What is an electronic lab notebook (ELN), and what should it do in a GxP lab?

An electronic lab notebook (ELN) is software that replaces the paper notebook: scientists record methods, samples, materials, observations, calculations and results as they work. In a GxP lab its entries are regulated records, so it should identify who entered each value and when, keep every change and its reason in an audit trail, hold or link the original instrument data, let reviewers check the data and audit trail and sign electronically, and be validated for each GxP workflow it runs.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal reading a laboratory notebook ruled as a table, with a red thread from one entry to a beaker.

1Which rules apply to an ELN.

Neither Part 11 nor Annex 11 mentions a lab notebook.1,2 Part 11 covers electronic records kept under FDA’s other record rules, which FDA calls predicate rules, and electronic records submitted to FDA (Part 11 requirements);1,3 Annex 11 to all computerized systems used in GMP-regulated activities.2 FDA says the requirements for record retention and review “do not differ depending on the data format”.4 Which apply depends on the work:

  • GLP studies. 21 CFR Part 58 covers nonclinical laboratory studies that support applications for FDA research or marketing permits.5
  • GMP testing. Laboratory records under drug CGMP must contain what 21 CFR 211.194 lists.6 EU GMP names laboratory notebooks among the raw data a QC laboratory retains and keeps readily available.7
  • Development and research. A record that is not submitted but is “used in generating a submission” is “not a part 11 record unless it is otherwise required to be maintained under a predicate rule”, FDA says.3 So a development notebook may sit outside Part 11 and still be audited: FDA tells pre-approval inspectors to “audit research and development notebooks” and to compare analysts’ notebooks and other raw data with the summary data filed.8

2GxP ELN entry template.

Build these fields into every template, required where the system allows.

Table 1. A GxP ELN entry template.
FieldWhat to recordRequirement
Entry number, statusNumber assigned by the system; draft, signed, reviewed or amendedControlled, reconciled forms: FDA Q&A 64
Author, date, timeTaken from the login and clock, not typed21 CFR 58.130(e);5 11.10(e)1
Purpose and linkObjective; the protocol, study or batch the work supports58.130(a);5 EU GMP 4.109
MethodProcedure and version followed; any departure, with its reason211.160(a), 211.194(a)(2) and (b)6
Samples, materialsSample identity; for a sample received, its source, quantity, lot and dates taken and received; standard and reagent lots and expiry dates, traceable to their standardization records211.194(a)(1) and (c);6 EU GMP 6.217
EquipmentInstrument identity and calibration status211.160(b)(4);6 EU GMP 6.177
Weights and volumesCaptured from the instrument, or typed and checked211.194(a)(3);6 Annex 11, 62
ObservationsWhat happened, recorded when it happenedEU GMP 4.8;9 58.130(e)5
Raw dataThe original instrument files, attached or linked211.194(a)(4);6 FDA Q&A 104
CalculationsFormula, units, conversion and equivalency factors211.194(a)(5)6
ResultsResult against the acceptance criterion or objective211.194(a)(6)6
CorrectionsOriginal left readable; reason, date and person, kept in the audit trail58.130(e);5 EU GMP 4.9;9 11.10(e);1 Annex 11, 92
SignaturesAuthor and reviewer: printed name, date and time, meaning211.194(a)(7) and (8);6 11.50 and 11.701

3Example: a completed ELN entry.

This entry is invented for illustration: one day of a method validation study in analytical development.

