When is a medical device complaint MDR reportable, and when are FDA’s 30-day, 5-day and supplemental reports due?

A complaint is MDR reportable when information reasonably suggests your device may have caused or contributed to a death or serious injury, or malfunctioned in a way likely to cause or contribute to one if it recurred; FDA presumes recurrence. Report within 30 calendar days of any employee becoming aware, within 5 work days when remedial action is needed to prevent an unreasonable risk of substantial harm or FDA asks, and supplement within 30 calendar days of new information.

Illustration of a seal examining a circled change in a trace alongside run, sample and procedure records.

1Is the complaint reportable? Decide in this order.

FDA uses MDR for a medical device report under 21 CFR Part 803; in Europe the letters name Regulation (EU) 2017/745, compared at the end with Great Britain and Canada. Evaluate every device-related complaint for reportability.¹²

Table 1. Reportability for a manufacturer. Information on one patient and event from several sources may go in one report (803.22(a)).²
StepAskIf yesIf no
1Did you make or import the device?Step 2Forward the information to FDA with a cover letter (803.22(b)(2)).²
2Did a device-related adverse event occur?Step 3Not reportable; keep the documentation in the MDR files (803.22(b)(1)).²
3Did a death or serious injury occur that the device may have caused or contributed to?Step 5Step 4
4Did the device malfunction, and would that be likely to cause or contribute to a death or serious injury if it recurred?Step 5Not reportable; record why (803.18(b)(1)(i)).²
5Would a qualified person, such as a physician, nurse, risk manager or biomedical engineer, reasonably conclude that the device did not cause or contribute, or that the malfunction would not on recurrence?Not reportable; keep the information that person used (803.20(c)(2)).²Reportable: step 6
6Is remedial action needed to prevent an unreasonable risk of substantial harm, or has FDA asked for 5-day reports?5-day report (803.53).²30-day report (803.50).²

Manufacturer includes specification developers, relabellers and US agents of foreign manufacturers (803.3(l)).² The calls that decide most files:

  • Serious injury is life-threatening, causes permanent impairment or damage, or “necessitates medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure” (803.3(w)).² A temporary threat to life still counts, and no malfunction is needed.¹
  • Malfunction is a failure to meet performance specifications, which “include all claims made in the labeling”, or to perform as intended (803.3(k)).² “We presume that once the malfunction has occurred it will recur.”¹
  • Once harm has happened, the presumption sticks. After a malfunction has caused or contributed to a death or serious injury, later events of the same malfunction stay reportable until two years pass with no further death or serious injury, or valid data show the risk is remote. FDA recommends notifying it with the data at two years; stopping sooner needs an exemption under 803.19.¹²

2The deadlines: 30-day, 5-day and supplemental.

Table 2. Device reporting clocks in the United States.
ReportDueClock starts the day afterRule
30-day report30 calendar daysAny employee becomes aware803.50(a)²
5-day, remedial action5 work daysA manager or adverse-event reporter becomes aware the event needs remedial action803.53(a);² guidance 2.20¹
5-day, FDA request5 work days, for later events of the same nature in the period FDA setsAny employee becomes aware803.53(b)²
Supplemental report30 calendar daysYou receive the new information803.56²
Importers, user facilitiesImporters 30 calendar days; user facilities 10 work daysAwareness803.40, 803.30²
Correction or removal10 working days, unless already reported under Part 803You initiate the correction or removal806.10(b) and (f)³

Awareness is the day any employee, in any role, has information that “reasonably suggests a reportable adverse event has occurred” (803.3(b)).²¹ The clock does not wait for the investigation: report what is “reasonably known”, explain any gap, and supplement later (803.50(b)).² FDA expects “at least one written (including email) request for information”.¹

A work day is Monday to Friday, except federal holidays (803.3(y)).² Part 803 counts calendar days for the 30-day clock and has no rule moving a deadline off a weekend, so a 30th day that falls on a Saturday is still the deadline.²

Remedial action is “any action other than routine maintenance or servicing” needed to prevent a reportable event recurring (803.3(v)).² Fixing only the device involved does not count. If FDA would consider the action a Class I recall, the MDR is due “within 5 work days of the decision to initiate the remedial action”.¹ Once the 5-day reports for the events that prompted the action are in, later events tied to it go back to 30-day reports, and an event already reported gets a supplemental report identifying the action.¹

