1Is there FDA-certified Part 11 software?
No. FDA’s 2024 questions and answers on electronic systems in clinical investigations state that “FDA does not certify electronic systems and methods used to obtain electronic signatures”, and that signatures meeting Part 11 are acceptable “regardless of the particular technology or brand used”. FDA also “does not perform preliminary evaluations of electronic systems” against Part 11: “These systems will be evaluated during an inspection.”1 A vendor’s compliance badge describes features, not a status a regulator granted.
Part 11 places its controls on “persons who use closed systems”, not on products.2 Some duties only your organization can carry out:
- confirming that people who develop, maintain or use the system have the education, training and experience for their tasks (11.10(i));
- written policies that hold people accountable for actions under their electronic signatures (11.10(j));
- verifying a person’s identity before issuing an electronic signature (11.100(b));
- certifying to FDA that your electronic signatures are the legally binding equivalent of handwritten ones (11.100(c)).2
Validation is shared: the vendor tests its product, and you validate each CGMP workflow you build in it, which FDA calls “an intended use of a computer system to be checked through validation”.3
2What a Part 11 compliant system should be able to do.
Check these in a sandbox set up like your process, not only in a questionnaire. PIC/S expects systems to be “appropriately evaluated for data integrity controls prior to purchase”, and the MHRA says accepting vendor-supplied validation data “in isolation of system configuration and users intended use is not acceptable”.4,
| Control | How to check it | What good looks like | Rule |
|---|---|---|---|
| Audit trail | Change a value; as administrator, try to edit the entry, switch the trail off or change the clock | Who, when, old and new value; users can’t change it; administrator changes, clock included, are recorded | 11.10(e); MHRA 6.13; draft Annex 11 12.3 |
| Reason for change | Change GMP data without giving a reason | The system asks for one | Annex 11 section 9 |
| Audit trail review | Pull one record’s trail and filter it to GMP changes | A readable list or exception report a reviewer can check with the record before approval | Annex 11 section 9; FDA data integrity Q&A 7 and 8; MHRA 6.13 |
| Signature display | Sign; view on screen and as a printout | Printed name, date and time, and meaning, in both | 11.50 |
| Signature link | Copy a signature; edit a signed record | Can’t be copied; the edit is blocked or shows as unsigned | 11.70; draft Annex 11 13.8 |
| Signing components | Sign twice in a session, then after logout | ID and password first, then at least the password; both after logout | 11.200(a) |
| Unique accounts | Export users and roles | One person per account; no shared account can change data | 11.300(a); FDA data integrity Q&A 5 |
| Access changes | Grant, change and remove a role, then find each record; compare the user list with current staff | Each change recorded; the list is current, with no leavers | Annex 11 section 12.3; FDA data integrity Q&A 4; PIC/S 9.3 |
| Password controls | Enter wrong passwords; read the expiry rules; disable a token reported lost | Failed attempts limited and reported to security at once; passwords periodically checked or revised; lost tokens disabled | 11.300(b) to (d) |
| Authority checks | Sign or edit without the role, or as a disabled user | Blocked | 11.10(d) and (g) |
| Sequence checks | Approve before review is complete | Blocked | 11.10(f) |
| Administrators | Ask who holds them, vendor staff included | Named, independent of record content; vendor access in writing | FDA data integrity Q&A 4; Annex 11 section 3.1 |
| Copies | Export a record with audit trail and signatures | Complete, electronic and human-readable | 11.10(b); Annex 11 section 8 |
| Retention and exit | Restore a backup; ask how data leaves at contract end | Records and metadata retrievable for the retention period | 11.10(c); Annex 11 sections 7 and 17; MHRA 6.20 |
| Vendor releases | Ask how releases reach you and whether you can test them first; change a setting and find its record | Release notes and test results per update; a test window before production; setting changes controlled and recorded | Annex 11 section 10; draft Annex 11 7.5(ix); FDA CSA guidance, Example 4 |
| Supplier evidence | Ask for its quality system, audit reports and right to audit | Supplier assessed; its quality and audit information available to inspectors | Annex 11 sections 3.2, 3.4 and 4.5 |
Passing shows a product can be configured to comply, not that your system does.2,
A cloud system is not open just because a vendor hosts it. Part 11 calls a system closed when access is controlled by the people responsible for the content of its records, and open when it isn’t (11.3(b)(4) and (9)). A vendor-hosted eQMS is closed if your organization decides who gets access, vendor staff included. If it can’t, 11.30 adds measures “such as document encryption and use of appropriate digital signature standards”.2
3Is SharePoint, Box, Excel or DocuSign Part 11 compliant?
