1What review by exception means.
In a review by exception, QA reviews a batch record through a report that brings forward the entries needing judgment, instead of reading every line. PIC/S and MHRA define the tool the same way: an exception report is “a validated search tool that identifies and documents predetermined ‘abnormal’ data or actions” that need further attention or investigation by the data reviewer.1,
| Point | Line-by-line review | Review by exception |
|---|---|---|
| What the reviewer reads | Every entry | The exceptions, the parts of the record they touch, and whatever the report doesn’t check |
| Depth of review | The same for every batch | Can vary with the exceptions and with experience of the process |
| What QA approves | The whole record, under 211.192 | Unchanged, and the record stays open to the reviewer |
| Who releases | The quality unit; a QP in the EU | Unchanged |
2Do FDA and EU inspectors accept it?
In the EU, yes, in writing. EMA’s answer to “Is a risk-based review of electronic data acceptable?” is: “Yes. The principles of quality risk management may be applied during the review of electronic data and review by exception is permitted, when scientifically justified.” The automated system and its batch exception report must be tested and validated against business and GMP requirements.3
In the US, neither 21 CFR Part 211 nor FDA’s 2018 data integrity guidance mentions it: a search of both full texts finds the word “exception” nowhere.6,
| Source | What it says | Status |
|---|---|---|
| EMA Q&A 8 | Validated audit trail exception reports can replace “an onerous line-by-line review” | EU guidance |
| Chapter 4, 4.20 | For a validated, continuously monitored process, reports “may be limited to compliance summaries and exception/ out- | EU GMP, since 2011 |
| MHRA, 2018, 6.15 | Validated, locked reports “may reduce the checks required to assure the integrity of the data” | UK guidance |
The 2025 drafts of Chapter 4 and Annex 11, consulted on until 7 October 2025 and not yet in force,9,
3What the exception report should flag.
The rules come from a risk assessment, made during qualification, of which data and audit trail entries in each system are GMP-relevant and how critical each is. Critical audit trails are reviewed with the rest of the operation’s records before release; non-critical ones can be reviewed at system reviews at a set frequency.1 Audit trail review covers how.
| Data or audit trail | Criticality | How it is reviewed | Rule |
|---|---|---|---|
| Fill weights, B-07 | Critical: the dose in each vial | Each value checked against its limits; any change flagged before release | FW-01; CD-01 |
| Filler set points | Critical process parameters | Edits during a batch flagged; edit rights limited | AT-03 |
| Stopper bowl alarm | Critical: affects stoppering | Each acknowledgment and its comment flagged | AL-02 |
| Log-ins and log-outs | Not critical | Periodic system review, not each batch | None |
Each row below except alarms, which rest on the draft Annex 11, is a check that GMP rules or guidance already ask someone to make, so one the report leaves out has to be made another way.
| Exception | Flag when | Source |
|---|---|---|
| Out of limits | An in-process result, process parameter or yield is outside its limit | 211.188(b)(5), (7); 211.192; Chapter 4, 4.20 |
| Changed data | A value changed after entry, with who, when and why | Annex 11, 8.2 and 9; FDA Q&A 7 |
| Manual entry | Critical data typed rather than captured, needing its accuracy check | Annex 11, 6 |
| Missing signature | A significant step lacks its performer or checker | 211.188(b)(11); 211.68(c) |
| Alarms | A critical alarm was acknowledged, with its comment | Draft Annex 11, 8.4 and 8.7 |
| Deviations | Any deviation, investigation or OOS linked to the batch, open or closed | ICH Q7, 6.72; Annex 16, 1.7.16 |
| Audit trail | Anomalous or unauthorized activity, deletions, or data generated but not reported | PIC/S 9.6 and 9.8 |
| Outside the system | Attachments, paper logs and results held in other systems | EMA Q&A 16; draft Chapter 4, 4.85 |
4Example: an exception report and its review.
This record is invented.
