1What a certificate can and can’t replace.
| Test or material | Can the supplier’s certificate replace your test? | Basis |
|---|---|---|
| Identity | No. A specific identity test on each lot, by you, after receipt; in the EU and Canada, normally on every container | 211.84(d)(1);1 EU 5.35 and Annex 8;2, |
| Other specification tests | Yes, once you have established the reliability of the supplier’s results | 211.84(d)(2);1 EU 5.36;2 Q7 7.30 and 7.314 |
| DEG and EG limit | No. Part of identity for glycerin, propylene glycol and similar components in the US; in the EU, the Ph. Eur. limit test with identity at receipt of each batch | FDA, May 2023;6 EMA Annex 8 Q&As7 |
| Methanol limit | No. In the US, part of the identity test for ethanol and isopropyl alcohol, on each container | FDA, October 20238 |
| Containers and closures | Yes, with at least a visual identification and periodic validation | 211.84(d)(3);1 EU 5.452 |
| Q7 special materials | Yes, all tests, for processing aids, hazardous or highly toxic materials and in-company transfers; visual checks support identity; justify it | Q7 7.324 |
| Reissued by a broker | Only if it names the testing laboratory and the original maker, with the original attached | Q7 11.44;4 Health Canada5 |
| ATMP raw materials | Yes, once the risks are understood and mitigated, for example by qualifying the supplier; none needed for authorised medicinal products | EU Part IV 7.159 |
In the EU, “each batch according to Annex 8” means a sample from every container unless a validated procedure justifies fewer, and Health Canada says the same; the rules and the exceptions are in raw material sampling plans.2,
Section 211.84(d)(2) is the US permission:1
In lieu of such testing by the manufacturer, a report of analysis may be accepted from the supplier of a component, provided that at least one specific identity test is conducted on such component by the manufacturer, and provided that the manufacturer establishes the reliability of the supplier’s analyses through appropriate validation of the supplier’s test results at appropriate intervals.
FDA leaves the validation procedure to the finished product manufacturer, “consistent with the principles of CGMP and risk management”.10 The other rulebooks fill in the detail.
| Point | FDA | EU GMP | ICH Q7 | Health Canada |
|---|---|---|---|---|
| Before relying | Validate the supplier’s results; no number set | Experience of batches received and compliance history | Full analysis of at least three batches | Full testing of the first three lots |
| Audit | Not stated in 211.84 | Of the sites that test and sample, by risk | A system to evaluate suppliers | Audit report; for APIs from a regulator, EDQM or WHO, or on site |
| Certificate | Supplier’s report of analysis | Signed by a designated person with appropriate qualifications | Numerical results, signed by quality staff | Numerical results, batch, specification, methods |
| Periodic check | Validation “at appropriate intervals” | Full analysis by risk, in-house or approved contract laboratory | Full analysis at appropriate intervals | One lot a year per vendor; each material every five years |
| On a mismatch | Not set in 211.84; warning letters ask for the programme’s procedure | Investigate; stop accepting certificates until measures are complete | Not stated | Programme says how failures and requalification are handled |
The PIC/S GMP guide (PE 009-18, September 2026) carries EU 5.35 and 5.36 with the same numbers and requirements, and MHRA’s GMP guidance points to the same EudraLex text.12,
2A validation programme, step by step.
- Decide what is eligible. Exclude identity, the DEG, EG and methanol limit tests,6,
7, 8 and anything your procedure lists, such as reprocessed lots.5 Map the supply chain back to the original manufacturer; FDA recommends current knowledge of the original manufacturer and every repacker or distributor of high-risk components, and EU GMP asks for distribution controls that keep the supplier’s results applicable to what arrives.6, 2 - Qualify the supplier and the testing site. Supplier approval needs evidence, such as past quality history, that the maker can consistently meet specifications.4 For EU products, audit the sites that test and sample the material.2 EMA expects you to know the systems that generate the certificate and to review data integrity at the supplier, for example during on-site audits.14 The full process is in supplier qualification.