Not as products, either way: it depends on the record, whether people rely on the electronic version, and your configuration.
| Question | If yes | If no |
|---|---|---|
| Required by an FDA rule, or submitted to FDA? | Next question | Not a Part 11 record; document the decision |
| Relied on electronically? | Part 11 applies; next question | A complete signed printout is the record, if it is the original or a true copy; write that down |
| Passes the checks above? | Validate it for this use | Add controls, choose another tool or keep that record on paper |
Under FDA’s 2003 guidance, a record a predicate rule requires is a Part 11 record when it is kept electronically instead of on paper, or kept both ways with the electronic version relied on. FDA recommends that you “determine in advance whether you plan to rely on the electronic record or paper record” and document it.9 A printout can be the record only if it is the original or a true copy: for dynamic data, such as an FT-IR spectral file, FDA says a printout “would not satisfy CGMP requirements to retain original records or true copies”.3 Part 11 also covers electronic records submitted to FDA “even if such records are not specifically identified in agency regulations”.2
3.1Cases that catch teams out.
- A word processor used only to draft SOPs is FDA’s own example of a system for which validation “would not be important”.9
- A spreadsheet that calculates GMP results is still a computer system under 211.68, even when the result is printed and signed.10 See spreadsheet validation.
- In SharePoint, test whether version history records who changed what and when, and whether site owners can delete it. See eQMS or document management system.
- An e-signature service passes only if the signed file shows each signature’s meaning and stays linked to it. See DocuSign and Adobe Sign under Part 11.
- FDA has exercised enforcement discretion over Part 11’s validation, audit trail, retention and copying requirements since 2003, but enforces the predicate rules and the other controls, including signatures and the 11.300 password controls.9 Don’t buy on the strength of it: see the letters below.
4Example: checking a vendor’s claim before purchase.
This record is invented: the company, product, people and numbers are illustrative. Corrabell Therapeutics shortlisted Orvelline, a cloud eQMS whose datasheet claimed Part 11 compliance. On 15 and 16 September 2025 its QA and IT leads ran eight of the checks above in a vendor sandbox set up with Corrabell’s SOP approval workflow.
| Check | Result | Action | Rule |
|---|---|---|---|
| Audit trail | Pass: who, when, old and new value; administrator couldn’t edit it | None | 11.10(e) |
| Reason for change | Gap: optional by default | Made mandatory; setting recorded in the configuration specification | Annex 11 section 9 |
| Signature display | Gap: the PDF showed name, date and time, not meaning | Vendor fix due in release 4.2, written into the contract | 11.50(b) |
| Signature link | Pass: an edited SOP showed as unsigned | None | 11.70 |
| Signing components | Pass: ID and password first, password at each later signing | None | 11.200(a) |
| Sequence checks | Pass: approval blocked until review was complete | None | 11.10(f) |
| Administrators | Gap: a standing vendor support account with full rights | Replaced by named, time-limited access Corrabell grants, in the quality agreement | 11.10(d); Annex 11 section 3.1 |
| Copies | Pass: export with audit trail and signatures, as PDF and CSV | Exit terms added to the contract | 11.10(b) and (c) |
Of eight checks, five passed and three found gaps: one closed by configuration, one by a vendor fix and one by contract. Orvelline’s SOC 2 Type 2 report and release test summaries went into Corrabell’s supplier assessment as evidence for infrastructure and standard screens, not for Corrabell’s workflow. Corrabell signed on 3 October 2025. Its validation, approved in February 2026, retested the signature printout on release 4.2: it passed.
5What GMP adds, and what inspectors cite.
