1Sample management process, step by step.
US drug CGMP requires “scientifically sound and appropriate specifications, standards, sampling plans, and test procedures”, and samples that are representative and identified: “adequately identified” for components, “properly identified” for in-process materials and drug products. Any deviation from a written sampling plan or procedure “shall be recorded and justified”.1 With the rules on laboratory records and reserve samples, these make one chain from container to reported result, and a sample management SOP can give each of its seven steps a section.
| Step | Who | What happens | Record it leaves | Source |
|---|---|---|---|---|
| 1. Plan | QC writes, QA approves | Decide what to sample, how, and how much: enough for every test, repeat, OOS investigation and reserve | Sampling procedure or plan | 21 CFR 211.84(b) and 211.160;2, |
| 2. Take | Trained sampler | Sample, label at once, mark the containers sampled, note anything abnormal | Sampling record, labeled containers | 21 CFR 211.84(c);2 EU GMP 6.13 and Annex 8, point 1;3, |
| 3. Receive | Sample custodian | Check identity, quantity, seals, condition and storage against the request; accept, hold or reject | Receipt entry with date received | 21 CFR 211.194(a)(1)7 |
| 4. Store | Sample custodian | Keep it secure, under the material’s storage conditions and within its hold time, apart from look-alikes | Location, conditions, time | EU GMP 6.13;3 WHO 2.3;6 ICH Q2(R2) 3.48 |
| 5. Test | Analysts | Issue portions per analyst and per test; record the amount each test uses and all data | Test record | 21 CFR 211.194(a)(2) to (7);7 EU GMP 6.173 |
| 6. Review | Second person, then approver | Check the record, the calculations and the status decision | Reviewed, approved record | 21 CFR 211.194(a)(8);7 EU GMP 6.17(vii) and (viii)3 |
| 7. Retain or destroy | Custodian, on authorization | Keep reserves apart; destroy the rest once the data are checked and no investigation needs it | Retention or disposal entry | 21 CFR 211.170;9 FDA OOS guidance III.A4 |
Steps 3, 4 and 7 belong to the sample custodian. FDA’s current manual for its own ISO/IEC 17025 testing laboratories uses the sample number “to track and uniquely identify samples and their progress from the time the sample is collected in the field until the analysis is completed and the sample is disposed”, and “to provide traceability between the sample and the data”.10 Its archived ORA manual11 gives the custodian receipt of every sample, storage under the right conditions until transfer to the analyst, and disposition of every sample “whether or not the sample is analyzed”.12 It calls the whole chain sample accountability: “proof of sample receipt, storage, transfer of sample or sample portions between individuals, analysis, disposition authorization and destruction.”12 They are FDA’s rules for its own testing, not for manufacturers, but they are a fair model.
2Sample tracking log template.
No regulation prescribes the form. These fields cover what the texts ask for; use them as the columns of a sample tracking sheet or the fields of a LIMS record. Section 211.84 governs components; it is cited where it spells out a field the drug product rules leave implicit.
| Field | What to record | Requirement |
|---|---|---|
| Sample ID | A unique code on every container and every record | 211.160(b): “properly identified”;1 EU GMP 6.13: minimize mix-up;3 FDA manual 7.410 |
| Material and batch | Name, dosage form, batch or lot, and supplier where relevant | 211.194(a)(1): “lot number or other distinctive code”;7 211.84(c)(5);2 EU GMP 6.13 and 6.17(i) and (ii)3 |
| Source | The container, line, packaging run or location it came from | 211.194(a)(1): “location from where sample was obtained”;7 211.84(c)(5)2 |
| Taken by and when | Sampler’s name and the date of sampling | 211.194(a)(1) and 211.84(c)(5);7, |
| Procedure | The sampling procedure or plan used | EU GMP 6.11;3 WHO sampling record6 |
| Quantity taken | Units, mass or volume | 211.194(a)(1)7 |
| Containers marked | That each container sampled shows a sample was removed | 211.84(c)(6);2 EU GMP 6.11(vi)3 |
| Received | Date, and time where hold time matters; condition and any abnormality | 211.194(a)(1);7 WHO sampling record;6 EMA clinical laboratory paper 6.913 |
| Storage | Location, conditions and hold-time limit | EU GMP 6.11(viii) and 6.13;3 WHO 2.3;6 ICH Q2(R2) 3.48 |
| Issued and used | Who took each portion, for which test, and the weight or measure it used | 211.194(a)(3);7 FDA’s accountability model12 |
| Tests and results | Method, equipment, all data, results against specification | 211.194(a)(2), (4) and (6);7 EU GMP 6.17(iii), (iv) and (ix)3 |
| People and dates | Who tested, when, and the second-person review | 211.194(a)(7) and (8);7 EU GMP 6.17(v) to (vii)3 |
| Status | Approval, rejection or other status, signed and dated | EU GMP 6.17(viii)3 |
| What is left | Quantity remaining, where it went, and who authorized disposal | 211.194(a)(3);7 211.170 for reserves;9 FDA’s accountability model12 |
3Example: one release sample, from bottle to disposal.
