What are common data integrity violations in pharma?

FDA’s 2026 warning letters cite familiar violations: shared logins and analysts with administrator rights; audit trails off, or changes made outside them; trial injections of real samples; failing results dropped while a passing retest is reported; deleted data with no backup; original printouts thrown away; backdated or fabricated records; and uncontrolled blank forms. PIC/S grades falsification, or reporting a desired result instead of an actual out-of-specification QC result, as critical.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal beside a balance and a spectrometer, each cabled to one box of raw data.

1What counts as a data integrity violation.

FDA defines data integrity as “the completeness, consistency, and accuracy of data”, and expects data to be attributable, legible, contemporaneously recorded, original or a true copy, and accurate (ALCOA).1 EMA names “data deletion, amendment, duplication, reusing and fabrication” as “common data integrity failures”.2 A violation is any practice that breaks those expectations, whether it comes from intent, bad habit or a system that allows it. For the principles themselves, see what data integrity means.

Warning letters cite computer controls under 21 CFR 211.68(b), which requires changes “instituted only by authorized personnel”, and missing results under 211.194(a).3,4 All eight letters below find drugs adulterated under section 501(a)(2)(B) of the Federal Food, Drug, and Cosmetic Act; those to API makers, such as Curia and Shimoga, cite CGMP for APIs rather than Part 211.5,6,7

2The violations FDA cited in 2026.

Each row is a pattern from a warning letter of March to September 2026, what is expected, and where to look before an investigator does.

Table 1. Violations in FDA warning letters of March to September 2026, and where to look for each. Full letters are in the sources.
Violation2026 exampleWhat is expectedWhere to look
Shared loginsAva used one username and password on an impurity-testing HPLC;5 a Curia analyst used a colleague’s login;6 GC America kept a meter password on a sticker on the laptop8Actions must be attributable to one person; shared accounts only for read-only viewing1User lists of every CDS and stand-alone instrument; notes beside keyboards
Too much accessAva analysts could modify and delete data;5 Fareva operators could change system dates and alarm limits9Administrators independent of the record content1Role lists: who can delete data or change clocks and limits
Trial injectionsCuria analysts made thousands of unexplained Empower injections over 30 months;6 Ava kept an HPLC location for trial injections5A “violative practice” if an actual sample is used to disguise testing into compliance1CDS injections not tied to a reported result; aborted or incomplete injections, which FDA says should be “investigated and justified”1
Passing results onlyFareva kept only conforming particle count retests from Grade A areas;9 Shimoga reported a passing injection, not the out-of-specification one7Complete data from all tests4Sequences with more injections than results reported; retests with no OOS record
Audit trail gapsJaloma’s GC audit trails were not enabled;10 Intas’s vendor changed an electronic batch record outside the audit trail11Audit trails “enabled and locked at all times”12Audit trail settings and their history; vendor service tickets and emails
Deleted or lost dataAva’s deleted GC sequences sat in the recycle bin;5 a vendor calibration at Fareva deleted all earlier electronic data, with no backup;9 GC America’s meter data were deleted at shutdown8Backup exact, complete and secure3Recycle bins, local drives, backup logs, deletion events
Discarded originalsCuria’s original pH and balance printouts were in shred bins;6 Shimoga acknowledged discarded printouts7Originals or true copies kept1Shred bins and waste near balances and pH meters
Backdated or fabricatedHuons manipulated camera timestamps and replaced logbook pages;13 a Jaloma microbiologist recorded samples never taken10Falsification is a critical deficiency12Clock-change events; timestamps against camera and door logs; logbook pagination
Uncontrolled blank formsHuons had 1,897 blank uncontrolled CGMP-related forms13Forms controlled by the quality unit and reconciled1Form issue and reconciliation logs; loose forms in laboratories
Data not reviewedShimoga released a batch because its quality unit did not review electronic data7Second-person review of original records, audit trails included4,1Whether reviewers saw electronic data and audit trails or only printouts2

The patterns cluster: Ava’s one letter covers at least five of them, so one finding is a reason to check for the others.5 For the routine review that catches them, see audit trail review.

