How do you perform an audit trail review, and how often should audit trails be reviewed?

An audit trail review is the documented check that the creations, changes and deletions behind GxP data are complete, attributable and justified. FDA asks whoever reviews a record to review its audit trail with it, at the record’s own review points, such as each significant production step and batch release. PIC/S and the draft revised Annex 11 add independent or peer review. Review system trails, such as access, settings and alerts, periodically, at a frequency set by risk.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal following successive procedure versions and the version used for a batch.

1How to review an audit trail: steps and checklist.

  1. Open the trail in the system, beside the original electronic data, not a printout or summary report.1,2
  2. Run the validated report or filter your procedure sets for that system, and note what it leaves out.2,3
  3. Compare the entries with the other records: notebook, logbooks, instrument use log and reported results. Ask whether this is the only data the activity generated.3,2
  4. Check each change, deletion and exception against the lines below, verifying the reason against the source record.4
  5. Investigate and record any significant variation from the expected outcome.3,4
  6. Sign a positive statement of whether issues were found, with the date, before anyone relies on the data.2,3

The first table is written for a chromatography data system, FDA’s example of an audit trail;5 the second for batch records and LIMS; the third for the system’s own trails, reviewed on a schedule. Each line names the text it satisfies.

1.1With each laboratory record, before the data are used.

CheckWhat to look forSource
All injectionsEvery acquired injection, aborted and failed ones included, is reported or explained in the notebookFDA Q&A 12;5 21 CFR 211.194(a)6
Unexplained injectionsTrial, test or equilibration injections of samples, or injections no sample accounts forFDA Q&A 135
ReprocessingEvery processing run kept with its result; parameters not changed step by step until a result passesFDA Q&A 14;5 MHRA 6.92
Manual integrationDone as an approved procedure allows, with the reason, and the original keptPIC/S 9.7;3 Annex 11, 97
Changed valuesOld value, new value, who, when, and a reason the source record supportsAnnex 11, 9;7 draft Annex 11, 12.2 and 12.74
Deleted or excluded dataDeleted, renamed or invalidated files and results, each scientifically justified, with the original data keptFDA Q&A 25
Method and sequence editsSequence or processing method changes, and the science behind eachFDA Q&A 12;5 EMA Q&A 7 and 81
User and roleThe analyst in the notebook, on a personal account without administrator rightsFDA Q&A 4 and 55
TimeEntries match the notebook, balance printouts and instrument log, with no gap while the instrument ranPIC/S 9.8;3 Ava warning letter8

1.2In eBR, MES and LIMS records, step by step.

In production, FDA reads 21 CFR 211.188(b) as requiring review after each significant step, which the person supervising or checking the step may do; the quality unit then reviews the whole record before release.5,9,10 The draft Annex 11 expects the trail to log every manual user interaction, alarm acknowledgments and electronic signatures included.4

CheckWhat to look forSource
Manual entriesValues keyed in, not captured, with old and new value, user, time and reason, and the second check critical entries needdraft Annex 11, 12.1 and 12.2;4 Annex 11, 67
Weighing overridesWeights keyed in instead of read from the balance, and the second person’s check of each component21 CFR 211.101(c) and (d);11 Annex 11, 67
Alarm acknowledgmentsCritical alarms acknowledged by authorized users, in time, with a comment and the action takendraft Annex 11, 8.4, 8.5 and 8.74
Recipe and set pointsSet points changed during the batch, each approved or justified as a deviation21 CFR 211.100;12 draft Annex 11, 12.14
Electronic signaturesEach significant step signed by its performer and checker, with meaning, date and time21 CFR 211.188(b)(11);9 draft Annex 11, 13.4 to 13.64
LIMS specificationsLimits, test methods or sampling plans changed only through a change the quality unit approved21 CFR 211.160(a)13

1.3At system level, on a set schedule.

