1Deviation report template.
Copy these sections; neither 21 CFR 211.192 nor EU GMP sets a layout.
| Section | What to write | Basis |
|---|---|---|
| Header | Number, product, batch, site, equipment, dates raised and closed, owner and investigator, classification | 21 CFR 211.192: a written record1 |
| Summary and background | Event, cause, impact, recommendation and CAPA, written last; the process step, equipment and key documents | None: a reader who was not there needs it |
| 1. Event | What should have happened (procedure, version, step, limit), what did, when, where, who found it | 21 CFR 211.100(b): recorded and justified;2 FDA out- |
| 2. Immediate actions | What was held or stopped, by whom, when; who authorized any work under hold, within what hold time | FDA 2006: segregate nonconforming product;4 EU GMP 8.14: risk-reducing actions5 |
| 3. Scope | Batches, products, equipment and materials that could be affected; where the review starts and stops, and why | 21 CFR 211.192: other batches and products;1 FDA 2026 draft: justify exclusions6 |
| 4. Investigation | Records, interviews, tests and data; each hypothesis with the evidence for and against | FDA 2026 draft: “investigating each potential cause individually”6 |
| 5. History | Whether a similar event happened before, what was done then, why it came back | FDA OOS guidance: “has occurred previously”;3 EU GMP 8.125 |
| 6. Cause | Root cause, or the most likely cause and why it is unconfirmed; contributing causes; any human error justified | EU GMP 1.4(xiv)7 |
| 7. Impact and decision | Effect on each batch, with evidence; data integrity and validated state; the recommendation, who decides, any report due | 21 CFR 211.22(a);8 EU GMP Annex 16, 3.19 |
| 8. CAPA | Corrections, corrective and preventive actions, each with owner, due date and effectiveness criteria | EU GMP 1.8(vii); 1.4(xiv): effectiveness “monitored and assessed”7 |
| 9. Conclusion and sign-off | Conclusions and follow-up; prepared, reviewed and approved, each signed and dated, approval by the quality unit | 21 CFR 211.192: “the conclusions and followup”;1 21 CFR 211.22(a)8 |
EU GMP 8.12 and 8.14 cover quality defects; apply them to other deviations by analogy. Whoever found the event writes sections 1 and 2 that day; QA classifies it. The investigator, with subject experts, writes 3 to 6 and proposes 8; the quality unit approves 7 to 9 and decides the batch.8
Define the causes in your procedure: the direct cause produced the event; removing the root cause stops it recurring; a contributing cause made it likelier or let it continue.
2Example: a completed deviation investigation report.
This report is invented: product, site, batch, people and numbers are illustrative. It continues as a CAPA record and an effectiveness check.
2.1Header, summary and background.
DEV-26-0412, Exemplol 50 mg tablets, Batch 26-017, Wenmoor plant, blender B-2. Raised 3 March 2026 at 10:14, classed major because it could affect content uniformity, closed 24 March 2026. Owner and investigator: the production manager. Summary: the wrong recipe ran, blending for 12 minutes against 15; impact negligible; release recommended once the barcode interlock in CAPA-26-0087 is live. Background: since 1 March 2025, blenders B-1 to B-3 have loaded recipes from a control screen. Documents: master batch record MBR-EX50 version 9, the blender procedure, validation report VR-118.
2.21. Event.
Step 7.4 requires 15 minutes’ blending. At 09:41 the operator selected recipe EX25-BL, not EX50-BL, and the blender ran for 12. The second operator found this at 10:02, at step 7.5.
2.32. Immediate actions.
| When | Action | Who |
|---|---|---|
| 3 March, 10:14 | Batch on QA hold in the bin; B-2 stopped at 10:20. No extra blending, which is not a validated step. | Shift lead |
| 11:30 | All three blenders: a second person confirms the recipe before each start | Production manager |
| 14:10 | B-2 restarts on that condition | J. Marsh, QA manager |
| 15:30 | Ten blend samples taken before lubrication, under BS-EX50-03 | Authorized by J. Marsh |
| 6 March | On the sample results, lubricant mix and compression authorized, output quarantined, within the 14-day bulk hold validated in VR-118. Both ran on 9 March. | J. Marsh |
2.43. Scope.
Audit trails were compared with batch records for every GMP run on B-1 to B-3 since the screen went live: 240 runs to 28 February 2026, 80 of them Exemplol, and the runs since. Earlier runs loaded recipes another way. Six earlier wrong-recipe events were found, all on Exemplol, the only product with look-alike names. The 310 granulator and coater runs use the same pick list but show none, because both strengths share one recipe there.
