What is the FDA QMSR, and what changed from the Quality System Regulation (QSR)?

The Quality Management System Regulation (QMSR) is 21 CFR Part 820, FDA’s good manufacturing practice rule for medical devices, effective 2 February 2026. It replaced most of the Quality System Regulation with ISO 13485:2016, incorporated by reference, plus FDA additions on complaint, servicing and unique device identifier records, labeling checks and links to FDA’s reporting rules. Risk management now runs through the whole system, management review and audit reports are open to inspectors, and FDA inspects under Compliance Program 7382.850.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal studying an auto-injector laid out in assembly order, threaded to its device history folder.

1What changed from the QSR.

FDA judged ISO 13485, the international quality management system (QMS) standard for medical devices, “when taken in totality”, to be “substantially similar” to the Quality System Regulation (QSR).1 For most requirements, what changed is where they are written, not what they demand. The first three rows below move requirements; the last three change what a site must do or show.

Table 1. Sources: the QSR as in force on 1 February 2026, the QMSR, the final rule and FDA’s QMSR pages.2,3,4,5
QSR, to 1 February 2026QMSR, from 2 February 2026
RequirementsFDA’s own text, subparts A to OISO 13485:2016 by reference, plus 820.10, 820.35 and 820.45
DefinitionsFDA’s own, in 820.3ISO 13485, clause 3 of ISO 9000:2015 and a few kept in 820.3; the Food, Drug, and Cosmetic Act’s own prevail
Record typesDesign history file (DHF), device master record (DMR), device history record (DHR) and quality system recordNot retained; ISO 13485 largely requires their contents
Risk managementNamed only in design validation, 820.30(g)Required throughout, for example clauses 4.1, 7.1, 7.3, 7.4, 7.5, 7.6 and 8.2
Audit and review reportsExempt from FDA review under 820.180(c)Open to FDA
Inspection methodQuality System Inspection Technique (QSIT)Compliance Program 7382.850

Does it apply? A site making only drugs or biologics is outside Part 820, even if another facility later adds a device; a site making a combination product with a device constituent part meets some or all of it through 21 CFR Part 4, set out below.3,6

2QSR-to-⁠QMSR gap checklist.

Run it against procedures written for the QSR and a sample of live records. FDA does not expect records made before 2 February 2026 to be revised or recreated, but investigators may review them; it suggests “some type of comparative analysis” and expects a plan for any gap the analysis finds.1,7

Table 2. Clause numbers refer to ISO 13485:2016, incorporated by 820.7.
CheckWhat to look forRequirement
Procedure referencesNone still rests on removed sections such as 820.30, 820.100 or 820.198820.10(a)3
Quality manualA quality manual exists; the QSR never required oneClause 4.2.28,2
Training effectivenessChecked in proportion to risk, which FDA says “goes far beyond just evaluating whether or not training was completed”Clause 6.28,7
Work environmentConditions that can affect product quality documented, monitored and controlledClause 6.4.18,9
Customer propertyControlled as far as device safety and effectiveness needClause 7.5.108,4
Suppliers and contractorsSelection and re-evaluation scaled to each purchase’s effect on the device; contract design controlled, design file accessible; change notification bindingClauses 4.1.5, 7.4.1, 7.4.24,8,10,11
CAPA proceduresCorrective and preventive action (CAPA) procedures define how causes are found and effectiveness reviewed, and cover preventive action, which FDA inspects separatelyClauses 8.5.2, 8.5.38,4,9
Risk in realizationOne or more documented risk management processes for product realization, with recordsClause 7.17,9
Risk files kept liveComplaints, returns and postmarket data feed a risk file that is revised, not left as first writtenClause 8.2.1; CP Part V, Situation 1(g)8,12
Risk-based processesDecisions such as retraining frequency documented with their risk basisClause 4.1.2(b)7
Renamed recordsDevice file, design file and per-batch record contain or reference what the DMR, DHF and DHR heldClauses 4.2.3, 7.3.10, 7.5.14
SoftwareQMS and production software validated for its intended use, effort set by riskClauses 4.1.6, 7.5.6, 7.6; Computer Software Assurance guidance, February 202613
Reports FDA may readManagement reviews and audit reports, earlier ones included, written knowing an investigator may read themFAQ 81
Internal auditsCover Part 820’s additions as well as ISO 13485; a notified body audit is no substituteClause 8.2.47

Part 820’s own additions, on links to FDA’s reporting, unique device identifier (UDI) and tracking rules, complaint and servicing records and labeling checks, have a line-by-line check in ISO 13485 vs 21 CFR 820.3

2.1Example: gap records for one procedure.