CheckWhat to look forSource
Audit trail settingsTrail on and locked; any change to its settings or the clock recorded and explainedPIC/S 9.6;3 WHO 11.11;14 draft Annex 11, 12.34
Accounts and rolesAccess created, changed or canceled, administrator activity, accounts that should have goneAnnex 11, 12.3;7 draft Annex 11, 11.114
Log-insSuccessful and failed log-in attempts, with user, role, date and timePIC/S 9.53
Instrument logsInstrument operational status, communication logs and alert recordsFDA Q&A 8, note 135

2Example: an audit trail review log.

This log is invented for illustration: the product, batch, instrument, procedures and people are not real. It applies the laboratory checklist to one HPLC sequence.

Table 4. Example. An audit trail review log, invented for illustration.
FieldEntry
RecordAssay and related substances, Exemplol 50 mg tablets, Batch 26-⁠052, sequence SEQ-⁠26-⁠0388 on HPLC-⁠07, acquired 12 May 2026
ReviewerSecond-person QC reviewer, not involved in the testing
ScopeThe sequence’s full audit trail, its results, the processing method history and the instrument alert log for 12 May. Log-ons and print events left out by validated report AT-⁠RPT-⁠03
Injections14 planned, 15 acquired: injection 9 aborted at 14:32 on a logged pressure alarm and repeated from the same vial at 14:51. All 14 complete injections are in the result set
Changed valuesPreparation B sample weight changed from 50.12 mg to 50.21 mg at 15:10 by the analyst, reason “transcription error”. Matches the balance printout in notebook NB-⁠0447, page 31
Sequence and methodSequence edited once, at 14:49, to add the repeat of injection 9, reason “repeat after pressure alarm”. Processing method unchanged since its approval on 3 March 2026
Deleted or excludedNone. No file deleted, renamed or invalidated; aborted injection 9 kept with its alarm record
Manual integrationInjection 13, once, reason “shoulder not resolved, SOP-⁠QC-⁠045 6.3”. Impurity B 0.09% automatic, 0.14% manual, limit 0.20%; both kept, and the manual result reported as the procedure requires
Users and rolesTwo users, both in the analyst role. No administrator actions, clock changes or audit trail setting changes
TimeEntries agree with the notebook and the instrument use log, 08:55 to 17:20
FindingsOne: the aborted injection, recorded as laboratory incident LI-⁠26-⁠112 with the alarm record and the repeat
ConclusionIssues found: one, resolved as above. No effect on the reported results. Data acceptable for batch disposition
Sign-offReviewer, 13 May 2026. QA reviews it with the batch record before release on 15 May

Three calls are worth copying. The weight change was checked against the balance printout, not the reason alone, because the draft Annex 11 makes verifying the reason “a key element”.4 The log says which way manual integration moved the result, because MHRA wants processing visible so parameters are not “manipulated to achieve a more desirable result”.2 And the aborted injection became an incident although it gave no result, because FDA says aborted or incomplete injections “should be investigated and justified”.5

3How often to review audit trails.

FDA ties the frequency to the data: where the CGMP regulations set one for the record, such as 211.188(b) for each significant step and 211.22 before release, the audit trail follows it; otherwise a risk assessment of “data criticality, control mechanisms, and impact on product quality” decides.5,9 Annex 11, in force since 2011, says only “regularly reviewed”; MHRA asks for review “where this is determined by a risk assessment”; WHO sets the frequency by system criticality and trail category.7,2,14 None of these, PIC/S PI 041-⁠1 or the draft Annex 11 sets a number of days or months.

