What is cleaning validation, and what do FDA and EU GMP require?

Cleaning validation is documented evidence that an approved cleaning procedure reproducibly removes the previous product, cleaning agents and microbial contamination from product-contact equipment to below a scientifically justified limit, in the EU one based on toxicology. It has three parts: design the cleaning process, prove it in a predetermined number of protocol runs at worst case, and monitor it. In the US it rests on 21 CFR 211.67; EU GMP Annex 15 and PIC/S PI 006-⁠4 set out the detail.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal comparing two equipment versions, each linked to its retained evidence.

1Cleaning validation, step by step.

Residues must fall below a maximum allowable carryover (MACO), which in the EU rests on a health-based exposure limit (HBEL) such as the permitted daily exposure (PDE).1,2,3 MACO and PDE limits shows the calculation.

StepWhat happensRecordSource
1. Risk assessmentProducts, equipment, exposure limits and cleaning difficulty set the scope and effortRisk assessment, product–equipment matrixPI 006-⁠4 7.3.1 and 7.4.1;2 Annex 15 10.51
2. DesignWrite the procedure; find the critical parameters and hardest-to-clean partsCleaning SOP, equipment assessmentPI 006-⁠4 7.1.5(a) and 7.6.12
3. Limits and methodsSet residue, cleaning agent and microbial limits; validate methods; show recovery on every materialLimit calculation, method validation, recovery reportAnnex 15 10.6 and 10.12;1 PI 006-⁠4 7.11 and 7.122
4. ProtocolApproved before any runProtocolFDA 1993 guide;4 PI 006-⁠4 7.4.2 and 7.4.32
5. RunsClean at worst case; inspect dry; sample and test; investigate any failureCleaning records, results, deviationsAnnex 15 10.13;1 PI 006-⁠4 7.3.11 and 7.8.12
6. ReportWhether the process met the criteria, any limitations, and the ongoing checks that followApproved final reportFDA 1993 guide;4 PI 006-⁠4 7.4.42
7. MonitoringRisk-based residue monitoring and trending; changes through change controlMonitoring results and reviewsFDA Q&A 8 and 12;5 PI 006-⁠4 7.13;2 Annex 15 111

1.1How many runs, and when.

Neither FDA nor EU GMP sets a number: Annex 15 asks for “an appropriate number of times based on a risk assessment”, though Health Canada notes that “a three-clean assessment has long been the industry norm”.1,6 PI 006-⁠3 asked for “at least three consecutive applications”, as WHO’s 2019 guideline still does; PI 006-⁠4 wants a predetermined number, set in the protocol by documented risk assessment (7.1.5(b), 7.1.6, 7.4.2(f)), so an SOP citing PIC/S for three runs needs a new rationale.7,8,2

Scale matters too. FDA’s Q&A says firms “should not pre-clean”, and “batches significantly smaller than full-scale would not offer sufficient assurance”.5 PI 006-⁠4 lets studies start in development only if the process and formulation “are equivalent to the intended commercial production”, and by commercial production at the latest; meanwhile, a risk assessment decides whether equipment is released on verification data.2 So a clean after a much smaller engineering or clinical batch is verification, not a validation run; Annex 15 may expect verification after each batch for investigational products (10.3).1

2What FDA and EU GMP require.

The US regulations, 21 CFR 211.67 and 211.182, never say “validation”; FDA’s nonbinding Q&A does: “Firms must validate that cleaning procedures are adequate to ensure that cross-contamination does not occur”. Its 1993 guide for investigators “does not bind FDA”.9,10,5,4 EU Annex 15 is guidance for interpreting the GMP principles, and MHRA lists the same guide.1,11 PI 006-⁠4’s cleaning section covers finished products and is “supplementary, optional guidance for APIs”; cells marked PIC/S rest on it alone.2

QuestionFDAEU GMP, and PIC/S where marked
Cleaning procedureWho cleans, schedules, methods with disassembly, removing previous batch identification, protecting clean equipment, inspection just before use9Written cleaning procedures; processing instructions check that equipment is clean12
Cleaning recordsEquipment log: date, time, product and lot, signed by the cleaner and a second person, or the verifier of automated cleaning10Chronological logbooks;12 PIC/S: records of equipment, person, time, SOP and previous product, signed by cleaner, any supervisor and visual inspector2
Basis of limits“logical”, and “practical, achievable, and verifiable”;4 documented to be safe5A toxicological evaluation (HBEL), plus cleaning agents and the whole train1,3
Visual checkEnough between batches of the same product4Not the only criterion1
Rinse samplesNot acceptable alone; measure the surface where feasible5Swab and/or rinse, depending on the equipment;1 PIC/S: swab where feasible2
Retesting until clean“acceptable only in rare cases”4“not considered an acceptable approach”1
HoldsSample equipment stored uncleaned longer than validated5Define dirty and clean hold times1
After validationNo test after every clean, except perhaps manual cleaning; risk-based monitoring5Testing at each changeover in shared facilities unless risk assessment justifies less;13 manual cleaning confirmed at a justified frequency1
Potent productsDedicated equipment not specifically required5Dedicated equipment or other appropriate measures where cleaning is ineffective;1 PIC/S: may be required for a very low HBEL2