Part 11 covers records and signatures. For drug manufacture, 21 CFR 211.68(b) adds controls over any “computer or related systems”: only authorized people change master or other records, input and output are checked for accuracy, and backup data must be “exact and complete” and secure from alteration or loss.10 FDA reads those systems broadly, to include software, cloud infrastructure, personnel and SOPs.3
Recent warning letters show gaps a pre-purchase check would have found:
“For example, your ultraviolet-visible (UV-vis) and infrared (IR) spectroscopy equipment was found to lack audit-trail functionality, and your analysts were able to alter and delete data, files, and folders. This is a repeat observation from your 2016 inspection.” Guangxi Yulin Pharmaceutical Group Co. Ltd., warning letter, 30 September 202511
“Our investigator documented an uncontrolled password for the (b)(4) meter affixed to a laptop by a sticker used for processing (b)(4) meter test results. In addition, the data from the (b)(4) meter is automatically deleted when the laptop is shutdown, so no records were available to show that your QU reviewed this data.” GC America, Inc., warning letter, 14 May 202612
“The review of aborted air sampler data was limited by its unacceptable data retention capability of only 115 entries.” Somerset Therapeutics Private Limited, warning letter, 4 September 202513
FDA cited Guangxi Yulin under 211.68(b), and GC America under 211.22(d) and 211.68(b); neither letter cites Part 11.11,
6How FDA, EU, MHRA and PIC/S differ.
| Topic | FDA | EU GMP, MHRA and PIC/S |
|---|---|---|
| Certification | None; systems are evaluated at inspection1 | No scheme; users review supplier documents against their requirements (Annex 11 3.3)6 |
| Vendor evidence | Device makers may leverage it after a risk-based vendor assessment8 | Not on its own, say MHRA and PIC/S5, |
| Audit trail | 11.10(e), under enforcement discretion since 2003;9 cited under 211.68(b)11 | Based on risk, with a reason for each GMP change (Annex 11 section 9);6 users can’t switch it off (MHRA 6.13)5 |
| Signatures | Name, date and time, meaning; one letter can cover the organization2, | Time and date, permanently linked (Annex 11 section 14);6 no letter |
7What changed, 2023 to 2026.
- March 2023. 11.100(c) now lets the handwritten signature certification be submitted to FDA in electronic or paper form.2
- October 2024. Revision 1 of FDA’s clinical investigation questions and answers listed what to agree with and check in IT service providers.1
- February 2026. For device makers, the QMSR took effect on 2 February, and FDA’s final computer software assurance guidance of 3 February replaced the September 2025 version. It says Part 11’s validation discretion “expressly does not apply” to the software validation that ISO 13485 subclauses 4.1.6, 7.5.6 and 7.6 require.8 See CSV and CSA.
- EU, still draft. The draft Annex 11 consulted on from 7 July to 7 October 2025 would require audit trails “enabled and locked at all times” (12.3), multifactor authentication for remote access to critical systems (11.6), automatic logout after a set period of inactivity (11.8) and contracts with an exit strategy (7.5).7,
14 On 26 November 2026 EudraLex still lists the January 2011 Annex 11.15 Our judgment: ask vendors about these now; a system bought today will likely still run when the revision applies.
8How it runs in Seal.
Seal connects records with version history, audit trails and role-based access, and an electronic signature keeps the signer, the time and the meaning of the approval, linked to the record. The validation and AI assurance guide sets out what Seal provides, the platform controls, assurance capabilities and supplier evidence for the version in scope, and what your team owns: the intended use, risk acceptance and production approval, including qualification of integrations and migrated data, user procedures and training. It also lists what your security team should review, from support access and backup recovery to data export. When your quality and IT reviewers assess Seal as a supplier, they can request its security and validation evidence or work through the free Part 11 and Annex 11 readiness checklist.
As the regulatory overview puts it, no software alone makes an operation compliant, and it suggests agreeing an exit plan before production use: the records, formats, attachments, version history, signatures and audit evidence to export. On the Seal platform, every change runs its tests before your team releases it, and failures stay on the record.
AQuestions and answers
Does Part 11 require multi-factor authentication?
Not by that name. Part 11 limits access to authorized people and requires electronic signatures that are not biometric to use at least two distinct components, such as an ID and a password. The EU’s 2025 draft of Annex 11 would require multifactor authentication for remote access to critical systems from outside controlled perimeters, but it is still a draft.