- Put it in writing. Health Canada’s minimum agreement covers the certificate’s content and format, with numerical results, and obliges the supplier to report process or specification changes and significant deviations.5 Where a contract manufacturer buys the material, FDA says the quality agreement should state which party audits, qualifies and monitors component suppliers.15 Health Canada holds a contract fabricator responsible for testing even the raw materials its client supplies, unless the contract excludes it and C.02.009 to C.02.010 are fulfilled.5 See quality agreements.
- Test the first lots in full and compare. At least three, under Q7 and Health Canada.4,
5 Set the agreement criteria before the results arrive, as described below. - Approve in writing. Health Canada wants a document for each vendor, approved by quality control and updated periodically, showing it meets the criteria.5 State which tests the certificate covers and which you always run.
- At each receipt. Run the identity test, and check each certificate result against your own specification, not the range the supplier printed: Part 211 tests components against “all appropriate written specifications”, and the EU certificate’s signature assures compliance with “the agreed product specification”.1,
2 Check it is the real thing: dated and signed by the maker’s quality unit, naming the original manufacturer, not a copy on a distributor’s letterhead.4, 2, 6 The certificate of analysis page lists what else to check. - Repeat the full test. At risk-based intervals, compared with the certificate.2,
4 Health Canada’s floor is one lot a year from each vendor, in rotation, and each material at least every five years.5 - Act on a mismatch. Investigate, and stop accepting that supplier’s certificates until the measures are complete.2 Test in full meanwhile, and look back at the lots already accepted on its certificates, the step FDA found missing from one firm’s response.16
- Retest after significant change. Health Canada asks for full confirmatory testing after any significant change to the manufacturing process; EU GMP asks you to consider such changes.5,
2
3CoA validation record template.
| Field | What to record | Source |
|---|---|---|
| Material | Name, code, specification and version | Q7 7.12;4 C.02.0095 |
| Supplier | Original manufacturer, manufacturing site, any distributor | Q7 7.13 and 11.434 |
| Supply chain | Route from maker to you, and transport and storage conditions | EU 5.36(i);2 C.02.0105 |
| Basis | Audit date and report, quality history, compliance record | EU 5.36(ii) and (iv);2 Q7 7.314 |
| Agreement | Certificate content with numerical results; notice of changes and significant deviations | C.02.010, interpretation 15 |
| Profiles | Residual solvent profile; impurity profile for medicinal ingredients | C.02.010, interpretation 15 |
| Tests accepted | Each specification test the certificate replaces | 211.84(d)(2);1 EU 5.352 |
| Always tested | Identity and any limit test in it; materials and lots excluded from the programme | 211.84(d)(1);1 C.02.0105 |
| Initial lots | At least three: lot, certificate result and your result for each test | Q7 7.31;4 C.02.0105 |
| Agreement criteria | Largest difference allowed per test, and its basis, set beforehand | Site’s choice; EU 6.39(v) by analogy17 |
| Approval | Quality control approver and date; document updated periodically | C.02.0105 |
| Periodic test | Frequency, last lot tested, next due | EU 5.36(v);2 Q7 7.31;4 C.02.0105 |
| Suspension | What stops acceptance; how failures and requalification are handled | EU 5.36(v);2 C.02.0105 |
4How to set the agreement criteria.
None of the four rulebooks says how close your result and the certificate’s must be: the criteria are your documented choice. The nearest rule is EU GMP’s for transferring a test method between laboratories, where acceptance criteria are “based upon the current validation study of the methodology”.17 Applied here, start from the method’s validated precision: reproducibility, “the precision between laboratories”, where an inter-laboratory study gives it, or otherwise each laboratory’s intermediate precision.18 If both laboratories are unbiased and their results are normally distributed with standard deviations s1 and s2, two results differ by more than 1.96 × √(s1² + s2²) only one time in twenty by chance. With one standard deviation s for both, that is 1.96 × √2 × s, about 2.8s. Even then, a reliable supplier fails a two-attribute comparison about one time in ten by chance (1 − 0.95² ≈ 0.10); 99% limits, 2.576 × √2 × s ≈ 3.6s, make that one in fifty but let larger real differences through. Where the supplier’s precision is unknown, use your own, and check at the audit of its testing site that it runs the methods in your dossier.2
Three judgements follow:
- Your result decides the lot. Agreement with the certificate does not rescue an in-house result outside the specification: Part 211 rejects a lot that fails it,1 and the result needs an out-of-specification investigation.
