1The difference in one table.
| ISO 9001:2026 | ISO 13485:2016 | |
|---|---|---|
| Purpose | Consistently meeting customer, statutory and regulatory requirements; customer satisfaction1 | Medical devices that “consistently meet customer and applicable regulatory requirements”2 |
| Edition | Sixth, 16 September 2026, replacing ISO 9001:20153, | Third, 2016, confirmed on 31 October 20254 |
| Structure | ISO’s harmonized approach; Clauses 4 to 101, | ISO 9001:2008’s; Clauses 4 to 82, |
| “Risk” means | “effect of uncertainty”, positive or negative1 | “combination of the probability of occurrence of harm and the severity of that harm”, applied to device safety, performance and regulatory requirements2 |
| Vocabulary | ISO 9000, undated, so now ISO 9000:20261, | ISO 9000:2015, a dated reference2 |
| In regulation | Not in 21 CFR Part 8208 | Incorporated into Part 820; EU harmonized; GB designated8, |
ISO 13485 is not a superset of ISO 9001. It “excludes some of the requirements of ISO 9001 that are not appropriate as regulatory requirements”, so organizations that conform to it “cannot claim conformity to ISO 9001 unless their quality management system meets all the requirements of ISO 9001”.2 Their aims differ too: ISO 9001 aims to enhance customer satisfaction, with clauses on customer satisfaction (9.1.2) and continual improvement (10.1), while ISO 13485 gathers feedback (8.2.1) and its quality policy commits to maintaining the system’s effectiveness (5.3 b)).1,
2Which standard you need.
| Situation | What applies |
|---|---|
| No device involvement | ISO 9001, which suits organizations “of all sizes and across all sectors”; certification is a choice3 |
| Supplier to device firms | ISO 13485 by choice, or because a customer’s contract requires it2 |
| Claiming both | One system meeting both in full; auditing both together adds at least 25% to the initial ISO 13485 audit days2, |
| US finished devices | Part 820, which incorporates ISO 13485; FDA neither requires nor issues certificates14, |
| Devices for the EU | A quality system under MDR Article 10(9) or IVDR Article 10(8); EN ISO 13485:2016/A11:2021 harmonized under both16, |
| Canada | For class II to IV licenses, an ISO 13485 certificate from an MDSAP-recognized auditing organization18 |
| Great Britain | UK MDR 2002, with EN ISO 13485:2016+A11:2021 designated from 17 January 202611 |
| Drug-led combination | Either each CGMP rule in full, or the streamlined option: drug CGMP plus ISO 13485 4.1, 5, 6.1, 7.3 with documented risk management, 7.4, 7.5.3, 7.5.4, 8.2.2, 8.4 and 8.5, ISO 9000’s Clause 3 definitions, 820.10 and 820.35(a) and (b)19 |
3Clause matrix: ISO 9001:2026 to ISO 13485:2016.