2.54 and 5. Investigation and history.
Records were reviewed, three staff interviewed and the selection walked through on 4 March. On 5 March the history search linked the six earlier events (7.5 per 100 Exemplol runs), by four operators, each closed by QA as operator error with retraining. All six were reopened on 23 March; each batch’s blend time was inside the proven range and its uniformity results met specification, so the dispositions stand.
| Hypothesis | Evidence for | Evidence against | Outcome |
|---|---|---|---|
| Recipe EX50-BL altered | None | Audit trail: no edit since version 3 on 12 November 2025 | Ruled out |
| Timer fault | None | Calibrated 18 February 2026; log shows 12:00 against a 12:00 setpoint | Ruled out |
| Wrong recipe selected | Audit trail: EX25-BL selected at 09:41 | None | Direct cause |
| Screen and procedure | Error reproduced at the walk-through | None | Root cause |
2.66. Cause.
Root cause: recipes picked by hand from names two digits apart, nothing tying the list to the batch record, and the only check after blending. Contributing: six earlier deviations closed one at a time, with no repeat search in QA review. Human error is not accepted, because four operators made the same error.
2.77. Impact and decision.
The two recipes differ only in blend time, and 12 minutes is inside the 10 to 20 minutes proven in VR-118 and registered in the marketing authorization. Audit trail and batch record agree, so data integrity is intact. Test results support this but do not decide it:
- Ten blend samples: 98.6 to 101.0% of label claim, mean 99.75%, relative standard deviation 0.77%, against the plant’s own limits in MBR-EX50 of 90.0 to 110.0% and 5.0% (blend uniformity criteria).
- Ten tablets: 98.9 to 101.1%, mean 100.0%, standard deviation 0.73, acceptance value 2.4 × 0.73 = 1.75 against a limit of 15.0.10
Registered specifications are met, so under Annex 16 section 3 the Qualified Person (QP) may certify once the root cause is corrected; risk assessment QRA-26-014 records negligible impact and adds the batch to the stability program.9 The QP judged this deviation still unexpected: the six earlier events were closed as one-off errors, so the shared cause came to light only on 5 March, after this batch was blended, and EMA lets a QP consider all batches made before discovery.11 That call is open to challenge: EMA speaks of discovering the deviation, not its cause. Every later batch ran under the second-person check, then the interlock.
No report is due: this batch was not distributed; the six earlier ones met specification, were not mistaken for another article and, blended inside the proven range, showed no “significant chemical, physical, or other change” (21 CFR 314.81(b)(1));12 and no defect could lead to a recall or an abnormal restriction in supply (EU GMP 8.15).5
2.88 and 9. CAPA, conclusion and sign-off.
CAPA-26-0087, corrective unless marked: barcode interlock on B-1 to B-3 (automation lead, due 17 April); revised SOP and training (production manager, due 17 April); monthly trending by failure mode and a repeat search in QA review (QA manager, due 30 April); extension to granulators and coaters (preventive; automation lead, due 31 July). Effectiveness criterion: no wrong-recipe event or undocumented override in the first 40 Exemplol runs after go-live.
Conclusion: release once the interlock is live. Prepared by the production manager, reviewed by the automation lead, approved on 24 March 2026, 15 business days after the event, by the QA manager and by corporate QA, which approved none of the earlier closures. The QP certified on 14 April, a day after the interlock went live.
3How it runs in Seal.
In Seal’s deviation blueprint, the sections of a report like this one are records of their own. The deviation links the batch, sample and procedure version behind the event. The QA hold is an immediate action with its scope, actor, time and reason, each hypothesis keeps its evidence for and against, and the impact assessment and the batch decision each carry their own evidence and authority. A repeat search by failure mode, equipment or material proposes earlier events as matches, linked to their prior actions, and your team decides whether they are a pattern. Approved causes create a linked CAPA that keeps the source deviation, and configured gates can require an approved impact assessment, a cause or inconclusive rationale, and linked actions before closure.
Neil, Seal’s AI agent, can draft a deviation investigation from the deviation description, execution timestamps, equipment events and procedures: a timeline with its sources, and open questions where evidence is missing. The investigator determines root cause, and QA approves the investigation and any CAPA.
4The first hour.
Whoever finds a problem contains it and keeps the evidence intact.
