1Part 11 requirements checklist.
Each line is one requirement of the regulation, in section order.1 “Discretion” marks the four areas (validation, audit trails, retention and copies) where FDA’s 2003 guidance does not intend to enforce Part 11 itself; the predicate rules, the regulations that require the record, such as Part 211, still apply.2 It also does not intend to enforce any of Part 11 for systems operational before 20 August 1997 that meet its conditions.2
| Section | Requirement | Evidence to keep | 2003 policy |
|---|---|---|---|
| 11.1(e) | Systems, controls and their documentation readily available for FDA inspection | Document index; who demonstrates the system | Enforced |
| 11.10(a) | Validation for accuracy, reliability, consistent intended performance and spotting invalid or altered records | Report tracing each intended use to its tests and results | Discretion |
| 11.10(b) | Accurate, complete copies, human readable and electronic, for FDA | Tested export, with metadata and audit trail | Discretion |
| 11.10(c) | Records retrievable throughout the retention period | Restore test: what was restored, when, and the comparison | Discretion |
| 11.10(d) | Access limited to authorized people | Access list: user, role, date granted, approver | Enforced |
| 11.10(e) | Secure, time-stamped audit trail of creation, change and deletion that keeps earlier entries visible, kept as long as the record | Settings showing it on and locked; an entry with user, date and time, old and new value | Discretion |
| 11.10(f) | Operational checks enforce the permitted sequence of steps, as appropriate | Test of an out-of-order step | Enforced |
| 11.10(g) | Authority checks on who can use the system, sign, alter a record or operate | Role matrix and refused-permission tests | Enforced |
| 11.10(h) | Device checks on the source of input, as appropriate | Instrument check, or why none is needed | Enforced |
| 11.10(i) | Developers, maintainers and users have the education, training and experience for their tasks | Training records and CVs, including IT and supplier staff | Enforced |
| 11.10(j) | Written policy holding people accountable for their signatures | Policy and each signer’s dated acknowledgment | Enforced |
| 11.10(k)(1) | Control over system documentation | Controlled manuals and procedures | Enforced |
| 11.10(k)(2) | Change control with an audit trail of documentation changes | Change records | Discretion |
| 11.30 | Open systems: the 11.10 controls plus measures such as encryption | Open or closed rationale; encryption in transit | As for 11.10 |
| 11.50 | Signed record shows printed name, date and time, and meaning | Screenshot and printout of a signed record | Enforced |
| 11.70 | Signatures cannot be excised, copied or transferred by ordinary means | Test that a copied record carries no valid signature | Enforced |
| 11.100(a), (b) | Each signature unique to one person, never reused by or reassigned to anyone else; identity verified first | Rule against reusing IDs; identity check at set-up | Enforced |
| 11.100(c) | Certification to FDA, before or at first use, that the signatures are the legally binding equivalent of handwritten ones; the letter is signed by hand, sent electronically or on paper | Copy of the letter | Enforced |
| 11.100(c)(2) | Further certification or testimony on a specific signature, if FDA asks | Who answers, and how | Enforced |
| 11.200(a)(1) | At least two components, such as ID and password. In one continuous session, all at the first signing, then at least one “only executable by, and designed to be used only by, the individual”; outside one, all at every signing | Sign twice in a session; time out, sign again | Enforced |
| 11.200(a)(2) | Used only by their genuine owners | Policy against sharing; signers’ acknowledgments | Enforced |
| 11.200(a)(3) | Using someone else’s signature needs two or more people to collude | Reset procedure: no one person can sign as another | Enforced |
| 11.200(b) | Biometric signatures usable only by their owner | Enrollment and rejection tests | Enforced |
| 11.300(a) | No two people share an ID and password combination | Account uniqueness check | Enforced |
| 11.300(b) | Codes and passwords periodically checked, recalled or revised | Password aging settings | Enforced |
| 11.300(c), (e) | Lost tokens deauthorized; tokens tested initially and periodically | Loss procedure; test records | Enforced |
| 11.300(d) | Safeguards against misuse; attempts reported urgently | Lockout and alert settings | Enforced |
Discretion is narrower than it sounds. FDA’s 2018 Q&A calls each CGMP workflow “an intended use of a computer system to be checked through validation”.3 For device makers’ production and quality system software, FDA says the discretion “expressly does not apply” to validation under ISO 13485, which the QMSR made part of Part 820.4,
1.1Recording the assessment.