This record is invented: Exemplol 50 mg tablets is not a real product, and the people, codes and numbers are illustrative. The same batch appears in the reserve sample example.
Sample S-26-0815 is the release sample for Batch 26-041, packed at Plant 2 in two runs. On 15 September 2026 a QA sampler took 5 bottles of 30 tablets (150 tablets) under sampling procedure SOP-QC-031: bottles 1 to 3 from run P-1, bottles 4 and 5 from run P-2. Each bottle carries the label EU GMP 6.13 asks for;3 21 CFR 211.84(c)(5) asks the same of component samples, plus the sampler’s name.2
| Label field | Bottle 4 | Rule |
|---|---|---|
| Sample | S-26-0815, bottle 4 of 5 | 211.160(b)(3);1 EU GMP 6.133 |
| Contents | Exemplol 50 mg tablets, 30, for release testing | EU GMP 6.13;3 WHO 2.26 |
| Batch | 26-041 | EU GMP 6.133 |
| Container sampled | Run P-2, shipper 12 | EU GMP 6.133 |
| Date of sampling | 15 Sep 2026 | EU GMP 6.133 |
| Sampled by | J. Marsh, QA | 211.84(c)(5)2 |
| Storage | 15 to 25 °C | WHO 2.26 |
| When | Event | By | Out | Left |
|---|---|---|---|---|
| 15 Sep, 09:40 | Taken and labeled; run and shipper recorded for each bottle | QA sampler | 0 | 150 |
| 15 Sep, 10:05 | Received; labels, bottle count, seals and condition checked against the request | Custodian | 0 | 150 |
| 15 Sep, 10:10 | Stored in cabinet RT-2 at 15 to 25 °C | Custodian | 0 | 150 |
| 16 Sep, 08:30 | Issued first, from unopened bottles 1 and 4, one per run, for microbial enumeration | Analyst 4 | 40 | 110 |
| 16 Sep, 08:50 | Issued for assay and related substances | Analyst 1 | 30 | 80 |
| 16 Sep, 08:55 | Issued for uniformity of dosage units | Analyst 2 | 10 | 70 |
| 16 Sep, 08:56 | Issued for dissolution, stage 1 | Analyst 2 | 6 | 64 |
| 16 Sep, 09:05 | Issued for water content | Analyst 3 | 5 | 59 |
| 17 Sep, 11:20 | Issued for dissolution, stage 2 | Analyst 2 | 6 | 53 |
| 18 Sep, 15:00 | Remaining stock counted: 51 found, 53 in the log | Custodian | 0 | 53 |
| 18 Sep, 16:30 | Lab incident LI-26-112 opened; dated entry for 2 tablets taken for micro | QC reviewer | 2 | 51 |
| 2 Oct, 14:00 | Residue destroyed after batch release and closure of LI-26-112, on QC manager’s authorization | Custodian | 51 | 0 |
The arithmetic closes: 40 + 30 + 10 + 6 + 5 + 6 + 2 + 51 = 150. Three calls are worth copying.
- Micro goes first, from bottles nobody has opened. Tablets that chemistry analysts have handled no longer represent the batch’s microbial quality, and 211.160(b) requires representative samples.1
- The count after testing is not bookkeeping. 21 CFR 211.194(a)(3) asks for the measure of sample each test used,7 so tablets that no record explains are a question about unrecorded testing. LI-26-112 found the cause: the micro method sets its portion by mass, the laboratory issued it by count, and Analyst 4 took 2 more tablets to reach the mass. The deviation is recorded and justified,1 and CAPA-26-047 has micro portions issued by mass.
- One entry per person and per test. The accountability model above covers each transfer between individuals,12 and a shared portion leaves no trail of who used what.