- Stop what could make it worse and hold the material: FDA calls it “essential to identify and/or segregate the product”.4
- Do not reprocess it without quality unit approval.13
- Leave the screen, printouts, samples and logbook as they are, and write down what you saw: a deviation “shall be recorded and justified”.2
- Tell your supervisor and QA at once, even if unsure it matters.
QA then asks whether affected product has shipped: a reporting clock may be running.
5Reporting clocks.
| Trigger | Deadline and form | Source |
|---|---|---|
| NDA or ANDA, distributed | A batch fails specification, shows contamination or significant deterioration, or is “mistaken for, or applied to, another article”: field alert report within 3 working days (FDA recommends Form FDA 3331a) | 21 CFR 314.81(b)(1) and 314.98;12, |
| Licensed biologic | A deviation that may affect the safety, purity or potency of a distributed product: Form FDA 3486 within 45 calendar days | 21 CFR 600.14;16 606.171 for blood17 |
| Nonreproductive 361 HCT/P | A deviation relating to a core requirement on a distributed product: Form FDA 3486 within 45 days of discovery. Section 351 products, such as licensed cell and gene therapies, “are not subject to” this subpart. | 21 CFR 1271.350(b) and 1271.33018, |
| EU quality defect | A defect that may lead to a recall or an abnormal restriction in supply: tell the authorization holder or sponsor and the authorities “in a timely manner” | EU GMP 8.155 |
| UK license holder | “any defect which could result in a recall”: tell MHRA immediately on the online form, and by phone for a critical or major defect or out of hours | MHRA 202620 |
A field alert is due even when the root cause is found and fixed within the 3 working days.15 See field alert report vs BPDR for when the clock starts.
6How deep to go.
ICH Q10 says effort, formality and documentation should be “commensurate with the level of risk”.21 Classify early by possible effect on quality and patients, against your procedure’s deviation classes.
| What happened | What the report needs | Basis |
|---|---|---|
| Minor, cause known | No possible effect on quality: a short, justified record | 21 CFR 211.100(b);2 ICH Q7 2.1622 |
| Unexplained or failing | Thorough investigation, covering associated batches, distributed or not | 21 CFR 211.1921 |
| Significant deviation | Full investigation aimed at the root cause, with CAPA | EU GMP 1.8(vii);7 ICH Q7 2.1622 |
| Minor but recurring | Investigate the pattern; one-at-a-time closure hides it | ICH Q10 3.2.2: CAPA from trends;21 MHRA23 |
| Batch rejected anyway | Still investigate: “Batch rejection does not negate the need to perform the investigation.” | FDA OOS guidance3 |
| Planned in advance | Not this report: use a planned deviation or change control | EMA: planned deviations fall outside Annex 16 3.111 |
| Laboratory OOS result | Start with an OOS investigation; its full-scale phase feeds this report | FDA OOS guidance3 |
7How long it should take.
21 CFR 211.192 and EU GMP Chapter 1 set no number of days. One number comes from Barr: failure investigations “must be performed promptly, within thirty business days of the problem’s occurrence”.24 That is a 1993 court ruling, not a regulation. Set your own target, and before any overrun record the reason, progress, interim control and new date, approved by QA.
The batch sets the real deadline. In the US, the quality control unit approves production records “before a batch is released or distributed”.1 In the EU, a QP may certify a batch that meets registered specifications despite an unexpected deviation once it is “thoroughly investigated and the root cause corrected” and the impact negligible (Annex 16 section 3).9 EMA treats a deviation as unexpected only “until the time of discovery”; repeats after that “should be considered changes”, and a QP may go on certifying them while CAPA is in progress only “Where the relevant authorities have confirmed the need to avoid supply disruption”.11 See what a QP certifies.
Producing with the cause unknown has its own risk:
“You opened the investigation on June 4, 2025, and the status of the investigation on November 3, 2025, five months later, was still in progress. You continued to manufacture additional batches on the same processing line without definitive root causes identified.” Jubilant HollisterStier General Partnership, warning letter, 28 May 202625
8Investigational products, ATMPs and devices.
| Scope | Source | What it adds |
|---|---|---|
| Investigational | 21 CFR 210.2(c); Regulation (EU) 2017/1569, Article 9(1) | US phase 1 is exempt from Part 211, but FDA expects records of “Deviations and investigations”.26, |
| EU ATMPs | EU GMP Part IV, 11.52, guideline | A QP may certify despite an unexpected deviation only after “an in-depth assessment of the impact of the deviation”.29 |
| Medical devices | 21 CFR 820.10, regulation, since 2 February 2026; FDA’s preamble | Nonconforming product (ISO 13485 clause 8.3) and corrective action (8.5.2).30, |
9When a contract manufacturer is involved.
