1How to calculate the RPN.
WHO’s quality risk management guideline defines it: “Each failure mode gets a numeric score that quantifies likelihood of occurrence, likelihood of detection and severity of impact. The product of these three scores is the RPN for that failure mode.”1
RPN = severity × occurrence × detection
| Factor | What it rates | Source definition |
|---|---|---|
| Severity (S) | The consequence if the failure happens | “A measure of the possible consequences of a hazard”2 |
| Occurrence (O) | How often the cause arises | “likelihood of occurrence”1 |
| Detection (D) | How likely it is to be found before harm | “The ability to discover or determine the existence … of a hazard”2 |
Detection is scored by how hard the failure is to find: hard to find scores high, so on all three scales a higher number means a higher priority. Every score needs evidence: ICH Q9(R1) says “Risk scores, ratings and assessments should be based on an appropriate use of evidence, science and knowledge”.2
The maximum RPN is 125 on 1 to 5 scales and 1,000 on 1 to 10 scales.
2RPN rating scales and action thresholds.
2.1Example rating scales.
Neither ICH Q9(R1) nor WHO prescribes a scale; these, the invented set QRM-SC-02 v3, are illustrative.2,
| Score | On 1 to 10 | Severity | Occurrence: the cause arises | Detection before release |
|---|---|---|---|---|
| 5 | 9 or 10 | Could harm a patient, such as loss of sterility | More than once in 10 batches | Not detectable |
| 4 | 7 or 8 | Batch fails a specification | Once in 10 to 100 batches | Found only by chance, or by a check not proven for it |
| 3 | 5 or 6 | Within specification, a quality attribute affected | Once in 100 to 1,000 batches | Likely found by a periodic or sampled check |
| 2 | 3 or 4 | GMP non-conformance, no product effect | Rarer than once in 1,000 batches | Found by a validated check of every unit or batch |
| 1 | 1 or 2 | No noticeable effect | None in over 5,000 batches of relevant data | Always found: a validated automatic check stops the line |
2.2Example action rules.
- Controls a regulation or GMP guideline requires are in place before scoring and are never accepted away by an RPN.
- Act when the RPN is 40 or more, or when severity is 5 and detection 4 or 5.
- From 20 to 39, act or record a rationale for accepting, approved by QA.
- Below 20, accept and review at the next periodic review.
No regulation sets these numbers. WHO says that scoring systems and trigger points for action “are subjective so the rationale for score categorization should be defined in as much detail as possible”1 and ICH Q9(R1) that the acceptable level of risk “should be decided on a case-by-case basis”.2 Rule 1 follows PIC/S, which asks inspectors to check that risk assessments are “not being used to override GMP and other regulatory requirements”.3 Annex 1 sets such minimums: RABS gloves “should be visually examined with each use, and integrity testing should be performed at periodic intervals”.4
3Example: an FMEA of a sterile filling line.
The product, line and numbers are invented. The risk question: which failures on Line 3 could let a non-sterile or under-filled vial of Exemplar 10 mg/mL injection reach a patient? The team scores with scale set QRM-SC-02 v3 and the rules above; QA approves every decision and rationale.
