What are the dissolution acceptance criteria at S1, S2 and S3, and for extended- and delayed-release products?

At S1, each of 6 units must dissolve at least Q + 5%. Otherwise test 6 more (S2): the mean of 12 must reach Q, no unit below Q − 15%. Otherwise test 12 more (S3): the mean of 24 must reach Q, at most 2 units below Q − 15%, none below Q − 25%. Q is a percentage of label claim. Extended release uses L1 to L3; delayed release, A1 to A3 and B1 to B3.Checked against its sources on 25 November 2026. How these answers are made

1The S1, S2 and S3 acceptance table, and what Q means.

An immediate-release dissolution test runs in up to three stages, and each stage adds units to those already tested. JP 6.10 is harmonized with Ph. Eur. 2.9.3 and USP <711>, and ICH accepts the three texts as interchangeable, with the exceptions given below, so the table, from the Japanese Pharmacopoeia, is the one all three apply.1,2

Table 1. Immediate-release acceptance table, from JP 6.10 Acceptance Table 6.10-⁠1.1
StageUnits testedAcceptance criteria
S16Each unit not less than Q + 5%
S26 more, 12 in allMean of the 12 is Q or more, and no unit is below Q − 15%
S312 more, 24 in allMean of the 24 is Q or more, no more than 2 units are below Q − 15%, and no unit is below Q − 25%

“Continue testing through the three stages unless the results conform at either S1 or S2.”1 Q is the amount of active ingredient dissolved that the monograph or approved specification requires at the stated time, “expressed as a percentage of the labeled content of the dosage unit”. The 5%, 15% and 25% are also percentages of label claim, “so that three values and Q are in the same terms”: they are added to or taken from Q, not calculated as a fraction of it.1

Table 2. The limits for common values of Q, computed from Table 6.10-⁠1.
QS1: every unit at leastS2 and S3: mean at leastS2: no unit below; S3: at most 2 units belowS3: no unit below
70%75%70%55%45%
75%80%75%60%50%
80%85%80%65%55%
85%90%85%70%60%

So a specification of Q = 80% in 30 minutes asks for 85% from each of the first six units. FDA’s 1997 guidance writes it both ways: “a single-point dissolution test specification of NLT 85% (Q=80%) in 60 minutes or less is sufficient as a routine quality control test for batch-to-batch uniformity”.3 Q comes from the monograph or the approved specification; ICH says acceptance criteria “should be specified in the application dossier”.2,1 Where a Japanese monograph gives no Q, JP’s own Interpretation 2 applies instead (six units each meeting the monograph’s limit; if one or two of them fail, six more are tested and at least 10 of the 12 must meet it), and ICH does not treat it as interchangeable.1,2

2Extended-release: levels L1, L2 and L3.

An extended-release specification gives a range at each earlier time point and a minimum at the last, all as percentages of label claim. The levels work like the stages, with margins of 10 and 20 percentage points of label claim:1

Table 3. Extended-release acceptance table, from JP 6.10 Acceptance Table 6.10-⁠2.1
LevelUnits testedAcceptance criteria
L16No unit outside any stated range, and none below the stated amount at the final time
L26 more, 12 in allMean of 12 within every range and not below the final amount; no unit more than 10% of label claim outside any range or below the final amount
L312 more, 24 in allMean of 24 within every range and not below the final amount; at most 2 of the 24 more than 10% outside each range, and at most 2 more than 10% below the final amount; no unit more than 20% outside any range or below the final amount

“Where more than one range is specified, the acceptance criteria apply individually to each range.”1 Count the units outside each time point’s range separately.

The ranges are set at registration. ICH Q6A normally allows a total spread of 20% at any time point, so “a requirement of 50 +/- 10% thus means an acceptable range from 40% to 60%”, unless bioequivalence data support more.4,5 EMA asks, in general, for at least three points: early (typically 20 to 30% dissolved) to exclude dose dumping, around 50%, and Q = 80% at the end.6 FDA recommends ±10% around the mean profile of the clinical or bioavailability lots, set on “average dissolution data for each lot under study, equivalent to USP Stage 2 testing”, and accepts “reasonable deviations from ± 10%” if “the range at any time point does not exceed 25%”.7

3Delayed-release: acid stage A1 to A3, buffer stage B1 to B3.