Who investigates depends on the quality agreement; who answers for it does not. The agreement should say “how deviations will be investigated, documented, and resolved in compliance with CGMP”;32 in the EU it “should describe clearly who undertakes each step”.33 The owner’s quality unit still approves or rejects the product.8 The applicant owns the field alert clock, so the agreement “should establish a procedure” for contractors to notify in time.15 Each report should name who wrote each section and who approved the disposition.
10What inspectors look for.
FDA’s drug inspection program checks that investigations are “Documented; investigated in a timely manner using scientific evidence to identify the root cause”.34 Every NDA or ANDA pre-approval inspection reviews investigations on development, exhibit and validation batches and on similar marketed products, to “determine if the establishment is prepared for the proposed commercial manufacturing process at commercial scale”. Development reports may be less comprehensive, but the firm should still “objectively assess all development data and information”.35
A Dutch statement rated the OOS process a critical deficiency:
“An Advanced Therapy Medicinal Product (ATMP) was released despite a failed sterility test which was performed as an in-process control. In addition, the procedure for OOS is not unambiguous and can lead to batch certification and release without thorough investigation of sufficient justification.” Antoni van Leeuwenhoek hospital, statement of non-compliance NL/H 26/2008484A, 7 October 202636
10.1Pre-inspection check.
Pull these from your deviation log; regulators have flagged each:
- investigations open while production continued (Jubilant);25
- canceled deviations and their reasons (Genzyme’s own review found 36 of 74 needed investigation);37
- operator error closures without in-depth analysis of the conditions behind the error (Pharmathen);38
- no-impact conclusions resting on passing test results (Fresenius);39
- repeat failure modes, which MHRA reads as a sign of an ineffective system;23
- extensions past target, since MHRA flags investigations “taking an unreasonable time to complete”.23
11Write it so a reviewer can follow it.
| Weak wording | What regulators said | Stronger wording |
|---|---|---|
| “Operator error.” | FDA: no “in-depth analysis of the conditions that create the circumstances for human error”.38 | “Four operators made the same error on look-alike names. Human error is not accepted as the root cause.” |
| “No product impact.” | FDA: “overreliance on passing finished drug product test results”.39 | “12 minutes is inside the range proven in VR-118. Test results support this but do not decide it.” |
| “Root cause undetermined.” | FDA: the investigation “did not thoroughly examine all possible root causes”.40 EU GMP accepts a most likely cause.7 | “True cause not confirmed. Most likely cause: X, because Y. A and B were ruled out by records C and D.” |
See human error as a root cause.
AQuestions and answers
What is the difference between a deviation report and an investigation report?
The deviation report records the event when it is found: what was expected, what happened and what was done straight away. The investigation report records the work that followed: the cause, the scope, the impact on product, the decision and the CAPA. They can be one record with separate sections, as long as the first observation is kept as it was recorded.
Who approves a deviation investigation?
The quality unit. In the US, the quality control unit reviews production records, makes sure errors have been fully investigated, and approves or rejects the batch. In the EU, the assessment of deviations is part of the product assessment before a Qualified Person certifies the batch.
Does the investigation need to cover batches already shipped?
Yes. 21 CFR 211.192 requires the investigation whether or not the batch has been distributed, and it extends to other batches and products that may be associated. For an approved NDA or ANDA, a distributed batch that fails its specification, or a product or label mistaken for another article, also triggers a field alert report within three working days.
Can a deviation be canceled after it is raised?
Only rarely, and never quietly. If the record really was raised in error, for example a duplicate, QA should review it and keep the reason and the record. Do not cancel a real event to avoid an investigation. FDA cited one firm in 2026 for canceling deviations without looking at root cause or product impact, and asked for a review of all of them.
What tools are used in a deviation investigation?
Before any root cause tool, reconstruct the event: a timeline from batch records, audit trails and alarm logs, a walk-through at the place it happened, and interviews with those involved. No regulation requires a particular tool. MHRA recommends approaches such as a fishbone diagram and the 5 Whys; a hypothesis table then tests each candidate cause against the evidence, and a search of past deviations checks for repeats.
What if new evidence turns up after the report is approved?