4Receipt and handling decisions.
| Situation | What to do | Grounded in |
|---|---|---|
| Label unreadable | Record the container as received and hold it. Assign identity only on evidence, or reject and resample | 21 CFR 211.160(b);1 EU GMP 6.133 |
| ID differs from request | Hold until the sampler clarifies; if they cannot, store it temporarily or return it | FDA ORA-LAB.5.8, 6.2, for FDA’s own laboratories14 |
| Unstable, ID unconfirmed | For a clinical sample that would degrade, analyze it but quarantine the result until identity is confirmed | EMA clinical laboratory paper 6.913 |
| Seal broken or leaking | Record the condition, assess integrity, and reject if it cannot be shown | WHO 2.2;6 ORA-LAB.5.8, 6.214 |
| Shipped out of condition | Record it and assess against stability data before testing. See temperature excursions | EU GMP 6.13;3 ORA-LAB.5.8, 6.214 |
| Waiting to be tested | Store under the material’s own conditions; test the sample and its preparations within a hold time stability data support | WHO 2.3;6 ICH Q2(R2) 3.4;8 EMA clinical laboratory paper 6.1013 |
| Too little material | Release cannot rest on partial testing. Record and justify the deviation, then resample by procedure | 21 CFR 211.165(a) and 211.160(a)15, |
| Top, middle and bottom | For a component, test each subdivision; do not composite | 21 CFR 211.84(c)(4)2 |
| Non-uniform portions | Sample and test them separately; do not pool | WHO 2.26 |
| Unexpected result | Keep the test preparations, if stable, and the rest of the sample; tell the supervisor | FDA OOS guidance III.A and IV.B.24 |
| Retest or resample | Retest the original, homogeneous sample. Resample if the investigation shows it was unrepresentative or wrongly prepared | FDA OOS guidance IV.B, citing 211.160(b)(3) and 211.165(c);4 ICH Q7 11.1516 |
| Leftover after testing | Never return it to the bulk; not a reserve unless in the marketed pack or its equivalent | WHO 2.2;6 21 CFR 211.170(b)9 |
| Trial injection | Not permitted: FDA answers no to a “trial injection” of samples | FDA Q&A on laboratory controls, question 1717 |
FDA’s OOS guidance sets the sizing rule: “The original sample from a batch should be sufficiently large to accommodate additional testing in the event an OOS result is obtained.”4 How many retests, by whom, and when a new sample is justified are covered under OOS investigations.
A hold time needs data behind it. ICH Q2(R2) expects robustness work to show “the stability of the sample preparations and reagents for the duration of the procedure, if appropriate”,8 and EMA’s paper for clinical trial laboratories says “Relevant storage stability data must be available if samples are to be stored prior to analysis.”13 The storage must not change what is measured: in August 2026 FDA cited R3 Medical under 211.160(b) because product samples were frozen and thawed before sterility testing, which “has the potential to destroy or compromise any microbial content in the samples before testing”.18
5Moving samples between sites.
A sample sent to a contract laboratory or another site needs a record the receiving custodian can check line by line.
| Field | What to record | Source |
|---|---|---|
| Contents | Sample IDs, number of containers, size of each sample | WHO 2.36 |
| Origin | Product, consignment, date and place of sampling, sampler | WHO 2.36 |
| Seal | Seal number and integrity, at dispatch and on arrival | WHO 2.2;6 ORA-LAB.5.8, 6.214 |
| Conditions | Required range, logger and its reading, time in transit | EMA clinical laboratory paper 6.9;13 ORA-LAB.5.8, 6.214 |
| Handoffs | Name, signature, date and time for each person who handles it | Eurofins warning letter, 202619 |
| Receipt | Date and time, condition, discrepancies, and the sender consulted before testing | EMA clinical laboratory paper 6.9;13 FDA manual 7.410 |
6What inspectors look for.
FDA’s drug inspection program lists, under the laboratory control system, “Specifications, standards, and representative sampling plans” and “Required testing performed on the correct samples”.20 The pre-approval program implemented on 10 August 2026 has investigators confirm “that representative samples are collected and tested/examined to verify product quality”, compare raw data with the summary data filed in the application, and treat “Multiple analyses of assay using the same sample without adequate justification” as a possible sign of a data integrity problem.21 So every submitted result should trace to a sample ID, its portions and each injection made from them; see pre-approval inspections.