Example, invented. Site: Brantmoor Devices, a single-site maker of class II devices. Procedure: SOP-⁠QA-⁠012 Complaint handling, revision B, written for the QSR. Reviewer: the QA manager, 16 March 2026.

Table 3. Illustrative. The first two rows were QSR duties (820.198) under new citations, as Koven’s 820.35(a) violation shows. The last two are new: the QSR named risk only in design validation, and required training (820.25(b)) but not a check that it worked.3,2,8,7,10
GapFoundRequirementAction and ownerPriority and dueEvidence of closure
Returns unscreenedReturns go straight to repair with no complaint decisionClause 8.2.2; 820.35(a)Add a complaint screen to the returns form; service managerHigh, 15 April 2026The next 30 returns each show a recorded complaint decision
Old referencesSection 2 cites 820.198 and 820.198(a)(3)Clauses 8.2.2 and 8.2.3; 820.35(a); 820.10(b)(3)Revise references and the medical device reporting (MDR) step; QA managerHigh, 30 April 2026Revision C approved, staff trained before it takes effect
Risk file not fedComplaint trends go to management review only; the risk file was last revised at design transferClause 8.2.1; CP Part V, Situation 1(g)Send each investigated complaint for a check against the risk file’s hazards and estimates; risk managerHigh, 30 April 2026Risk file revised with the quarter’s complaints, review signed
Training by signatureSection 9 trains complaint coordinators by read-and-sign onlyClause 6.2Add a graded case exercise; training leadMedium, 29 May 2026Each coordinator passes before handling complaints alone

3Where each QSR section went.

Table 4. All 31 QSR sections, mapped by subject; 820.1 and 820.3 survive as scope and definitions. Groups follow Attachment A of Compliance Program 7382.850 and the final rule; for statistics, they are the clauses whose plans FDA’s October 2025 draft guidance, not yet final, says should describe statistical techniques or methods.2,8,3,4,14,11
QSR sectionWhere FDA now looks
820.5 and 820.186Clauses 4.1 and 4.2
820.20 ManagementClauses 5.1 to 5.6; resources, 6.1
820.22 Quality auditClause 8.2.4
820.25 PersonnelClause 6.2
820.30 Design controlsClause 7.3, for the devices in 820.10(c)
820.30(j) DHFDesign and development file, clause 7.3.10
820.40 DocumentsClause 4.2.4
820.50 PurchasingClauses 4.1.5 and 7.4
820.60 and 820.65Clauses 7.5.8 and 7.5.9, with 820.10(b) and (d)
820.70 ProductionClauses 6.3, 6.4, 7.5.1 and 7.5.2
820.70(i) SoftwareClauses 4.1.6, 7.5.6 and 7.6
820.72 Test equipmentClause 7.6
820.75 Process validationClauses 7.5.6 and 7.5.7
820.80 and 820.86Clauses 7.4.3 and 8.2.6; status, 7.5.8
820.90 NonconformingClause 8.3
820.100 CAPAClauses 8.5.2 and 8.5.3
820.120 and 820.130820.45, with clause 7.5.1
820.140 and 820.150Preservation of product, clause 7.5.11
820.160 DistributionClauses 7.5.1 and 7.5.9
820.170 InstallationClause 7.5.3, with 820.35(b) where serviced
820.180 RecordsClause 4.2.5, with 820.35
820.181 DMRMedical device file, clause 4.2.3
820.184 DHRRecord per device or batch, clause 7.5.1
820.198 ComplaintsClauses 8.2.2 and 8.2.3, with 820.35(a)
820.200 ServicingClause 7.5.4, with 820.35(b)
820.250 StatisticsClauses 7.3.6, 7.3.7 and 7.5.6; cited under 8.2.6 in Unilam’s letter

FDA declined to publish its own map, calling a direct comparison “cumbersome and not a useful tool”, and points to AAMI TIR102:2019.4,5 To check a clause yourself, ANSI’s incorporated-by-reference portal shows ISO 13485:2016 read-only, “at no cost to the user”.4,1

4Who the QMSR covers.