| Topic | ISO 9001:2026 | ISO 13485:2016 |
|---|---|---|
| Context, parties | 4.1, 4.2 | None |
| Scope of the system | 4.3 | 1, and 4.2.2 quality manual |
| QMS and processes | 4.4 | 4.1 |
| Leadership, customer | 5.1.1, 5.1.2 | 5.1, 5.2 |
| Quality policy | 5.2 | 5.3 |
| Roles, authorities | 5.3 | 5.5.1 |
| Management representative | None | 5.5.2 |
| Risks | 6.1.1, 6.1.2 | 4.1.2 b), 7.1 |
| Opportunities | 6.1.3 | None |
| Objectives, planning | 6.2, 6.3 | 5.4.1, 5.4.2 |
| Resources | 7.1.1 | 6.1 |
| People, infrastructure | 7.1.2, 7.1.3 | 6.2, 6.3 |
| Work environment | 7.1.4 | 6.4.1, with 6.4.2 contamination control |
| Measuring resources | 7.1.5 | 7.6 |
| Organisational knowledge | 7.1.6 | None |
| Competence, awareness | 7.2, 7.3 | 6.2 |
| Communication | 7.4 | 5.5.3 internal; 7.2.3 with customers |
| Documented information | 7.5 | 4.2.1, 4.2.4, 4.2.5 |
| Manual, device file | None | 4.2.2, 4.2.3 |
| Operational planning | 8.1 | 7.1 |
| Product requirements | 8.2 | 7.2 |
| Design, development | 8.3 | 7.3.1 to 7.3.10 |
| External providers | 8.4 | 4.1.5, 7.4 |
| Production, service | 8.5.1 | 7.5.1 to 7.5.5; 7.5.6 process validation; 7.5.7 sterilization |
| ID and traceability | 8.5.2 | 7.5.8, 7.5.9 |
| Property, preservation | 8.5.3, 8.5.4 | 7.5.10, 7.5.11 |
| Post-delivery | 8.5.5 | 7.5.3 installation, 7.5.4 servicing |
| Control of changes | 8.5.6 | No single clause: 4.1.4, 7.3.9, 7.4.2, 7.5.6 |
| Release | 8.6 | 8.2.6 |
| Nonconforming output | 8.7 | 8.3, with 8.3.4 rework |
| Monitoring, general | 9.1.1 | 8.1, 8.2.5 |
| Satisfaction, feedback | 9.1.2 | 8.2.1 |
| Complaints, reporting | No clause of its own | 8.2.2, 8.2.3 |
| Analysis | 9.1.3 | 8.4 |
| Internal audit | 9.2 | 8.2.4 |
| Management review | 9.3 | 5.6 |
| Improvement | 10.1 | 8.5.1 |
| Corrective action | 10.2 | 8.5.2 |
| Preventive action | None | 8.5.3 |
4Moving from ISO 9001 to ISO 13485: gap checklist.
| Clause | Evidence an auditor asks for |
|---|---|
| 0.1 Regulatory role | Your role under each market’s rules identified, and those rules built in2 |
| 3 Definitions | Procedures revised for ISO 13485’s “complaint”, “product” and “risk”, each of which differs from ISO 9000:2015’s2 |
| 4.1.2 b), 7.1 Risk | Documented risk management in product realization; a risk-based approach to controlling QMS processes20 |
| 4.1.6, 7.6 Software | Documented procedures for validating software used in the quality system and for monitoring and measurement21, |
| 4.2.2 Quality manual | Scope; justified exclusions, design only where a regulation permits; how processes interact; references to required procedures, such as risk management and traceability2, |
| 4.2.3 Device file | For each device type or family, a file that contains or references its description, intended use, labeling and instructions for use; product specifications; specifications or procedures for manufacturing, packaging, storage, handling and distribution; measuring and monitoring procedures; and, as appropriate, installation requirements and servicing procedures. In the US it takes over what the device master record held23, |
| 5.5.2 Representative | A management representative, with responsibilities and authority clearly stated12, |
| 6.4.1 Environment | Documented work environment requirements, monitored and controlled where they can affect product quality22 |
| 7.3.1 to 7.3.8 Design | Planning, inputs, outputs, review, verification, validation, software validation and transfer; in the US, required for class II, III and some class I devices21, |
| 7.3.9, 7.3.10 Changes | Change procedures that judge each change’s significance; a design file holding or referencing the plan and procedures25, |
| 7.5.1 Batch record | A record for each device or batch, carrying much of what the device history record held24 |
| 7.5.6 Process validation | A procedure covering 7.5.6 a) to g); processes not fully verified afterwards, such as injection molding, validated against criteria set in advance; changes assessed for revalidation26, |
| 7.5.9 Traceability | Documented traceability procedures; 7.5.9.2’s implantable-device rules, which FDA also applies to devices that support or sustain life; Part 821 tracking where it applies14, |
| 8.2.2 Complaints | Complaint review and investigation records, with the reason when one is not investigated27, |
| 8.2.1, 8.2.3 Post-market | Feedback and post-market surveillance data used as input to risk management; procedures for reporting to regulators, such as FDA under Part 803, and for advisory notices under Part 80621, |
| 8.3.4 Rework | Documented rework procedures that weigh its possible adverse effect on the product22 |
| 8.5.2, 8.5.3 CAPA | Procedures that define how causes are found and effectiveness reviewed; a separate preventive action process22, |
FDA cites these clauses by number:
“Failure to document procedures for the validation of the application of computer software used for the monitoring and measurement of requirements, as required by ISO 13485:2016, Clause 7.6” Linemaster Switch Corporation, warning letter, 27 May 202622
4.1Example: an ISO 9001 molder supplying a syringe stopper.