Quellotin 20 mg/mL injection, Batch 26-034, was filled at the Fenmoor fill suite on 12 October 2026, with 11,400 vials theoretical. Who decided each flag follows the rule after the table.
| Flag | What the system found | Disposition | Decided by |
|---|---|---|---|
| Manual entry | Balance B-07 interface down 10:30 to 11:15; nine fill weights typed, each second-person verified | Matched the nine attached printouts. Accepted. | Production supervisor |
| Changed data | One entry changed from 10.65 g to 10.56 g at 10:44, reason “typing error” | Printout reads 10.56 g. Accepted. | Production supervisor |
| Out of limits | Rule FW-01: fill weight 10.24 g at 13:20, lower limit 10.30 g | DEV-26-0588: the 600 vials filled since the 13:00 pass rejected, pump recalibrated. Closed 21 October. | QA |
| Alarm | Stopper bowl low level, acknowledged four times, each with a comment | Within procedure, no product impact. Accepted. | Production supervisor |
| Audit trail | Stopper insertion set point edited and restored within three minutes from an engineering account | Not authorized during a batch. DEV-26-0592, no product impact found. Closed 28 October. | QA |
| Outside the system | Paper intervention log; sterility result in the laboratory information management system (LIMS) | Log checked entry by entry: 14 interventions, all of permitted types. Sterility result reviewed in LIMS: complies. | QA |
| Yield, no flag | 10,612 good vials of 11,400: 600 rejected under DEV-26-0588, 188 at visual inspection | 93.1% against limits of 90.0% to 101.0% | QA |
The QA reviewer opened the record around each exception and signed on 30 October: “Review complete. Six exceptions, all dispositioned. Deviations DEV-26-0588 and DEV-26-0592 closed. No other issues found.” MHRA expects a data review record to carry “a positive statement regarding whether issues were found or not”, the date and the reviewer’s signature.2 The QP then certified the batch.5
4.1Who may close which exception.
PIC/S has records of critical steps checked in production, then reviewed and approved by the quality unit before release, while “Batch production records of non-critical process steps is generally reviewed by production personnel according to an approved procedure” (8.8).1 FDA lets “a person directly supervising or checking information” review some activities, but the quality unit approves all records, audit trails included (Q&A 7).7
- Production may close an exception whose response an approved procedure already defines, such as an acknowledged alarm or a manual entry verified against its printout. QA still sees it in the record.
- QA decides anything outside a limit or a procedure, any unauthorized or unexplained change, and anything the report can’t check.
5What an exception report can’t see.
An exception report checks only the data the system holds. It cannot flag an event nobody recorded, a change made outside the audit trail, or a result left in an instrument. Inspectors have found all three.
For example, the batch record for (b)(4) injection (b)(4) mg/(b)(4)ml, on February 6, 2026, documented that 12 interventions had occurred during filling. However, records indicated approximately 200 interventions were conducted.
Jabil Inc., FDA warning letter 320-26-120, 27 August 2026.17
Your quality assurance employee instructed your software vendor to make changes to your electronic batch record which were not captured in the audit trail or managed through your quality system.
Intas Pharmaceuticals Limited, FDA warning letter 320-26-57, 30 March 2026.18
Your Production and QU did not review electronic raw data and audit trails to ensure data integrity prior to batch release. Your operators obtained and printed a passing result after multiple production filter integrity tests failed.
Aspen Pharmacare Holdings Limited, FDA warning letter 320-25-46, 24 February 2025.19
Whatever the report does not check needs another defined route.
| Part of the record | How to review it | Source |
|---|---|---|
| Paper and hybrid parts | By a defined reviewer, every field checked for completion and against its acceptance criteria, under a written procedure for the hybrid record | Draft Chapter 4, 4.85; PIC/S 8.8 |
| Other systems’ data | Through that system’s own validated review, or as the original electronic data, not from printouts alone | EMA Q&A 16 |
| Unaudited periods | Extra controls, because changes there are invisible to the reviewer | EMA Q&A 8 |
6How to validate the exception report.
- Assess the data, as in the criticality example above.1
- Settle who owns the rules. Process and system owners write them with IT (Annex 11, 2), and QA approves them, as it approves procedures that affect quality (211.22(c)).15,
6 A vendor can supply rules and scripts, but the regulated user “is fully accountable” (draft 9.9).12 - Write each rule as a requirement traceable through the life cycle (Annex 11, 4.4). FDA treats each CGMP workflow as an intended use to validate.15,
7 - Test with seeded exceptions, just outside and just inside each limit, as in the example below.