| ID | Failure mode | Cause | Current controls |
|---|---|---|---|
| F1 | RABS glove pinhole: ingress into Grade A | Wear; gloves replaced every 10 production days | Visual check at each use; weekly integrity test |
| F2 | Stopper not seated: closure integrity lost | Head misaligned after changeover | Validated 100% camera reject |
| F3 | Low fill volume: under-filled vials | Pump wear | Checkweigh every 30 minutes |
| F4 | Vial neck crack: closure integrity lost | Thermal shock in the tunnel | 100% visual inspection; validated integrity test on a 20-vial sample |
| ID | S × O × D | Evidence for O and D | Decision |
|---|---|---|---|
| F1 | 5 × 4 × 3 = 60 | 3 failed glove tests in 2026, about 260 batches; a breach can go a week unseen | Act: RPN 40 or more |
| F2 | 5 × 2 × 2 = 20 | 1 misaligned head in 3,400 batches; camera rejected 300 of 300 challenge vials | Accept: QA rationale QRA-26-017 |
| F3 | 4 × 3 × 3 = 36 | 4 pump-wear drifts in 600 batches, each caught at a 30-minute check | Accept: QA rationale QRA-26-018 |
| F4 | 5 × 1 × 4 = 20 | Tunnel temperature records within validated limits in all 6,200 batches; a 20-vial sample finds a random crack only by chance | Act: severity 5 with detection 4 |
F4’s occurrence rests on the tunnel records, not on visual inspection, which the team does not trust for cracks. The record names the assumption left, that the tunnel is the only source of cracks, since ICH Q9(R1) says revealing “assumptions and reasonable sources of uncertainty will enhance confidence in this output”.2
| ID | Action | Owner | Evidence for the rescore | New S × O × D |
|---|---|---|---|---|
| F1 | Glove test after each batch and intervention; replacement interval halved from 10 to 5 production days | Sterility assurance lead | Test qualified with holed gloves; Line 2 at 5 days: 1 failure in 410 batches | 5 × 3 × 2 = 30 |
| F4 | 100% automated leak test after capping | Engineering manager | 120 of 120 cracked challenge vials rejected | 5 × 1 × 2 = 10 |
Severity stays 5: the controls do not change what happens if the failure gets through. F1’s occurrence falls only because Line 2 data show the shorter interval works; ICH Q9(R1) says uncertainty may be reduced using “accumulated and new information (both internal and external)”.2 F1 at 30 is accepted with QA rationale QRA-26-019, and the team adds a row for the leak tester failing in use, since risk reduction can bring new risks.2
4FMEA worksheet template: the steps, column by column.
Fill it from top to bottom: the first two rows head the sheet, and the other nine name its columns, left to right, as the header row to copy.
| Field | What goes in it | Requirement |
|---|---|---|
| Risk question | Problem, scope and assumptions; leader, deliverables and timeline | ICH Q9(R1) 4.2 and 4.32 |
| Team | Quality, production and engineering: “experienced staff trained in the use of relevant risk assessment tools” | ICH Q9(R1) 4.1;2 PIC/S PI 038-2, 2.63 |
| Step and failure mode | One row per failure mode and cause, step by step along a process map whose critical process parameters are already defined | WHO 2013, Table 3;1 ICH Q9(R1) Annex I.22 |
| Effect and cause | What it does to the patient, and why it would happen | ICH Q9(R1) Annex I.22 |
| Current controls | Prevention and detection in place now, minimums included | ICH Q9(R1) 5;2 Annex 14 |
| S, O, D | Scores on the approved scales, with the scale version | WHO 2013, 3.41 |
| Evidence | The data behind each score; for a new process, data from similar ones, with the uncertainty | ICH Q9(R1) 4.3, 5.1 and 5.3;2 PIC/S PI 038-2, 2.33 |
| RPN and decision | S × O × D; act or accept, and why, endorsed by QA | WHO 2013, 3.8 and glossary;1 ICH Q9(R1) 4.42 |
| Action and owner | A control to eliminate, contain, reduce or control the failure; who and by when, under change control | ICH Q9(R1) 4.4 and Annex I.22 |
| Rescore and new risks | New scores once actions “have really been implemented in the manner they appear in the risk assessment”; risks the action adds | ICH Q9(R1) 4.4;2 PIC/S PI 038-2, 4.13 |
| Review triggers | Deviations, complaints and changes that reopen the row; a frequency “based upon the level of risk” | ICH Q9(R1) 4.62 |
5If your scales run from 1 to 10.
Split each level in two, as the example scales’ second column does; Rule 2 then reads severity 9 or 10 with detection 7 to 10. Set the RPN bands afresh, because they do not scale in proportion: 40 of 125 is 320 of 1,000, yet a row scored 5, 4 and 2, which acts at 40, can become 9 × 7 × 3 = 189, while one scored 4, 3 and 3, below the line at 36, can become 8 × 6 × 6 = 288. No single cut-off keeps both decisions, so test new bands on rows already decided.
