What is the difference between ISO 13485 and 21 CFR Part 820 (the QMSR)?

Since 2 February 2026, 21 CFR Part 820, the Quality Management System Regulation, incorporates ISO 13485:2016, so the core requirements are the same. The differences are legal force and additions. ISO 13485 is a voluntary standard checked by certification audits; Part 820 is US law enforced by FDA inspection, and FDA neither requires nor relies on certificates. Part 820 adds FDA definitions, UDI, tracking, adverse event and correction reporting, set complaint and servicing record fields, and labeling checks.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal studying an auto-injector laid out in assembly order, threaded to its device history folder.

1The difference in one table.

Since 2 February 2026, Part 820 has incorporated ISO 13485:2016 and Clause 3 of ISO 9000:2015 by reference.1 A manufacturer must document a quality management system that complies with ISO 13485 and the rest of Part 820.2

ISO 13485:201621 CFR Part 820 (QMSR)
What it isA voluntary standard from ISO/TC 210, confirmed on 31 October 20253US law. Failing it makes a device adulterated under section 501(h) of the FD&C Act.2
Who follows itAny organization in the device life cycle, and suppliers by choice or by contract4Finished-device manufacturers, contract manufacturers and specification developers included, each for the operations it performs; not component makers, unless FDA treats the part as an accessory, such as blood tubing5,6,7
Who checksCertification bodies, annually on a three-year plan8FDA investigators, scheduled on risk factors8
ProofA certificateNone. FDA “will not require” and “will not issue” ISO 13485 certificates.8
DefinitionsISO 13485 Clause 3 and ISO 90004Those, plus six terms added in 820.3(a); under 820.3(b), the FD&C Act’s section 201 definitions, such as device and labeling, and five more terms override the matching ISO terms6
Design controlsClause 7.3, which can be excluded where regulation permits4Required for class II and III devices and some class I devices2

FDA judged ISO 13485, “when taken in totality”, “substantially similar” to the old Quality System (QS) regulation, and found that weaving risk management through its clauses “more explicitly establishes a requirement for risk management to occur throughout a QMS”.8,9 FDA’s compliance program makes four ISO 13485 requirements inspection elements of their own: a risk-based approach to QMS processes (Clause 4.1.2 b)), control of outsourced processes (4.1.5), validation of QMS software (4.1.6) and the medical device file (4.2.3). Every baseline inspection covers the outsourcing and medical device file elements.10 Apart from Clause 3 of ISO 9000, FDA did not incorporate ISO 14971 or the other standards ISO 13485 references.9 The QMSR page maps each old section to its new home.

2Gap checklist: what a certified system adds for FDA.

FDA will not rely on certificates partly because it “cannot ensure that other FDA medical device requirements, such as parts 803, 806, 821, 830, are audited during independent ISO 13485 audits”.9 Its investigators follow Compliance Program 7382.850, which sorts the QMSR into six QMS areas: change control; design and development; management oversight; measurement, analysis and improvement; outsourcing and purchasing; and production and service provision. Beside them sit four other applicable FDA requirements: medical device reporting (MDR), corrections and removals, tracking, and unique device identification (UDI). Both inspection models cover every area and, where they apply, all four; only a PMA preapproval inspection leaves those four out.10

Table 2. Clause numbers are ISO 13485’s. Devices exempt from CGMP still keep complaint files and 820.35 records.7
CheckEvidence to holdRequirement
Design scopeA list of devices under Clause 7.3: class II and III, class I automated with software, five named class I devices, and devices under an investigational device exemption (IDE) among them820.10(c); 812.1(a)2,11
UDIA documented UDI system under Part 830; the UDI recorded for each device or batch820.10(b)(1), 820.35(c)12
TraceabilityPart 821 tracking procedures if it applies; Clause 7.5.9.2 traceability for devices that support or sustain life820.10(b)(2), (d)
Device reporting (MDR)Written MDR procedures; an event file recording each reportability decision, which may sit in the complaint file if prominently identified820.10(b)(3); 803.17, 803.1813
CorrectionsAdvisory notices handled under Part 806820.10(b)(4)
ComplaintsA review, evaluation and investigation record for each possible specification failure; the reason when one is not investigated; one unit “ultimately responsible for coordinating all complaint handling functions”820.35(a); comment 549
Complaint fieldsThe seven 820.35(a) items on each reportable or investigated complaint, as in the example below820.35(a)(1) to (7)
ServicingDevice, UDI or Universal Product Code (UPC), date, who serviced it, service performed, any test and inspection data820.35(b)
LabelingExamination for UDI or UPC, expiry date, storage, handling and processing instructions; release documented; inspection before use820.4514
ApprovalsA signature and date wherever ISO 13485 requires approval; compliant electronic signaturesComment 53
ReportsManagement review, internal audit and supplier audit reports written for an investigator to readFAQ 88

The details are in MDR reporting, corrections and removals, DHF, DMR and DHR under the QMSR, inspection access to audit and review reports and combination products.