Invented. Molder M, ISO 9001:2015 certified, wins a plunger stopper order from a drug maker whose prefilled syringe is a combination product. Part 820 does not cover component makers,28 but the drug maker’s streamlined route under 21 CFR 4.4(b)(1) includes ISO 13485 Clause 7.4 on purchasing,19 so its supply agreement asks for these.
| Clause | Today | Gap | Action | Evidence |
|---|---|---|---|---|
| 6.4.2 | General housekeeping | No defined particle controls | Enclosed molding cell with set limits | Monitoring records |
| 7.4.2 | Changes under ISO 9001 8.5.6 | Customer not told of resin or tool changes | Change notice clause in the quality agreement | Signed agreement, change log |
| 7.5.6 | Lots inspected; settings from first-article approval | Molding not validated | Validate with criteria set in advance | Protocol, report, revalidation triggers |
| 7.5.8, 7.5.9 | Lot number on each carton | No link to resin lot or press | Record resin lot, press and tool per lot | Lot genealogy |
5How it runs in Seal.
In Seal, the records of an ISO 9001 system, such as procedures, training, audits and corrective actions, carry on into ISO 13485. The design and development file (7.3.10) is the chain from requirements to verification, the medical device file (4.2.3) is the released configuration, and the record for each device or batch (7.5.1) links each unit to the versions and lots it was made from. In design control, a requirement with no linked test shows as a gap while the design is developed.
When a supplier changes a material or method, as in Moulder M’s example, the impact assessment starts from the specifications it reaches, including the manufacturing validations that may need review, and keeps the source notice and decisions. Supplier change notices follows a change to the parts and customers it reaches and the notice each customer is owed, and each received lot links to the units and distributions that used it. Complaints carry a reportability assessment, and CAPA has a preventive action template beside corrective action.
6ISO 9001:2026: what changed and transition dates.
ISO 13485:2016 is unchanged; FDA will evaluate any revision and decide whether to amend the QMSR.4,
- 5.1.1 i), 7.1.4, 7.3 e). Quality culture and ethical behavior, in leadership, the environment for operating processes and awareness.1,
29 - 6.1.1 to 6.1.3. Risks and opportunities are determined, then acted on separately.
- 9.3.2 c). Management review inputs “shall include” changes in relevant interested parties’ needs and expectations.
- 31 March 2028. No initial ISO 9001:2015 certifications are made after this date.29
- 30 September 2029. Accredited ISO 9001:2015 certifications stop being valid. The transition audit can be part of a surveillance or recertification audit, or a special audit, and may be remote.29
AQuestions and answers
Can one quality system cover ISO 9001, ISO 14001 and ISO 13485?
Yes. ISO 9001 and ISO 14001 share ISO’s common structure for management system standards, which makes them easy to combine. ISO 13485 keeps its older structure and has no environmental requirements of its own, but it says an organization can align or integrate its quality system with other management systems.
Does ISO 9001:2026 change ISO 13485?
No. ISO 9001:2026 revises the general standard only. ISO 13485:2016 was reviewed and confirmed in October 2025, still points to ISO 9000:2015 for its vocabulary, and is the edition FDA’s device regulation incorporates.
Do pharmaceutical companies need ISO 9001 or ISO 13485?
