1eQMS vs QMS.
The QMS is what the rules require; the eQMS is software that runs it. For medicines, GMP and ICH Q10 describe the quality system;1,
| Question | QMS | eQMS |
|---|---|---|
| What is it? | Procedures, responsibilities and records | Software that runs those procedures and links their records |
| Is it required? | Yes, by GMP and the QMSR | No: paper and electronic records face the same requirements4 |
| Who answers for it? | Senior management, through management review2 | A system owner, with a process owner for each process5 |
| How is it checked? | Monitoring, audits and management review2 | Validation for its intended use, then periodic evaluation5 |
2eQMS modules and the records each keeps.
Each module is a quality process with records the regulations describe. Configure each form so it cannot close without them.
| Module | What the record must show | Where required |
|---|---|---|
| Document control | Approved, signed and dated by authorized people; uniquely identified; an effective date; reviewed regularly | EU GMP 4.3 and 4.5;6 21 CFR 211.100(a)7 |
| Training | Training in each person’s operations and in CGMP, by qualified people, continuing | 211.25(a)7 |
| Deviations | Each departure from a written procedure recorded and justified | 211.100(b)7 |
| Investigations | Thorough, extended to other batches that may be affected, written with conclusions and follow-up | 211.1927 |
| CAPA | Root cause found in proportion to risk; effectiveness monitored and assessed | ICH Q10 3.2.2;2 EU GMP 1.4(xiv)1 |
| Change control | Evaluated and approved before; evaluated after for unintended impact | ICH Q10 3.2.3;2 EU GMP 1.4(xii) and (xiii)1 |
| Complaints | Quality unit review, a decision on investigation, a written record | 211.198;7 devices 820.35(a)3 |
| Audits | Self-inspections on a pre-arranged program; each report records the observations, proposed corrective measures and actions taken | EU GMP 1.4(xvii), 9.1 and 9.3;1, |
| Suppliers | Outsourced activities and purchased materials controlled and reviewed | ICH Q10 2.72 |
| Risk | Effort and documentation in proportion to the risk; sets the depth of investigations, change evaluations and validation | EU GMP 1.12, 1.13 and 1.4(xiv);1 ICH Q10 3.2.3;2 Annex 11, 15 |
| Management review | Inspection and audit results, commitments to regulators and follow-up from earlier reviews; periodic quality reviews, which can cover complaints and recalls, monitoring conclusions and whether changes and CAPAs worked | ICH Q10 3.2.42 |
3Example: one complaint through an eQMS.
This trail is invented for illustration: product, site, numbers and dates. It follows one complaint through the linked records an inspector would ask to see.
| Record | Entry | Rule it meets |
|---|---|---|
| Complaint | COMP-26-0081, received 12 January 2026 from Quellford Pharmacy: chipped Exemplol 50 mg tablets, Batch 25-117. Reply with the findings sent 9 February. | 211.198(b)7 |
| Quality unit review | 13 January: a possible failure of the appearance specification, so investigate; not a serious and unexpected adverse drug experience to report. | 211.198(a)7 |
| Investigation | INV-26-0034. Cause: a worn filling-chute liner on packaging line 2 at the Wenmoor plant. Retained samples of every batch packed on line 2 since the liner was last changed examined. Closed 6 February with conclusions. | 211.1927 |
| CAPA | CAPA-26-0021. Corrective: replace the liner and add a liner wear check to planned maintenance. Preventive: the same check on the other packaging lines. Opened 6 February. | ICH Q10 3.2.2 and glossary2 |
| Change control | CC-26-0040. Maintenance procedure SOP-ENG-021 revised to version 7. Approved 20 February; effective 2 March. | EU GMP 1.4(xii);1 EU GMP 4.36 |
| Training | Nine technicians trained on version 7 by 27 February, before it took effect. | 211.25(a)7 |
| Effectiveness check | Criterion set on 6 February: no chipped-tablet complaints and no appearance failures in the next 20 batches. Met on 30 June. | EU GMP 1.4(xiv)1 |
| Management review | July 2026 review: complaint trend and CAPA result reported. | ICH Q10 3.2.42 |
In an eQMS each record opens from the one before it, so a reviewer can go from the complaint to the effectiveness result without searching. For a full CAPA record, see what is CAPA.
4What the rules expect of the software.
When eQMS records are kept in place of paper, or relied on to perform regulated work, Part 11 applies.10 The system must be validated, limit access, keep secure time-stamped audit trails, use authority checks and enforce the permitted sequence of steps where that matters, for example so that a CAPA cannot close before its effectiveness check.11
Buying the eQMS as a service does not hand over that validation: the company stays “ultimately responsible” for control of what it outsources.2 Annex 11 requires a formal agreement with “clear statements of the responsibilities of the third party”, and adds: “The supplier should be assessed appropriately.”5 Validating SaaS for GxP lists what to agree with the vendor.