A supplemental report, needed when new facts would alter what was filed, carries only the new, changed or corrected information and the original report number (803.56).²¹

For eligible product codes, FDA’s Voluntary Malfunction Summary Reporting Program allows most malfunction reports as quarterly summaries, though new types of malfunction and those tied to a 5-day report or a Part 806 correction still go individually.⁴⁵

A combination product approved under an NDA, ANDA or BLA owes five-day, malfunction and correction or removal reports for its device constituent part (4.102(c)(1)); see combination products under Part 4.⁶

3How to file: eMDR, Form 3500A and codes.

Manufacturers and importers must file initial and supplemental reports electronically (803.12(a)), a requirement in force since 14 August 2015.²⁷ A report carries the content of Form FDA 3500A, “if known or reasonably known”: patient (block A), event (B), device (D), initial reporter (E), reporting details (G) and manufacturer information (H) (803.52).² Investigators also check that reports carry the device’s UDI when it is known.⁸

  • Low volume. Fill in the MedWatch 3500A form in FDA’s free eSubmitter tool and upload it through the ESG NextGen Unified Submission Portal. Test submissions are recommended, not required.⁷
  • High volume. Generate the XML from your own system to FDA’s specification and send it by AS2 or API. Test submissions are required.⁷
  • Codes. Block H needs evaluation codes (803.52(f)(6)).² FDA’s adverse event codes come in seven sets mapped to IMDRF annexes A to G: device problem, component, type of investigation, findings, conclusion, clinical signs and health impact. Choose the most specific code in each set; new codes now go through IMDRF, not FDA.⁹
  • Two checks before sending. The report number is the site’s FDA registration number, the four-digit year of submission and a five-digit sequence number (803.3(m)), kept unchanged on supplements. Block H1, type of reportable event, must match the evaluation (803.52(f)(1)).² FDA cited Pensar Medical for a number with the wrong year and a serious injury filed as a malfunction.¹⁰

4Example: one complaint through every clock.

Example, invented. A US manufacturer makes Pump A, an infusion pump whose labelling claims an alarm whenever delivery stops.

Table 3. Example: the clocks for one malfunction that turned out to have harmed a patient.
Date (2026)What happenedClock and due date
Monday 5 OctoberA field service engineer hears from a hospital nurse that a Pump A stopped infusing without alarming. Patient outcome unknown.Awareness: an employee knows. 30-day report due Wednesday 4 November.²
Thursday 8 OctoberThe complaint unit logs it.No change: awareness stays 5 October.¹
Tuesday 13 OctoberFailing to alarm breaks a labelled claim: a malfunction that could contribute to serious injury if it recurred, which is presumed.Reportable malfunction.²¹
Monday 2 November30-day malfunction report filed, saying the outcome is unknown and recording the emailed request to the hospital.On time.²¹
Monday 16 NovemberThe hospital replies: the patient’s blood pressure infusion stopped, pressure fell and the patient was resuscitated. Life-threatening, so a serious injury.Supplemental report changing the event type to serious injury, due Wednesday 16 December.²¹
10 to 19 NovemberThree more silent-stop complaints, on Tuesday 10, Thursday 12 and Thursday 19 November; no harm reported.Each a 30-day report, due Thursday 10, Saturday 12 and Saturday 19 December: no extension under Part 803, where MDCG would move an EU deadline to Monday.²¹¹
Monday 23 NovemberThe quality director concludes from the four events that a firmware correction is needed to prevent an unreasonable risk of substantial harm.5-day reports for the three unfiled events, due Tuesday 1 December, as Thanksgiving on 26 November is not a work day. A supplemental report for the first event identifies the remedial action.²¹²¹
Monday 30 NovemberThe firmware correction starts in the field.Part 806 report due Monday 14 December, unless already reported under Part 803.³

5The MDR procedure FDA tests.