A delayed-release (enteric-coated or gastro-resistant) product has an acid stage, where little may dissolve, and a buffer stage, judged like an immediate-release test against Q.1

Table 4. Delayed-release acceptance tables, from JP 6.10 Acceptance Tables 6.10-⁠3 and 6.10-⁠4.1
LevelUnits testedAcid stageBuffer stage
A1 and B16No unit above 10% dissolvedNo unit below Q + 5%
A2 and B26 more, 12 in allMean of 12 not more than 10%, and no unit above 25%Mean of 12 is Q or more, and no unit below Q − 15%
A3 and B312 more, 24 in allMean of 24 not more than 10%, and no unit above 25%Mean of 24 is Q or more, at most 2 units below Q − 15%, and no unit below Q − 25%

“Continue testing through the three levels unless the results of both stages conform at an earlier level.”1 Here the pharmacopeias part. JP marks its acid-stage procedure and interpretation as its own, generally 2 hours in its 1st fluid for tablets and capsules.1 ICH does not treat the test as interchangeable for products “referred to in the regional compendia as delayed-release, gastro-resistant, or enteric-coated”, though the EU says it could accept USP’s approach and Health Canada treats USP and Ph. Eur. as interchangeable.2 EMA wants “less than 10% dissolved after 2 hours” in acid and, for continued testing, refers to Ph. Eur.6 Name in the specification the pharmacopeia whose tables you apply.

4Dissolution results record and stage decision checklist.

Keep every unit. Laboratory records hold “complete data derived from all tests necessary to assure compliance with established specifications and standards”, and FDA asked one firm for “All individual and average results from each stage of dissolution testing (i.e., Stage 1, Stage 2, Stage 3).”8,9

Table 5. Dissolution results record. For delayed release, add the acid-stage rows.
FieldWhat to recordSource
SampleProduct, strength, batch, purpose (release, or stability condition and time point), date sampled and received211.194(a)(1)8
SpecificationVersion; Q and time, or the ranges and final amount for each time point, or the acid and buffer limits; the pharmacopeia whose tables applyJP 6.10, 4;1 Q4B 2.22
MethodMethod and version; apparatus, rotation speed, sinker, medium and volume211.194(a)(2); JP 6.10, 1 and 3.1
Apparatus suitabilityDissolution unit, last performance check or mechanical calibration, next due dateJP 6.10, 2; Q4B 2.1.3
Run conditionsMedium volume within ±1%, buffer pH within 0.05 unit, deaeration, 37 ± 0.5 °C, sampling times within ±2%, filterJP 6.10, 3.1.1
Unit resultsUnit number, stage, vessel and percentage of label claim at each time point, for every unit testedJP 6.10, 4; FDA OOS IV.C10
S1 or L1 checkLowest unit against Q + 5%, or every unit against each range and the final amountTables 6.10-⁠1 and 6.10-⁠2
S2 or L2 checkMean of 12 against Q or the ranges; lowest unit against Q − 15%, or units more than 10% outsideTables 6.10-⁠1 and 6.10-⁠2
S3 or L3 checkMean of 24; number of units below Q − 15% (2 or fewer); lowest unit against Q − 25%; or the L3 countsTables 6.10-⁠1 and 6.10-⁠2
OutcomeConforms at S1, S2 or S3, or fails at S3JP 6.10;1 EDQM11
Trend flagDid a product that usually passes at the first stage need the next? Investigation referenceMHRA12
FailureOOS investigation reference; no outlier test; any retest only under the approved procedureEDQM; FDA OOS IV.B and IV.C
SignaturesAnalyst and date; second-person review of the original records211.194(a)(7) and (8)

5How it runs in Seal.

In Seal’s pharmaceutical QC, the dissolution specification is approved data, not a PDF beside the result: tests, acceptance criteria, calculations, units, rounding and effective versions. When a release or stability sample is created, the applicable version resolves from its material, stage, market, method and effective date, and a result outside limits is flagged for assessment rather than passing silently through review. In the LIMS, you configure the S2 and S3 means and unit counts as calculations in the method, with explicit inputs, units and reporting rules, and verify boundary values, missing inputs and rounding whenever the calculation changes; each result keeps the specification version it was evaluated against.

If S3 fails, the OOS investigation opens from the result with its sample, raw data, method, instrument and batch linked. The first reportable value stays permanent, and each retest runs under a plan approved before its result is known, so exploratory testing stays visible. Out-of-trend results use the same connected history, so a reviewer can investigate drift while results are still within specification.