FDA’s computer software assurance guidance for device makers, issued in September 2025 and revised for the QMSR on 3 February 2026, says a manufacturer could start from the developer’s validation work and “determine what additional activities may be needed”. It counts CAPA routing, complaint logging, change control and procedure management as generally not high process risk. In its worked example, a nonconformance system, only the prompt to decide whether distributed product needs a correction or removal gets a detailed scripted protocol; initiation and electronic signatures get exploratory and scenario testing. For each function the manufacturer keeps:12
- Intended use
- Result of the risk-based analysis
- Testing performed, and any issues found
- Conclusion on acceptability, including how issues were resolved
- Who tested, and when
- Review and approval, where appropriate
See CSV vs CSA.
5Who does what in an eQMS.
| Role | What they do | Source |
|---|---|---|
| Senior management | Answers for an effective quality system; governs it through management review | ICH Q10 2.1 and 2.62 |
| Quality unit | Approves or rejects procedures, specifications and products; reviews investigations | 21 CFR 211.22(a) and (c)7 |
| Process owner | “The person responsible for the business process”, such as CAPA or complaints | Annex 11, glossary5 |
| System owner | Availability and maintenance of the system and security of its data | Annex 11, glossary5 |
| Administrator | Accounts, roles and configuration, each change controlled and recorded; independent of the record content | FDA Q&A 4;4 Annex 11, 10 and 12.35 |
| Supplier | Hosting, upgrades and support, under a written agreement | Annex 11, 3.15 |
| Users | Their own login; signatures that show name, date, time and meaning | FDA Q&A 5;4 21 CFR 11.5011 |
6eQMS, EDMS, LIMS, MES, ERP or PLM.
Most companies run several of these. Decide which system holds the original of each record.
| Record | Usual home | Where required |
|---|---|---|
| Quality events | eQMS: deviations, investigations, CAPA, change, complaints, audits | 211.192 and 211.1987 |
| Procedures, training | eQMS or a document management system | EU GMP 4.3;6 211.257 |
| Test results | LIMS | 211.1947 |
| Batch execution | MES or electronic batch record | 211.1887 |
| Material status | ERP: quarantined, approved or rejected | 211.80(d)7 |
| Device design | PLM or a design control module | 820.10(c)3 |
See also how LIMS, MES and QMS fit together.
7What inspectors find in eQMS records.
Software does not make a record complete. FDA wrote to a device maker:
“Records of complaint investigations do not include required information, including any unique device identifier (UDI) or universal product code (UPC).” Sol-Millennium Medical, Inc., warning letter, 18 March 202413
Its procedures required the UDI on records printed from its “electronic Quality Management System (e-QMS)”.13 Under the QMSR, 820.35(a) still asks for any UDI or UPC.3
A new system inherits the gaps in the old one’s data. Olympus moved to TrackWise in November 2024 and dropped its complaint investigation tiers, but FDA found:
“your trending is based on 24 months of historical data which is still limited by the lack of investigation results on potential device performance issues.” Olympus Medical Systems Corporation, warning letter, 7 October 202614
Nor is software a response to an open finding. A sterile drug maker told FDA it was requiring all deviations to be initiated in TrackWise; FDA found the response inadequate because it did “not confirm whether the particulate investigation is complete”.15
8What changed in 2025 and 2026.
- QMSR. The QMSR took effect on 2 February 2026, and FDA now inspects device makers under Compliance Program 7382.850 instead of the Quality System Inspection Technique.16 Former sections such as 820.100 and 820.198 are gone; eQMS forms built on them need remapping.3 Quality audit, supplier audit and management review reports, exempt under the old 820.180(c), are now open to FDA inspection.9
- EU. Revised Chapter 4 and Annex 11 were consulted on until 7 October 2025.17 The draft Chapter 4 says records that exist electronically “should be signed electronically”, and a hybrid system “should be avoided”.18 On 26 November 2026 EudraLex still lists the 2011 texts.19 See the Annex 11 revision.
9How it runs in Seal.
Seal’s QMS covers deviations, CAPA, change control, documents, training, audits, suppliers, complaints and risk. Each event opens with its evidence, read from Seal or from the ERP, LIMS, MES and training systems you already run, and a corrective action can become a control in the workflow, released through change control with its own verification. Neil, Seal’s AI agent, assembles the evidence for an investigation and drafts the corrective change for your team to review; required people and signatures still control the decisions.
On the Seal platform, laboratory, manufacturing and quality work run on one record, and you can start with one quality workflow alongside the eQMS you have. The eQMS comparison lists what to check when choosing.
AQuestions and answers
What does eQMS stand for?
Electronic quality management system: software that runs and records a company’s quality processes, such as document control, training, deviations, CAPA, change control, complaints and audits, in place of paper forms and spreadsheets.
Is an eQMS required by GMP or the QMSR?
No. The regulations require a documented quality system and its records, not software, and FDA holds paper and electronic drug CGMP records to the same requirements. An eQMS is one way to meet them, with enforced workflows, linked records, audit trails and electronic signatures.
Does an eQMS have to be validated?