2.1What follows the findings.

  • Import alert. FDA put the sites of three of the eight letters on Import Alert 66-⁠40, detention without physical examination, each before its warning letter was issued: Huons on 3 April, Jaloma on 23 April and Shimoga on 2 June 2026.14
  • Approvals. All eight letters warn that FDA “may withhold approval of new applications or supplements listing your firm as a drug manufacturer”; see FDA 483 vs warning letter.11,6
  • Applications. Under the Application Integrity Policy, FDA can notify a firm that it is “deferring substantive scientific review” of its applications or proceeding to withdraw approved ones.15 Every pre-approval inspection audits the raw data behind an application; see pre-approval inspection.16

3What to do when you find one.

  1. Record it the day it is found. PIC/S says lapses should be handled “as any deviation would be”, and FDA says suspected falsification must be investigated under the quality system “Regardless of intent or how or from whom the information was received”.12,1
  2. Secure the evidence and stop the practice. Copy the electronic data, audit trails and paper records before anyone corrects anything, then close the gap, for example by disabling a shared account.
  3. Grade it. Decide whether the facts show falsification, misreporting or poor capture, or a bad practice confined to one area, using the PIC/S scale below.12
  4. Scope it beyond the incident. Check every system, product and period with the same weakness. FDA found Curia’s review of 12 stand-alone instruments over a year, and Intas’s review of emails to one software vendor, too narrow.6,11
  5. Decide who must be told. In the US, a distributed batch that fails its application specification needs a Field Alert Report within 3 working days of the applicant receiving the information, and a deviation that may affect a distributed licensed biologic needs a BPDR within 45 calendar days.17,18 In the EU, the manufacturer should report a quality defect “in a timely manner” to the marketing authorization holder and “all concerned Competent Authorities” where it may lead to a recall or “an abnormal restriction in the supply of the product” (Chapter 8.15).19 PIC/S lists client notification and reporting to regulators among corrections, and MHRA expects notification of significant incidents.12,20 Disclose fully at once: Huons first gave the investigator only part of the data, and its management reported “critical breaches in data integrity” the following day.13
  6. Work through FDA’s standard request. Most 2026 letters close their data integrity findings with the request below, and Q18 of FDA’s Q&A sets out the same approach, which it says mirrors the Application Integrity Policy.1 Use it as the checklist for an internal finding too.
Table 2. FDA’s standard data integrity remediation request in its 2026 warning letters.5,8,10,13,7,6,1
ItemWhat to includeLetters that ask
Investigation protocolEvery laboratory, operation and system covered, with a justification for anything excludedAva, Jaloma, Huons, Shimoga, Curia
InterviewsCurrent and former employees, preferably by a qualified third partyAva, Jaloma, Huons, Shimoga, Curia
Extent of lapsesOmissions, alterations, deletions, record destruction and records completed lateAva, Jaloma, Huons, Shimoga, Curia
Risk assessmentRisk to patients from released drugs and from ongoing operations; FDA’s Q&A adds the effect on data in submissionsAva, GC America, Jaloma, Huons, Shimoga, Curia
Root causesWhether those responsible can still influence CGMP or application dataAva, Jaloma, Huons, Shimoga, Curia
Interim measuresCustomer notification, recall, added testing, stability lots, application actions, complaint monitoringAva, Jaloma, Huons, Shimoga, Curia
CAPA effectivenessAnnual audits by a qualified consultant for at least two yearsAva, Jaloma, Huons, Shimoga
Integrity officerEmpowered to receive anonymous complaints from employeesAva, Jaloma, Huons, Shimoga, Curia

Keep the assessment reviewable. Huons told FDA its third-party assessment would be classified as an “internal audit” and its reports would not be provided. FDA disagreed: the assessment answered inspection findings, so it “is subject to routine FDA review”.13

4How inspectors word it.

