1PUPSIT and the post-use test, side by side.
PUPSIT stands for pre-use post-sterilization integrity test. EU GMP Annex 1 point 8.87 asks for two integrity tests on a sterilizing grade filter that sterilizes a fluid. The first is on the sterilized assembly before use, “to check for damage and loss of integrity caused by the filter preparation prior to use”. The second is non-destructive, after use, before the filter is removed from its housing.1
| Point | PUPSIT | Post-use test |
|---|---|---|
| When | After sterilization, before filtration | After filtration, with the filter still in its housing |
| What it can catch | Damage from sterilization, transport, handling or installation | Damage from before or during filtration, unless the product has masked it |
| EU GMP Annex 1 | Expected for final product sterilizing filters whenever feasible | Expected |
| FDA 2004 guidance | Optional: it “can be performed prior to processing” | Expected: it “should be routinely performed post-use” |
The reason for both is masking. Annex 1 asks a risk assessment to consider whether the product could “alter integrity-testing values and therefore prevent the detection of a non-integral filter during a post-use filter integrity test”.1 A filter damaged before use and then partly blocked by product might pass afterwards. PUPSIT tests it before product can hide anything.
2Which filters need PUPSIT.
PIC/S answered this in PI 058-1, written by a joint PIC/S, EMA and WHO working group so that all three read Annex 1 the same way. It came into force on 28 October 2026 and adds no new requirements.2
| Filter | Tests expected | If a test fails | Source |
|---|---|---|---|
| Final sterilizing filter | PUPSIT whenever feasible, and a post-use test | Failed PUPSIT: not used. Failed post-use test: no evidence of a sterile filtration, investigated before certification. | 8.87; 3.2 |
| Redundant backup filter | PUPSIT if it might be used to support batch release | When the primary fails, the backup is tested after use, and supports release only if it passed its PUPSIT | 8.92; PI 058-1 |
| Filters validated as one | For the final product, PUPSIT whenever feasible; a post-use test on every filter in the system | One failure fails the whole system | 8.87; 8.91; PI 058-1 |
| Other sterilizing filter | Based on risk and the CCS, such as upstream bioburden control; PUPSIT not mandatory | Investigated as a non-conformity | PI 058-1; 3.2 |
| Critical gas or vent | Post-use test with the filter still in its housing | Investigated as a non-conformity before certification | 8.88; 6.19; 3.2 |
| Non-critical gas or vent | At set intervals, and at installation and before replacement when in place for long periods | Investigated as a non-conformity | 8.89; 3.2 |
2.1When PUPSIT can be left out.
Annex 1 recognizes that PUPSIT “may not always be possible after sterilisation due to process constraints”, with very small volumes as its example. An alternative is then acceptable after a thorough risk assessment, with controls that mitigate the risk of a non-integral filtration system.1 The EMA sterilization guideline agrees: the sterilized filter is tested before use “unless specifically justified and validated”.3
PI 058-1 sets three limits on that exception.2
- The constraint must not be “linked to design flaws”, and manufacturers should first consider process design changes that would make PUPSIT possible.
- The justification sits in the contamination control strategy, supported by “a comprehensive product and patient safety focused risk evaluation” and mitigation.
- It warns that “risk assessments should not be used inappropriately to justify the omission of PUPSIT”.
Annex 1 8.82 already expects the system to permit in-place testing of the final filter, “preferably as a closed system”.1 Read with PI 058-1, a missing test path is a design flaw to fix, not a constraint to justify.
2.2PUPSIT risk assessment template.
Use this when a final sterilizing filter can’t have PUPSIT.