Do not edit the approved report. Reopen it or add a dated addendum under your procedure that states what changed, what it means for the cause, the scope and the batch decision, and who approved it. Keep the original conclusion visible, and tell any regulator or contract partner who relied on it.
References40 sources
References
- 121 CFR 211.192, Production record review, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 192 (read 25 November 2026) - 221 CFR 211.100, Written procedures; deviations, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 100 (read 25 November 2026) - 3Investigating Out-
of- Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, Revision 1, FDA, May 2022, sections III, IV and V. https:/ / www. fda. gov/ media/ 158416/ download (read 25 November 2026) - 4Guidance for Industry: Quality Systems Approach to Pharmaceutical CGMP Regulations, FDA, September 2006, section IV.C.4. https:/
/ www. fda. gov/ media/ 71023/ download (read 25 November 2026) - 5EudraLex Volume 4, Part I, Chapter 8: Complaints, Quality Defects and Product Recalls, European Commission, 13 August 2014, in operation since 1 March 2015, sections 8.12, 8.14 and 8.15. https:/
/ health. ec. europa. eu/ document/ download/ b1eb2292- cb0d- 4e3f- aea9- e3fe79faf6e3_ en?filename=2014- 08_ gmp_ chap8. pdf (read 25 November 2026) - 6Responding to FDA Form 483 Observations at the Conclusion of a Drug CGMP Inspection, draft guidance for industry, FDA, March 2026, section IV. https:/
/ www. fda. gov/ media/ 191427/ download (read 25 November 2026) - 7EudraLex Volume 4, Part I, Chapter 1: Pharmaceutical Quality System, European Commission, in operation since 31 January 2013, sections 1.4(xiv) and 1.8(vii). https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap1_ 2013- 01_ en_ 0. pdf (read 25 November 2026) - 821 CFR 211.22, Responsibilities of quality control unit, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 22 (read 25 November 2026) - 9EudraLex Volume 4, Annex 16: Certification by a Qualified Person and Batch Release, European Commission, 12 October 2015, sections 1.7.16, 1.7.17, 3 and 3.1. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ v4_ an16_ 201510_ en_ 0. pdf (read 25 November 2026) - 10The Japanese Pharmacopoeia, Eighteenth Edition, General Test 6.02, Uniformity of Dosage Units, a test harmonised with the European Pharmacopoeia and the US Pharmacopeia, MHLW, English version. https:/
/ www. mhlw. go. jp/ content/ 11120000/ 001372337. pdf (read 25 November 2026) - 11Guidance on good manufacturing practice and good distribution practice: Questions and answers, EU GMP guide annexes, Annex 16, question 3, EMA. https:/
/ www. ema. europa. eu/ en/ human- regulatory- overview/ research- development/ compliance- research- development/ good- manufacturing- practice/ guidance- good- manufacturing- practice- good- distribution- practice- questions- answers (read 25 November 2026) - 1221 CFR 314.81, Other postmarketing reports, paragraph (b)(1), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 314. 81 (read 25 November 2026) - 1321 CFR 211.115, Reprocessing, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 115 (read 25 November 2026) - 1421 CFR 314.98, Postmarketing reports, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 314. 98 (read 25 November 2026) - 15Field Alert Report Submission: Questions and Answers, guidance for industry, FDA, July 2021, questions 1h, 1j, 2, 3 and 4a. https:/
/ www. fda. gov/ media/ 114549/ download (read 25 November 2026) - 1621 CFR 600.14, Reporting of biological product deviations by licensed manufacturers, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 600. 14 (read 25 November 2026) - 1721 CFR 606.171, Reporting of product deviations by licensed manufacturers, unlicensed registered blood establishments, and transfusion services, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 606. 171 (read 25 November 2026) - 1821 CFR 1271.350, Reporting, paragraph (b), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 1271. 350 (read 25 November 2026) - 1921 CFR 1271.330, Applicability, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 1271. 330 (read 25 November 2026) - 20A Guide to Defective Medicinal Products, MHRA, published February 2026, sections 4 and 5. https:/
/ assets. publishing. service. gov. uk/ media/ 698f286b43eb912f2e15b3fd/ DMRC_ Guide- to- Defective- Medicinal_ Products- Feb2026. pdf (read 25 November 2026) - 21ICH Q10, Pharmaceutical Quality System, ICH, Step 4 version of 4 June 2008, section 3.2.2. https:/