Part 803 requires a written MDR procedure, and since 2 February 2026 FDA’s device inspection programme tells investigators: “Verify the firm follows their MDR procedures and they are effective in identifying MDR reportable deaths, serious injuries, and malfunctions.”²⁸ Under the QMSR, ISO 13485 clause 8.2.3, reporting to regulatory authorities, means notifying FDA of complaints that meet Part 803’s criteria (820.10(b)(3)), so quality system auditors look for the MDR procedure too.¹³ At Pensar Medical, FDA found that the procedure did not define what “reasonably suggests” a reportable event, and did not require documenting the deliberations behind each decision.¹⁰ Check yours:

Table 4. MDR procedure checklist, each line tied to Part 803.
Procedure coversWhat it should sayRule
IntakeEvery employee and channel forwards device events to the complaint unit at once; the first employee’s date is the awareness date803.17(a)(1);² guidance¹
DefinitionsReasonably suggests, become aware, serious injury, malfunction, remedial action, work day803.17(a)(1);² Pensar¹⁰
DecisionThe steps above, who decides, who the qualified person is, and how fast803.17(a)(2);² 803.20(c)(2)²
FilingWho files each report type by which clock; eMDR accounts and a back-up803.17(a)(3);² 803.12(a)²
Evaluation recordThe information evaluated and the deliberations, for every event, reported or not803.17(b)(1);² 803.18(b)(1)(i)²
AccessEvent files findable for follow-up and inspection, kept 2 years or the device’s expected life, whichever is longer803.17(b)(4);² 803.18(c)²

6Not reportable? Write the rationale so it survives inspection.

The MDR event file must hold “all documentation of your deliberations and decision making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable” (803.18(b)(1)(i)).² Copy these fields into the complaint record:

Table 5. Not-reportable rationale template.
FieldWhat to writeRequirement
Complaint and deviceComplaint number; device name, UDI and lot; date received; date any employee first knew; the event as reported, and its source803.3(b);² 820.35(a)¹³
Death or serious injuryYes, no or unknown, and why, against 803.3(w)803.50(a)(1)²
MalfunctionYes or no; if yes, why recurrence would not be likely to cause or contribute to death or serious injury803.50(a)(2);² guidance 2.14¹
Earlier harmHas this malfunction ever caused or contributed to a death or serious injury? Report numbersGuidance 2.15¹
Evidence relied onInvestigation results, records and medical opinion, each referenced; the qualified person’s name and role803.20(c)(2)²
Missing informationWhat was requested, when and how; the replies803.50(b)(3);² guidance 2.23¹
DecisionNot reportable, the reason in one sentence, date, approver803.17(a)(2)²
Reopen ifThe facts that would change the answer803.56²

7How FDA enforces it.

Investigators review a firm’s MAUDE reports before a device inspection.⁸

Absent additional information regarding the wound location, severity, or your firm’s rationale for determining that the event was not reportable, the available information reasonably suggests that the patient sustained an infection in the treated wound area that necessitated medical or surgical intervention, including removal of the retained foam, to preclude permanent impairment of a body function or permanent damage to a body structure.

Lion Street Medical, d/b/a Pensar Medical LLC, warning letter of 21 August 2026, on complaint C-029, a negative pressure wound therapy event filed as a malfunction.¹⁰

A serious injury MDR 2951238-2019-00404, submitted by your firm on February 1, 2019, documents a serious injury resulting from the same malfunction involving the same device.

Olympus Medical Systems Corporation, warning letter of 7 October 2026. Two February 2025 complaints of broken ceramic insulation on a resection set were reported three and seven months late, though the 2019 report made each reportable within 30 days.¹⁴

8FDA, EU MDR, Great Britain and Canada compared.

Table 6. Manufacturer vigilance deadlines. Sources: Part 803, Regulation (EU) 2017/745 and the GB regulations.²¹⁵¹⁶
TopicFDA, 21 CFR 803EU MDR, Article 87Great Britain
What is reportedDeaths, serious injuries, and malfunctions likely to cause either if they recurredSerious incidents, including those that “might have led or might lead” to death or serious deterioration, and field safety corrective actionsThe same
Standard deadline30 calendar daysImmediately once a causal link is established or reasonably possible, and within 15 days15 days
Faster deadlines5 work days for remedial action or an FDA requestImmediately, and within 10 days for death or unanticipated serious deterioration, or 2 days for a serious public health threat10 days; 2 days
Unsure whether reportableNot reportable only on a qualified person’s reasonable conclusionReport within the deadline anywayNo express rule; report once a causal link is established or reasonably possible
After the first reportSupplemental report within 30 days of new informationFinal report after the investigation (Article 89(5))Final report (regulation 44ZI)