2.1Example: one complaint record that meets both.

Invented device, lot and people.

Table 3. Example: one complaint record for an invented device.
FieldEntryRequirement
Device nameHVD-⁠9 handheld vascular Doppler820.35(a)(1)12
Date received12 March 2026, by email to service820.35(a)(2)
IdentificationUDI from the label; lot 24-⁠017; serial HVD9-⁠1180820.35(a)(3)
ComplainantClinic nurse; name, address and phone on file820.35(a)(4)
Nature, detailsNo waveform from the toe probe after a battery change, with the probe-fault message shown; unit returned under returned goods authorization (RGA) 26-⁠114820.35(a)(5)
Complaint?Yes: it alleges a performance deficiency after releaseClause 3.44
Investigate?Yes: a possible failure to meet the waveform specification820.35(a); comment 549
Part 803 reportabilityNot reportable: no injury, and a recurrence would not be likely to cause or contribute to a death or serious injury; rationale, reviewer and date in the MDR event file803.50(a), 803.18(b)(1)(i)13
InvestigationCracked probe connector, the same crack as in 2 of the 14 other HVD-⁠9 units returned since 1 January820.35(a)
Correction, actionProbe replaced, a correction in ISO 9000’s sense; CAPA-⁠26-⁠031 on the connector design (corrective action)820.35(a)(6); comment 299
Part 806 actionA removal, as the unit came back for repair. Identical repairs of 3 units since 1 January are not routine servicing, so that exemption is closed.806.1(b)(3); 806.2(j), (l)15
Part 806 reportNot reportable: the unit announces the fault and gives no false reading, so the record concludes its use could cause no serious, temporary or reversible harm, not even remotely. Justification reviewed by a designated person.806.2(k), 806.10(a), 806.20(b)(4)
ReplyLetter of 2 April 2026 with the finding and replacement820.35(a)(7)

3Two words the texts use differently.

Where an ISO 13485 clause conflicts with the FD&C Act or FDA’s regulations, the Act and the regulations control.5 Two words catch teams out.

  • Rework. FDA declined ISO 9000’s definition and kept its own: action on a nonconforming product “so that it will fulfill the specified requirements in the medical device file (MDF) before it is released for distribution”.6,9 FDA cited Linemaster Switch in May 2026 under Clause 8.3.4 for undocumented rework: torquing screws and recalibrating foot pedals during manufacture.16 Adjusting the same pedals after they ship is not rework: FDA wanted those actions under its rules on corrections, removals and recalls, Parts 7, 806 and 810.9
  • Correction. In ISO 13485 and 820.35(a)(6) it has ISO 9000’s meaning; FDA says it “relates to eliminating a nonconformity” and may be scrap, repair, rework or adjustment.9 FDA considers Part 806 to apply to corrections made after release.9 There the word is narrower: a repair at the device’s point of use is a correction, and taking the device elsewhere for repair is a removal.15

4Where an ISO 13485 certificate does count.

MarketWhat the certificate or standard doesSource
United StatesNone: a certificate “will not exempt a manufacturer from an FDA inspection”. Firms in the Medical Device Single Audit Program (MDSAP) skip surveillance inspections; FDA uses their audit reports, “rather than the certificate”.FAQ;8 comment 79;9 CP 7382.85010
European UnionConformity with EN ISO 13485:2016/A11:2021 is presumed to meet the quality system requirements of the EU MDR (Regulation (EU) 2017/745) that it coversMDR Articles 8, 10(9);17 Decision 2022/618
CanadaClass II, III and IV licenses need an ISO 13485 certificate from an MDSAP-recognized auditing organizationMedical Devices Regulations, section 3219
Great BritainEN ISO 13485:2016 has been a designated standard under the UK MDR 2002 since 1 January 2021; the +A11:2021 text was added on 17 January 2026Designated standards notices20

Getting and keeping one is covered in ISO 13485 certification.

5What FDA has cited since February 2026.

Recent letters cite what FDA adds, not only ISO 13485 clauses:

Failure to maintain records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications, as required by 21 CFR 820.35(a).

