic lab notebook (ELN) for life sciences?
Choose an ELN with your own experiments, because every product looks good on its own.
- How do you choose an electronic trial master file (eTMF) for GCP?
Choose an eTMF by testing how it proves trial completeness in real time rather than how many folders it holds.
- How do you choose a clinical trial management system (CTMS)?
Choose a CTMS by checking whether its study milestones rest on live operational evidence rather than manual date updates.
- How do you choose a regulatory information management system (RIMS)?
Choose a RIMS by checking how cleanly it connects registered product details back to the lab, stability, manufacturing and change-control records that keep dossiers accurate.
AI in GxP13 answers
AI in GxP
- Can AI be used in GMP?
Yes.
- What is EU GMP Annex 22 on artificial intelligence, and what does it allow?
Annex 22 is a draft EU GMP annex on AI, not yet in force.
- What does FDA guidance say about using AI in drug development and manufacturing?
FDA’s guidance on AI for drugs is still a draft: the January 2025 guidance on AI that supports regulatory decisions, which recommends a seven-step credibility assessment scaled to the model’s risk.
- How do you validate an AI tool or machine learning model for GxP use?
Validate it as a GxP computerized system, scaled to risk, then add what a trained model needs: an intended use describing its input data, a measured baseline for the process it replaces, acceptance criteria approved before testing, independent, carefully labeled test data, results with confidence intervals by subgroup, and change control and drift monitoring in use.
- What did FDA’s first warning letter about AI in drug manufacturing say?
Warning letter 320-26-58, issued to Purolea Cosmetics Lab on 2 April 2026, is the first FDA warning letter to cite a drug maker’s use of AI.
- Can ChatGPT and other generative AI or LLMs be used in GMP work?
Yes, in an approved tool, for non-critical GMP work that a qualified reviewer checks, such as drafting, summarizing or translating, but not for uses with direct impact on patient safety, product quality or data integrity.
- Can AI review batch records?
Yes, as a preparer, not as the reviewer of record.
- Can AI write deviation reports and help with root cause analysis and CAPA?
Yes, as a drafting and evidence-gathering aid, but people own the investigation.
- Can you use AI to write SOPs and other GMP documents?
Yes, as an approved drafting aid, if a qualified person checks every line against the regulation, the product file and what the site actually does, and it is approved like any other procedure, by the quality unit under FDA rules.
- Does the EU AI Act apply to AI used in pharmaceutical manufacturing and quality?
Partly.
- What does human in the loop mean for AI in GMP, and is it enough?
Human in the loop means a qualified person checks AI output and makes the decision while the work waits.
- Can AI be used for automated visual inspection of injectable products, and how do you qualify it?
Yes.
- Can smart glasses and computer vision be used in GMP manufacturing?
Yes, when smart glasses and computer vision draft observations or capture visual evidence for a qualified operator to verify and sign, rather than making unverified GMP decisions on their own.
See your process in Seal.