The EU counts calendar days from the day after awareness: the date the first employee or representative of the manufacturer, outsourced complaint handlers included, receives the information. A deadline ending on a Saturday, Sunday or public holiday moves to the next working day. Reports go on the Manufacturer Incident Report (MIR) form; where new information makes an incident first judged non-reportable reportable, the clock runs from the day after it arrives.¹¹

Health Canada wants a preliminary report within 10 days of awareness for a Canadian incident that led to death or a serious deterioration in health, or 30 days for one that could if it recurred, then a final report.¹⁷

9What changed in 2024 to 2026.

  • 2 February 2026: the QMSR took effect, with complaint records under 820.35(a); FDA’s 2016 MDR guidance, which still cites the old 820.198, now opens with a note that it predates the QMSR.¹³¹
  • 16 June 2025: Great Britain’s post-market surveillance regulations came into force.¹⁶
  • January 2025: MDCG 2023-3 Rev.2, the current revision, covers the awareness date, calendar-day counting and the “Final (Non-reportable incident)” report type.¹¹
  • Unchanged: Article 87’s deadlines, in the EU MDR consolidated to 9 November 2026, and the clocks in 803.50, 803.53 and 803.56, unamended since 14 February 2014.¹⁵²

For how the complaint itself is handled, see complaint handling; for the CAPA a reportable trend usually opens, see what CAPA is.

10How it runs in Seal.

In Seal’s complaint handling workflow, a reportability assessment keeps the criteria applied, the facts known at the time, what is still missing and the reviewer’s rationale; when new facts arrive it is revised, and the earlier decision stays in its history. Seal keeps each receipt and awareness event with its source, so the qualified reviewer can see when the organisation could have known, not only when the complaint unit logged the report. A regulatory report carries its deadline basis, dispatch, acknowledgement, follow-up and amendments, linked to the assessment that required it.

For regulatory obligations, awareness date, statutory due date and internal target are separate fields, and date calculations retain the rule and calendar used. Neil can compare recurring complaints across lots and distribution exposure; complaint and quality owners assess significance and the reporting obligations that apply. For device makers, the medical device blueprint prepares serious incident reporting and FSCA documentation from the complaint trend, affected lots and distribution records, with notification tracking.