Part 820 covers manufacturers of finished devices, including contract sterilizers, installers, relabelers, remanufacturers, repackers, specification developers and initial distributors of foreign entities that perform these functions. A site doing only some regulated operations meets only those requirements. Makers of components, including software, firmware and labeling meant for the device, are excluded, but “A manufacturer of accessories is subject to the QMSR”, and FDA treats blood tubing and diagnostic x-ray components as finished devices.3,5

  • Contract manufacturers. Those making finished devices “are considered manufacturers” for the operations they perform. They and the finished device manufacturer “are all legally responsible”, though the finished device manufacturer bears overall responsibility. FDA may read written agreements to see how the parties divided their activities and responsibilities.8
  • Exempt devices. Makers of devices exempt from current good manufacturing practice (CGMP) still keep complaint files and the records required by 820.35.5
  • Investigational devices. Those made under an investigational device exemption still need the design controls in 820.10(c).5
  • Cell and tissue products. Human cell and tissue products regulated as devices are inside Part 820.3
  • EU and UK. The QMSR is US law, so their obligations do not change: an EU manufacturer’s quality system must still cover what Article 10(9) of the Medical Device Regulation lists, such as a regulatory compliance strategy and clinical evaluation.15 What an ISO 13485 certificate does in the US, EU, UK and Canada is set out separately.

4.1Combination products.

Table 5. Under 21 CFR 4.4 a site can run one system on either set of rules, plus named provisions from the other; a biological constituent part adds Parts 600 to 680. Compliance Program 7356.000 still prints QSR section numbers.6,16,8
SituationWhat appliesInspected under
Drug-⁠CGMP routeParts 210 and 211, plus ten ISO 13485 clauses, 820.10, 820.35(a) and (b), and risk management in product realizationCompliance Program 7356.000 with the drug program
QMSR routeAll of Part 820, plus eight Part 211 sections, from 211.84 to 211.1707356.000 with the device program
Separate facilityAll the rules for the constituent part made thereThat product’s own program
CBER-led productEither route aboveCBER’s own compliance programs

5How FDA inspects under the QMSR.

Compliance Program 7382.850 groups the requirements into six QMS areas, from change control to production and service provision, and four other FDA requirements: medical device reporting, corrections and removals, tracking and UDI. Investigators read your risk management documentation, select elements by product risk, and follow threads between areas.8

Against QSIT, FDA says “the inspection process will stay largely the same”: a Form FDA 482 at a domestic site, a tour, records, interviews, then a Form 483 the site can annotate or answer in writing. What changes is coverage: “on every inspection, FDA will evaluate, to some extent requirements in each of the main parts of your quality management system”.7

Two inspection types follow model 2, which evaluates named minimum elements: baseline surveillance, for a site with no FDA inspection or Medical Device Single Audit Program (MDSAP) audit history or with risk factors that call for one, and premarket approval (PMA) preapproval, which skips the four other requirements if the device is not yet on the market. The rest follow model 1, at least one element in every area: other surveillance, compliance follow-up, for-cause, specific product risk and PMA postmarket.8,7

What to prepare:

  • The last review and audit cycle. Management review, internal audit and supplier audit are expected to be reviewed in baseline and PMA preapproval inspections, and may be in others.7 Have the latest of each ready, with actions closed.
  • Your highest-risk products first. Investigators select records by product risk rather than by a statistical sampling plan, and may ask for “risk management files and reports” and “evidence that risk controls are effective”.7,8 FDA’s suggested internal audit drill is “tracing them through your processes, starting with identified patient risk, then following the associated risk controls through each impacted process”.7

6What inspectors have cited so far.

On 26 November 2026 FDA’s warning letter index listed seven letters citing ISO 13485 clauses, all from inspections that began on or after 2 February 2026. Letters from earlier inspections, such as 3B Medical’s (December 2025) and ZIIP’s (January 2026), still cite QSR sections, so the citation has followed the inspection date.17,18,19 Five of the seven, to Linemaster Switch, Koven Technologies, Nipro Renal Solutions, Pensar Medical and Unilam, cite clause 7.1 on risk management.9,10,12,20,11

“Your firm’s Risk Management procedure TM-⁠112 does not define how risk management activities are performed and documented, who is responsible for conducting and approving risk management activities, when risk management documentation must be updated, and how post-market feedback data (including complaints, adverse events, and recalls) is incorporated into risk management.” Linemaster Switch Corporation, warning letter, 27 May 20269

Nipro’s risk analysis reports “have not been updated since their initial creation”; one still rated all risks “either negligible or acceptable” during a Class I recall.12 Unilam, a South Korean sunlamp maker, had risk controls that leaned on labeling and on what the sunlamp bed maker would do.11

FDA also enforces its own additions. Koven’s violation 4, on returned goods authorization (RGA) forms, is cited under 820.35(a), not an ISO clause:

“The investigator reviewed 29 completed RGA forms … and found that 13 of these forms reported device failures but were not evaluated as complaints.” Koven Technologies, Inc., warning letter, 21 July 202610

Two more lessons:

  • Contract design. Koven’s contract manufacturer, which designs and makes its devices, changed a Doppler’s indications to fetal use, and Koven had no record of when. Under clause 7.4.1 FDA cited a quality agreement that did not reach design changes beyond labeling, and no site assessment of a supplier Koven’s own procedure rated “Highest/Critical Impact”.10
  • Inherited designs. Pro-⁠Optics told FDA “We don’t have device drawings” and could not produce design and development records for its contact lens cases (clause 7.3.1). A 2021 move away from a contract manufacturer was neither documented nor assessed (clause 4.1.4).21

A firm cited under the QSR still corrects to the QMSR: Specialty Health Products, inspected in December 2025, was told “any corrective actions you propose or implement must be pursuant to the QMSR requirements in effect as of February 2, 2026”.22

7What is settled and what is still draft.

  • 2 February 2024. Final rule published, two years before it took effect.4,1
  • October 2025. Draft guidance on QMS information in premarket submissions, which when final will supersede the 2003 guidance. Still a draft, not for implementation.14
  • 2 February 2026. QMSR in force. QSIT and Compliance Programs 7382.845 and 7383.001 retired; 7382.850 in use.5
  • 1 April 2026. At a CDRH town hall, FDA said the transition period is over and it does not intend to issue another document to supplement Compliance Program 7382.850.7

The QMSR fixes the 2016 edition of ISO 13485, and a new edition would apply only if FDA amends the rule.1 ISO last reviewed and confirmed that edition in 2025, so it remains current.23

8How it runs in Seal.

The medical device blueprint keeps the records the QSR called the design history file, device master record and device history record as connected records, whichever names your procedures use: requirements to verification for the design and development file, the released configuration for the medical device file, and each unit’s actual build for the record per device or batch. A design change shows the requirements, verification and built units it reaches through those links.

In risk management, a failed verification, complaint or maintenance finding links to the scenario and control it challenges, so the risk file is reopened by the evidence rather than at the next scheduled review. Complaint handling keeps the report as received, with reportability recorded as criteria, facts and rationale. Map the configured system to the QMSR and ISO 13485 yourself: a workflow does not by itself demonstrate conformity.

AQuestions and answers

Does the QMSR require ISO 14971?

No. Part 820 incorporates only ISO 13485:2016 and clause 3 of ISO 9000:2015, and the final rule says FDA does not incorporate ISO 14971 or any other standard that ISO 13485 references. Risk management is still required, through clause 7.1 and the risk-based approach that runs through the standard, and FDA acknowledges that standards such as ISO 14971 may help in applying ISO 13485.

How does the QMSR define a complaint?

Part 820 no longer defines it. Section 820.3 applies the definitions in ISO 13485 and clause 3 of ISO 9000, so the QSR’s own definition, which covered deficiencies in identity, quality, durability, reliability, safety, effectiveness or performance after release for distribution, has gone. When FDA cited Koven Technologies in 2026 it quoted the firm’s definition, which covered usability, devices released from the organization’s control and services that affect performance.

What labeling checks does the QMSR add to ISO 13485?

Section 820.45 adds to clause 7.5.1. Procedures must cover the integrity, inspection, storage and operations for labeling and packaging. Before release or storage they are examined for accuracy, including the correct unique device identifier or Universal Product Code, expiry date, and storage, handling and processing instructions. Releasing labeling for use is documented, and operations must prevent mix-ups, with labeling inspected before use against the medical device file and the results recorded.

Does taking part in MDSAP replace FDA inspections under the QMSR?

Only routine surveillance inspections. FDA uses MDSAP audit reports in their place, but still makes for-cause, premarket approval and other non-surveillance inspections, and still inspects electronic product radiation control activities. Its own inspections do not follow the MDSAP audit plan, FDA neither requires nor issues ISO 13485 certificates, and a certificate does not exempt a site from inspection. MDSAP stays voluntary, and combination products are outside it.

Do corrective and preventive action need separate procedures under the QMSR?

The requirements are separate: ISO 13485 clause 8.5.2 covers corrective action and 8.5.3 preventive action, FDA inspects them as separate elements, and the final rule says FDA did not combine them. Warning letters so far cite what procedures leave out and what records fail to show, not whether one document holds both. Pensar Medical’s combined procedure was cited for five records with blank root causes, missing actions and unverified effectiveness.