Table 5. A decision aid: which trail to review when, and the status of each source.
TrailWhen to review itSourceStatus
Release-decision dataWith the record, before the decisionFDA Q&A 8;5 PIC/S 9.8;3 draft Annex 11, 12.8 and 12.104FDA and PIC/S guidance; EU draft, not yet in force
Batch record stepsAs each significant step is checkedFDA Q&A 8, on 21 CFR 211.188(b)5FDA guidance on a binding rule
Non-critical trailsDuring system reviews at a set frequencyPIC/S 9.63PIC/S guidance
System, access and alertsPeriodically, at a frequency set by riskFDA Q&A 8, note 13;5 WHO 12.314FDA guidance; WHO guideline
Quality unit oversightRandom and targeted reviews within self-inspectionPIC/S 9.83PIC/S guidance
The system as a wholeAt periodic reviewAnnex 11, 11;7 draft Annex 11, 14.24EU GMP in force; EU draft

4Who reviews audit trails.

FDA: “Personnel responsible for record review under CGMP should review the audit trails that capture changes to data associated with the record as they review the rest of the record.” The quality unit reviews and approves all production and control records, which FDA says include audit trails, and laboratory records need a second person’s review.5,10,6 For production steps, “The regulations provide flexibility to have some activities reviewed by a person directly supervising or checking information (e.g., § 211.188).”5

PIC/S goes further: critical audit trails “should be independently reviewed” with the other records of the operation, by the originating department, with the quality unit verifying where necessary.3 The draft Annex 11 would ask for review by “personnel not directly involved in the activities covered by the review (a peer review)”.4 MHRA asks that reviewers have “sufficient knowledge and system access” to see the trails, raw data and metadata.2

4.1At a CDMO or contract laboratory.

MHRA: “Where data review is not conducted by the organisation that generated the data, the responsibilities for data review must be documented and agreed by both parties.” Summary reports “are limited and critical supporting data and metadata may not be included”, it warns.2 So the quality agreement should name who reviews the trails behind each released result, and how the contract giver checks that review.

5What the procedure has to decide.

FDA asks for the procedure itself: Ava Inc. was told to “Provide your procedures for audit trail review.”8 Beyond the checklist rows, it settles:

  • Scope and depth. Which trails and entries matter for each system, assessed during qualification, and what the validated report leaves out and why.3 Review need not include “every system activity (e.g. user log on/off, keystrokes etc.)”, says MHRA; the draft Annex 11 aims it at “repetition of activities, errors, omissions, unauthorised process deviations and loss of data integrity”.2,4
  • Who reviews what, and when, for each system or type of system.4
  • Filters and exception reports, each validated as a search tool.2,1 Review by exception covers how they are set up.
  • Escalation, up to senior management and, where necessary, national authorities.3

What the trail must capture is in audit trail requirements.

6Where reviews go wrong.

  • Reviewing a printout or PDF. EMA: electronic data “is the original record which must be reviewed and evaluated prior to making batch release decisions”, and a review based only on printouts can miss records holding uninvestigated failures.1 The draft Annex 11 says of electronic copies: “Flat and locked files are not acceptable”.4
  • Reviewing only what was reported. EMA’s example of an inspection citing: “Raw data for HPLC/GC runs which had been invalidated was stored separately to the QC raw data packages and had not been included in the review process.”1
  • Procedures that create the problem. Ava’s gas chromatography procedure told analysts a trial injection “is not processed as a part of raw data”, which FDA called “particularly concerning”.8

7What inspectors look for.

On every pre-approval inspection, Compliance Program 7346.832, reissued on 29 June 2026, has investigators compare raw data with the summary data filed, names stability, dissolution, content uniformity and API impurity data as “good candidates”, and lists “possible indications of data integrity problems”.15 Run the matching query on the trails behind each filed data set first.