“Your quality assurance employee instructed your software vendor to make changes to your electronic batch record which were not captured in the audit trail or managed through your quality system.”

That is Intas Pharmaceuticals’ warning letter of 30 March 2026, under 21 CFR 211.22. One change replaced the employee number recorded for “Dispensed by” in a batch record, and FDA asked for a review of all software-related communications with support vendors, correlated with audit trails.11

“Initial failing particle count results in Grade A areas were not recorded in batch records. Instead, retests were performed, and only conforming retest results were retained and included in production documentation.”

That is Fareva Amboise’s letter of 10 April 2026, under 21 CFR 211.68(b).9 EU inspectors cite the same failures:

“Audit trail was disabled from chromatographic sequences in QC analysis. The password complexity policy for software was not enforced with high protection requirements. Deliberate false information was given to the inspector for the password length setting set by IT administrator in the Agilent Open Labs software.”

That is the Malta Medicines Authority’s statement of non-compliance MT/002NCR/2026 of 5 May 2026, on Zenzi Pharmaceutical Industries in Thane, India, after an inspection ending 31 March 2026. It also found backdating, reissued logbook pages and sequence files analysts could rename or delete, and asked member states to consider not authorizing new marketing authorizations with the site listed as a finished product manufacturer.21

5How inspectors grade them.

PIC/S PI 041-⁠1 gives inspectors an indicative scale in section 11.2, by impact on the product and the patient.12

Table 3. Indicative data integrity grades in PIC/S PI 041-⁠1, section 11.2.
GradeExamples PIC/S gives
Critical“Product failing to meet Marketing Authorisation specification at release or within shelf life”; a ‘desired’ result reported instead of an actual out-of-specification result for QC tests or critical parameters
Critical, falsificationFraud, misrepresentation or falsification; wide-ranging falsification that “erodes all confidence” in the site’s medicines
MajorIn-specification results altered to give a more favorable trend; poorly designed capture, such as “using scraps of paper to record info for later transcription”; bad practices across a limited number of functional areas
OtherBad practice or poor design in a discrete area; “manipulation of non-critical data by an individual” in an otherwise acceptable system

Inspectors judge from data governance, system design, audit trails and user access whether a failure is company-wide or limited, and exacerbating or mitigating circumstances can change the grade.12

6Example: scoping a shared login found at self-inspection.

Example, invented: the company, systems, account and batch are illustrative. A contract manufacturer makes tablets for a client under the client’s ANDA. On 2 October 2026 a self-inspection finds a generic “QC-⁠LAB” account in use on HPLC-⁠04. The head of QA owns each decision below.

  • Scope. User lists of all 21 systems with local or CDS logins, 9 chromatography systems and 12 stand-alone instruments, show the account on 3 HPLCs. The CDS audit trail dates its creation to 1 April 2026, so the review covers the 185 days to 2 October. Systems that accept only site directory logins are excluded, with that justification recorded.10
  • Matching. Of 2,400 injections under the account, 2,280 match a sequence in an approved test record. Of the other 120, 5% of the total, 96 are labeled blanks and standards and 24 are sample injections.
  • Checking the labels. FDA asked Curia for “supporting documentation showing that the trial injections were not sample injections”, so each of the 96 chromatograms is compared with the method’s blank and standard profiles.6 92 match; 4 labeled as standards show a sample’s impurity peaks and join the sample injections, making 28.
  • Results. The 28 sample injections come from 6 batches, 4 of them released. 2 injections for batch L-⁠2291, released and distributed, are out of specification, though its reported result passed.
  • Grade. If confirmed, L-⁠2291 failed its specification at release and a ‘desired’ result was reported instead of an actual one: both critical examples in PIC/S 11.2.12
  • Actions. The account is disabled that day and individual logins confirmed on all 21 systems. A deviation covers the 3 systems and 185 days, and an OOS investigation under 211.192 covers the 2 injections, “whether or not the batch has already been distributed”.22 The client is told the day the 2 injections are found, under the quality agreement, since its Field Alert Report clock runs from when it receives the information.23,17 L-⁠2291 goes to a product impact assessment and recall decision, and the analysts are interviewed.