ICH Q9(R1) names “poorly designed risk scoring scales” as a source of subjectivity,2 and detection is the score WHO says inspectors check for overrating,1 so define its ten levels by the check that would find the failure.
| D | What would find the failure before release |
|---|---|
| 10 | Nothing: no check covers it |
| 9 | Only a final test that cannot assure it, such as the sterility test |
| 8 | Chance: someone happens to notice |
| 7 | A routine check not proven for it, such as visual inspection for vial cracks |
| 6 | A proven check with long gaps, such as a weekly glove integrity test |
| 5 | A proven check on samples from each batch that would likely catch it, such as a checkweigh every 30 minutes |
| 4 | A qualified manual check of every unit or batch, such as visual inspection for particles |
| 3 | A validated automated check of every unit or batch, challenged with known defects |
| 2 | A validated automatic check of every unit that stops the line |
| 1 | Two independent validated checks of every unit, either able to stop the line |
Three levels rest on Annex 1. The sterility test should be regarded only as “the last in a series of critical control measures by which sterility is assured”, so it is no proven check for lost sterility. Manual inspectors should be qualified at least annually, and automated inspection validated to detect known defects and “equal to, or better than, manual inspection methods”.4
Level 7 rests partly on Annex 1. For containers closed by fusion, such as ampoules, Annex 1 says “visual inspection is not considered as an acceptable integrity test method”; other closures are sampled and checked “using validated methods”.4 For a stoppered vial the rest is judgment: FDA has told one sterile manufacturer that visual examinations of its aseptic equipment “provide far lower detectability than, and cannot be substituted for,” integrity testing devices,5 so F4’s team does not count its 100% visual inspection as a proven check for cracks.
6How it runs in Seal.
Whichever scale you settle on, you configure it in Seal’s quality risk management blueprint as versioned criteria, so a reviewer can see what an assigned 7 means. A process FMEA there holds each step’s failure modes, effects, causes, existing controls, estimates, actions, verification and residual decisions, and each estimate, before and after the action, keeps its sources, assumptions, uncertainty, rationale and reviewer. The action goes through change control with the verification the rescore relies on, and a later complaint or failed verification links back to the row it reopens.
To try the example 1 to 5 scales and rules first, the free FMEA worksheet opens on Line 3’s four failure modes. Replace them with your own, score each, and it shows each RPN, ranks the rows by what the rules say to act on, and saves the sheet for Word.
7FMEA, risk matrix or ranking: which to use.
Strictly, a scored FMEA is an FMECA, as IEC 60812 and ICH Q9(R1) use the terms, but WHO’s glossary places the RPN within FMEA, so either name works if your procedure defines the method.6,
A product of ordinal scores hides how it was reached, which is why F2 and F4 tie. The 125 combinations on 1 to 5 scales give only 30 different RPNs, and the 1,000 on 1 to 10 scales only 120, so unlike risks share a number. IEC 60812:2018 added “alternative means of calculating risk priority numbers (RPN)” and a criticality matrix method.6
Action Priority versus RPN. The automotive AIAG & VDA FMEA Handbook of 2019 brought in “the Action Priority (AP) methodology and Tables to replace RPN”, and a process that prioritizes actions “considering severity, occurrence, and detection individually and in combination with risk-reducing factors”.7 A supplier FMEA built to it arrives with priorities read from tables, not products. ICH Q9(R1) and WHO’s 2013 guideline do not mention it; within an RPN procedure, Rule 2 is what stops a low product hiding a severe, hard-to-detect failure.2,
8What inspectors look for.