Table 6. A pre-approval self-check: concerns and indications from CP 7346.832, and the audit-trail query that would reveal each.
IndicationCP 7346.832 wordingAudit-trail queryWhere to look
Gaps in a sequence“absences in a submitted chromatographic sequence”Injections acquired against those reportedCDS sequence trail
Repeat testing“Multiple analyses of assay using the same sample without adequate justification”Injections per sample ID against the method’s countCDS trail, LIMS result history
Missing stability lots“Exclusion of specific lots from the stability program to avoid submitting failed results”Lots, pulls or sample sets canceled, deleted or invalidatedLIMS stability module, CDS deletions
Backdating“Backdating stability test results to meet required commitments”Entry and signature times against acquisition times; clock changesCDS and LIMS trails, clock log
Reprocessing to pass“Manipulation of a poorly defined analytical procedure and associated data analysis to obtain passing results”Method versions and reprocessing runs per injectionCDS method history
Results without tests“Fabrication of acceptable test results without performing the test”Results with no instrument activity behind themInstrument trail and use log
“The audit trail for the Fourier Transform Infrared (FTIR) spectroscopy equipment showed no activity from September 23 to 30, 2025. However, your testing records, including the FTIR daily usage log, showed drug testing was performed with this equipment during this time frame.” Ava Inc., warning letter 320-⁠26-⁠64, 14 April 20268
“Your laboratory analysts performed thousands of unexplained injections using Empower over the last 30 months, and over 70 trial injections using TotalChrom in October 2025 alone.” Curia New York, Inc., warning letter 320-⁠26-⁠127, 18 September 202616

EU inspectors cite the same failures. At an Indian tablet site, Malta’s inspectors found the audit trail disabled on chromatographic sequences, and:

“Sequence files for HPLC and GC were in folders with read/write access allowing renaming and deleting of data by analysts. Sequence files whose analysis was reviewed complete in one month showed modified dates in another month.” Zenzi Pharmaceutical Industries Private Limited, statement of non-compliance MT/002NCR/2026, Malta Medicines Authority, 5 May 202617

A trail that stays quiet while the logbook shows work is a finding, not a pass, and a file modified after its review needs reviewing again.

8What changed from 2024 to 2026.

  • Annex 11 is still the 2011 text. The Commission consulted on the draft quoted above from 7 July to 7 October 2025, and its Volume 4 page still lists the 2011 version; see Annex 11 revision.18,19,4
  • ICH E6(R3), adopted on 6 January 2025, makes review of clinical trial data, audit trails and metadata “a planned activity” whose extent is risk-based.20

9How it runs in Seal.

In Seal, each record keeps its own audit log, and the audit-trail export from a vendor system, such as a chromatography data system, is attached and reviewed beside the batch, test or result it belongs to. Events are prioritized by criticality, rules that suppress expected events are versioned, justified, tested and approved, and a required review can gate the result or disposition, so the review comes before the decision. The data integrity and audit trail review blueprint follows an integration change made after an out-of-specification result: it holds the result, opens DI-⁠026 and places the batch on hold.

Neil can prepare an audit-trail review: it reads the export, the records in scope and the review criteria, and drafts a brief with an indexed set of exceptions. The authorized reviewer decides what is significant and signs the review.

AQuestions and answers

Can the person who made a change review its audit trail?

Not as the review the record relies on: FDA wants the record’s reviewer to review its audit trail, and PIC/S and the draft revised Annex 11 expect that review to be independent of the work, though analysts can still check their own entries first.

Can audit trails be reviewed from a PDF or printout?

Not as a substitute for the electronic data, which EMA treats as the original record to review before release, and the draft revised Annex 11 rejects flat, locked files as electronic copies.

Does audit trail review apply to paper records?

The paper equivalent is the crossed-out entry, which should leave the original legible and show who changed it, when and why. FDA compares audit trail review to assessing cross-outs when reviewing paper data, and MHRA sets the same risk-based expectation for reviewing cross-outs and audit trails.