/ database. ich. org/ sites/ default/ files/ Q10%20Guideline. pdf (read 25 November 2026) - 22ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, ICH, 10 November 2000, section 2.16. https:/
/ database. ich. org/ sites/ default/ files/ Q7%20Guideline. pdf (read 25 November 2026) - 23A fresh look at an old topic: Investigations in the GMDP environment, MHRA Inspectorate blog, 14 September 2020. https:/
/ mhrainspectorate. blog. gov. uk/ 2020/ 09/ 14/ a- fresh- look- at- an- old- topic- investigations- in- the- gmdp- environment/ (read 25 November 2026) - 24United States v. Barr Laboratories, Inc., 812 F. Supp. 458, paragraph 32, US District Court, District of New Jersey, 5 February 1993, as amended 30 March 1993, Caselaw Access Project. https:/
/ static. case. law/ f- supp/ 812/ html/ 0458- 01. html (read 25 November 2026) - 25Jubilant HollisterStier General Partnership, warning letter, MARCS-CMS 723537, FDA, 28 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ jubilant- hollisterstier- general- partnership- 723537- 05282026 (read 25 November 2026) - 2621 CFR 210.2, Applicability of current good manufacturing practice regulations, paragraph (c), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 210. 2 (read 25 November 2026) - 27Guidance for Industry: CGMP for Phase 1 Investigational Drugs, FDA, July 2008, section V.H. https:/
/ www. fda. gov/ media/ 70975/ download (read 25 November 2026) - 28Commission Delegated Regulation (EU) 2017/1569 of 23 May 2017, Article 9(1), EUR-Lex, in force. https:/
/ eur- lex. europa. eu/ legal- content/ EN/ TXT/ ?uri=CELEX:32017R1569 (read 25 November 2026) - 29Guidelines on Good Manufacturing Practice specific to Advanced Therapy Medicinal Products, C(2017) 7694 final, European Commission, 22 November 2017, paragraph 11.52. https:/
/ health. ec. europa. eu/ document/ download/ ad33d9dd- 03f0- 4bef- af53- 21308ce2187d_ en?filename=2017_ 11_ 22_ guidelines_ gmp_ for_ atmps. pdf (read 25 November 2026) - 3021 CFR 820.10, Requirements for a quality management system, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 820. 10 (read 25 November 2026) - 31Medical Devices; Quality System Regulation Amendments, final rule, FDA, 89 FR 7496, 2 February 2024, effective 2 February 2026, preamble on nonconforming product. https:/
/ www. govinfo. gov/ content/ pkg/ FR- 2024- 02- 02/ html/ 2024- 01709. htm (read 25 November 2026) - 32Contract Manufacturing Arrangements for Drugs: Quality Agreements, guidance for industry, FDA, November 2016, sections IV.B and V.B. https:/
/ www. fda. gov/ media/ 86193/ download (read 25 November 2026) - 33EudraLex Volume 4, Part I, Chapter 7: Outsourced Activities, European Commission, in operation since 31 January 2013, section 7.15. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap7_ 2012- 06_ en_ 0. pdf (read 25 November 2026) - 34Compliance Program 7356.002, Drug Manufacturing Inspections, FDA, implemented 17 October 2022. https:/
/ www. fda. gov/ media/ 75167/ download (read 25 November 2026) - 35Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026, implemented 10 August 2026, Objective 1a. https:/
/ www. fda. gov/ media/ 193382/ download (read 25 November 2026) - 36Statement of non-compliance with GMP, report NL/H 26/2008484A, Het Nederlands Kanker Instituut-Antoni van Leeuwenhoek Ziekenhuis Stichting, Health and Youth Care Inspectorate (Netherlands), 7 October 2026, EudraGMDP reference 187962, reached from the GMP non-compliance search. https:/
/ eudragmdp. ema. europa. eu/ inspections/ gmpc/ searchGMPNonCompliance. do (read 25 November 2026) - 37Genzyme Ireland Limited, warning letter, MARCS-CMS 728681, FDA, 22 June 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ genzyme- ireland- limited- 728681- 06222026 (read 25 November 2026) - 38Pharmathen International S.A., warning letter, MARCS-CMS 723680, FDA, 27 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ pharmathen- international- sa- 723680- 05272026 (read 25 November 2026) - 39Fresenius Kabi USA LLC, warning letter, MARCS-CMS 733107, FDA, 22 September 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ fresenius- kabi- usa- llc- 733107- 09222026 (read 25 November 2026) - 40Jabil Inc., warning letter, MARCS-CMS 731037, FDA, 27 August 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ jabil- inc- 731037- 08272026 (read 25 November 2026)