7Where violations start.

FDA wants a culture where “data integrity is an organizational core value”, and MHRA warns that the effect of “the behaviour driven by performance indicators, objectives and senior management behaviour” on data governance “should not be underestimated”.1,20 At Huons a team leader “had instructed staff to discard the bioburden plates with significant growth and had manipulated camera timestamps to create backdated documentation”; when a supervisor directs the practice, retraining analysts does not reach the cause, and FDA’s letters ask whether those responsible can still influence CGMP data.13,10

8How it runs in Seal.

In Seal, each record carries its own audit log, and the vendor audit-trail export for the same work, from a chromatography data system for example, is attached and reviewed beside it. The data integrity and audit trail review blueprint prioritizes events by criticality and can make the required review a gate on the result or disposition, so a change such as reintegration after an unexpected result is seen before approval and can be traced to every record and decision it affects.

Neil, Seal’s AI agent, can prepare an audit-trail review around defined review questions and exceptions, and compare chromatography integration changes, keeping the original and reprocessed versions in one source-linked review package. Neil does not select the most favorable result or approve reprocessing; an authorized reviewer decides.

AQuestions and answers

Is a shared read-only login acceptable?

For viewing data, FDA says yes: it accepts shared read-only accounts that cannot modify data or settings. Any action that must be attributable, such as a second-person review, needs an individual login. PIC/S, MHRA and WHO give no read-only exception, and PIC/S says shared passwords should be prohibited even to save money.

Is reprocessing chromatography a violation?

Not in itself. For most analyses, reprocessing should not be needed regularly, and when it is, it must follow written procedures with every result kept for review. The violation is reprocessing until a result passes and keeping only that result, or changing integration without a recorded, reviewed justification.

Is a trial injection ever allowed?

Not of a real sample to preview the result. FDA treats using an actual sample in test, prep or equilibration runs to disguise testing into compliance as a violative practice. System suitability should use standards; if an actual sample is used, it should be a properly characterized secondary standard from a different batch, under written procedures.

What should we do if a supplier gets a data integrity warning letter?

EMA expects you to assess the risk to your products made or released with that contractor, using quality risk management, and to make the assessment available to inspectors. Depending on the outcome, that can mean removing the contractor from your approved list. If supply could be disrupted, consult the relevant regulators.

How far back should a data integrity investigation go?

Back to when the weakness began, which an audit trail or an account’s creation date often shows, and across every system with the same weakness. FDA judges the period against the facts. In September 2026 it found Curia’s review of 12 stand-alone instruments over about a year too narrow and asked for an independent assessment of all GMP-integrated computerized systems over three years. In June 2026 it asked Huons for an independent review of every invalidated OOS result for US products from the three years before the inspection.

How do you protect data integrity in a microbiology laboratory?

Treat a count read by one person as critical data. PIC/S says added controls should be considered when one individual makes a critical test interpretation, such as counting colonies on agar plates: a second review, sometimes in real time, or electronic evidence such as photographs of the plates. MHRA asks for manual readings such as environmental monitoring plates to be risk assessed, with a second contemporaneous check or another way of capturing the result where justified. In 2026 FDA cited Huons for discarding bioburden plates with significant growth, and Jaloma for a microbiologist who recorded samples never taken and counted colonies before the required incubation time.