Keep one table per system, headed with its name, version and the predicate rules behind its records. Give every N/A a written reason, as FDA recommends documenting scope decisions: 11.10(f) and (h) apply only “as appropriate”, and 11.30 only to open systems.1,
| Section | Applies, or N/A and why | Evidence | Status | Owner | Reviewer, date |
|---|---|---|---|---|---|
| 11.10(e) | Applies | Audit trail settings export; validation test reference | Met | System owner | QA, dated |
| 11.30 | N/A: closed system; QC and IT control access | System description | N/A | System owner | QA, dated |
| 11.50(a)(1) | Applies | Printout of a signed report | Gap: shows user ID, not name; CC-26-031 | Process owner | QA, dated |
Hosted is not the same as open: a system is closed when the people responsible for its records control access.1 For cloud and other IT services, FDA’s clinical Q&A says the regulated entity remains responsible for Part 11, and recommends a written agreement after evaluating the provider’s validation, access controls, audit trails, backup and retention.6 See what makes software Part 11 compliant.
2What to have ready for an inspector.
For each system, have these to hand.
- The record inventory: each record, its predicate rule, and whether people rely on the electronic record or a printout, decided in advance and documented.2
- The user list, with roles, dates granted and administrator rights. FDA wants administrators “independent from those responsible for the record content”.3
- Proof that every login is unique, and that any shared account is read-only.3
- The audit trail settings, and the audit trail review for a released batch, done by the record’s reviewer before release.3
- Who can change the system date and time; the Fareva letter below cites it.7
- Deleted data, each item explained. At Ava Inc., investigators found “multiple deleted gas chromatography (GC) analytical sequences in the recycle bin”.8
- The validation report for each workflow in use.3
- A copy of the certification letter. One letter can cover “all the electronic signatures used by that organization”,6 including people “located anywhere in the world”, and a mailed copy is now optional.9 A CDMO whose staff sign in its own system certifies for them; ask for its copy.1
- Before a pre-approval inspection, the raw data behind each result filed in the application. A data integrity audit is an objective of every NDA and ANDA PAI; FDA names stability, dissolution, content uniformity and API impurity data as “good candidates”, and unjustified “absences in a submitted chromatographic sequence” as a concern.10 CP 7346.832M does the same on every prelicense and pre-approval inspection of a CDER-regulated biologic, and has inspectors verify “that electronic signatures are not shared”.11
2.1What inspectors cite.
On Form 483s, inspectors usually cite the predicate rule. In FDA’s FY2025 data on 483s generated by its inspection software, no observation in any program area cites Part 11. On drug 483s, 21 CFR 211.68(b) appears 110 times, 87 of them for not limiting changes to records to authorized people. Manually prepared 483s are not counted.12 Both letters below cite 211.68(b).
“your firm utilized a common username and password to access high performance liquid chromatography (HPLC) equipment used for impurity testing of drug products. Additionally, analysts had administrator privileges to modify and delete data.” Ava Inc., warning letter, 14 April 20268
The firm said a software update had let users bypass its access restrictions, a reason to retest access after every update.8
“operators could modify system dates, adjust alarm limits, and access settings without restriction. The equipment displayed that network user authentication was not enforced, and management confirmed that no controls were in place to ensure the integrity of printed data records.” Fareva Amboise, warning letter, 10 April 20267
Both letters ask for an independent assessment of every system’s configuration, administrator rights, password controls, audit trail, validation, deviation history and backup.8,
3Does Part 11 apply to this record?