Koven Technologies, warning letter of 21 July 2026: 13 of the 29 returned goods authorization forms the investigator reviewed reported device failures but were not evaluated as complaints.21

Your organization’s MDR event files do not contain all documentation of its deliberations and decision-making processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable under 21 CFR Part 803, as required under 21 CFR 803.18(b)(1)(i).

O.N. Diagnostics, maker of VirtuOst software, warning letter of 1 September 2026. It had offered its complaint procedure, SOP-⁠014, as its MDR guidance; FDA found that did not meet 803.17, which requires written MDR procedures.22

6What changed in 2025 and 2026.

  • 31 October 2025: ISO confirmed ISO 13485:2016, the edition Part 820 names. A revision would reach the QMSR through FDA rulemaking.3,8
  • 4 December 2025: technical amendments, effective 2 February 2026 with no new requirements, repointed 162 class I classification regulations from 820.180 and 820.198 to 820.35.23
  • 2 February 2026: the QMSR took effect, and Compliance Program 7382.850 replaced QSIT.8
  • Since then: 820.5, 820.20 to 820.30, 820.40 and subparts C to O are reserved, so guides citing 820.30, 820.100 or 820.198 are out of date.24
  • Still draft: FDA’s October 2025 guidance, docket FDA-⁠2025-⁠D-⁠4051, listing by ISO 13485 clause the QMS information a premarket approval (PMA) or humanitarian device exemption (HDE) application should give. When final, it will replace the 2003 guidance.25
  • In development: ISO/CD TS 23485, guidance on applying ISO 13485:2016, a committee draft at stage 30.99.26

7How it runs in Seal.

Seal’s medical device blueprint records traceability as the work is done: requirements link to design outputs, verification and risk controls, and each unit links to the versions and lots it was built from. Complaints are entered as structured data, and the blueprint lists the fields the QMSR asks of an investigated or reported complaint. In complaint handling the report stays as received, and its technical investigation, safety assessment and reportability assessment against 21 CFR 803.50 are separate linked records, each with its own owner or reviewer. Electronic device history records carry the UDI and label version into each unit’s as-built record, and rework is a controlled branch that keeps the original failed state.

The blueprint’s own advice holds here: map the configured quality system to the QMSR, ISO 13485 and the other requirements that apply to your device, because the presence of a workflow does not itself demonstrate conformity.

AQuestions and answers

Does the QMSR apply to component manufacturers?

Not as a rule. Part 820 does not apply to makers of components or parts of finished devices, though FDA encourages them to consider its provisions as appropriate. A component FDA treats as an accessory, such as blood tubing or a diagnostic x-ray component, is a finished device, and its maker is subject to the QMSR. The device maker still controls what it buys under ISO 13485’s purchasing clauses, and can require a supplier to hold ISO 13485 by contract.

Does a contract manufacturer’s ISO 13485 certificate replace a supplier audit?

Not by itself. Clause 7.4.1 asks for supplier criteria based on the supplier’s ability to meet requirements, its performance and the effect of what it supplies on the device, and your own procedure sets what evidence that takes. Koven Technologies was cited because its procedure required a site assessment of its critical contract manufacturer and there was no record of one.

Do our procedures have to follow ISO 13485’s clause numbering?

No. ISO 13485 says it does not imply uniform documentation or alignment to its clause structure. FDA asks firms to be ready to discuss and provide records that show compliance with the QMSR, so a cross-reference from each procedure to the clauses and Part 820 sections it meets makes an inspection easier.

What is the difference between MDSAP and ISO 13485?

ISO 13485 is the quality system standard. The Medical Device Single Audit Program (MDSAP) is a single regulatory audit, by a recognized auditing organization, against ISO 13485 and the added requirements of the participating regulators: Australia’s TGA, Brazil’s ANVISA, Health Canada, Japan’s MHLW and PMDA, and FDA. FDA uses MDSAP audit reports in place of routine surveillance inspections, which a plain ISO 13485 certificate cannot do, and Health Canada requires a certificate issued under MDSAP for class II to IV licenses.