Medicines are made under GMP regulations, and neither certificate replaces them. ICH Q10, the model pharmaceutical quality system, says it is based on ISO quality concepts and sits alongside regional GMP rules. ISO 13485 comes in when the product includes a device part: FDA’s combination product rule names the ISO 13485 clauses a drug manufacturer meets on its streamlined route.
Is ISO 13485 equivalent to ISO 9001?
No. Neither is a stricter version of the other. ISO 13485 adds device requirements, such as a medical device file and reporting to regulators, and has no clause on opportunities or organisational knowledge. ISO puts the difference down to user group and application: ISO 9001 suits any organization, ISO 13485 the medical device sector.
References29 sources
References
- 1ISO 9001:2026, Foreword, Introduction and Clauses 1 to 3, public preview, ISO Online Browsing Platform. https:/
/ www. iso. org/ obp/ ui/ #iso:std:iso:9001:ed- 6:v1:en (read 25 November 2026) - 2ISO 13485:2016, Introduction (0.1 to 0.5) and Clauses 1 to 3, public preview, ISO Online Browsing Platform. https:/
/ www. iso. org/ obp/ ui/ #iso:std:iso:13485:ed- 3:v1:en (read 25 November 2026) - 3ISO 9001:2026, Quality management systems, Requirements, ISO, standard page and FAQ, publication date 2026-09. https:/
/ www. iso. org/ standard/ 9001 (read 25 November 2026) - 4ISO 13485:2016, Medical devices, Quality management systems, Requirements for regulatory purposes, ISO, standard page and life cycle (90.93 confirmed, 31 October 2025). https:/
/ www. iso. org/ standard/ 59752. html (read 25 November 2026) - 5ISO 9001:2026, table of contents expanded to every subclause, ISO Online Browsing Platform. https:/
/ www. iso. org/ obp/ ui/ #iso:std:iso:9001:ed- 6:v1:en (read 25 November 2026) - 6ISO 13485:2016, table of contents expanded to every subclause, ISO Online Browsing Platform. https:/
/ www. iso. org/ obp/ ui/ #iso:std:iso:13485:ed- 3:v1:en (read 25 November 2026) - 7ISO 9000:2026, Quality management, Fundamentals and vocabulary, ISO, publication date 2026-05. https:/
/ www. iso. org/ standard/ 9000 (read 25 November 2026) - 821 CFR 820.7, Incorporation by reference, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- A/ section- 820. 7 (read 25 November 2026) - 9Commission Implementing Decision (EU) 2022/6 amending Implementing Decision (EU) 2021/1182 on harmonised standards for medical devices, under Regulation (EU) 2017/745, EUR-Lex, 4 January 2022. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:32022D0006 (read 25 November 2026) - 10Commission Implementing Decision (EU) 2022/15 amending Implementing Decision (EU) 2021/1195 on harmonised standards for in vitro diagnostic medical devices, under Regulation (EU) 2017/746, EUR-Lex, 6 January 2022. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:32022D0015 (read 25 November 2026) - 11Designated standards: medical devices, consolidated list version 2 and notice of publication 0129/26, Department of Health and Social Care, GOV.UK, 17 January 2026. https:/
/ www. gov. uk/ government/ publications/ designated- standards- medical- devices (read 25 November 2026) - 12Common Findings: ISO 13485:2016 Stage 1 Audit, MTF-2031-B Rev 1.0, National Standards Authority of Ireland (NSAI), findings for clauses 4.2.2 and 5.5.2. https:/
/ www. nsai. ie/ images/ uploads/ general/ MTF- 2031- B. pdf (read 25 November 2026) - 13Global ACI-TECH-3-007, Application of ISO/IEC 17021-1 in the Field of Medical Device Quality Management Systems (ISO 13485), version 1.0, Global ACI, issued 10 July 2026, implemented 28 August 2026, Annex D, Conducting ISO 9001 and ISO 13485 Together. https:/
