1What CAPA means.
A deviation records that something went wrong; a CAPA records what the site will change because of it. ICH Q10 makes CAPA one of four elements of the pharmaceutical quality system.1 Four kinds of action get confused; the last column follows the invented example below.
| Term | Acts on | What the sources say | In the example |
|---|---|---|---|
| Containment | The risk, right now | FDA: “identify and/or segregate the product”;2 EU GMP: “risk-reducing actions”3 | Batch 26-017 held, blender stopped |
| Correction | The problem itself | FDA: “Repair, rework, or adjustment relating to the disposition of an existing discrepancy”2 | None: extra blending was ruled out |
| Corrective action | The cause of a problem that has happened | FDA: “Action taken to eliminate the causes of an existing discrepancy or other undesirable situation to prevent recurrence”2 | A barcode interlock on the three blenders |
| Preventive action | The cause of a problem that has not happened yet | FDA: “Action taken to eliminate the cause of a potential discrepancy or other undesirable situation to prevent such an occurrence”2 | The same fix on granulators and coaters, where no event has happened |
More in correction vs corrective action.
2The CAPA process, step by step.
No regulation fixes the number of steps; FDA’s 2006 guidance describes five parts.2 Roles vary, but the quality unit’s authority over the outcome does not.4
| Step | Who | What happens | Record it leaves | Source |
|---|---|---|---|---|
| 1. Raise | Whoever finds it; QA triages | Describe the problem, its source and what was contained. | CAPA request linked to its source | ICH Q10 3.2.21 |
| 2. Evaluate | QA and the area owner | Weigh the need for action against the consequences; add interim controls if the risk lasts. | Risk rating and rationale, including any decision not to act | FDA 2006;2 FDA 2026 draft5 |
| 3. Investigate | Investigator and experts | Find the root or most likely cause, with evidence, in proportion to risk. | Investigation report, such as a deviation report | ICH Q10: “commensurate with the level of risk”1 |
| 4. Plan | CAPA owner; QA approves | Choose actions that address the cause, each with owner and date; set effectiveness criteria now. | Approved CAPA plan | FDA 20062 |
| 5. Act | Action owners | Complete the actions, through change control where they alter a process, system or document. | Implementation evidence, change records | FDA 20062 |
| 6. Verify | Action owners; QA reviews | Confirm each action works in all affected areas and harms nothing else. | Verification evidence | FDA 2026 draft5 |
| 7. Check | CAPA owner assesses; QA approves | Measure whether the problem stopped, against the step 4 criteria. | Effectiveness check | FDA 20062 |
| 8. Close | QA | Close; feed trending, product quality review and management review. | Closed CAPA with conclusions and follow-up | 21 CFR 211.192;6 EU GMP 1.10;7 ICH Q10 3.2.41 |
When an action changes a validated system, as the example’s interlock does, change control approves the change and the CAPA keeps the reason and the effectiveness check. Test what the barrier must reject, not only what it must pass: Annex 11 asks tests to consider “system (process) parameter limits, data limits and error handling”.8 More in Annex 11 requirements.
2.1How long a CAPA takes.
No regulation sets a number of days. FDA’s draft asks for commitments that are “realistic, measurable, and achievable”,5 MHRA for “realistic and risk-based target dates”.9 Each part runs on its own clock.
| Part of the work | What sets the date | In the example |
|---|---|---|
| Interim control | The next exposure to the risk | Same day |
| Investigation and plan | The batch decision and your procedure | 15 business days |
| SOP, training, systems | Change control, testing and a production window | 17 to 36 days after the plan |
| Rollout to other equipment | Equipment access and repeat testing | 122 days after the plan |
| Effectiveness check | Exposure, not the calendar | Last run needed 199 days after the plan; closed at 210 |
3CAPA form template.