Part 11 applies record by record, not system by system, and FDA reads its scope narrowly. Answer these questions for each record a system holds, and document the decisions, for example in an SOP, as FDA recommends.2 A record an AI tool creates or changes is in scope like any other (validating AI for GxP).1 For drug makers, Part 211 is the main predicate rule, with its own computer controls in 211.68.13
- Does a predicate rule require the record, or will it be submitted to FDA? If neither, it is not a Part 11 record.1,
2 FDA’s 2026 software assurance guidance for device makers gives a Part 820 example: activity logs a program saves at start-up, not needed as evidence of a validated state, fall outside Part 11.4 - Is it kept electronically in place of paper? Then Part 11 applies.2
- Is it kept both ways? Part 11 applies if people rely on the electronic record to perform regulated work, and generally not if they rely on a printout that meets the predicate rule. FDA “may take your business practices into account”, so what people actually do decides.2
- Is the original dynamic? An FT-IR spectral file can be reprocessed, so FDA says a printout of it “would not satisfy CGMP requirements to retain original records or true copies”. A balance printout can be the original.3
- Does a predicate rule require a signature on it? Electronic signatures recording that something was approved, reviewed or verified are Part 11 signatures.2 A signature drawn with a finger or stylus counts as handwritten, but must still be linked to its record.6
4Example: a Part 11 assessment of one system.
This assessment is invented: site, system, product, people, numbers and dates are illustrative. Each finding names the predicate rule, which is what inspectors cite, beside the Part 11 section.
| Field | Entry |
|---|---|
| System | CDS-02, chromatography data system, Quarnley QC laboratory: release and stability testing of Exemplol 50 mg tablets. Closed: QC and IT control access.1 |
| Roles | Process owner: QC manager. System owner: IT lead, as Annex 11 defines them.14 QA approves. |
| Predicate rules | 21 CFR 211.194(a), laboratory records;15 211.180(d), originals or true copies;16 211.68(b), computer controls13 |
| Decision | Part 11 records and signatures: chromatograms can be reprocessed, so a printout cannot be the original.3 |
| Validation | VR-CDS-02 revision 3 covers the release and stability workflows: 11.10(a), 211.68(b).3 |
| Audit trail | On for every project and locked; entries show user, time, old and new value: 11.10(e). |
| Clock | Synchronized to the site time server; only the 2 IT administrators can change it, each change logged: 211.68(b). |
| Backup | Nightly. A full restore of one project on 12 February 2026 matched record counts: 11.10(c), 211.68(b). |
| Signatures | “Performed by” and “Reviewed by”, per 211.194(a)(7) and (8);15 company-wide certification letter on file: 11.100(c). |
| Accounts | 15: 12 named analysts, 2 IT administrators and 1 shared “LAB” account used to start instruments |
| Finding 1 | LAB could open and run sequences, so those actions had no single author: 211.68(b), 211.194(a)(7), 11.10(d) and (g).3 Disabled on 10 February under CAPA-26-014. |
| Finding 2 | 3 analysts held administrator rights and could change processing methods: 211.68(b), 11.10(g).3 Removed under CAPA-26-015. |
| Finding 3 | Printed reports show the user ID, not the printed name: 11.50(a)(1). Template changed under CC-26-031. |
| Finding 4 | Reason for change optional. Where an edit modifies an established test method, 211.194(b) requires records to “include the reason for the modification”;15 Annex 11, section 9, expects one for any change to GMP-relevant data.14 Made mandatory under CC-26-032. |
| Not a finding | A second signing in one session asks only for the password, the component only the signer can execute, as 11.200(a)(1)(i) allows. |
| Reviews | Audit trail with each result, before release;3 access quarterly against training records: 11.10(d), (i). |
| Closed | 11 May 2026, by QA, after the look-back and a re-review found no shared or excess accounts |
The look-back, DEV-26-022, ran because 211.192 extends an investigation to other batches:17
- Scope: 1 August 2025, when LAB was created, to 10 February 2026, when it was disabled.
- Work: the 212 LAB sequences were matched to the analyst on each day’s instrument log; all 47 reprocessings by the 3 analysts had a second-person review.
- Outcome: no reportable result changed for the 18 batches released. Effectiveness check: the re-review on 11 May 2026, 90 days after Finding 1.