Recent warning letters cite the sample record itself. FDA told a contract testing laboratory to:
“Establish a system that ensures uninterrupted custody and full reconciliation of all microbiological samples” Eurofins SF Analytical Laboratories, Inc., warning letter 320-27-05, 6 October 202619
The list that follows asks for unique labeling of each sample, signatures of all staff who handle it, and the date and time of each activity.19 At a manufacturer, the sampling record was the problem:
“For example, your microbiologist documented the collection of routine (b)(4) samples, but after questioning, your microbiologist admitted that the samples were never actually taken.” Laboratorios Jaloma S.A. de C.V., warning letter 320-26-78, 22 May 202622
“Your analyst confirmed these trial injections could not be traced to any product, sample, or analysis.” Curia New York, Inc., warning letter 320-26-127, 18 September 202623
FDA asked Curia for an investigation showing “that the trial injections were not sample injections”, and the impact on distributed API for any that were.23 A consumption log like the example’s answers half of it by accounting for every portion. The other half is the chromatography sequence: FDA’s answer on trial injections asks for “consistent and unambiguous injection nomenclature”.17 See data integrity violations for the wider pattern.
7Which rules apply to your laboratory.
| Laboratory | Main rules | What they add for samples |
|---|---|---|
| Drug QC, US | 21 CFR 211.84, 211.160, 211.165 and 211.1942, | Written sampling plans with the number of units per batch to test;15 the record fields above |
| Drug QC, EU | EU GMP Chapter 6, in operation since 1 October 20143 | Approved sampling procedures, a risk-based sampling plan, labeled containers |
| UK and PIC/S | MHRA points to EudraLex Volume 4;24 PIC/S PE 009-18, Part I25 | The same Chapter 6 sampling text, word for word |
| Starting materials | EU GMP Annex 8;5 21 CFR 211.84(d)2 | EU: an identity test on every container, unless a validated procedure allows fewer. US: at least one identity test of each component. See raw material sampling plans |
| API | ICH Q7 7.3 and 11.116 | Containers and sample size set by criticality, variability, supplier history and quantity needed; a documented procedure for any resampling or retesting after OOS |
| Contract testing | 21 CFR 211, since testing is part of manufacture;26 EU GMP for release testing; ISO/IEC 17025:2017 where accredited27 | FDA cites contract labs under Part 211, as with Eurofins above;19 MHRA inspects release-testing labs against EU GMP28 |
| GCP bioanalysis | EMA reflection paper, 201213 | Unequivocal labels, tracking from arrival to analysis, transit temperature, receipt time reported to the sponsor |
| Device QC | QMSR, 21 CFR 820, in effect since 2 February 202629, | Incorporates ISO 13485:2016; the text of Part 820 itself does not mention sampling. See QMSR |
| GLP study, US | 21 CFR Part 5831 | Specimen labels, test article accounting, reserves and archives; see below |
| Cell and gene therapy | Chain of identity and ATMP rules | See chain of identity |
Under GLP, specimens carry test system, study, nature and date of collection; each test article batch’s receipt and distribution is documented; and studies over 4 weeks keep reserve samples.31 Retention periods are compared under GLP, GMP and GCP. GLP does not govern routine QC: MHRA notes there is “no legal requirement for the quality control testing of medicinal product to be conducted in accordance with the OECD Principles of GLP”.28
8How it runs in Seal.
In Seal’s sample management, each sample’s collection, custody, aliquot lineage, reservations and consumption are recorded against it and connected to the experiment, method, result and quality assessment that depend on them. Every custody event keeps who moved the sample, when, from where to where, its quantity and condition, and the original amount must reconcile with the children, consumption and remaining material. Invalidating a result does not restore the material it used. In the LIMS, each sample is registered with its material, batch, collection point and required tests, with the method and specification versions assigned before execution.
When a receipt doesn’t match, Seal records the physical container before asserting its sample identity and keeps an unreadable or conflicting label unresolved until evidence supports the assignment, as the sample receipt discrepancy guide shows. Neil, Seal’s AI agent, can reconstruct a sample’s chain of custody as a timeline with unresolved handoffs, or check whether enough sample remains for a repeat test against its aliquots and reservations. The sample owner assesses any impact on testing, and the laboratory owner authorizes any repeat.
AQuestions and answers
Do the US drug CGMP regulations require a chain of custody record?
Part 211 does not use the term. It requires a description of each sample received, with its source, quantity, lot, the date it was taken and the date it was received, a statement of the amount each test used, and a second person’s review of the records. Together these show where a sample came from and where it went, which is what a custody record does. A formal custody form is still the clearest record for samples that leave the site.