No regulation prescribes a form. Copy these fields; the example fills them in.
| Field | What to enter | Requirement it serves |
|---|---|---|
| Number, dates, source | Number, date opened, link to the deviation, complaint, audit, inspection or trend | ICH Q10 3.2.21 |
| Problem statement | What is wrong, where, how often, against what exposure | 21 CFR 211.192: thorough investigation6 |
| Scope | Other batches, products, lines and sites considered, and why any were excluded | 21 CFR 211.192;6 FDA 2026 draft: justify exclusions5 |
| Risk and depth | Consequences, rating, and how the rating set the effort | ICH Q10: effort commensurate with risk1 |
| Containment | What was held or stopped, any correction, and the interim control | EU GMP 8.14;3 FDA 2026 draft: interim CAPAs5 |
| Root cause | Root or most likely cause, with evidence; what let it continue; the justification if it is human error | EU GMP 1.4(xiv)7 |
| Actions | For each: corrective or preventive, the cause it addresses, owner, due and done dates, change control number, verification | FDA 2006: “within a defined timeframe”;2 FDA 2026 draft: “no unintended consequences”5 |
| Effectiveness criteria | Measure, population, exposure, window and pass mark, set before the action takes effect, or the risk reason verification alone is enough | ICH Q10 Table II;1 EU GMP 1.4(xiv)7 |
| Effectiveness result | Data, conclusion, who assessed, who approved | EU GMP 1.4(xiv): “monitored and assessed”7 |
| Conclusions and approval | Conclusions, follow-up, and the owner’s and quality unit’s dated approval | 21 CFR 211.192;6 21 CFR 211.22(a)4 |
4Example: a completed CAPA record.
This record is invented: the product, site, batch, people and numbers are illustrative. The same event appears as a deviation report and an effectiveness check.
| Field | Entry |
|---|---|
| Number, dates, source | CAPA-26-0087, Wenmoor plant. Opened 24 March 2026 from deviation DEV-26-0412, Exemplol 50 mg tablets, Batch 26-017. |
| Problem statement | Wrong blend recipe loaded seven times on blenders B-1 to B-3: six in the 240 runs from 1 March 2025 to 28 February 2026, then Batch 26-017 on 3 March 2026 (12 minutes’ blending against 15). All on Exemplol, the only product with look-alike recipe names: 6 in its 80 runs, or 7.5 per 100. |
| Scope | Every blender run since the recipe screen went live, all products, and 310 granulator and coater runs (same pick list, no look-alike pair): no other mismatch. |
| Risk and depth | Major: blend time can affect content uniformity. Full investigation. |
| Containment | Batch 26-017 held at 10:14 on 3 March; blender B-2 stopped. No correction: nothing reworked. Until the interlock is live, a second person confirms each recipe. |
| Root cause | Root: recipes picked by hand from a list where the two strengths differ by two digits, with nothing tying the choice to the batch record. Four operators made the same error, so human error was not accepted. Contributing: QA closed the six earlier events one at a time as operator error. |
| Corrective action 1 | Barcode interlock on B-1 to B-3: a recipe loads only if the master recipe table links the scanned batch record and blender. Automation lead, CC-26-0119, due 17 April, done 13 April. Verified by VP-26-031: 20 mismatched pairs rejected, 20 valid pairs loaded, other recipes unchanged. |
| Corrective action 2 | SOP-PRD-114 and master batch records revised; 14 operators trained. Production manager, due 17 April, done 10 April. Verified: QA watched each operator’s first scan, 14 of 14 correct. |
| Corrective actions 3 and 4 | Corrective, as the gap already let six repeats through. Monthly trending of deviations by failure mode for management review; SOP-QA-021 makes reviewers search for repeats and needs the QA manager’s written reason for retraining alone. Six closures reopened on 23 March. QA manager, due 30 April, done 29 April. |
| Preventive action | Interlock extended to granulators and coaters before a new strength creates a look-alike pair there. Automation lead, due 31 July, done 24 July. Verified by VP-26-058 on each machine. |
| Effectiveness criteria | Set 24 March. Actions 1 and 2: no wrong-recipe event or undocumented override in the first 40 Exemplol runs after go-live, where the old rate gives 3.0. Actions 3 and 4: every repeat flagged by its second event; 59 closures sampled across the window all record the search, bounding missed searches below 5% with 95% confidence. Preventive action: verification alone, by recorded risk decision, as no events means no rate to compare. |
| Effectiveness result | 40th Exemplol run on 9 October, 120 runs in all: no events, no overrides. Two other failure modes recurred, each flagged at its second event. All 59 sampled closures recorded the search. |
| Conclusions and approval | Effective. Assessed by the CAPA owner; approved and closed by QA on 20 October 2026, with the risk assessment updated. |
5Does this need a CAPA?