References23 sources

References

  1. 1Quality Management System Regulation, Frequently Asked Questions, FDA, introduction and questions 5 to 8, 12 and 13, content current as of 5 November 2026. https://www.fda.gov/medical-devices/quality-management-system-regulation-qmsr/quality-management-system-regulation-frequently-asked-questions (read 26 November 2026)
  2. 221 CFR Part 820, Quality System Regulation, as in force on 1 February 2026, eCFR. https://www.ecfr.gov/on/2026-02-01/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  3. 321 CFR Part 820, Quality Management System Regulation, sections 820.1, 820.3, 820.7, 820.10, 820.35 and 820.45, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  4. 4Medical Devices; Quality System Regulation Amendments, final rule, FDA, 89 FR 7496, 2 February 2024, responses to comments 9, 14, 18, 19, 20, 21, 26, 31, 40 and 48. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  5. 5Quality Management System Regulation (QMSR), FDA, content current as of 2 February 2026. https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr (read 26 November 2026)
  6. 621 CFR 4.4, current good manufacturing practice for co-packaged or single-entity combination products, paragraphs (b)(1), (b)(2), (b)(4) and (c), eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-4/subpart-A/section-4.4 (read 26 November 2026)
  7. 7FDA’s Quality Management System Regulation (QMSR): Medical Device Risk-⁠Based Inspections, CDRH town hall transcript, FDA, 1 April 2026. https://www.fda.gov/media/191960/download (read 26 November 2026)
  8. 8Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Parts I.9, II.2.C, III and V and Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)
  9. 9Linemaster Switch Corporation, warning letter, MARCS-⁠CMS 730215, FDA, 27 May 2026, violations 2 to 4. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/linemaster-switch-corporation-730215-05272026 (read 26 November 2026)
  10. 10Koven Technologies, Inc., warning letter, MARCS-⁠CMS 734643, FDA, 21 July 2026, violations 2 to 4. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/koven-technologies-inc-734643-07212026 (read 26 November 2026)
  11. 11Unilam Co., Ltd., warning letter, MARCS-⁠CMS 738562, FDA, 22 September 2026, violations 1, 4 and 6. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/unilam-co-ltd-738562-09222026 (read 26 November 2026)
  12. 12Nipro Renal Solutions USA Corporation, warning letter, MARCS-⁠CMS 732874, FDA, dated 16 July 2026, issued 24 July 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/nipro-renal-solutions-usa-corporation-732874-07242026 (read 26 November 2026)
  13. 13Computer Software Assurance for Production and Quality Management System Software, final guidance, FDA, issued 3 February 2026, docket FDA-⁠2022-⁠D-⁠0795, sections I and V.A(1). https://www.fda.gov/regulatory-information/search-fda-guidance-documents/computer-software-assurance-production-and-quality-management-system-software (read 26 November 2026)
  14. 14Quality Management System Information for Certain Premarket Submission Reviews, draft guidance, FDA, October 2025, docket FDA-⁠2025-⁠D-⁠4051. https://www.fda.gov/media/189345/download (read 26 November 2026)
  15. 15Regulation (EU) 2017/745 on medical devices, Article 10(9), EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745 (read 26 November 2026)
  16. 16Compliance Program 7356.000, Inspections of CDER-led or CDRH-led Combination Products, FDA, implemented 4 June 2020, note on page 2, Part II and Tables 1 and 2. https://www.fda.gov/media/138592/download (read 26 November 2026)
  17. 17Warning Letters, FDA, index searched for “ISO 13485” and “Clause 7.1” on 26 November 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/compliance-actions-and-activities/warning-letters (read 26 November 2026)
  18. 183B Medical, Inc. dba Reach Health, Inc., warning letter, MARCS-⁠CMS 725108, FDA, 20 May 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/3b-medical-inc-dba-reach-health-inc-725108-05202026 (read 26 November 2026)
  19. 19ZIIP, Inc., warning letter, MARCS-⁠CMS 725870, FDA, 22 May 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/ziip-inc-725870-05222026 (read 26 November 2026)
  20. 20Lion Street Medical, d/b/a Pensar Medical LLC, warning letter, MARCS-⁠CMS 733015, FDA, 21 August 2026, QMSR violations 1 and 3. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/lion-street-medical-dba-pensar-medical-llc-733015-08212026 (read 26 November 2026)
  21. 21Pro-⁠Optics, LLC, warning letter, MARCS-⁠CMS 734689, FDA, 5 October 2026, QMSR violations 1 and 2. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/pro-optics-llc-734689-10052026 (read 26 November 2026)
  22. 22Specialty Health Products Co., LLC, warning letter, MARCS-⁠CMS 726564, FDA, 7 July 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/specialty-health-products-co-llc-726564-07072026 (read 26 November 2026)
  23. 23ISO 13485:2016, Medical devices, Quality management systems, Requirements for regulatory purposes, ISO, stage 90.93, confirmed. https://www.iso.org/standard/59752.html (read 25 November 2026)