- Build the experience EMA refers to. Compare what the report flagged with what a full review found on the first batches, and record the result before reducing the depth of review.3
- Control changes. Rules are configuration: change them only through change control, by authorized people, and retest (Annex 11, 10; 211.68(b)).15,
13 - Check it in self-inspection with a risk-based sample of logs and audit trails behind the routine review, which PIC/S asks for where data is “reviewed manually or by a validated ‘exception report’” (5.6.2).1
6.1Example: one rule, specified and tested.
Invented, and the rule that caught the low fill weight above. The draft Annex 11 asks for test cases traceable to requirements (9.5), executed scripts as evidence (9.4) and “boundary and negative testing” (9.6).12
| Field | Entry |
|---|---|
| Rule | FW-01, fill weight below lower limit |
| Requirement | URS-EBR-112: flag any fill weight below the master record’s lower limit, with time, balance and vials since the last pass |
| Limit source | Master record MR-QLT-20 version 4: 10.30 g, compared as “less than” |
| Test FW-01-T1 | Seeded 10.29 g: flag raised, showing time, balance and vials since the last pass. Pass. |
| Test FW-01-T2 | Seeded 10.30 g: no flag. Pass. |
| Test FW-01-T3 | Seeded 10.29 g typed by hand during a simulated balance outage: FW-01 and the manual-entry flag both raised. Pass. |
| Executed | Script with a screenshot of each result, run by a validation engineer on 14 September 2026; approved by QA on 16 September 2026 |
| Trace | URS-EBR-112 to functional specification 4.3.2 to tests T1 to T3 |
7How it runs in Seal.
In Seal’s electronic batch records, changed data, failures, alarms, overrides, missing evidence, open samples and deviations are brought to QA as the batch runs, and every instruction, value and signature stays available, so review by exception directs attention without hiding the batch. Each controlled value has a range and a configured response, from continuing or requiring verification to placing the batch on hold or opening a deviation, and the original value and the system’s evaluation stay visible even when the process continues.
Quantify the QA review hours tied up in line-by-line checks with the batch review time estimator. Neil, Seal’s AI agent, can prepare a batch for review: it checks execution, laboratory and quality evidence against your batch-review requirements and drafts a source-linked brief with missing evidence requested. In batch release, times and limits are calculated in code rather than judged by the model. QA disposes and the QP certifies; Neil never releases a lot or changes the record.
AQuestions and answers
Can review by exception be used for paper batch records?
Not as EMA and PIC/S describe it, because an exception report is a validated search tool run by a computerized system. A paper record is reviewed by checking that every field is complete and compared with its acceptance criteria and that calculations are verified, with the extent of the second check set by how critical the data is. In a hybrid record the paper parts get that review, under a written procedure, because the report cannot check them.
Can review by exception be used for laboratory data?
Yes, where scientifically justified and once the system and its report are validated. EMA describes exception reporting as a tool for reviewing electronic data such as, but not limited to, electronic batch records, and PIC/S names review by exception as one way to focus audit trail review. In the US, original laboratory records still need the second-person review that 211.194(a)(8) requires, so the report directs that review rather than replacing it.
What happens if the exception report misses something?
Treat it as a failure of a validated system. Investigate it as a deviation, extend the investigation to the other batches the same rule covered, correct the rule through change control and retest it before relying on it again. In the US, an unexplained discrepancy has to be investigated, and the investigation extends to other batches that may have been affected.
How many batches before the depth of review can be reduced?
EMA, PIC/S and MHRA guidance set no number. EMA links the depth of review to the exceptions and to confidence and experience with the process, and permits review by exception only when it is scientifically justified. Set the number in the validation plan, compare the report with a full review on those batches, and reduce the depth only once the report caught everything the full review found, with that comparison recorded.
Can a sponsor rely on a CDMO’s review by exception?
Only on terms both sides have written down. MHRA expects the responsibilities for data review to be documented and agreed where the review is done by an organization other than the one that generated the data, and warns that summary reports may leave out critical supporting data. The sponsor should know which checks the CDMO’s system makes, what its exception reports show and how it handles them.