Decide on risk, recurrence and whether a cause remains.
| Situation | What to do | Rule |
|---|---|---|
| Minor, cause known | No possible quality impact: record and justify it. A correction can be enough. | 21 CFR 211.100(b);10 ICH Q10 on proportion1 |
| Significant deviation | Investigate to root cause and implement CAPA. | EU GMP 1.8(vii)7 |
| Minor but recurring | Open a CAPA on the trend, even if each event was harmless. | ICH Q10 lists trends;1 MHRA: recurring issues show an ineffective system9 |
| Form 483 observation | Start the CAPA plan during or straight after the inspection; update it once the scope is investigated. Extend it to other drugs, processes and facilities the cause could affect. | FDA 2026 draft5 |
| Potential failure | Data point to a failure that has not happened: preventive action. | FDA 20062 |
6CAPA at a CDMO: who does what.
At a contract development and manufacturing organization (CDMO), the quality agreement sets the split. FDA’s biologics program tells investigators to “Confirm that the contract agreement describes each party’s involvement in investigations, CAPAs, and product disposition.”11
| Task | Contract site | Product owner | Source |
|---|---|---|---|
| Raise and investigate | Records the event, tells the owner on the agreed clock, investigates its own operations | Supplies product knowledge, reviews the report, opens its own CAPA if the cause is in its process | FDA quality agreement guidance12 |
| Approve | Its quality unit approves the investigation and actions | Its quality unit approves or rejects the product | 21 CFR 211.22(a)4 |
| Track effectiveness | Holds the data, runs the check | Reviews the result and the contractor | EU GMP 7.7 and 7.813 |
Neither side can hand off the duty. In FDA’s scenarios, a contract site that blames the owner’s method is still responsible, and FDA could cite the owner too.12 See quality agreements.
7Where the rules require CAPA.
The words “corrective action”, “preventive action”, “CAPA” and “root cause” appear nowhere in Part 211.14 FDA ties corrective action to the discrepancy investigations of 21 CFR 211.22(a) and 211.192, and lists no CGMP regulation for preventive action.2
| Scope | Source and status | What it asks |
|---|---|---|
| US drugs | 21 CFR 211.192, regulation | Unexplained discrepancies and specification failures “thoroughly investigated, whether or not the batch has already been distributed”, extended to associated batches, with a written record.6 |
| EU human medicines | Directive (EU) 2017/1572, law through each member state’s transposition | Deviations and defects “documented and thoroughly investigated” (Article 10(1)); self-inspections propose “any necessary corrective measures and/or preventive actions”, and actions taken are recorded (Article 14).15 |
| EU GMP Chapter 1 | 1.8(vii) and 1.4(xiv), guideline, since 31 January 2013 | Significant deviations investigated to root cause, with CAPA; CAPA effectiveness “monitored and assessed”.7 |
| ICH regions | ICH Q10, 3.2.2, guideline, in EU GMP Part III16 | In commercial manufacturing, “CAPA should be used and the effectiveness of the actions should be evaluated”.1 |
| Clinical supplies | US: 21 CFR 210.2(c). EU: Regulation (EU) 2017/1569, law | US phase 1 production is subject to statutory CGMP but exempt from Part 211 until the drug is made available for phase 2 or 3 studies.17 EU: deviations “documented and thoroughly investigated” (Article 9(1)) and “any necessary corrective action” after self-inspection (Article 15).18 |
| Great Britain, PIC/S | Human Medicines Regulations 2012, B17(3), law; PIC/S PE 009-18 | B17(3) keeps the GMP Directive’s principles as they stood at the end of 2020;19 MHRA lists EU GMP Volume 4 as guidance,20 and PIC/S carries the same Chapter 1.21 |
| APIs | ICH Q7, 2.16 and 2.51, guideline | Critical deviations investigated (2.16); corrective action from the product quality review “completed in a timely and effective manner” (2.51).22 |
| US devices | 21 CFR Part 820, regulation, since 2 February 2026 | ISO 13485:2016 clauses 8.5.2 and 8.5.3, incorporated by reference.23, |
| Other regimes | EU ATMPs: Part IV, guideline. EU veterinary: Regulations 2025/2091 and 2025/2154, law since 16 July 2026. US HCT/Ps: 21 CFR 1271.160(b)(3) | Corrective or preventive measures after deviations;25, |
What investigators ask before approval is in pre-approval inspection; cell and gene therapy by phase, in CGT GMP requirements.
7.1What changed from 2024 to 2026.