5How EU GMP and the MHRA differ.
Annex 11 is EU GMP guidance, and its 2011 text differs from Part 11 both ways. It asks more on reasons: “For change or deletion of GMP-relevant data the reason should be documented.” It asks less on the audit trail itself, to be considered “based on a risk assessment”, where 11.10(e) requires one. Its electronic signatures need no letter to a regulator.14 The draft revision consulted on from 7 July to 7 October 2025, not yet in force, would have audit trails “enabled and locked at all times”.18,
The MHRA says shared logins “should not be used” for systems that generate, amend or store GxP data,21 and PIC/S guidance for inspectors says shared passwords, “even for reasons of financial savings, should be prohibited”.22 On signatures, the MHRA says an inserted signature image, or a footnote saying a document was signed electronically, “is not adequate” unless it comes from the validated signing process. Its signature manifestation adds the signer’s title and needs the time only “if significant”; 11.50 asks no title but always the time.21,
6What has changed since 2003.
FDA’s 2003 guidance still sets the enforcement policy in the checklist.2
- 2 March 2023. Certification letters may be sent electronically, the only amendment to 11.10 to 11.300 since 1997.1
- 2 October 2024. FDA’s final clinical Q&A replaced its May 2007 guidance.23
- 2 February 2026. The QMSR took effect.5
- 3 February 2026. FDA’s final computer software assurance guidance, with a section on Part 11, superseded the September 2025 version.4
7How it runs in Seal.
Seal keeps version history, audit trails and role-based access with every record, and an electronic signature keeps the signer, the time and the meaning of the approval, linked to the record. The technical guide sets out what qualification should demonstrate in your configured workflow: who created or changed a record and when, required signatures and authority checks, and the effect of changing an approved record, including for restricted users.
No software alone makes an operation compliant: your quality system, configuration, procedures and training remain part of the assessment. Regulatory and validation covers Part 11, Annex 11 and GAMP 5 together, and the audit trail review blueprint reviews changes beside the records they affect.
AQuestions and answers
When did 21 CFR Part 11 take effect?
FDA published Part 11 on 20 March 1997 and it took effect on 20 August 1997. FDA’s narrower reading of its scope dates from guidance issued in August 2003.
Does Part 11 apply to medical devices?
Yes, to records the device regulations require that are kept electronically. FDA’s 2026 software assurance guidance says a document required under Part 820 and maintained in electronic form would generally be an electronic record under Part 11.
What counts as a biometric signature under Part 11?
Part 11 defines biometrics as verifying identity by measuring a physical feature or repeatable action that is unique to the person and measurable. FDA’s 2024 clinical Q&A gives fingerprints, hand geometry, iris and retinal patterns and voice prints as examples. A biometric signature must be designed so that no one but its owner can use it.
Does Part 11 require passwords to expire?
Not in those words. Section 11.300(b) requires identification codes and passwords to be periodically checked, recalled or revised, and gives password aging as an example. Because the rule allows checking, recalling or revising, a forced change on a fixed cycle is one way to meet it, not the only one.
Does Part 11 apply to paper records?
No. Part 11 says it does not apply to paper records that are, or have been, transmitted by electronic means, such as a faxed record. A printout can also keep a record outside Part 11 if it meets the predicate rule and people rely on the paper to do their work. If they rely on the electronic version instead, or the original is dynamic data such as a chromatogram, Part 11 applies.
Does Part 11 set a date and time format?
No. Section 11.50 requires a signed record to show the date and time the signature was executed, but sets no format. For systems that span time zones, the 1997 preamble to Part 11 advised recording the signer’s local time, but FDA’s 2003 guidance says it does not expect that and that system documentation should explain the time zone reference used. FDA’s 2024 clinical Q&A asks sponsors to state the time zone of each time stamp or record times in GMT. Pick one unambiguous format, state it in the system documentation and use it everywhere.