6Example: Exampladine 20 mg tablets reach S3.

Invented for illustration. Exampladine 20 mg immediate-release tablets are tested for release against Q = 80% at 30 minutes.

Table 6. Example. Product and results invented.
StageUnits, % of label claim at 30 minutesCheckResult
S188, 91, 84, 89, 86, 90Every unit 85% or moreOne unit at 84%: go to S2
S287, 63, 89, 85, 90, 88Mean of 12 at least 80%; no unit below 65%Mean 85.83%, but one unit at 63%: go to S3
S386, 89, 85, 91, 87, 88, 84, 90, 86, 89, 92, 85Mean of 24 at least 80%; at most 2 units below 65%; none below 55%Mean 86.75%, one unit below 65%, lowest 63%: conforms at S3

The mean was never in doubt; single units decided each stage. Had the 63% unit read 78%, the batch would have conformed at S2 (mean 87.08%). Had it read 54%, it would fail at S3 despite a mean of 86.38%, because no unit may fall below Q − 25%. Exampladine normally conforms at S1, so the record also carries a trend flag and an investigation.12 The first question is the 63% vessel: EMA names coning and tablet sticking as causes of variable results.13

6.1Extended release: Exampladine 200 mg reaches L3.

Exampladine 200 mg extended-release tablets (invented) must release 20 to 40% at 1 hour, 45 to 70% at 4 hours and not less than 80% at 12 hours. The 25-point range at 4 hours is wider than Q6A’s usual 20, so the dossier supports it with bioequivalence data. At L2 no unit may lie outside 10 to 50%, 35 to 80% or below 70%; at L3, outside 0 to 60%, 25 to 90% or below 60%.

Table 7. Example. Product and results invented.
LevelWhat the units showedResult
L1Unit 3 released 43% at 1 hour, above the 20 to 40% rangeGo to L2
L2Means of 12 were 32.58%, 57.33% and 92.33%, all within. Unit 8 released 33% at 4 hours, 12 points below 45%More than 10% outside: go to L3
L3Means of 24 were 32.25%, 58.17% and 92.42%. One unit of 24 was more than 10% outside the 4-hour range; none more than 20% outside, none below 70% at 12 hoursConforms at L3

7When a stage fails, and when the result is OOS.

Moving to the next stage is part of the test. EDQM’s worked example reaches S3 before it says: “If the results still do not comply, the medicinal product has failed the test and an out-of-specification investigation must be launched.”11 The other tables end the same way, at L3, A3 and B3.1 FDA’s OOS definition also covers criteria set “by the manufacturer”, so a tighter internal limit creates its own OOS.10

Needing a later stage is still a signal. MHRA: “if the sample test criteria is usually the first level of testing and a sample has to be tested to the next level this should be investigated as it is not following the normal trend.”12 Treat it as an out-of-trend result.

Do not start again. The stages accumulate units; a fresh six at S1 after an S2 failure discards results. FDA has seen firms “use a strategy of repeated testing until a passing result is obtained, then disregarding the OOS results without scientific justification.” Retests follow a written procedure that fixes their number in advance, and outlier tests “have no applicability in cases where the variability in the product is what is being assessed, such as for content uniformity, dissolution, or release rate determinations.”10 See OOS investigation for the phases.

Frequent S3 means a marginal specification or process. EMA: “The specification should be set in such a way so that during routine manufacture and testing it would be expected that compliance with S2 is attained.”13 And it holds to the end of shelf life: “Once a dissolution specification is set, the drug product should comply with that specification throughout its shelf life.”3 Profile comparisons after a change are judged by the f2 similarity factor, not these tables.

8What inspectors cite.

A November 2026 letter faults a firm for missing the signals before the failure:

“Despite these quality signals, you did not initiate an investigation until you obtained out-of-specification (OOS) dissolution results during release testing in March 2025”

Alkem Laboratories Limited, warning letter 320-⁠27-⁠25, 18 November 2026.14 A 2021 study had shown slower dissolution with excipients from new suppliers; the firm later took market action, including recalls, on batches made from July 2024 to August 2025.

In 2020 FDA cited a firm whose temozolomide capsules failed on stability with a Stage 3 average below even a relaxed, approved specification:

“you did not determine a root cause for dissolution variability, nor implement adequate corrective action and preventive action (CAPA) to prevent future dissolution failures using the new specification.”