Yes, for its intended use, because it holds GMP or device quality records. The company that uses it answers for that validation even when the vendor hosts the system and supplies test evidence, which can be its starting point. FDA’s device guidance scales testing to the risk of each function, from unscripted testing for routing records to a scripted protocol where a failure could compromise safety.
What are examples of eQMS software?
Widely used eQMS products in life sciences include MasterControl, Veeva Vault QMS, TrackWise Digital and Qualio, and Greenlight Guru is built for medical devices. Some platforms, Seal among them, run quality processes on the same records as laboratory and manufacturing work. Compare them on how each record is linked and validated, not on the list of modules.
What does an eQMS administrator do?
Sets up accounts and roles, maintains configuration such as workflows and form fields, and supports users, with every access and configuration change approved and recorded. The administrator should not also own the records being controlled, so that no one can change both the data and the rules around it.
Can one eQMS serve both pharmaceutical and medical device work?
Yes, if it is configured for each set of rules. Medicines follow GMP and ICH Q10; devices follow ISO 13485 and, in the US, the QMSR, which adds its own requirements such as the contents of complaint records. The workflows, records and review steps differ even where the processes share a name.
References19 sources
References
- 1EudraLex Volume 4, Chapter 1: Pharmaceutical Quality System, European Commission, in operation from 31 January 2013, sections 1.4, 1.12 and 1.13. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap1_ 2013- 01_ en_ 0. pdf (read 26 November 2026) - 2ICH Q10, Pharmaceutical Quality System, ICH, Step 4, 4 June 2008, sections 2.1, 2.6, 2.7 and 3.2, and glossary. https:/
/ database. ich. org/ sites/ default/ files/ Q10%20Guideline. pdf (read 26 November 2026) - 321 CFR Part 820, Quality Management System Regulation, sections 820.10 and 820.35, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820 (read 26 November 2026) - 4Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, FDA, December 2018, questions 2, 4 and 5. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 5EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, in operation from 30 June 2011, principle, sections 1, 3.1, 4.5, 10, 11 and 12, and glossary. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 6EudraLex Volume 4, Chapter 4: Documentation, European Commission, January 2011, sections 4.3 and 4.5. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ chapter4_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 721 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, sections 211.22, 211.25, 211.80, 211.100, 211.188, 211.192, 211.194 and 211.198, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211 (read 26 November 2026) - 8EudraLex Volume 4, Chapter 9: Self Inspection, European Commission, sections 9.1 and 9.3. https:/
/ health. ec. europa. eu/ document/ download/ 07195808- d02e- 4d7a- b8f4- f84a83278b62_ en?filename=cap9_ en. pdf (read 26 November 2026) - 9Quality Management System Regulation: Frequently Asked Questions, FDA, content current as of 5 November 2026, question 8. https:/
/ www. fda. gov/ medical- devices/ quality- management- system- regulation- qmsr/ quality- management- system- regulation- frequently- asked- questions (read 26 November 2026) - 10Guidance for Industry: Part 11, Electronic Records; Electronic Signatures, Scope and Application, FDA, August 2003, section III.B. https:/
/ www. fda. gov/ media/ 75414/ download (read 26 November 2026) - 1121 CFR Part 11, Electronic Records; Electronic Signatures, sections 11.10 and 11.50, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 11 (read 26 November 2026) - 12Computer Software Assurance for Production and Quality Management System Software, Guidance for Industry and FDA Staff, FDA, 3 February 2026, sections V.A.2, V.A.5 and V.A.6, and Appendix A, Example 1. https:/
/ www. fda. gov/ media/ 188844/ download (read 26 November 2026) - 13Sol-Millennium Medical, Inc., warning letter, FDA, 18 March 2024. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ sol- millennium- medical- inc- 677524- 03182024 (read 26 November 2026) - 14Olympus Medical Systems Corporation, warning letter, FDA, 7 October 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ olympus- medical- systems- corporation- 729208- 10072026 (read 26 November 2026) - 15Jubilant HollisterStier General Partnership, warning letter, FDA, 28 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ jubilant- hollisterstier- general- partnership- 723537- 05282026 (read 26 November 2026) - 16Quality Management System Regulation (QMSR), FDA, content current as of 2 February 2026. https:/
/ www. fda. gov/ medical- devices/ postmarket- requirements- devices/ quality- management- system- regulation- qmsr (read 26 November 2026) - 17Stakeholders’ Consultation on EudraLex Volume 4, Chapter 4, Annex 11 and new Annex 22, European Commission, open 7 July to 7 October 2025. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 4- annex_ en (read 26 November 2026) - 18EudraLex Volume 4, Chapter 4: Documentation, draft for consultation, European Commission, July 2025, section 4.70. https:/
/ health. ec. europa. eu/ document/ download/ fa336a6e- 753b- 46fc- b53b- 9178bacd8878_ en?filename=mp_ vol4_ chap4_ consultation_ guideline_ en. pdf (read 26 November 2026) - 19EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026)