| Section | What to record | Source |
|---|---|---|
| Filter and role | Position, product, batch size, and whether this is the final product sterilizing filter | PI 058-1 |
| The constraint | What in the process prevents PUPSIT, and evidence that it is not a design flaw | 8.87; PI 058-1 |
| Design options | Changes considered to make PUPSIT possible, such as an in-place closed test path, and why each was rejected | 8.82 vi; PI 058-1 |
| Filter sterilization | The validated method and evidence it does not damage this filter | 8.87 i |
| Supply chain | Contract sterilization, transport and the packaging that protects the sterilized filter | 8.87 ii; 2.5 vii, viii |
| Masking risk | Particle burden, and whether the product could alter post-use integrity values | 8.87 iii |
| Pre-filtration | Steps that remove particles and clarify the product before the final filter | 8.87 iii; 8.80 |
| Bioburden | The limit and recent results immediately before the final filter | 8.93; 10.3; EMA 4.1.5 |
| Mitigations | Each control, and the failure it detects or prevents | 8.87 |
| Patient safety | A conclusion focused on product and patient safety | PI 058-1 |
| CCS and dossier | Where the rationale sits in the CCS, when it is reviewed, and what the marketing authorization says | 2.6; PI 058-1; EMA 4.1.5 |
2.3Example: one justification that holds and one that doesn’t.
Both cases are invented. The test is whether the constraint comes from the process or the design.
| Case | What stops PUPSIT | Passes PI 058-1? |
|---|---|---|
| PET tracer, 15 mL batch | Wetting the filter and flushing the fluid out would dilute or lose a real share of a batch filtered minutes before dispatch | Can pass, as very small volume is Annex 1’s own example. Record supplier sterilization and transport controls (8.87 i, ii) and test every filter after use. |
| Bulk injection, 120 L | The single-use assembly has no downstream drain or vent, so a test would need an open aseptic connection | No. The constraint is the design. Change the assembly so the filter can be tested in place (8.82 vi). |
3What FDA and other regulators expect.
The US regulation is general: under 21 CFR 211.113(b), procedures to prevent microbiological contamination of sterile drugs “shall include validation of all aseptic and sterilization processes”.5 FDA’s 2004 aseptic processing guidance, still the current version, says integrity testing “can be performed prior to processing, and should be routinely performed post-use”, and that “Use of redundant sterilizing filters should be considered in many cases.”4
For a product made only for the US, PUPSIT is therefore a choice made on risk; the post-use test, its validation and the tester’s data are not. A human medicine supplied to the EU is held to Annex 1, a veterinary one since 16 July 2026 to the same test in Implementing Regulation (EU) 2025/2091,6 and either to what its marketing authorization says about pre-use testing.3
| Authority | The 2022 Annex 1 text |
|---|---|
| PIC/S | In its GMP guide since 25 August 2023 |
| WHO | The same 8.87, in its 2022 sterile GMP |
| Health Canada | GUI-0119, the PIC/S wording, since 1 April 2024 |
| Australia, TGA | Through PIC/S PE 009-17, for manufacture from 1 September 2025 |
| UK, MHRA | In the Orange Guide, which MHRA lists among its GMP and GDP guidance, in effect from 25 August 2023 |
4How PUPSIT is run.
- Design the train so the filter can be tested in place before and after filtration, with few aseptic connections after it and test methods that don’t affect product quality (8.82).1
- Sterilize the assembly by the validated process, on site or by the supplier, and keep its record (8.87 i).1
- Wet the membrane with the fluid the validated method names (8.85 i).1
- Test in place against the limit established in that fluid and correlated with bacterial retention (8.87).1,
3 - Remove the wetting fluid as validated, so it can’t affect the product (8.85 i).1
- Sample bioburden immediately before the final filter, or before the first filter in a redundant set-up (8.93). In most situations no more than 10 CFU/100 mL is acceptable.1,
3 - Filter, recording the time to filter a known volume and the pressure difference across the filter (8.86).1
- Test after use before the filter leaves its housing (8.87).1
- Review before certification, with the tester’s raw data and audit trail, investigating any non-conformity first (3.2).1
4.1Keeping the sterile side closed.
Steps 3 to 5 wet and pressurize a filter whose downstream side is already sterile, and Annex 1 lists pre- and post-use filter testing among the risks of single-use systems (8.132).1 A closed single-use assembly has three features.