References23 sources
References
- 121 CFR Part 11, Electronic Records; Electronic Signatures, sections 11.1 to 11.300, as amended to 2 March 2023 (88 FR 13018), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 11 (read 26 November 2026) - 2Guidance for Industry: Part 11, Electronic Records; Electronic Signatures, Scope and Application, FDA, August 2003. https:/
/ www. fda. gov/ regulatory- information/ search- fda- guidance- documents/ part- 11- electronic- records- electronic- signatures- scope- and- application (read 26 November 2026) - 3Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, FDA, December 2018, questions 3, 4, 5, 7, 8 and 10. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 4Computer Software Assurance for Production and Quality Management System Software, guidance for industry and FDA staff, FDA, 3 February 2026, section V.B. https:/
/ www. fda. gov/ media/ 188844/ download (read 26 November 2026) - 5Quality Management System Regulation (QMSR), FDA, updated 2 February 2026. https:/
/ www. fda. gov/ medical- devices/ postmarket- requirements- devices/ quality- management- system- regulation- qmsr (read 26 November 2026) - 6Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers, guidance for industry, FDA, October 2024, section III.C and questions 12, 17, 25 and 29. https:/
/ www. fda. gov/ media/ 166215/ download (read 26 November 2026) - 7Fareva Amboise, MARCS-CMS 723502, warning letter, FDA, 10 April 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ fareva- amboise- 723502- 04102026 (read 26 November 2026) - 8Ava Inc., MARCS-CMS 721180, warning letter, FDA, 14 April 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ ava- inc- 721180- 04142026 (read 26 November 2026) - 9Letters of Non-Repudiation Agreement, FDA, content current as of 16 July 2026. https:/
/ www. fda. gov/ industry/ getting- started- esg- nextgen/ letters- non- repudiation- agreement (read 26 November 2026) - 10Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026, implemented 10 August 2026, Objective 3 and its coverage table. https:/
/ www. fda. gov/ media/ 193382/ download (read 26 November 2026) - 11Compliance Program 7346.832M, Prelicense and Preapproval Inspections of CDER-Regulated Biological Product Manufacturers, FDA, issued 14 April 2026, implemented 14 May 2026, Part II.1B(4) and Part III, Objective 2. https:/
/ www. fda. gov/ media/ 191983/ download (read 26 November 2026) - 12Inspection Observations, FY 2025 data set, FDA, all program area tabs. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ inspection- references/ inspection- observations (read 26 November 2026) - 1321 CFR 211.68, Automatic, mechanical, and electronic equipment, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 68 (read 26 November 2026) - 14EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, January 2011, sections 2, 9 and 14 and glossary. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 1521 CFR 211.194, Laboratory records, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 194 (read 26 November 2026) - 1621 CFR 211.180, General requirements (records and reports), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 180 (read 26 November 2026) - 1721 CFR 211.192, Production record review, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 192 (read 26 November 2026) - 18Stakeholders’ Consultation on EudraLex Volume 4: Chapter 4, Annex 11 and New Annex 22, European Commission, 7 July to 7 October 2025. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 4- annex_ en (read 26 November 2026) - 19EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 26 November 2026) - 20Annex 11: Computerised Systems, draft for consultation, European Commission, July 2025, section 12. https:/
/ health. ec. europa. eu/ document/ download/ 40231f18- e564- 4043- 94de- c031f813d38b_ en?filename=mp_ vol4_ chap4_ annex11_ consultation_ guideline_ en. pdf (read 26 November 2026) - 21GXP Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018, sections 6.14 and 6.16. https:/
/ assets. publishing. service. gov. uk/ media/ 5aa2b9ede5274a3e391e37f3/ MHRA_ GxP_ data_ integrity_ guide_ March_ edited_ Final. pdf (read 26 November 2026) - 22PI 041-1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, PIC/S, 1 July 2021, section 9.5. https:/
/ picscheme. org/ docview/ 4234 (read 26 November 2026) - 23Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers; Guidance for Industry; Availability, Federal Register, 2 October 2024. https:/
/ www. govinfo. gov/ content/ pkg/ FR- 2024- 10- 02/ html/ 2024- 22562. htm (read 26 November 2026)