Table 2. Example. ELN entry EXP-⁠26-⁠0193.
FieldEntry
Entry number, statusEXP-⁠26-⁠0193, signed and reviewed.
Author, date, timeR. Lane (R.L.), Wenmoor analytical development. Opened 5 March 2026 at 09:15 from the login.
Purpose and linkAccuracy at 100% of label claim for assay method AM-⁠118 draft 2, Exemplol 25 mg tablets (in development). Validation protocol VP-⁠26-⁠007, approved 2 March 2026; acceptance 98.0 to 102.0% for each preparation.
MethodAM-⁠118 draft 2, steps 1 to 6. No departures.
Samples, materialsPlacebo blend PB-⁠26-⁠003. Exemplol reference standard RS-⁠EXE-⁠04, potency 99.6%, expires 31 December 2026.
EquipmentBalance BAL-⁠03 and HPLC-⁠07, both within calibration. Column C-⁠2219.
Weights and volumesThree spiked placebo preparations, weights captured from BAL-⁠03 (below). Each made up to 50.0 mL.
ObservationsAll three solutions clear at make-up.
Raw dataChromatography sequence SEQ-⁠26-⁠0311 linked, with processing method PM-⁠118-⁠2.
CalculationsAdded = weight × 0.996, the potency of RS-⁠EXE-⁠04. Recovery = found ÷ added × 100.
ResultsRecoveries 100.28%, 99.36% and 99.68%; mean 99.77%. All three within 98.0 to 102.0%.
CorrectionsColumn serial typed as C-⁠2291, corrected to C-⁠2219 on 5 March 2026 at 11:02 by R.L.; reason: digits transposed, checked against the column label. The audit trail keeps both values.
SignaturesAuthor R. Lane, 5 March 2026 at 16:02, meaning authorship. Reviewer A. Mehta, 9 March 2026 at 10:05, meaning review against the original data and audit trail.
Table 3. Example. The three preparations in EXP-⁠26-⁠0193.
PreparationWeighed (mg)Added (mg)Found (mg)Recovery
125.14125.0425.11100.28%
225.07024.9724.8199.36%
325.20125.1025.0299.68%

4How to keep an electronic lab notebook.

  1. Record as you work, in the notebook. FDA says it is not acceptable to record data “on pieces of paper that will be discarded after the data are transcribed to a permanent laboratory notebook”.4 EU GMP asks for records made “at the time each action is taken”.9
  2. Use your own login. When credentials are shared, “a unique individual cannot be identified”.4
  3. Keep the original data with the entry. A printout of a file that can be reprocessed “would not satisfy CGMP requirements to retain original records or true copies”.4 FDA and MHRA both make an exception for simple instruments that only print, such as a balance or pH meter: the printout is the original record, so keep it or a true copy.4,10
  4. Keep every record retrievable for its retention period. Part 11 requires “accurate and ready retrieval throughout the records retention period”, and audit trails kept as long as their records.1 Annex 11 asks for restores from backup to be “checked during validation and monitored periodically”, and retrieval “ensured and tested” when the system changes, as in a migration.2 FDA counts as a backup only a true copy, with its metadata, kept for the whole period: for US GMP records tied to a batch, at least a year past its expiry; for GLP, as 58.195 sets.4,6,5
  5. Review the original data and its audit trail. In GMP a second person reviews the original records;6 FDA expects the audit trails of the record’s data to be reviewed with the rest of the record,4 and Annex 11 asks for audit trails to be “regularly reviewed”.2 Check the changes to the entry’s data, not every login: MHRA says audit trail review need not include “every system activity”.10 See audit trail review.
  6. Control templates like blank forms. FDA expects electronic worksheets and laboratory notebooks to be “controlled by the quality unit or by another document control method”.4
  7. Validate each GMP template for its intended use. FDA treats each CGMP workflow as an intended use to validate, with effort “commensurate with the risk”; where one system also runs research, the risk to GMP records should be “assessed and mitigated appropriately”.4 Validation steps: computer system validation.

5What inspectors find in lab notebooks.