- Wetting fluid supplied upstream, for example from a sterile bag, so nothing is opened to wet the membrane. If the fluid is not the product, its effect on the product is controlled (8.85 i).1
- A closed destination downstream: a waste bag with its own sterilizing grade vent, so fluid and test gas leave without opening the sterile side (8.82 ii).1
- The fewest connections between the final filter and filling, made with intrinsic sterile connectors or welds (8.82 iii; 8.128).1 PI 058-1 accepts a validated weld from qualified equipment as one, if the aseptic process simulation covers it and the system stays integral at every use.2
A fixed stainless steel train can be closed by design: sterilized in place (8.82 v), kept integral under positive pressure or behind a sterilizing vent filter until use (8.63), and tested in place (8.82 vi).1
5Example: a completed PUPSIT and post-use record.
This record is invented. The product, site, batch, filter and limits are illustrative; real limits come from your own filter validation.
| Field | Entry | Requirement |
|---|---|---|
| Batch | Quellotin 20 mg/mL injection, Batch 26-031, Fenmoor fill suite, 5 October 2026 | Batch record, 8.86 |
| Filter | F-2, final sterilizing filter, 0.2 µm, lot FL-4417; no redundant filter in this train | 8.79; 8.87 |
| Sterilization | Autoclave cycle AC-1188, validated load pattern 4, cycle passed | 8.87 i |
| Wetting fluid | Water for injections from a sterile bag upstream, as the filter maker recommends | 8.85 i; 8.82 vi |
| PUPSIT | Bubble point 3,780 mbar against a minimum of 3,450 mbar water-wet. Pass at 07:52, first attempt, raw trace attached. | 8.87; EMA 4.1.5 |
| Fluid removal | Downstream side drained to the vented waste bag, first 2.0 L of filtrate discarded, as validated | 8.85 i |
| Bioburden | 1 CFU/100 mL, sampled at the end of a 17-hour hold (validated maximum 24 hours); limit 10 CFU/100 mL | 8.93; 10.3; EMA 4.1.5 |
| Filtration | 120 L from 08:10 to 09:05: 55 minutes (validated maximum 90), 2.2 L/min on average, peak differential 0.9 bar (maximum 2.0) | 8.85 ii; 8.86 |
| Post-use test | Product-wet bubble point 3,320 mbar against a minimum of 3,100 mbar. Pass at 09:15, filter in housing. | 8.87; EMA 4.1.5 |
| Vent filter | VF-1 on the receiving vessel, water intrusion 0.12 mL/min against a maximum of 0.30. Pass. | 6.19; 8.88 |
| Tester audit trail | Tester IT-03, 07:30 to 09:20: two F-2 tests, both reported above; none aborted, repeated or deleted | Annex 11, 9; FDA Q&A 7 |
| Review | QA reviewed every result above, and the audit trail, with the batch record before certification | 3.2; 6.19; FDA Q&A 2 |
6What inspectors look at.
FDA’s compliance program for sterile drugs, 7356.002A, tells investigators to verify that production filters are identical to those used in validation, to “observe filter integrity testing to verify procedures are followed” and to “review investigations of any integrity test failures”.14 Annex 1 and PIC/S look for the same evidence.
- Bacterial retention validation with the product or a justified surrogate at a challenge of at least 10⁷ CFU/cm², and how the routine limit was derived from it (8.84; 8.87; EMA table 3).1,
3 - The filtration design, with the arrangement of components justified and the in-place test path shown (8.81; 8.82 vi).1
- Every test the integrity tester ran, matched against the batch record. The tester is a computerized system: Annex 11 expects its audit trail to be “regularly reviewed” (9), and FDA expects the people who review a record to review its audit trail with it, which for batch data means before release (Q&A 7 and 8).12,
13 - Each failure investigation, with its reasoning for including or excluding other batches (3.2; 211.192).1,
5 - The risk assessment and CCS entry wherever PUPSIT is omitted.2
This is what an unreviewed tester can hide.
On August 8, 2024, two operators performed four post-use sterilizing filter integrity tests for (b)(4) batch (b)(4). The four tests included three failing results and the final passing result. Your operators only reported the passing result.
Aspen Pharmacare Holdings Limited, FDA warning letter 320-25-46, 24 February 2025. FDA cited 211.68(b), on controls over computer systems, and asked for procedures “regarding the review of all critical electronic data (e.g., filter integrity testing)”.15
7When an integrity test fails.