References19 sources
References
- 1PI 041-1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, sections 5.6.2, 8.8, 9.6, 9.8 and glossary, PIC/S, 1 July 2021. https:/
/ picscheme. org/ docview/ 4234 (read 26 November 2026) - 2‘GXP’ Data Integrity Guidance and Definitions, Revision 1, sections 6.13 and 6.15, MHRA, March 2018. https:/
/ assets. publishing. service. gov. uk/ media/ 5aa2b9ede5274a3e391e37f3/ MHRA_ GxP_ data_ integrity_ guide_ March_ edited_ Final. pdf (read 26 November 2026) - 3Guidance on good manufacturing practice and good distribution practice: questions and answers, data integrity, questions 8, 16 and 17, European Medicines Agency. https:/
/ www. ema. europa. eu/ en/ human- regulatory- overview/ research- development/ compliance- research- development/ good- manufacturing- practice/ guidance- good- manufacturing- practice- good- distribution- practice- questions- answers (read 26 November 2026) - 421 CFR 211.192, Production record review, eCFR, as of 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211/ subpart- J/ section- 211. 192 (read 26 November 2026) - 5EudraLex Volume 4, Annex 16: Certification by a Qualified Person and Batch Release, sections 1.7 and 1.7.16, European Commission, 12 October 2015, in operation since 15 April 2016. https:/
/ health. ec. europa. eu/ document/ download/ 20c41532- 33d5- 4635- ae80- 8735d3d09fe0_ en?filename=v4_ an16_ 201510_ en. pdf (read 26 November 2026) - 621 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, full text, eCFR, as of 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211 (read 26 November 2026) - 7Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, questions 3 and 7, FDA, December 2018. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 8EudraLex Volume 4, Part I, Chapter 4: Documentation, section 4.20 with its note, European Commission, in operation since 30 June 2011. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ chapter4_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 9Stakeholders’ Consultation on EudraLex Volume 4, Good Manufacturing Practice Guidelines: Chapter 4, Annex 11 and New Annex 22, European Commission, open 7 July to 7 October 2025. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 4- annex_ en (read 26 November 2026) - 10EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026) - 11Chapter 4: Documentation, draft for consultation, sections 4.36 and 4.85, European Commission and PIC/S, July 2025. https:/
/ health. ec. europa. eu/ document/ download/ fa336a6e- 753b- 46fc- b53b- 9178bacd8878_ en?filename=mp_ vol4_ chap4_ consultation_ guideline_ en. pdf (read 26 November 2026) - 12Annex 11: Computerised Systems, draft for consultation, sections 8.4, 8.7, 9.4 to 9.6, 9.9 and 12.8, European Commission and PIC/S, July 2025. https:/
/ health. ec. europa. eu/ document/ download/ 40231f18- e564- 4043- 94de- c031f813d38b_ en?filename=mp_ vol4_ chap4_ annex11_ consultation_ guideline_ en. pdf (read 26 November 2026) - 1321 CFR 211.68, Automatic, mechanical, and electronic equipment, eCFR, as of 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211/ subpart- D/ section- 211. 68 (read 26 November 2026) - 1421 CFR 211.188, Batch production and control records, eCFR, as of 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211/ subpart- J/ section- 211. 188 (read 26 November 2026) - 15EudraLex Volume 4, Annex 11: Computerised Systems, sections 2, 4.4, 6, 8.2, 9 and 10, European Commission, in operation since 30 June 2011. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 16ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, section 6.72, ICH, Step 4, 10 November 2000. https:/
/ database. ich. org/ sites/ default/ files/ Q7%20Guideline. pdf (read 26 November 2026) - 17Warning Letter 320-26-120 to Jabil Inc., MARCS-CMS 731037, FDA, 27 August 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ jabil- inc- 731037- 08272026 (read 26 November 2026) - 18Warning Letter 320-26-57 to Intas Pharmaceuticals Limited, MARCS-CMS 721151, FDA, 30 March 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ intas- pharmaceuticals- limited- 721151- 03302026 (read 26 November 2026) - 19Warning Letter 320-25-46 to Aspen Pharmacare Holdings Limited, MARCS-CMS 701671, item 3, FDA, 24 February 2025. https:/