At a suppository maker, sample weights were on an adhesive note and the notebook’s last entry predated the bench preparations:

“All raw data must be recorded in a controlled document at the time they are observed or measured to ensure data integrity.” Unipack LLC, warning letter, 19 December 202511

At an API site, reviewers checked only copies:

“Your laboratory personnel reviewed photocopies of the original data rather than the original records, and subsequently they signed off on the original analyst notebooks without verifying any changes to the underlying data.” Curia New York, Inc., warning letter, 18 September 202612

And where trial injections gave out-of-specification results that were never reported:

“Furthermore, your laboratory notebook contains no record of the sample preparations associated with the OOS results.” Ava Inc., warning letter, 14 April 202613

6ELN, LES, LIMS or CDS: where each record goes.

Table 4. Where each laboratory record belongs. The LIMS boundary: LIMS vs ELN. To compare products on your own work: how to choose an ELN.
WorkWhere to record itWhy
Exploratory researchELN, with free text allowedKnowledge kept for later development, ICH Q1014
GLP studyELN validated for GLP useDirect, prompt, attributable entries; indexed archives, 58.130(e) and 58.1905
Method validationELN, structured templatesMethod accuracy and reproducibility documented, 211.165(e)6
Routine QC testLIMS worksheet or a laboratory execution system (LES) that enforces the stepsOperational checks on the sequence of steps, 11.10(f)1
ChromatographyChromatography data system, linked from the entryEvery result of reprocessing kept, FDA Q&A 144
Paper printoutsAttached to the entry as a verified true copyTrue copies keep content and meaning, FDA Q&A 94

7What changed in 2024 to 2026.

  • NIH. Since 30 June 2024 its intramural scientists must document new and ongoing research only electronically.15
  • EU drafts. The European Commission consulted until 7 October 2025 on a revised Chapter 4 and Annex 11.16 The draft Chapter 4 says records that exist electronically “should be signed electronically” and a hybrid system “should be avoided” (4.70).17 On 26 November 2026 EudraLex still lists the 2011 texts as current.18 See the Annex 11 revision.
  • FDA. Its revised pre-approval inspection program, implemented 10 August 2026, contains both notebook audits quoted above (pre-approval inspection).8

8How it runs in Seal.

In Seal the notebook is structured. Each protocol defines its fields, units, calculations and required inputs, and each experiment keeps the protocol version, the values entered and the original results, linked to the samples, reagent lots and equipment it used. A later revision of a protocol doesn’t rewrite earlier runs. See the electronic lab notebook.

The notebook sits on the same platform as the LIMS, batch records and quality system, so a protocol from development can become the proposed controlled method with its supporting studies still linked; analytical development shows that path. The ELN comparison lists what to check when choosing one.

AQuestions and answers

How do you move from a paper lab notebook to an ELN?

List what your notebooks record and decide which of it is raw data. Build templates for the methods you repeat most, with fields, units and calculations, and validate them for their intended use. Decide how instrument files reach the record. Keep existing paper notebooks as original records for their retention period, or make verified true copies, and set a date after which new work goes only into the ELN so no record is split across both.

Is a free ELN good enough for a GxP lab?

Price matters less than controls and evidence. Check for individual accounts, an audit trail you can review, electronic signatures linked to their records, complete exports, and supplier documentation you can use in validation. EU GMP also expects a formal agreement with any third party that supplies, configures, maintains or retains the system.

Can scanned notebook pages replace the paper originals?

Yes, if each scan is a verified true copy that preserves the content and meaning of the original. Define how copies are checked before any original is destroyed. A scan or printout of electronic data that can be reprocessed, such as a chromatogram, doesn’t replace the original electronic file.

Does NIH require electronic lab notebooks?

For its own intramural researchers, yes: since 30 June 2024 new and ongoing research must be documented only electronically, in NIH-approved notebooks. Notes that exist only on paper, such as a benchtop printout, must be captured into the notebook within 72 hours. Most notebooks are kept at least 7 years after a project ends, or 30 years after a patent is filed.

Do ELN entries need a witness signature?

Not under the GMP and GLP record rules, which don’t ask for one. A US GMP laboratory record needs a second person’s review for accuracy, completeness and compliance, and a GLP entry needs the person entering it identified and the entry dated. Witness signatures come from patent practice and your company’s IP policy. If your ELN offers both, give each signature its own stated meaning.