The filter table above says what a failure means. Annex 1 contains no retest rule for filters, so the validated test process (8.87) has to settle in advance what happens after a failing result.1
- Invalid or failed. What makes a run invalid, such as an aborted test or a leak in the set-up, and the evidence needed to say so.
- Rewetting. Whether the filter may be rewetted or flushed again before a repeat, within the wetting conditions set in validation (8.85 i).1
- Maximum repeats. A fixed number, set in validation rather than decided at the skid.
- When an investigation starts. A confirmed failure is investigated before certification (3.2),1 and in the US “shall be thoroughly investigated, whether or not the batch has already been distributed” (211.192).5
- Every result kept. Invalidating a result needs “a valid, documented, scientifically sound justification”, and the original data stays in the batch record the quality unit reviews (FDA Q&A 2).13
Failed tests also end in recalls, such as AMIVAS’s Artesunate for Injection in 2021 after an air filter failed its integrity test (D-0002-2022)16 and six patient doses of Sofie’s Florbetaben F-18 in 2023 (D-1177-2023).17
8How it runs in Seal.
In Seal, sterile filtration binds the approved filtration design, product-specific validation, the assembly and its sterilization state, installation, PUPSIT, process conditions and post-use integrity into one evidence chain for each filter position. Each integrity test keeps its method, wetting, instrument, raw trace, result and any repeat. Where PUPSIT is not performed under an approved exceptional rationale, the risk assessment, alternative controls, scope and approval are recorded as such rather than shown as a passing test. Interventions and filter replacements divide the run into distinct filtration intervals, so a failed post-use test is bounded to the material and filled units it can affect, and integrity values are trended by filter lot, assembly and product so near-limit drift can prompt action before a failure.
For sterile injectables, the batch record is linked to environmental conditions, interventions, qualification, equipment state, materials, microbiology and release on one timeline, so a filter lot concern traces forward to each filling batch.
AQuestions and answers
What is filter flaw masking?
It is the risk that a filter damaged before use passes its post-use test because the product has partly blocked the damaged area. Annex 1 asks a risk assessment to consider whether the product could alter integrity test values and so hide a non-integral filter after use. Testing the filter before product reaches it removes that blind spot.
Is PUPSIT the same as a filter integrity test?
PUPSIT is one filter integrity test, defined by when it is done: after the filter is sterilized and before product passes through it. The same kinds of test, such as bubble point, diffusive flow, water intrusion or pressure hold, are used after use and on vent filters. Each is validated, and its limits are set in the fluid actually used.
Can the filter supplier’s release test replace PUPSIT?
No. PUPSIT exists to find damage caused after the supplier’s test, by sterilization, transport, storage and installation.
What is a PUPSIT skid?
It is the equipment that lets a filter be wetted, tested and drained in place: an integrity tester with its gas supply, valves, the wetting fluid feed and a closed waste container. The regulators care about what it does, not what it is called. The test it runs is validated, the sterile side is preferably kept closed while it runs, and the tester is a computerized system whose results and audit trail are reviewed before release.
Is PUPSIT done before every batch?
It is done on each sterilized filter assembly before that assembly is used, which in most processes means each batch. A filter that is sterilized again for another use is tested again, because the damage PUPSIT looks for comes from preparing the filter: sterilization, handling and installation.