What if a system already in use has no audit trail?
Assess what could be changed without trace and add controls for those records, such as restricted access, administrative procedures and second-person checks of the data. PIC/S accepts alternative arrangements for very simple systems, but asks companies to work toward buying or upgrading to software with an electronic audit trail. Plan the replacement and record the risk until then.
Can software we build in-house be Part 11 compliant?
Yes. Part 11 sets the same controls whoever wrote the software, and FDA certifies neither bought nor built systems. The difference is the evidence: with no supplier to assess, you hold the whole development record yourself. EU GMP Annex 11 expects bespoke or customized systems to have a process for formally assessing and reporting quality and performance at every stage of their life cycle, so keep design, testing and change records from the start.
Is a validated system enough to be GMP compliant?
No. The records it holds must still meet the predicate rules, such as production and control records reviewed and approved by the quality unit before release, and the procedures around the system must be followed. The MHRA notes that a validated system may lower data integrity risk but does not eliminate it.
References15 sources
References
- 1Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers, guidance for industry, Revision 1, FDA, October 2024, questions 16, 17, 28 and 29 and section III.C. https:/
/ www. fda. gov/ media/ 166215/ download (read 26 November 2026) - 221 CFR Part 11, sections 11.1, 11.3, 11.10, 11.30, 11.50, 11.70, 11.100, 11.200 and 11.300, as amended 2 March 2023, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 11 (read 26 November 2026) - 3Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, FDA, December 2018, questions 1.f, 3, 4, 5, 7, 8 and 10. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 4PI 041-1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, PIC/S, 1 July 2021, sections 9.3 and 9.6. https:/
/ picscheme. org/ docview/ 4234 (read 26 November 2026) - 5‘GXP’ Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018, sections 4.5, 6.13, 6.19 and 6.20. https:/
/ assets. publishing. service. gov. uk/ media/ 5aa2b9ede5274a3e391e37f3/ MHRA_ GxP_ data_ integrity_ guide_ March_ edited_ Final. pdf (read 26 November 2026) - 6EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, in operation since 30 June 2011, sections 3, 4.5, 7, 8, 9, 10, 12, 14 and 17. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 7EudraLex Volume 4, Annex 11: Computerised Systems, draft for consultation, European Commission, consultation opened 7 July 2025, sections 7.5, 11.6, 11.8, 12.3 and 13.8. https:/
/ health. ec. europa. eu/ document/ download/ 40231f18- e564- 4043- 94de- c031f813d38b_ en?filename=mp_ vol4_ chap4_ annex11_ consultation_ guideline_ en. pdf (read 26 November 2026) - 8Computer Software Assurance for Production and Quality Management System Software, guidance for industry and FDA staff, FDA, 3 February 2026, sections V.A(5) and V.B and Appendix A, Example 4. https:/
/ www. fda. gov/ media/ 188844/ download (read 26 November 2026) - 9Guidance for Industry: Part 11, Electronic Records; Electronic Signatures, Scope and Application, FDA, August 2003, sections III.A, III.B.2 and III.C.1. https:/
/ www. fda. gov/ media/ 75414/ download (read 26 November 2026) - 1021 CFR Part 211, section 211.68, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211 (read 26 November 2026) - 11Guangxi Yulin Pharmaceutical Group Co. Ltd., warning letter, MARCS-CMS 710422, FDA, 30 September 2025, item 4. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ guangxi- yulin- pharmaceutical- group- co- ltd- 710422- 09302025 (read 26 November 2026) - 12GC America, Inc., warning letter, MARCS-CMS 727602, FDA, 14 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ gc- america- inc- 727602- 05142026 (read 26 November 2026) - 13Somerset Therapeutics Private Limited, warning letter, MARCS-CMS 711340, FDA, 4 September 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ somerset- therapeutics- private- limited- 711340- 09042025 (read 26 November 2026) - 14Stakeholders’ Consultation on EudraLex Volume 4, Good Manufacturing Practice Guidelines: Chapter 4, Annex 11 and New Annex 22, European Commission, open 7 July to 7 October 2025. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 4- annex_ en (read 26 November 2026) - 15EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026)