References26 sources

References

  1. 121 CFR 820.7, Incorporation by reference, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.7 (read 26 November 2026)
  2. 221 CFR 820.10, Requirements for a quality management system, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.10 (read 26 November 2026)
  3. 3ISO 13485:2016, Medical devices, Quality management systems, Requirements for regulatory purposes, ISO, standard page and life cycle (90.93 confirmed, 31 October 2025). https://www.iso.org/standard/59752.html (read 26 November 2026)
  4. 4ISO 13485:2016, Introduction (0.1, 0.2) and Clauses 1 and 3, public preview, ISO Online Browsing Platform. https://www.iso.org/obp/ui/#iso:std:iso:13485:ed-3:v1:en (read 25 November 2026)
  5. 521 CFR 820.1, Scope, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.1 (read 26 November 2026)
  6. 621 CFR 820.3, Definitions, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.3 (read 26 November 2026)
  7. 7Quality Management System Regulation (QMSR), FDA, applicability and CGMP exemptions, content current as of 2 February 2026. https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr (read 26 November 2026)
  8. 8Quality Management System Regulation, Frequently Asked Questions, FDA, introduction and questions 5 to 8 and 13, content current as of 5 November 2026. https://www.fda.gov/medical-devices/quality-management-system-regulation-qmsr/quality-management-system-regulation-frequently-asked-questions (read 26 November 2026)
  9. 9Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, FDA, 2 February 2024, responses to comments 9, 19, 29, 39, 53, 54 and 79. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 25 November 2026)
  10. 10Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Part I.9, Part III and Attachment A. https://www.fda.gov/media/80195/download (read 25 November 2026)
  11. 1121 CFR 812.1, Scope, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-812/subpart-A/section-812.1 (read 26 November 2026)
  12. 1221 CFR 820.35, Control of records, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-B/section-820.35 (read 26 November 2026)
  13. 1321 CFR Part 803, Medical Device Reporting, sections 803.17, 803.18 and 803.50(a), eCFR, up to date as of 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-803 (read 26 November 2026)
  14. 1421 CFR 820.45, Device labeling and packaging controls, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-B/section-820.45 (read 26 November 2026)
  15. 1521 CFR Part 806, Medical Devices; Reports of Corrections and Removals, sections 806.1, 806.2, 806.10 and 806.20, eCFR, up to date as of 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-806 (read 26 November 2026)
  16. 16Linemaster Switch Corporation, warning letter, MARCS-⁠CMS 730215, FDA, 27 May 2026, violation 1. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/linemaster-switch-corporation-730215-05272026 (read 26 November 2026)
  17. 17Regulation (EU) 2017/745 on medical devices, Articles 8 and 10(9), EUR-⁠Lex, 5 April 2017. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745 (read 25 November 2026)
  18. 18Commission Implementing Decision (EU) 2022/6 amending Implementing Decision (EU) 2021/1182 on harmonised standards for medical devices, EUR-⁠Lex, 4 January 2022. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32022D0006 (read 25 November 2026)
  19. 19Guidance Document: Requirements in the Recognition Process for Medical Device Single Audit Program (MDSAP) Auditing Organizations, Health Canada, sections 1.2 to 1.4. https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/quality-systems-13485/requirements-recognition-process-mdsap-auditing-organizations-profile/guidance-document.html (read 26 November 2026)
  20. 20Designated standards: medical devices, notices of publication 0034/21 of 1 January 2021 and 0129/26 of 17 January 2026, Department of Health and Social Care, GOV.UK, last updated 17 January 2026. https://www.gov.uk/government/publications/designated-standards-medical-devices (read 25 November 2026)
  21. 21Koven Technologies, Inc., warning letter, MARCS-⁠CMS 734643, FDA, 21 July 2026, violations 3 and 4. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/koven-technologies-inc-734643-07212026 (read 26 November 2026)
  22. 22O.N. Diagnostics, LLC, warning letter, MARCS-⁠CMS 730301, FDA, 1 September 2026, Medical Device Reporting violations. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/diagnostics-llc-730301-09012026 (read 26 November 2026)
  23. 23Medical Devices; Quality Management System Regulation Technical Amendments, final rule, 90 FR 55978, FDA, 4 December 2025, effective 2 February 2026, section II. https://www.federalregister.gov/documents/2025/12/04/2025-21955/medical-devices-quality-management-system-regulation-technical-amendments (read 26 November 2026)
  24. 2421 CFR Part 820, Quality Management System Regulation, eCFR, current text, up to date as of 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  25. 25Quality Management System Information for Certain Premarket Submission Reviews, draft guidance for industry and FDA staff, FDA, October 2025, docket FDA-⁠2025-⁠D-⁠4051, sections I and VI.A. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/quality-management-system-information-certain-premarket-submission-reviews (read 26 November 2026)
  26. 26ISO/CD TS 23485, Medical devices, Quality management systems, Guidance for the application of ISO 13485:2016, ISO, standard page, stage 30.99. https://www.iso.org/standard/90565.html (read 26 November 2026)