WHO warns manufacturers: “Inspectors will assess whether risk assessments underrate the likelihood of occurrence and the consequences of overrating detection such that the patient risk is underestimated. The factual evidence behind statements should be robust to challenge by inspectors.”1 Two 2026 FDA warning letters and a 2026 EU statement of non-compliance show what they find:
“Despite your Risk Management Matrix indicating peritonitis as a potential harm and recommending the highest severity level, you instead assigned the lowest severity level with the potential harm being ‘damage of property.’ … Moreover, users should not be relied upon to detect leaking bags.” Fresenius Medical Care, warning letter 320-26-119, 25 August 20268
“Visual (b)(4) examinations provide far lower detectability than, and cannot be substituted for, (b)(4) integrity testing devices.” Simtra BioPharma Solutions, warning letter 320-26-49, 3 March 20265
“The risk assessment QRM/EU2/061/2025 did not evaluate the need for revalidation of the Meropenem API manufacturing process following the installation of new RABS and modification of the filling line layout.” Eugia Pharma Specialities, statement of GMP non-compliance FT020/SA/002/2023/NCR, 27 March 20269
9Devices and combination products.
Device makers set their own thresholds too: ISO 14971:2019 “requires manufacturers to establish objective criteria for risk acceptability but does not specify acceptable risk levels”.10 Drug-device combination products made under Part 211 also need a documented risk management process under 21 CFR 4.4(b)(1)(ii).11 FDA has cited device makers for not using the RPNs their procedure required,12 calling a risk low in a health hazard evaluation when their design FMEA gave it the highest severity,13 and leaving about 29 design FMEA rows at Risk Priority “B” with no recommended action.14
10What changed, and what is still draft.
- ICH Q9(R1). Adopted on 18 January 2023, in effect in the EU since 26 July 2023 and FDA final guidance from May 2023, it added sections on formality, risk-based decision-making, product availability and managing subjectivity.15,
16, 2 - WHO. Draft working document QAS/26.994, published in September 2026, closed for comments on 15 November 2026, and WHO’s schedule prepares a version for possible adoption from July to September 2027. It keeps the 2013 warnings against underrating occurrence or severity and overrating detection, and adds a section on managing subjectivity (3.9). Until it is adopted, the 2013 text stands.17,
1 - Standards. IEC 60812:2018 remains current, with a stability date of 2029; ISO 14971:2019 was confirmed in 2025.6,
10
AQuestions and answers
What is an acceptable RPN?
There is no regulatory number. ICH Q9(R1) says the acceptable level of risk is decided case by case, and WHO says trigger points are subjective and need a written rationale. Set bands in your procedure, such as act at 40 or more on 1 to 5 scales, and add a severity rule so that a serious, hard-to-detect failure with a low product still gets action. A cut-off does not carry over between scales: 40 on 1 to 5 scales is not 320 on 1 to 10.
How do you reduce an RPN?
Lower occurrence by removing or controlling the cause, or lower the detection score with a check proven to find the failure. Severity usually only falls if the design changes. Rescore only after the control is in place and verified, because inspectors look for detection that has been overrated.
Should FMEA scales run from 1 to 5 or 1 to 10?
Either works if every level is defined, because ICH Q9(R1) names poorly designed scoring scales as a source of subjectivity. Ten levels give finer steps but are harder to tell apart. Use one scale throughout an FMEA, record its version, and set new action bands if you change scale rather than multiplying the old ones.
What is the difference between a process FMEA and a design FMEA?
A process FMEA looks at how each manufacturing or testing step can fail. A design FMEA looks at how a product’s design, such as a device component or a container closure, can fail to meet its requirements. A drug manufacturer’s FMEA is usually a process FMEA; device makers often run both, and FDA has cited design FMEAs in device warning letters.
Is FMEA required by GMP?
No. EU GMP requires quality risk management, FDA has adopted ICH Q9(R1) as guidance, and Q9(R1) lists FMEA as one of several tools that may be used. Pick FMEA for processes and equipment you understand well enough to break into steps.