References20 sources

References

  1. 1Guidance on good manufacturing practice and good distribution practice: questions and answers, data integrity, questions 7, 8, 16 and 17, European Medicines Agency. https://www.ema.europa.eu/en/human-regulatory-overview/research-development/compliance-research-development/good-manufacturing-practice/guidance-good-manufacturing-practice-good-distribution-practice-questions-answers (read 26 November 2026)
  2. 2‘GXP’ Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018, sections 6.9, 6.13, 6.15 and 6.16. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 26 November 2026)
  3. 3PI 041-⁠1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, PIC/S, 1 July 2021, sections 9.5 to 9.8. https://picscheme.org/docview/4234 (read 26 November 2026)
  4. 4Annex 11: Computerised Systems, draft revision for consultation, EMA GMP/GDP Inspectors Working Group and PIC/S, 2025, sections 8, 11.11, 12, 13 and 14.2. https://picscheme.org/docview/9714 (read 26 November 2026)
  5. 5Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, FDA, December 2018, questions 1.c, 2, 4, 5, 7, 8, 12, 13 and 14. https://www.fda.gov/media/119267/download (read 26 November 2026)
  6. 621 CFR 211.194, Laboratory records, eCFR. https://www.ecfr.gov/current/title-21/section-211.194 (read 26 November 2026)
  7. 7EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, in operation since 30 June 2011, sections 6, 9, 11 and 12. https://health.ec.europa.eu/system/files/2016-11/annex11_01-2011_en_0.pdf (read 26 November 2026)
  8. 8Ava Inc., warning letter 320-⁠26-⁠64, MARCS-⁠CMS 721180, FDA, 14 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/ava-inc-721180-04142026 (read 26 November 2026)
  9. 921 CFR 211.188, Batch production and control records, eCFR. https://www.ecfr.gov/current/title-21/section-211.188 (read 26 November 2026)
  10. 1021 CFR 211.192, Production record review, eCFR. https://www.ecfr.gov/current/title-21/section-211.192 (read 26 November 2026)
  11. 1121 CFR 211.101, Charge-in of components, eCFR. https://www.ecfr.gov/current/title-21/section-211.101 (read 26 November 2026)
  12. 1221 CFR 211.100, Written procedures; deviations, eCFR. https://www.ecfr.gov/current/title-21/section-211.100 (read 26 November 2026)
  13. 1321 CFR 211.160, General requirements (laboratory controls), eCFR. https://www.ecfr.gov/current/title-21/section-211.160 (read 26 November 2026)
  14. 14Guideline on data integrity, WHO Technical Report Series No. 1033, Annex 4, World Health Organization, 2021, sections 11 and 12. https://cdn.who.int/media/docs/default-source/medicines/norms-and-standards/guidelines/inspections/trs1033-annex4-guideline-on-data-integrity.pdf (read 26 November 2026)
  15. 15Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026, implementation date 10 August 2026, Objective 3: Data Integrity Audit. https://www.fda.gov/media/193382/download (read 26 November 2026)
  16. 16Curia New York, Inc., warning letter 320-⁠26-⁠127, MARCS-⁠CMS 728790, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 26 November 2026)
  17. 17Zenzi Pharmaceutical Industries Private Limited, statement of non-compliance with GMP MT/002NCR/2026, Malta Medicines Authority, EudraGMDP, inspection ending 31 March 2026, issued 5 May 2026. https://eudragmdp.ema.europa.eu/inspections/gmpc/searchGMPNonCompliance.do (read 26 November 2026)
  18. 18Stakeholders’ Consultation on EudraLex Volume 4, Good Manufacturing Practice Guidelines: Chapter 4, Annex 11 and New Annex 22, European Commission, open 7 July to 7 October 2025, closed. https://health.ec.europa.eu/consultations/stakeholders-consultation-eudralex-volume-4-good-manufacturing-practice-guidelines-chapter-4-annex_en (read 26 November 2026)
  19. 19EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https://health.ec.europa.eu/medicinal-products/eudralex/eudralex-volume-4_en (read 26 November 2026)
  20. 20ICH E6(R3), Guideline for Good Clinical Practice, ICH, adopted 6 January 2025, sections 4.2.2 and 4.2.3. https://database.ich.org/sites/default/files/ICH_E6%28R3%29_Step4_FinalGuideline_2025_0106.pdf (read 26 November 2026)