References17 sources
References
- 1EudraLex Volume 4, Annex 1: Manufacture of Sterile Medicinal Products, C(2022) 5938 final, points 2.5, 2.6, 3.2, 6.19, 8.63, 8.79 to 8.93, 8.128, 8.132 and 10.3, European Commission, 22 August 2022, in operation since 25 August 2023. https:/
/ health. ec. europa. eu/ system/ files/ 2022- 08/ 20220825_ gmp- an1_ en_ 0. pdf (read 26 November 2026) - 2PI 058-1, Interpretation of the most important changes in the Annex 1 on the manufacture of sterile products (2022), sections 1, 2 and 4.8 (questions on 8.87, 8.92 and 8.128), PIC/S, adopted 16 October 2026, in force 28 October 2026. https:/
/ picscheme. org/ docview/ 11495 (read 26 November 2026) - 3Guideline on the sterilisation of the medicinal product, active substance, excipient and primary container, EMA/
CHMP/ CVMP/ QWP/ 850374/ 2015, section 4.1.5 and table 3, European Medicines Agency, 6 March 2019, in effect from 1 October 2019. https:/ / www. ema. europa. eu/ en/ documents/ scientific- guideline/ guideline- sterilisation- medicinal- product- active- substance- excipient- and- primary- container_ en. pdf (read 26 November 2026) - 4Guidance for Industry: Sterile Drug Products Produced by Aseptic Processing, Current Good Manufacturing Practice, section IX.B, FDA, September 2004. https:/
/ www. fda. gov/ media/ 71026/ download (read 26 November 2026) - 521 CFR Part 211, sections 211.113 and 211.192, eCFR, current to 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211 (read 26 November 2026) - 6Commission Implementing Regulation (EU) 2025/2091 on good manufacturing practice for veterinary medicinal products, Official Journal, 27 October 2025, Article 46 and Annex I, point VII.4.7.9, applying from 16 July 2026. https:/
/ eur- lex. europa. eu/ eli/ reg_ impl/ 2025/ 2091/ oj/ eng (read 25 November 2026) - 7WHO good manufacturing practices for sterile pharmaceutical products, WHO Technical Report Series No. 1044, Annex 2, point 8.87, World Health Organization, 2022. https:/
/ cdn. who. int/ media/ docs/ default- source/ medicines/ norms- and- standards/ guidelines/ production/ trs1044- annex- 2- gmp- for- sterile- pharmaceutical- products. pdf (read 26 November 2026) - 8Publication of updated Annex 1 to the Good manufacturing practices guide, Manufacture of sterile drugs (GUI-0119), Drug Establishment Licensing Bulletin 148, Health Canada, 24 August 2023. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ establishment- licences/ drug- establishment- licensing- bulletin/ annex- 1- good- manufacturing- practices- guide- manufacture- sterile- drugs- gui- 0119. html (read 26 November 2026) - 9Therapeutic Goods (Manufacturing Principles) Amendment (PIC/S Guide) Determination 2025, F2025L00990, Schedule 1, items 1 and 2, Federal Register of Legislation, registered 28 August 2025, commenced 1 September 2025. https:/
/ www. legislation. gov. au/ F2025L00990/ asmade/ 2025- 08- 28/ text/ original/ pdf (read 26 November 2026) - 10The MHRA Orange Guide March 2023 Update, Pharmaceutical Press, publisher of the Orange Guide, March 2023. https:/
/ www. pharmaceuticalpress. com/ knowledge- updates/ the- mhra- orange- guide- march- 2023- update/ (read 26 November 2026) - 11Medicines: good manufacturing practice and good distribution practice, MHRA and Department of Health and Social Care, GOV.UK, last updated 13 May 2024. https:/
/ www. gov. uk/ guidance/ good- manufacturing- practice- and- good- distribution- practice (read 26 November 2026) - 12EudraLex Volume 4, Annex 11: Computerised Systems, section 9, European Commission, in operation since 30 June 2011. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 13Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, questions 2, 7 and 8, FDA, December 2018. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 14Compliance Program 7356.002A, Sterile Drug Process Inspections, part III, section 3.8 (sterile filtration), FDA, implemented 11 September 2015. https:/
/ www. fda. gov/ media/ 75174/ download (read 26 November 2026) - 15Warning Letter 320-25-46 to Aspen Pharmacare Holdings Limited, MARCS-CMS 701671, item 3, FDA, 24 February 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ aspen- pharmacare- holdings- limited- 701671- 02242025 (read 26 November 2026) - 16FDA Enforcement Report, recall D-0002-2022, AMIVAS (US), LLC, Artesunate for Injection, Class II, report dated 13 October 2021, through openFDA. https:/
/ api. fda. gov/ drug/ enforcement. json?search=recall_ number:%22D- 0002- 2022%22 (read 26 November 2026) - 17FDA Enforcement Report, recall D-1177-2023, Sofie Co dba Sofie, Florbetaben F-18 Injection, Class II, report dated 4 October 2023, through openFDA. https:/
/ api. fda. gov/ drug/ enforcement. json?search=recall_ number:%22D- 1177- 2023%22 (read 26 November 2026)