Stason Pharmaceuticals, Inc., warning letter, 8 July 2020.9

An investigator recorded at Ipca Laboratories in June 2023 that, with no assignable cause found: “Based on this inconclusive hypothesis, you retested the samples for dissolution and based on the passing results, you invalidated the initial OOS results.”15 A Form 483 lists observations, not FDA’s final determination.

9Pharmacopeias, regions and what is changing.

  • Harmonization. ICH Q4B Annex 7(R2), at Step 4 since 11 November 2010, makes JP 6.10, Ph. Eur. 2.9.3 and USP <711> interchangeable for the basket, paddle and flow-through cell, except with enzymes in the medium, for delayed release, for JP Interpretation 2 and for vessels over 1 liter.2
  • Editions. The 19th edition of the Japanese Pharmacopoeia took effect on 10 April 2026, but the English translations published by MHLW and PMDA still end at the 18th, which this page cites.16,17 USP <711> and Ph. Eur. 2.9.3 are subscription texts; EDQM’s public FAQ walks through the same S1 to S3 stages for Q = 75%.11
  • United States. For immediate-release products that meet all the conditions in its section III, among them a high-solubility drug substance, no narrow therapeutic index and no rapid-onset claim, FDA’s 2018 guidance recommends 500 mL of 0.1 N HCl, basket at 100 rpm or paddle at 50 rpm, and “the dissolution criterion is Q=80% in 30 minutes.”18 Other products follow the 1997 guidance.3
  • European Union. For generics, EMA recommends Q at the biobatch mean minus 10%, usually 75 to 85% in steps of 5%, at 15, 30 or 45 minutes, and says “A limit greater than 85% is not relevant.”13
  • ICH Q6(R1). ICH is revising its specifications guidelines, including Q6A, which has stood since 1999. Its July 2026 work plan expected Step 2 sign-off in November 2026,19 but the ICH Assembly of 17 to 18 November 2026 endorsed no Q6(R1) draft, so no consultation has opened and Q6A still applies.20,21

AQuestions and answers

What does Q mean in a dissolution test?

Q is the amount of active ingredient that must dissolve by the stated time, as a percentage of label claim, set by the monograph or the approved specification. It is a limit for the mean at S2 and S3. At S1 each unit must reach Q plus 5 percentage points, so Q = 80% in 30 minutes means 85% from each of the first six tablets. If that fails, the mean of 12 must reach 80% with no tablet below 65%.

Is a dissolution result that needs S2 testing out of specification?

Not under the pharmacopeial table: going to S2 or S3 is part of the test, and the batch fails only if it does not conform at S3, which then needs an out-of-specification investigation. Two cautions apply. An internal release limit tighter than the table creates its own OOS. And MHRA expects a product that usually passes at the first stage, but needed the next, to be investigated as a result not following the normal trend.

Can you discard S1 results and start the dissolution test again?

No. The stages accumulate: S2 is judged on all 12 units and S3 on all 24, so restarting with six fresh units throws away results and is the repeated testing FDA calls testing into compliance. Original results are replaced only for a clearly identified laboratory error, and any retest follows a written procedure that fixes the number of retests in advance.

Is there an RSD limit for dissolution results?

Not in the pharmacopeial acceptance tables, which limit individual units and the mean instead. Variability limits apply to profile comparisons: to use mean data for an f2 comparison, FDA’s 1997 guidance asks for a coefficient of variation of no more than 20% at early time points and 10% later. EMA treats more than 20% RSD up to 10 minutes, or more than 10% later, with 12 units, as high variability that may justify a higher stirring speed during method development.