- A constant offset is a bias, not a trend. A certificate that reads 0.2 points lower for water on every lot, even within the criteria, points to a difference in method or sample handling between the two laboratories. Investigate it: EU GMP asks for critical results to be “checked to make sure that they are consistent with each other”.17
- A trend is movement over time. Results drifting towards a limit across lots, on the certificate or in your own tests, are out-of-trend data to investigate; EU GMP asks for test results to be recorded in a way that permits trend evaluation.17
5Example: validating a certificate for Excipient-Y.
Example, all invented. Site A makes oral tablets for the US and Canada. It buys Excipient-Y from Supplier S, shipped direct from S’s manufacturing site in sealed drums. The specification covers identity by IR, assay (98.0 to 102.0%) and water (no more than 1.5%). After auditing the site in February 2026 and signing an agreement, Site A tested three lots in full. Its validation gives standard deviations of 0.35 percentage points for assay and 0.10 for water, and the audit found the supplier’s methods comparable. Its procedure therefore set the criteria in advance: each in-house result within specification, assay within 1.0 point of the certificate (2.8 × 0.35 = 0.98, rounded up) and water within 0.30 (2.8 × 0.10 = 0.28, rounded up).
| Lot | Assay, CoA | Assay, ours | Water, CoA | Water, ours | Agrees? |
|---|---|---|---|---|---|
| 26-031 | 99.6% | 99.2% | 0.42% | 0.51% | Yes: 0.4 and 0.09 |
| 26-047 | 100.1% | 99.8% | 0.38% | 0.45% | Yes: 0.3 and 0.07 |
| 26-062 | 99.4% | 99.9% | 0.47% | 0.40% | Yes: 0.5 and 0.07 |
On 14 April 2026 quality control approved acceptance of the certificate’s assay and water results, with identity by IR on a sample from every drum of every lot, Health Canada’s default, and a full test of one lot every six months.5 On 6 October 2026 the periodic test of Lot 26-139 found water of 1.12% against 0.44% on the certificate. Site A’s own result met the specification, so the lot was released on it. But the difference of 0.68 exceeds 0.30, so certificate acceptance was suspended that day and every lot was tested in full. The investigation found that Supplier S had changed its drum liner in July, letting moisture in after its own test, without telling Site A as the agreement required. Site A then tested the retained samples it keeps of every raw material lot under Health Canada’s rules: the three lots accepted on the certificate since July had water of 0.96, 1.04 and 1.09%, all within specification.5 Site A’s procedure required three further matching lots before reinstating acceptance.
6How inspectors enforce it.
FDA enforces this against API makers as well as finished-product makers. From the warning letter of 11 March 2026 to Flowchem Pharma Private Limited:16
You failed to conduct an identity test on each lot of raw material used in the manufacture of your API. You instead relied on the certificates of analysis (COAs) from your supplier without adequately qualifying them.
FDA found the response inadequate because “it fails to perform a retrospective analysis of materials that you accepted solely based on COAs”.16 FDA’s letters then set out what a credible response contains, here to TSP Solutions Pte. Ltd. on 22 October 2026:11
If you intend to accept any results from your supplier’s COA instead of testing each component lot for strength, quality, and purity, specify how you will robustly establish the reliability of your supplier’s results through initial validation as well as periodic re-validation. In addition, include a commitment to always conduct at least one specific identity test for each incoming component lot.
They also ask for a “summary of results obtained from testing all components to evaluate the reliability of the COA from each component manufacturer”, with the procedure for the programme.11 A plan to reduce testing is not enough on its own: FDA found SnugZ USA’s response inadequate because it lacked “sufficient details about your planned procedural improvements, the testing to be performed, your strategy for reduced testing”.19 FDA’s and EMA’s accounts of contaminated glycerin show why: manufacturers relied on the supplier’s certificate, often a distributor’s copy that obscured the material’s origin.6,
7How it runs in Seal.