References23 sources

References

  1. 1Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, FDA, December 2018. https://www.fda.gov/media/119267/download (read 26 November 2026)
  2. 2Guidance on good manufacturing practice and good distribution practice: questions and answers, data integrity section, European Medicines Agency. https://www.ema.europa.eu/en/human-regulatory-overview/research-development/compliance-research-development/good-manufacturing-practice/guidance-good-manufacturing-practice-good-distribution-practice-questions-answers (read 26 November 2026)
  3. 321 CFR 211.68, Automatic, mechanical, and electronic equipment, eCFR. https://www.ecfr.gov/current/title-21/section-211.68 (read 26 November 2026)
  4. 421 CFR 211.194, Laboratory records, eCFR. https://www.ecfr.gov/current/title-21/section-211.194 (read 26 November 2026)
  5. 5Warning Letter to Ava Inc., 320-⁠26-⁠64, MARCS-⁠CMS 721180, FDA, 14 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/ava-inc-721180-04142026 (read 26 November 2026)
  6. 6Warning Letter to Curia New York, Inc., 320-⁠26-⁠127, MARCS-⁠CMS 728790, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 26 November 2026)
  7. 7Warning Letter to Shimoga Chemicals, 320-⁠26-⁠101, MARCS-⁠CMS 727904, FDA, 13 July 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/shimoga-chemicals-727904-07132026 (read 26 November 2026)
  8. 8Warning Letter to GC America, Inc., 320-⁠26-⁠76, MARCS-⁠CMS 727602, FDA, 14 May 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/gc-america-inc-727602-05142026 (read 26 November 2026)
  9. 9Warning Letter to Fareva Amboise, MARCS-⁠CMS 723502, FDA, 10 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/fareva-amboise-723502-04102026 (read 26 November 2026)
  10. 10Warning Letter to Laboratorios Jaloma S.A. de C.V., 320-⁠26-⁠78, MARCS-⁠CMS 725939, FDA, 22 May 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/laboratorios-jaloma-sa-de-cv-725939-05222026 (read 26 November 2026)
  11. 11Warning Letter to Intas Pharmaceuticals Limited, 320-⁠26-⁠57, MARCS-⁠CMS 721151, FDA, 30 March 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/intas-pharmaceuticals-limited-721151-03302026 (read 26 November 2026)
  12. 12PI 041-⁠1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, PIC/S, 1 July 2021. https://picscheme.org/docview/4234 (read 26 November 2026)
  13. 13Warning Letter to Huons Co., Ltd., 320-⁠26-⁠95, MARCS-⁠CMS 724650, FDA, 15 June 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/huons-co-ltd-724650-06152026 (read 26 November 2026)
  14. 14Import Alert 66-⁠40, Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs, FDA, published 25 November 2026. https://www.accessdata.fda.gov/cms_ia/importalert_189.html (read 26 November 2026)
  15. 15Application Integrity Policy, FDA. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/compliance-actions-and-activities/application-integrity-policy (read 26 November 2026)
  16. 16Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026. https://www.fda.gov/media/193382/download (read 26 November 2026)
  17. 1721 CFR 314.81(b)(1), NDA field alert report, eCFR. https://www.ecfr.gov/current/title-21/section-314.81 (read 26 November 2026)
  18. 1821 CFR 600.14, Reporting of biological product deviations by licensed manufacturers, eCFR. https://www.ecfr.gov/current/title-21/section-600.14 (read 26 November 2026)
  19. 19EudraLex Volume 4, Part I, Chapter 8: Complaints, Quality Defects and Product Recalls, European Commission, 13 August 2014, in operation from 1 March 2015. https://health.ec.europa.eu/system/files/2016-11/2014-08_gmp_chap8_0.pdf (read 26 November 2026)
  20. 20‘GXP’ Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 26 November 2026)
  21. 21Statement of non-compliance with GMP, MT/002NCR/2026, Zenzi Pharmaceutical Industries Private Limited, Malta Medicines Authority, EudraGMDP, 5 May 2026; to open it, search non-compliance reports issued 1 to 10 May 2026 and choose that report number. https://eudragmdp.ema.europa.eu/inspections/gmpc/searchGMPNonCompliance.do (read 25 November 2026)
  22. 2221 CFR 211.192, Production record review, eCFR. https://www.ecfr.gov/current/title-21/section-211.192 (read 26 November 2026)
  23. 23Contract Manufacturing Arrangements for Drugs: Quality Agreements, Guidance for Industry, FDA, November 2016. https://www.fda.gov/media/86193/download (read 26 November 2026)