References18 sources

References

  1. 121 CFR Part 11, Electronic Records; Electronic Signatures, sections 11.1(b), 11.10(c), (e) and (f), 11.50 and 11.70, full text searched for “notebook”, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-11 (read 26 November 2026)
  2. 2EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, in operation from 30 June 2011, Principle and sections 6, 7, 9 and 17, full text searched for “notebook”. https://health.ec.europa.eu/system/files/2016-11/annex11_01-2011_en_0.pdf (read 26 November 2026)
  3. 3Part 11, Electronic Records; Electronic Signatures: Scope and Application, Guidance for Industry, FDA, August 2003, sections I and III.B.2. https://www.fda.gov/media/75414/download (read 26 November 2026)
  4. 4Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, FDA, December 2018, questions 1.e, 2, 3, 5, 6, 7, 9, 10, 12 and 14. https://www.fda.gov/media/119267/download (read 26 November 2026)
  5. 521 CFR Part 58, Good Laboratory Practice for Nonclinical Laboratory Studies, sections 58.1, 58.130, 58.190 and 58.195, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58 (read 26 November 2026)
  6. 621 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, sections 211.160, 211.165, 211.180 and 211.194, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 26 November 2026)
  7. 7EudraLex Volume 4, Chapter 6: Quality Control, European Commission, in operation from 1 October 2014, sections 6.7, 6.10, 6.17 and 6.21. https://health.ec.europa.eu/system/files/2016-11/2014-11_vol4_chapter_6_0.pdf (read 26 November 2026)
  8. 8Compliance Program 7346.832, Preapproval Inspections, Part III, Objectives 2 and 3, FDA, issued 29 June 2026, implemented 10 August 2026. https://www.fda.gov/media/193382/download (read 26 November 2026)
  9. 9EudraLex Volume 4, Chapter 4: Documentation, European Commission, January 2011, sections 4.8, 4.9 and 4.10. https://health.ec.europa.eu/system/files/2016-11/chapter4_01-2011_en_0.pdf (read 26 November 2026)
  10. 10‘GXP’ Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018, sections 6.2 and 6.13. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 26 November 2026)
  11. 11Unipack LLC, warning letter, FDA, 19 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/unipack-llc-716621-12192025 (read 26 November 2026)
  12. 12Curia New York, Inc., warning letter, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 26 November 2026)
  13. 13Ava Inc., warning letter, FDA, 14 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/ava-inc-721180-04142026 (read 26 November 2026)
  14. 14ICH Q10, Pharmaceutical Quality System, ICH, Step 4, 4 June 2008, section 1.6.1. https://database.ich.org/sites/default/files/Q10%20Guideline.pdf (read 26 November 2026)
  15. 15Intramural Electronic Lab Notebook Policy, NIH Office of Intramural Research, page last updated 10 July 2024. https://oir.nih.gov/sourcebook/intramural-program-oversight/electronic-lab-notebooks/intramural-electronic-lab-notebook-policy (read 26 November 2026)
  16. 16Stakeholders’ Consultation on EudraLex Volume 4, Chapter 4, Annex 11 and new Annex 22, European Commission, open 7 July to 7 October 2025. https://health.ec.europa.eu/consultations/stakeholders-consultation-eudralex-volume-4-good-manufacturing-practice-guidelines-chapter-4-annex_en (read 26 November 2026)
  17. 17EudraLex Volume 4, Chapter 4: Documentation, draft for consultation, European Commission, July 2025, section 4.70. https://health.ec.europa.eu/document/download/fa336a6e-753b-46fc-b53b-9178bacd8878_en?filename=mp_vol4_chap4_consultation_guideline_en.pdf (read 26 November 2026)
  18. 18EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https://health.ec.europa.eu/medicinal-products/eudralex/eudralex-volume-4_en (read 26 November 2026)