References17 sources
References
- 1WHO guidelines on quality risk management, WHO Technical Report Series No. 981, Annex 2, sections 3.4, 3.8, 5.3 and 6, glossary and Tables 1 and 3, World Health Organization, 2013. https:/
/ cdn. who. int/ media/ docs/ default- source/ medicines/ norms- and- standards/ guidelines/ production/ trs981- annex2- who- quality- risk- management. pdf (read 25 November 2026) - 2ICH Q9(R1), Quality Risk Management, sections 4, 5, 5.1, 5.3 and 7 and Annex I, I.2 to I.8, ICH, adopted 18 January 2023. https:/
/ database. ich. org/ sites/ default/ files/ ICH_ Q9%28R1%29_ Guideline_ Step4_ 2023_ 0126_ 0. pdf (read 25 November 2026) - 3PI 038-2, Aide-Memoire: Assessment of Quality Risk Management Implementation, sections 2, 3 and 4, PIC/S, 1 January 2021. https:/
/ picscheme. org/ docview/ 3823 (read 25 November 2026) - 4EudraLex Volume 4, Annex 1: Manufacture of Sterile Medicinal Products, sections 4.21, 8.22, 8.23, 8.31, 8.32 and 10.5, European Commission, 22 August 2022. https:/
/ health. ec. europa. eu/ system/ files/ 2022- 08/ 20220825_ gmp- an1_ en_ 0. pdf (read 25 November 2026) - 5Simtra BioPharma Solutions, warning letter 320-26-49, FDA, 3 March 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ simtra- biopharma- solutions- 720436- 03032026 (read 25 November 2026) - 6IEC 60812:2018, Failure modes and effects analysis (FMEA and FMECA), edition 3.0, IEC, 10 August 2018. https:/
/ webstore. iec. ch/ en/ publication/ 26359 (read 25 November 2026) - 7It’s here: claim your copy of the new AIAG & VDA FMEA Handbook today, announcement of the AIAG & VDA FMEA Handbook, 1st edition, AIAG, 3 June 2019. https:/
/ blog. aiag. org/ its- here. . . claim- your- copy- of- the- new- aiag- vda- fmea- handbook- today (read 25 November 2026) - 8Fresenius Medical Care AG & Co. KGaA, Ogden plant, warning letter 320-26-119, FDA, 25 August 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ fresenius- medical- care- ag- co- kgaa- 730319- 08252026 (read 25 November 2026) - 9Statement of non-compliance with GMP, Eugia Pharma Specialities Limited, Unit 2, Bhiwadi, India, report FT020/
SA/ 002/ 2023/ NCR, EudraGMDP document 184031, National Authority of Medicines and Health Products, Portugal, 27 March 2026; search issue date 2026-03-27. https:/ / eudragmdp. ema. europa. eu/ inspections/ gmpc/ searchGMPNonCompliance. do (read 25 November 2026) - 10ISO 14971:2019, Medical devices: Application of risk management to medical devices, ISO. https:/
/ www. iso. org/ standard/ 72704. html (read 25 November 2026) - 1121 CFR 4.4, current good manufacturing practice for co-packaged and single-entity combination products, paragraph (b)(1)(ii), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 4/ subpart- A/ section- 4. 4 (read 25 November 2026) - 12NeuroSync, Inc., warning letter, FDA, 9 May 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ neurosync- inc- 705489- 05092025 (read 25 November 2026) - 13Medline Industries, LP, warning letter, FDA, 25 March 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ medline- industries- lp- 723866- 03252026 (read 25 November 2026) - 14Xoran Technologies, LLC, warning letter, FDA, 9 December 2024. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ xoran- technologies- llc- 694229- 12092024 (read 25 November 2026) - 15Q9(R1) Quality Risk Management, guidance for industry, FDA, May 2023. https:/
/ www. fda. gov/ regulatory- information/ search- fda- guidance- documents/ q9r1- quality- risk- management (read 25 November 2026) - 16ICH Q9 Quality risk management, scientific guideline page, European Medicines Agency. https:/
/ www. ema. europa. eu/ en/ ich- q9- quality- risk- management- scientific- guideline (read 25 November 2026) - 17WHO guidelines on quality risk management, draft working document QAS/26.994, sections 3.8, 3.9 and 5.3, World Health Organization, September 2026. https:/
/ cdn. who. int/ media/ docs/ default- source/ mhp- hps/ 2026- 09- 18_ qas26_ 994_ qrm_ mna. pdf (read 25 November 2026)