What does a QC sample management technician do?
They receive samples into the laboratory, check identity, quantity, condition and storage against the request, store them, issue portions to analysts, keep the custody and consumption log, and retain or destroy what is left when that is authorized. FDA’s archived laboratory manual calls the role the sample custodian. Most of the job is the record: a used portion or an injection that no log explains is what an inspector will ask about.
Is a LIMS required for sample management?
No. Neither 21 CFR Part 211 nor EU GMP Chapter 6 requires software, and a controlled paper log can meet the same requirements. If the record is electronic, the rules for computerized systems apply: 21 CFR 211.68 and Part 11 in the US; in the EU, Annex 11, or Implementing Regulation (EU) 2025/2091 for veterinary medicines. A LIMS helps most where samples are split into aliquots, moved between sites or consumed by many tests, because quantities and custody have to reconcile across all of them.
How should laboratory samples be numbered?
Any scheme works if each number is unique, appears on every container and record, and is never reused. A simple pattern is a prefix, the year and a sequence, such as S-26-0815, with a suffix for each container or aliquot. FDA’s own laboratories use one sample number to track a sample from collection to disposal and to link it to its data. Keep facts that can change, such as storage location or assigned tests, in the record rather than in the number.
How long must laboratory test samples be kept?
Part 211 and EU GMP Chapter 6 set no period for test samples beyond the reserve and retention rules, which run to a year past expiry in most cases. Keep test samples until the batch decision and any investigation are closed, then destroy them as your procedure says. Nonclinical GLP specimens follow their own retention periods under 21 CFR 58.195.
References32 sources
References
- 121 CFR 211.160, Laboratory controls, General requirements, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 160 (read 26 November 2026) - 221 CFR 211.84, Testing and approval or rejection of components, drug product containers, and closures, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 84 (read 26 November 2026) - 3EudraLex Volume 4, Part I, Chapter 6: Quality Control, European Commission, in operation since 1 October 2014, sections 6.11 to 6.17. https:/
/ health. ec. europa. eu/ document/ download/ c74c8720- 27bf- 4252- 808f- d65a206a90bb_ en?filename=2014- 11_ vol4_ chapter_ 6. pdf (read 26 November 2026) - 4Investigating Out-
of- Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, Revision 1, FDA, May 2022, sections III.A and IV.B. https:/ / www. fda. gov/ media/ 158416/ download (read 26 November 2026) - 5EudraLex Volume 4, Annex 8: Sampling of Starting and Packaging Materials, European Commission, points 1 and 2. https:/
/ health. ec. europa. eu/ document/ download/ 22d03c04- 8512- 4336- 89dd- 556771fca388_ en?filename=anx08_ en. pdf (read 26 November 2026) - 6WHO guidelines for sampling of pharmaceutical products and related materials, WHO Technical Report Series No. 929, Annex 4, World Health Organization, 2005, sections 1.2, 1.6, 2.2 and 2.3. https:/
/ cdn. who. int/ media/ docs/ default- source/ medicines/ norms- and- standards/ guidelines/ quality- control/ trs929- annex4- guidelinessamplingpharmproducts. pdf (read 26 November 2026) - 721 CFR 211.194, Laboratory records, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 194 (read 26 November 2026) - 8ICH Q2(R2) Validation of Analytical Procedures, ICH, adopted 1 November 2023, error corrections 30 November 2023 and 4 April 2025, section 3.4. https:/
/ database. ich. org/ sites/ default/ files/ ICH_ Q2%28R2%29_ Guideline_ 2023_ 1130_ ErrorCorrection_ 2025. pdf (read 26 November 2026) - 921 CFR 211.170, Reserve samples, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 170 (read 26 November 2026) - 10Regulatory Testing Laboratory Manual of Quality Policies, MAN-000025, revision 07, FDA, 1 July 2025, section 7.4, Handling of Test or Calibration Items. https:/
/ www. fda. gov/ media/ 187901/ download (read 26 November 2026) - 11Field Science: Laboratory Manual, FDA, content current as of 29 July 2025. https:/
/ www. fda. gov/ science- research/ field- science- and- laboratories/ field- science- laboratory- manual (read 26 November 2026) - 12ORA Laboratory Manual Volume III, Section 2, Chain of Custody: Sample Handling, revision 02, FDA, 6 June 2019, section 1.2. https:/