- Device CAPA moved to ISO 13485 on 2 February 2026, so guides still citing 21 CFR 820.100 are out of date.23
- FDA’s March 2026 draft guidance on Form 483 responses, not for implementation, sets out the CAPA plan and effectiveness check FDA expects.5
- The European Commission consulted on a revised Chapter 1 from 3 September to 3 December 2025. The draft keeps the 1.4(xiv) CAPA sentences; 1.8(vii) becomes 1.8(viii). The 2013 chapter still applies.29,
16
8What inspectors look for.
FDA’s drug inspection program checks that discrepancies and failures are “investigated in a timely manner using scientific evidence to identify the root cause” and that “effectiveness of the CAPAs is evaluated”.30 In FDA’s FY2025 data, 21 CFR 211.192 drew 236 of 2,837 drug citations on system-generated Form 483s.31
Warning letters in 2025 and 2026 repeat one request about the CAPA program almost word for word,32 some asking for an “independent third-party” assessment.33,
“Provide a report that evaluates whether the program includes effective root cause analysis, ensures CAPA effectiveness, analyzes investigations trends, improves the CAPA program when needed, implements final quality unit decisions, and is fully supported by executive management.” CareFusion 213, LLC, warning letter, 30 April 202635
The Dutch inspectorate withdrew a hospital manufacturer’s GMP certificates in 2026, with this among the major deficiencies:
“The Pharmaceutical Quality System (PQS) is insufficient. This is shown by deficiencies identified in the management of deviations, CAPAs, changes, complaints, self-inspections, and the Product Quality Reviews (PQRs) that were assessed.” Antoni van Leeuwenhoek hospital, statement of non-compliance NL/H 26/2008484A, 7 October 202636
8.1Inspection self-check.
Pull these from your CAPA log before anyone asks; MHRA reads the first two as symptoms of a system out of control, the third as a sign it is ineffective.9
| Log field | Question to ask | Tied to |
|---|---|---|
| Root cause | How many say human error, and is each justified? | CareFusion;35 EU GMP 1.4(xiv)7 |
| Age and extensions | How many are past target or extended, for how long and why? | CP 7356.002: “in a timely manner”;30 MHRA9 |
| Repeat flag | Has the same failure mode returned after closure? | CareFusion35 |
| Effectiveness result | Were criteria set first, and what was the result? | CP 7356.00230 |
| Final approval | Did the quality unit decide, and could it say no? | CareFusion35 |
9How it runs in Seal.
In Seal, the CAPA blueprint keeps the source deviation, investigation, actions, controlled change and follow-up evidence in one record, and the corrective action can be a controlled change to the workflow itself rather than only a revised document and a training task. The effectiveness review keeps its baseline, population, exposure and follow-up window, and closure can be configured to require the approved review. Seal’s quality management system covers deviations, CAPA, change control, documents, training, audits, suppliers, complaints and risk.
Neil, Seal’s AI agent, can compare a CAPA’s event rates before and after the action against a defined population and exposure measure. The CAPA owner assesses effectiveness and QA approves the conclusion.
AQuestions and answers
What should a CAPA SOP cover?
No drug GMP regulation lists the contents, so build it from what FDA asks firms to show: root cause analysis, CAPA effectiveness, investigation trends, improvement of the CAPA system, final quality unit decisions and executive support. Add scope and sources, who can raise a CAPA, how risk sets the depth, how actions link to change control, and how due dates and extensions work.
Can we go into a pre-approval inspection with open CAPAs?
Yes. No text in FDA’s pre-approval program, Compliance Program 7346.832, or its biologics counterpart 7346.832M requires every CAPA to be closed first. On failures and investigations, investigators “Evaluate these events or investigations to determine if the establishment is prepared for the proposed commercial manufacturing process”, and changes from CAPAs “must be implemented promptly”. Our judgment, not a rule: for each open CAPA, show a dated plan, the interim control and the evidence so far.
How do you write a good CAPA plan?
FDA’s 2026 draft says a thorough plan addresses the root cause, fits the risks identified, and gives clear steps, timelines and deliverables. Give each action the cause it addresses, an owner, a due date and any change control number. Say how you will verify it, and set the effectiveness criteria before it takes effect.
What are the 7 steps of CAPA?
No regulation lists seven. FDA’s 2006 guidance describes five parts: evaluate, investigate, plan, act and check. Seven-step lists usually split those further, for example by adding problem identification, verification and closure. Your procedure can use any number, as long as each part leaves a record.
Can a CAPA due date be extended?