References

  1. 1Medical Device Reporting for Manufacturers, guidance for industry and FDA staff, sections 2.20 to 2.22, FDA, 8 November 2016, with its QMSR note. https://www.fda.gov/media/86420/download (read 8 November 2026)
  2. 221 CFR Part 803, Medical Device Reporting, sections 803.3, 803.12, 803.17, 803.18, 803.19, 803.20, 803.22, 803.30, 803.33, 803.40, 803.50, 803.52, 803.53 and 803.56, eCFR, current to 5 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-803 (read 8 November 2026)
  3. 321 CFR 806.10, Reports of corrections and removals, eCFR. https://www.ecfr.gov/current/title-21/section-806.10 (read 8 November 2026)
  4. 4Medical Devices and Device-Led Combination Products; Voluntary Malfunction Summary Reporting Program, notice, 83 FR 40973, FDA, 17 August 2018. https://www.federalregister.gov/documents/2018/08/17/2018-17770/medical-devices-and-device-led-combination-products-voluntary-malfunction-summary-reporting-program (read 8 November 2026)
  5. 5Mandatory Reporting Requirements: Manufacturers, Importers and Device User Facilities, FDA. https://www.fda.gov/medical-devices/postmarket-requirements-devices/mandatory-reporting-requirements-manufacturers-importers-and-device-user-facilities (read 8 November 2026)
  6. 621 CFR Part 4, Regulation of Combination Products, section 4.102, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-4 (read 8 November 2026)
  7. 7How to Enroll in eMDR Program, FDA, updated 13 April 2026. https://www.fda.gov/medical-devices/emdr-electronic-medical-device-reporting/how-enroll-emdr-program (read 8 November 2026)
  8. 8Compliance Program 7382.850, Inspection of Medical Device Manufacturers, Medical Device Reporting, FDA, issued and implemented 2 February 2026. https://www.fda.gov/media/80195/download (read 8 November 2026)
  9. 9MDR Adverse Event Codes, FDA, content current as of 21 August 2026. https://www.fda.gov/medical-devices/mandatory-reporting-requirements-manufacturers-importers-and-device-user-facilities/mdr-adverse-event-codes (read 8 November 2026)
  10. 10Warning letter to Lion Street Medical, d/b/a Pensar Medical LLC, MARCS-CMS 733015, FDA, 21 August 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/lion-street-medical-dba-pensar-medical-llc-733015-08212026 (read 8 November 2026)
  11. 11MDCG 2023-3 Rev.2, Questions and Answers on vigilance terms and concepts, questions 12, 14, 15 and 16, Medical Device Coordination Group, January 2025. https://health.ec.europa.eu/document/download/af1433fd-ed64-4c53-abc7-612a7f16f976_en?filename=mdcg_2023-3_en.pdf (read 8 November 2026)
  12. 125 U.S.C. 6103, Holidays, United States Code, govinfo. https://www.govinfo.gov/content/pkg/USCODE-2023-title5/html/USCODE-2023-title5-partIII-subpartE-chap61-subchapI-sec6103.htm (read 8 November 2026)
  13. 1321 CFR Part 820, Quality Management System Regulation, sections 820.7, 820.10(b) and 820.35(a), eCFR, current to 5 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 8 November 2026)
  14. 14Warning letter to Olympus Medical Systems Corporation, MARCS-CMS 729208, FDA, 7 October 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/olympus-medical-systems-corporation-729208-10072026 (read 8 November 2026)
  15. 15Regulation (EU) 2017/745 on medical devices, Articles 2(64) to 2(66), 87 and 89, consolidated version of 9 November 2026, EUR-Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:02017R0745-20261109 (read 8 November 2026)
  16. 16The Medical Devices (Post-market Surveillance Requirements) (Amendment) (Great Britain) Regulations 2024, SI 2024/1368, regulations 44ZH, 44ZI and 44ZJ, legislation.gov.uk, made 16 December 2024. https://www.legislation.gov.uk/uksi/2024/1368/made (read 8 November 2026)
  17. 17Medical Devices Regulations, SOR/98-282, sections 59 to 61, Justice Laws Website, Government of Canada, current to 19 October 2026. https://laws-lois.justice.gc.ca/eng/regulations/sor-98-282/FullText.html (read 8 November 2026)

AQuestions and answers

What counts as a serious incident under the EU MDR?

An incident that directly or indirectly led, might have led or might lead to a death, a temporary or permanent serious deterioration in someone’s health, or a serious public health threat. MDCG lists a life-threatening illness or injury, impairment of a body structure or function, hospitalisation, and medical or surgical intervention to prevent these among the deteriorations that can count. An incident with no harm is still serious unless the manufacturer can exclude that, in less favourable circumstances, it could have led to one of these outcomes.

When is a trend report needed under the EU MDR?

When there is a statistically significant increase in the frequency or severity of incidents that are not serious incidents, or of expected undesirable side-effects, that could significantly affect the benefit-risk analysis. The method for detecting the increase and the observation period belong in the post-market surveillance plan.

Is a use error MDR reportable?

It can be. FDA counts user error among the ways a device can cause or contribute to a death or serious injury, and wants those events reported, because they often reflect labelling or design problems. An event caused solely by user error, with no other performance issue and no death or serious injury, need not be reported, but the supporting information stays in the complaint file.

Does a contract manufacturer have to submit MDRs?

Not if it does not market or distribute the devices it makes for the specification developer. If both firms market the device, both have reporting duties, and they can jointly ask FDA for an exemption so that only one files.

Does each device need its own MDR when several were involved?

Yes, for a manufacturer. FDA asks for a separate report for each of your devices that may have contributed to the event when you cannot tell which one did. A user facility may send one report covering several suspect devices.