/ www. fda. gov/ media/ 74006/ download (read 26 November 2026) - 13Reflection paper for laboratories that perform the analysis or evaluation of clinical trial samples, EMA/
INS/ GCP/ 532137/ 2010, EMA GCP Inspectors Working Group, 28 February 2012, sections 6.9 and 6.10. https:/ / www. ema. europa. eu/ en/ documents/ regulatory- procedural- guideline/ reflection- paper- laboratories- perform- analysis- or- evaluation- clinical- trial- samples_ en. pdf (read 26 November 2026) - 14ORA Laboratory Manual Volume II, ORA-LAB.5.8, Sample Management, revision 02, FDA, 6 June 2019, section 6.2. https:/
/ www. fda. gov/ media/ 73966/ download (read 26 November 2026) - 1521 CFR 211.165, Testing and release for distribution, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 165 (read 26 November 2026) - 16ICH Q7 Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, ICH, Step 4, 10 November 2000, sections 7.3 and 11.1. https:/
/ database. ich. org/ sites/ default/ files/ Q7%20Guideline. pdf (read 26 November 2026) - 17Questions and Answers on Current Good Manufacturing Practice Requirements: Laboratory Controls, question 17, FDA, content current as of 16 November 2022. https:/
/ www. fda. gov/ drugs/ guidances- drugs/ questions- and- answers- current- good- manufacturing- practice- requirements- laboratory- controls (read 26 November 2026) - 18R3 Medical Companies, warning letter CBER 26-726330, MARCS-CMS 726330, FDA, 14 August 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ r3- medical- companies- 726330- 08142026 (read 26 November 2026) - 19Eurofins SF Analytical Laboratories, Inc., warning letter 320-27-05, MARCS-CMS 736056, FDA, 6 October 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ eurofins- sf- analytical- laboratories- inc- 736056- 10062026 (read 26 November 2026) - 20Compliance Program 7356.002, Drug Manufacturing Inspections, FDA, implemented 17 October 2022, Laboratory Control System. https:/
/ www. fda. gov/ media/ 75167/ download (read 26 November 2026) - 21Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026, implemented 10 August 2026, Objectives 1b and 3. https:/
/ www. fda. gov/ media/ 193382/ download (read 26 November 2026) - 22Laboratorios Jaloma S.A. de C.V., warning letter 320-26-78, MARCS-CMS 725939, FDA, 22 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ laboratorios- jaloma- sa- de- cv- 725939- 05222026 (read 26 November 2026) - 23Curia New York, Inc., warning letter 320-26-127, MARCS-CMS 728790, FDA, 18 September 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ curia- new- york- inc- 728790- 09182026 (read 26 November 2026) - 24Good manufacturing practice and good distribution practice, MHRA, GOV.UK, last updated 13 May 2024. https:/
/ www. gov. uk/ guidance/ good- manufacturing- practice- and- good- distribution- practice (read 26 November 2026) - 25PIC/S Guide to Good Manufacturing Practice for Medicinal Products, Part I, PE 009-18, PIC/S, 24 September 2026, sections 6.11 to 6.13. https:/
/ picscheme. org/ docview/ 11332 (read 26 November 2026) - 2621 CFR 210.3(b)(12), Definitions, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 210. 3 (read 26 November 2026) - 27ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories, ISO, confirmed 2023. https:/
/ www. iso. org/ standard/ 66912. html (read 26 November 2026) - 28Guidance for UK Manufacturer’s Licence and Manufacturer’s Authorisation (for Investigational Medicinal Products) Holders on the use of UK Stand Alone Contract Laboratories, MHRA, December 2015. https:/
/ assets. publishing. service. gov. uk/ media/ 5a80bf1c40f0b62305b8ceaf/ Guidance_ for_ mnfs_ on_ QC_ labs_ LG_ Dec_ 15_ version_ _ doc. pdf (read 26 November 2026) - 29Quality Management System Regulation (QMSR), FDA, update of 2 February 2026. https:/
/ www. fda. gov/ medical- devices/ postmarket- requirements- devices/ quality- management- system- regulation- qmsr (read 26 November 2026) - 3021 CFR Part 820, Quality Management System Regulation, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820 (read 26 November 2026) - 3121 CFR Part 58, Good Laboratory Practice for Nonclinical Laboratory Studies, sections 58.105, 58.107, 58.130, 58.190 and 58.195, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 58 (read 26 November 2026) - 32EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026)