Yes, with a recorded reason, the progress so far, any interim control and QA approval. Treat each extension as a signal, though. MHRA inspectors read a large share of overdue or extended CAPAs as a sign that the system is not in control, so set realistic dates from the start.
Is CAPA used in pharmacovigilance?
Yes, under good pharmacovigilance practices rather than GMP. In the EU, GVP Module I says corrective and preventive measures should follow when monitoring or audits show the pharmacovigilance system falling short, and after an audit, corrective action, including a follow-up audit of deficiencies, must be taken where necessary. It is the same discipline of cause, action and follow-up, applied to the pharmacovigilance system instead of manufacturing.
References36 sources
References
- 1ICH Q10, Pharmaceutical Quality System, sections 3.2.2, 3.2.3, 3.2.4, Table II and glossary, ICH, 4 June 2008. https:/
/ database. ich. org/ sites/ default/ files/ Q10%20Guideline. pdf (read 26 November 2026) - 2Guidance for Industry: Quality Systems Approach to Pharmaceutical CGMP Regulations, FDA, September 2006, sections IV.C.4, IV.D.4 and IV.D.5 and glossary. https:/
/ www. fda. gov/ media/ 71023/ download (read 26 November 2026) - 3EudraLex Volume 4, Part I, Chapter 8: Complaints, Quality Defects and Product Recalls, European Commission, 13 August 2014, in operation since 1 March 2015, section 8.14. https:/
/ health. ec. europa. eu/ document/ download/ b1eb2292- cb0d- 4e3f- aea9- e3fe79faf6e3_ en?filename=2014- 08_ gmp_ chap8. pdf (read 26 November 2026) - 421 CFR 211.22, Responsibilities of quality control unit, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 22 (read 26 November 2026) - 5Responding to FDA Form 483 Observations at the Conclusion of a Drug CGMP Inspection, draft guidance for industry, FDA, March 2026, sections III.B and IV. https:/
/ www. fda. gov/ media/ 191427/ download (read 26 November 2026) - 621 CFR 211.192, Production record review, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 192 (read 26 November 2026) - 7EudraLex Volume 4, Part I, Chapter 1: Pharmaceutical Quality System, sections 1.4, 1.8 and 1.10, European Commission, in operation since 31 January 2013. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap1_ 2013- 01_ en_ 0. pdf (read 26 November 2026) - 8EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, in operation since 30 June 2011, section 4.7. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 9A fresh look at an old topic: Investigations in the GMDP environment, MHRA Inspectorate blog, 14 September 2020. https:/
/ mhrainspectorate. blog. gov. uk/ 2020/ 09/ 14/ a- fresh- look- at- an- old- topic- investigations- in- the- gmdp- environment/ (read 26 November 2026) - 1021 CFR 211.100, Written procedures; deviations, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 100 (read 26 November 2026) - 11Compliance Program 7346.832M, Prelicense and Preapproval Inspections of CDER-Regulated Biological Product Manufacturers, FDA, issued 14 April 2026, implemented 14 May 2026, section on contract agreements. https:/
/ www. fda. gov/ media/ 191983/ download (read 26 November 2026) - 12Contract Manufacturing Arrangements for Drugs: Quality Agreements, guidance for industry, FDA, November 2016, sections IV.B and V.B. https:/
/ www. fda. gov/ media/ 86193/ download (read 26 November 2026) - 13EudraLex Volume 4, Part I, Chapter 7: Outsourced Activities, European Commission, in operation since 31 January 2013, sections 7.4, 7.7 and 7.8. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap7_ 2012- 06_ en_ 0. pdf (read 26 November 2026) - 1421 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, eCFR, full text searched. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211 (read 26 November 2026) - 15Commission Directive (EU) 2017/1572 of 15 September 2017 on the principles and guidelines of good manufacturing practice for medicinal products for human use, Articles 10(1) and 14, EUR-Lex. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:32017L1572 (read 26 November 2026) - 16EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026) - 1721 CFR 210.2, Applicability of current good manufacturing practice regulations, paragraph (c), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 210. 2 (read 26 November 2026) - 18Commission Delegated Regulation (EU) 2017/1569 of 23 May 2017 on good manufacturing practice for investigational medicinal products for human use, Articles 9(1) and 15, EUR-Lex. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:32017R1569 (read 26 November 2026) - 19The Human Medicines Regulations 2012 (SI 2012/1916), regulation B17, legislation.gov.uk. https:/