References21 sources

References

  1. 1The Japanese Pharmacopoeia, 18th edition, English version, General Tests 6.10 Dissolution Test, sections 1 to 4 and Acceptance Tables 6.10-⁠1 to 6.10-⁠4, MHLW Ministerial Notification No. 220, 7 June 2021. https://www.mhlw.go.jp/content/11120000/001372337.pdf (read 25 November 2026)
  2. 2ICH Q4B Annex 7(R2), Evaluation and Recommendation of Pharmacopoeial Texts for Use in the ICH Regions on Dissolution Test General Chapter, sections 2 and 4, ICH, Step 4, 11 November 2010. https://database.ich.org/sites/default/files/Q4B%20Annex%207%20%28R2%29%20Guideline.pdf (read 25 November 2026)
  3. 3Dissolution Testing of Immediate Release Solid Oral Dosage Forms, guidance for industry, section IV, FDA, August 1997. https://www.fda.gov/media/70936/download (read 25 November 2026)
  4. 4ICH Q6A Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and New Drug Products: Chemical Substances, section 3.3.2.1(a), ICH, Step 4, 6 October 1999. https://database.ich.org/sites/default/files/Q6A%20Guideline.pdf (read 25 November 2026)
  5. 5ICH Q6A Decision Trees, Decision Tree #7, Setting acceptance criteria for drug product dissolution, ICH, 6 October 1999. https://database.ich.org/sites/default/files/Q6A%20Decision%20Tree.pdf (read 25 November 2026)
  6. 6Guideline on quality of oral modified release products, EMA/CHMP/QWP/428693/2013, sections 2.2 and 3.3, European Medicines Agency, 20 March 2014. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-quality-oral-modified-release-products_en.pdf (read 25 November 2026)
  7. 7Extended Release Oral Dosage Forms: Development, Evaluation, and Application of In Vitro/In Vivo Correlations, guidance for industry, section VII.B.1, FDA, September 1997. https://www.fda.gov/media/70939/download (read 25 November 2026)
  8. 821 CFR 211.194, Laboratory records, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-211.194 (read 25 November 2026)
  9. 9Stason Pharmaceuticals, Inc., warning letter, MARCS-⁠CMS 604889, FDA, 8 July 2020. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/stason-pharmaceuticals-inc-604889-07082020 (read 25 November 2026)
  10. 10Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, Revision 1, sections I, IV.B.1 and IV.C, FDA, May 2022. https://www.fda.gov/media/158416/download (read 25 November 2026)
  11. 11Dissolution test for solid dosage forms: I do not understand how to interpret the acceptance criteria. Could you provide an example?, EDQM FAQ, last modified 15 September 2021. https://faq.edqm.eu/x/pwIV (read 25 November 2026)
  12. 12Out of Specification and Out of Trend Investigations, MHRA, October 2017, published on GOV.UK, last updated 26 February 2018. https://www.gov.uk/government/publications/out-of-specification-investigations (read 25 November 2026)
  13. 13Reflection paper on the dissolution specification for generic solid oral immediate release products with systemic action, EMA/CHMP/CVMP/QWP/336031/2017, sections 2.1 and 2.2, European Medicines Agency, 10 August 2017. https://www.ema.europa.eu/en/documents/scientific-guideline/reflection-paper-dissolution-specification-generic-solid-oral-immediate-release-products-systemic-action-first-version_en.pdf (read 25 November 2026)
  14. 14Alkem Laboratories Limited, warning letter 320-⁠27-⁠25, MARCS-⁠CMS 736113, FDA, 18 November 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/alkem-laboratories-limited-736113-11182026 (read 25 November 2026)
  15. 15Form FDA 483, Ipca Laboratories Limited, Pithampur, inspection of 15 to 23 June 2023, Observation 1A, FDA. https://www.fda.gov/media/170050/download (read 25 November 2026)
  16. 16JP History and Legal Status, Pharmaceuticals and Medical Devices Agency. https://www.pmda.go.jp/files/000280668.pdf (read 25 November 2026)
  17. 17The Japanese Pharmacopoeia, English version, editions and supplements listed, Ministry of Health, Labour and Welfare. https://www.mhlw.go.jp/stf/seisakunitsuite/bunya/0000066597.html (read 25 November 2026)
  18. 18Dissolution Testing and Acceptance Criteria for Immediate-⁠Release Solid Oral Dosage Form Drug Products Containing High Solubility Drug Substances, guidance for industry, sections III to V, FDA, August 2018. https://www.fda.gov/media/92988/download (read 25 November 2026)
  19. 19ICH Q6(R1) EWG Work Plan, 9 July 2026, sections 1.3 and 2, ICH. https://database.ich.org/sites/default/files/ICH53_Q6%28R1%29_EWG_WorkPlan_2026_0818.pdf (read 25 November 2026)
  20. 20Key Outcomes of Biannual ICH Assembly Meeting in Prague, press release, ICH, 19 November 2026. https://www.ich.org/pressrelease/key-outcomes-biannual-ich-assembly-meeting-prague (read 25 November 2026)
  21. 21Quality Guidelines, entries for Q4B Annex 7(R2), Q6A and Q6(R1), ICH. https://www.ich.org/page/quality-guidelines (read 25 November 2026)