Neil, Seal’s AI agent, reads each supplier certificate at goods receipt and compares every result in code with your specification rather than the range the supplier printed, checks the lot and expiry against the ERP, and looks for results walking towards a limit across lots, as the CoA review blueprint shows. It proposes release or hold, with the question to the supplier drafted; quality releases, holds or rejects the lot, and your sampling plan stays as it is. When checking a certificate, Neil also flags required tests the certificate doesn’t report, because an omitted result is not a pass.
In supplier quality, orders, shipments, lots and certificates stay connected to sampling, inspection and internal verification against the specification, and the qualification’s scope of site, material and specification is checked at receipt, so a new site or revised specification shows up as a difference before the material is used. Enhanced inspection or additional testing for a supplier takes effect by date, with its approvals.
AQuestions and answers
How many lots must be tested in full before relying on a CoA?
ICH Q7 asks for full analysis of at least three batches before reducing in-house testing, and Health Canada asks for full confirmatory testing of the first three lots of each raw material from each vendor. Part 211 and EU GMP set no number: FDA asks you to establish the reliability of the supplier’s results, and EU GMP asks for experience of batches previously received and the supplier’s compliance history.
Is reduced testing the same as skip-lot testing?
No. Under reduced testing every lot still carries the supplier’s full results and your own identity test; you validate the certificate rather than skip lots. Skip or periodic testing of a drug substance or drug product specification, where some batches are not tested for some attributes, is described in ICH Q6A and should be justified and approved by the regulatory authority before it is used.
Can a contract laboratory do the periodic full analysis?
Yes. EU GMP lets the medicinal product manufacturer perform the full analysis itself or through a separately approved contract laboratory. The laboratory must be qualified and overseen like any other contractor, and FDA’s warning letters ask firms to describe how they qualify and oversee contract facilities that test their components.
Can a distributor’s certificate be accepted?
Only if it leads back to the original manufacturer. ICH Q7 and Health Canada say a certificate reissued by a repacker, agent or broker should name the laboratory that did the analysis, refer to the original manufacturer and attach a copy of the original certificate. Health Canada adds that a broker generally can’t be certified itself, though the original source can be if labels, packaging and certificate reach you unchanged.
What happens if periodic testing doesn’t match the CoA?
EU GMP says the discrepancy is investigated, appropriate measures are taken, and certificates from that supplier are not accepted until the measures are complete. Test every lot in full meanwhile, look back at lots already accepted on that supplier’s certificates, and let your procedure say what evidence reinstates acceptance. FDA has faulted a firm for fixing its procedure without that look back.
Does the same apply to containers and closures?
In the US, a supplier’s certificate of testing can be accepted for containers and closures with at least a visual identification by the manufacturer and periodic validation of the supplier’s results. EU GMP says a similar approach should apply to packaging materials.
References19 sources
References
- 121 CFR 211.84, Testing and approval or rejection of components, drug product containers, and closures, eCFR, up to date as of 5 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211/ subpart- E/ section- 211. 84 (read 9 November 2026) - 2EudraLex Volume 4, Part I, Chapter 5, Production, sections 5.35, 5.36 and 5.45, European Commission, in operation from 1 March 2015. https:/
/ health. ec. europa. eu/ document/ download/ 4a1fdb4f- 6f6f- 49c4- b264- 8056e5bbe078_ en?filename=chapter_ 5. pdf (read 9 November 2026) - 3EudraLex Volume 4, Annex 8, Sampling of Starting and Packaging Materials, points 2 and 3, European Commission. https:/