/ www. legislation. gov. uk/ uksi/ 2012/ 1916/ regulation/ B17 (read 26 November 2026) - 20Medicines: good manufacturing practice and good distribution practice, MHRA, GOV.UK. https:/
/ www. gov. uk/ guidance/ good- manufacturing- practice- and- good- distribution- practice (read 26 November 2026) - 21PE 009-18, Guide to Good Manufacturing Practice for Medicinal Products, Part I, PIC/S, 24 September 2026, Chapter 1. https:/
/ picscheme. org/ docview/ 11332 (read 26 November 2026) - 22ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, ICH, 10 November 2000, sections 2.16 and 2.51. https:/
/ database. ich. org/ sites/ default/ files/ Q7%20Guideline. pdf (read 26 November 2026) - 2321 CFR Part 820, Quality Management System Regulation, sections 820.7 and 820.10, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820 (read 26 November 2026) - 24Medical Devices; Quality System Regulation Amendments, final rule, FDA, 89 FR 7496, 2 February 2024. https:/
/ www. govinfo. gov/ content/ pkg/ FR- 2024- 02- 02/ html/ 2024- 01709. htm (read 26 November 2026) - 25Guidelines on Good Manufacturing Practice specific to Advanced Therapy Medicinal Products, EudraLex Volume 4, Part IV, C(2017) 7694 final, European Commission, 22 November 2017, section 1.24. https:/
/ health. ec. europa. eu/ document/ download/ ad33d9dd- 03f0- 4bef- af53- 21308ce2187d_ en?filename=2017_ 11_ 22_ guidelines_ gmp_ for_ atmps. pdf (read 26 November 2026) - 26Commission Implementing Regulation (EU) 2025/2091 of 17 October 2025 laying down good manufacturing practice for veterinary medicinal products, applicable from 16 July 2026, Article 5(5)(i), EUR-Lex. https:/
/ eur- lex. europa. eu/ eli/ reg_ impl/ 2025/ 2091/ oj/ eng (read 26 November 2026) - 27Commission Implementing Regulation (EU) 2025/2154 of 17 October 2025 laying down good manufacturing practice for active substances used as starting materials in veterinary medicinal products, applicable from 16 July 2026, Article 5(4)(i), EUR-Lex. https:/
/ eur- lex. europa. eu/ eli/ reg_ impl/ 2025/ 2154/ oj/ eng (read 26 November 2026) - 2821 CFR 1271.160, Establishment and maintenance of a quality program, paragraph (b)(3), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 1271. 160 (read 26 November 2026) - 29Stakeholders’ Consultation on EudraLex Volume 4, Good Manufacturing Practice Guidelines: Chapter 1, European Commission, open 3 September to 3 December 2025, with the draft revised Chapter 1. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 1_ en (read 26 November 2026) - 30Compliance Program 7356.002, Drug Manufacturing Inspections, FDA, implemented 17 October 2022. https:/
/ www. fda. gov/ media/ 75167/ download (read 26 November 2026) - 31Inspection Observations, FY 2025 data set, drugs tab, inspections ending 1 October 2024 to 30 September 2025, FDA. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ inspection- references/ inspection- observations (read 26 November 2026) - 32Catalent Indiana, LLC, warning letter, MARCS-CMS 718189, FDA, 20 November 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ catalent- indiana- llc- 718189- 11202025 (read 26 November 2026) - 33Intas Pharmaceuticals Limited, warning letter, MARCS-CMS 721151, FDA, 30 March 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ intas- pharmaceuticals- limited- 721151- 03302026 (read 26 November 2026) - 34Baxter International Inc., warning letter, MARCS-CMS 733231, FDA, 29 September 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ baxter- international- inc- 733231- 09292026 (read 26 November 2026) - 35CareFusion 213, LLC, warning letter, MARCS-CMS 722729, FDA, 30 April 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ carefusion- 213- llc- 722729- 04302026 (read 26 November 2026) - 36Statement of non-compliance with GMP, report NL/H 26/2008484A, Het Nederlands Kanker Instituut-Antoni van Leeuwenhoek Ziekenhuis Stichting, Health and Youth Care Inspectorate (Netherlands), 7 October 2026, EudraGMDP reference 187962, reached from the GMP non-compliance search. https:/
/ eudragmdp. ema. europa. eu/ inspections/ gmpc/ searchGMPNonCompliance. do (read 25 November 2026)