/ sys. global- aci. org/ uploads/ documents/ Global_ ACI_ TECH- 3- 007_ %28M%29_ v1_ 0. pdf (read 25 November 2026) - 1421 CFR 820.10, Requirements for a quality management system, paragraphs (a), (b)(2) to (4), (c) and (d), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- A/ section- 820. 10 (read 25 November 2026) - 15Quality Management System Regulation, Frequently Asked Questions, FDA, questions 5 and 13, content current as of 5 November 2026. https:/
/ www. fda. gov/ medical- devices/ quality- management- system- regulation- qmsr/ quality- management- system- regulation- frequently- asked- questions (read 25 November 2026) - 16Regulation (EU) 2017/745 on medical devices, Article 10(9), consolidated text of 9 November 2026, EUR-Lex. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:02017R0745- 20261109 (read 25 November 2026) - 17Regulation (EU) 2017/746 on in vitro diagnostic medical devices, Article 10(8), consolidated text of 10 January 2025, EUR-Lex. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ HTML/ ?uri=CELEX:02017R0746- 20250110 (read 25 November 2026) - 18Guidance Document: Requirements in the Recognition Process for Medical Device Single Audit Program (MDSAP) Auditing Organizations, Health Canada, sections 1.2 to 1.4. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ medical- devices/ quality- systems- 13485/ requirements- recognition- process- mdsap- auditing- organizations- profile/ guidance- document. html (read 25 November 2026) - 1921 CFR 4.4, How can I comply with these current good manufacturing practice requirements for a co-packaged or single-entity combination product?, paragraphs (a)(1) and (b)(1), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 4/ subpart- A/ section- 4. 4 (read 25 November 2026) - 20FDA’s Quality Management System Regulation (QMSR): Medical Device Risk-Based Inspections, CDRH town hall transcript, FDA, 1 April 2026. https:/
/ www. fda. gov/ media/ 191960/ download (read 25 November 2026) - 21Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Part V section 1.A, Situation 1 (g), and Attachment A. https:/
/ www. fda. gov/ media/ 80195/ download (read 25 November 2026) - 22Linemaster Switch Corporation, warning letter, MARCS-CMS 730215, FDA, 27 May 2026, violations 1, 3, 4 and 5. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ linemaster- switch- corporation- 730215- 05272026 (read 25 November 2026) - 23MDSAP AU P0002.011, MDSAP Audit Approach, Medical Device Single Audit Program, revision date 3 August 2026, Chapter 6, Tasks 7 and 8, and Annex 1. https:/
/ www. mdsap. global/ sites/ default/ files/ 2026- 07/ MDSAP%20AU%20P0002. 011%20MDSAP%20Audit%20Approach. pdf (read 25 November 2026) - 24Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, FDA, 2 February 2024, responses to comments 31 and 54. https:/
/ www. federalregister. gov/ documents/ 2024/ 02/ 02/ 2024- 01709/ medical- devices- quality- system- regulation- amendments (read 25 November 2026) - 25Koven Technologies, Inc., warning letter, MARCS-CMS 734643, FDA, 21 July 2026, violations 1 and 3. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ koven- technologies- inc- 734643- 07212026 (read 25 November 2026) - 26Nipro Renal Solutions USA, Corporation, warning letter, MARCS-CMS 732874, FDA, dated 16 July 2026, issued 24 July 2026, violation 7. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ nipro- renal- solutions- usa- corporation- 732874- 07242026 (read 25 November 2026) - 2721 CFR 820.35, Control of records, paragraph (a), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- B/ section- 820. 35 (read 25 November 2026) - 2821 CFR 820.1, Scope, paragraph (a)(2), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- A/ section- 820. 1 (read 25 November 2026) - 29Global ACI-TECH-3-TR 2029-09-30 (M), Transition Requirements for ISO 9001:2026, version 1.1, Global Accreditation Cooperation Incorporated, issued 4 September 2026, sections 3 to 5. https:/
/ sys. global- aci. org/ uploads/ documents/ TECH- 3- TR_ 2029- 09- 30_ %28M%29_ ISO_ 9001_ 2026_ Transition_ v1_ 12. pdf (read 25 November 2026)