/ health. ec. europa. eu/ document/ download/ 22d03c04- 8512- 4336- 89dd- 556771fca388_ en?filename=anx08_ en. pdf (read 9 November 2026) - 4ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, sections 7.1, 7.3 and 11.4, ICH, 10 November 2000. https:/
/ database. ich. org/ sites/ default/ files/ Q7%20Guideline. pdf (read 9 November 2026) - 5Good manufacturing practices guide for drug products (GUI-0001), sections C.02.009, C.02.010 and C.02.025 and Appendix B, Health Canada, published July 2020, page modified 17 April 2026. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ gmp- guidelines- 0001/ document. html (read 9 November 2026) - 6Testing of Glycerin, Propylene Glycol, Maltitol Solution, Hydrogenated Starch Hydrolysate, Sorbitol Solution, and other High-Risk Drug Components for Diethylene Glycol and Ethylene Glycol, guidance for industry, Revision 1, FDA, May 2023. https:/
/ www. fda. gov/ media/ 167974/ download (read 9 November 2026) - 7Guidance on good manufacturing practice and good distribution practice: questions and answers, Annex 8: excipients at high risk of DEG/EG contamination, questions 1, 4 and 6 (April 2024 and January 2025), EMA. https:/
/ www. ema. europa. eu/ en/ human- regulatory- overview/ research- development/ compliance- research- development/ good- manufacturing- practice/ guidance- good- manufacturing- practice- good- distribution- practice- questions- answers (read 9 November 2026) - 8Policy for Testing of Alcohol (Ethanol) and Isopropyl Alcohol for Methanol, guidance for industry, FDA, October 2023. https:/
/ www. fda. gov/ media/ 173005/ download (read 9 November 2026) - 9EudraLex Volume 4, Part IV, Guidelines on Good Manufacturing Practice specific to Advanced Therapy Medicinal Products, sections 7.14 to 7.16, European Commission, adopted 22 November 2017, operational from 22 May 2018. https:/
/ health. ec. europa. eu/ system/ files/ 2017- 11/ 2017_ 11_ 22_ guidelines_ gmp_ for_ atmps_ 0. pdf (read 9 November 2026) - 10Questions and Answers on Current Good Manufacturing Practice Requirements: Control of Components and Drug Product Containers and Closures, question 15, FDA, content current as of 16 November 2022. https:/
/ www. fda. gov/ drugs/ guidances- drugs/ questions- and- answers- current- good- manufacturing- practice- requirements- control- components- and- drug (read 9 November 2026) - 11Warning letter to TSP Solutions Pte. Ltd., MARCS-CMS 735071, FDA, 22 October 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ tsp- solutions- pte- ltd- 735071- 10222026 (read 9 November 2026) - 12PIC/S Guide to Good Manufacturing Practice for Medicinal Products, PE 009-18, Part I, Chapter 5, sections 5.35 and 5.36, PIC/S, 24 September 2026. https:/
/ picscheme. org/ docview/ 11332 (read 9 November 2026) - 13Medicines: good manufacturing practice and good distribution practice, MHRA, last updated 13 May 2024. https:/
/ www. gov. uk/ guidance/ good- manufacturing- practice- and- good- distribution- practice (read 9 November 2026) - 14Guidance on good manufacturing practice and good distribution practice: questions and answers, data integrity, question 20, EMA. https:/
/ www. ema. europa. eu/ en/ human- regulatory- overview/ research- development/ compliance- research- development/ good- manufacturing- practice/ guidance- good- manufacturing- practice- good- distribution- practice- questions- answers#data- integrity (read 9 November 2026) - 15Contract Manufacturing Arrangements for Drugs: Quality Agreements, guidance for industry, section on materials management, FDA, November 2016. https:/
/ www. fda. gov/ media/ 86193/ download (read 9 November 2026) - 16Warning letter to Flowchem Pharma Private Limited, MARCS-CMS 720719, FDA, 11 March 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ flowchem- pharma- private- limited- 720719- 03112026 (read 9 November 2026) - 17EudraLex Volume 4, Part I, Chapter 6, Quality Control, sections 6.9, 6.16 and 6.39, European Commission, in operation from 1 October 2014. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ 2014- 11_ vol4_ chapter_ 6_ 0. pdf (read 9 November 2026) - 18ICH Q2(R2), Validation of Analytical Procedures, glossary, ICH, adopted 1 November 2023. https:/
/ database. ich. org/ sites/ default/ files/ ICH_ Q2%28R2%29_ Guideline_ 2023_ 1130. pdf (read 9 November 2026) - 19Warning letter to SnugZ USA, Inc., MARCS-CMS 735284, FDA, 28 September 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ snugz- usa- inc- 735284- 09282026 (read 9 November 2026)