For active substances, ICH Q7 12.7 asks for limits “based on the most deleterious residue”.14

3Cleaning validation protocol template.

SectionWhat to stateSource
Scope and rolesWhat is validated, for which products and equipment; who performs, reviews and approvesPI 006-⁠4 7.4.2(a) and (b);2 FDA 1993 guide4
EquipmentEach item, and the reason for any grouping of similar itemsPI 006-⁠4 7.4.2(c);2 Annex 15 10.11
Worst caseThe product or products chosen, why, and every product the study coversAnnex 15 10.10;1 PI 006-⁠4 7.3.4 and 7.4.2(l)2
Cleaning procedureThe procedure for each product or train, its critical parameters, and any worst-case challenges, such as minimum detergent contact timePI 006-⁠4 7.4.2(e);2 Health Canada 7.26
Automated or manualAutomated: the sequence and the validated operating ranges. Manual: step-by-step instructions, training, and how often effectiveness is confirmedAnnex 15 10.4 and 10.15;1 PI 006-⁠4 7.3.7 and 7.3.82
Holds and campaignsMaximum dirty hold, clean hold and campaign length, in time and batchesAnnex 15 10.8 and 10.9;1 PI 006-⁠4 7.3.9 and 7.4.2(d)2
Number of runsHow many consecutive cleans, and whyPI 006-⁠4 7.4.2(f);2 Annex 15 10.131
SamplingSwab, rinse or both, with the reason for eachPI 006-⁠4 7.4.2(g) and 7.9;2 Annex 15 10.121
LocationsHardest-to-clean and representative sites, and non-contact parts residue may migrate from, each with its reasonPI 006-⁠4 7.3.2, 7.4.2(h) and 7.9.2;2 Annex 15 10.111
RecoveryRecovery for every product-contact material and methodPI 006-⁠4 7.4.2(i);2 Annex 15 10.121
Analytical methodsValidated methods, with detection and quantitation limits; how long and how samples may be held before analysisPI 006-⁠4 7.4.2(j) and 7.11;2 FDA 1993 guide4
Acceptance criteriaResidue, cleaning agent, microbial, endotoxin where relevant, and visual criteria, the rationale for each, the visible threshold where visual inspection is relied on, train totals and unit-dose checksPI 006-⁠4 7.4.2(k), 7.8.3, 7.12.3 and 7.12.4;2 Annex 15 10.2, 10.6 and 10.71
FailuresHow every failure, visual ones included, is investigated; no testing until cleanPI 006-⁠4 7.3.11 and 7.3.12;2 Annex 15 10.21
ApprovalProduction, quality unit and other departmentsPI 006-⁠4 7.4.32

4Example: a sampling plan and its results.

This protocol extract is invented.

Table 4. Example. A cleaning validation sampling plan, invented for illustration.
FieldEntry
ProtocolCVP-⁠26-⁠014, line 3 granulation and compression train, approved 6 October 2026 by production, QC and QA
Worst caseExampladine 5 mg tablets: lowest PDE on the line, 10 µg/day. The least soluble product is a second worst case, in CVP-⁠26-⁠015, because Annex 15 weighs “solubility, cleanability, toxicity and potency”1
Next productExemplol tablets, the pair giving the lowest limit: smallest batch 100 kg, tablet weight 50 mg, maximum daily dose 8 tablets, or 400 mg
Health-based limit10 µg/day × 100 kg ÷ 400 mg/day = 2.5 g over 25 m² of shared surface = 10 µg/cm², or 250 µg per 25 cm² swab
Acceptance limit100 µg per swab: the site’s historical 10 ppm limit, which is tighter (10 mg/kg × 100 kg = 1 g over 25 m² = 4 µg/cm²), so the whole train carries at most 1 g
Press parts13 µg per swab on the punch tips and feed frame (1,200 cm²): half the PDE-based 26 µg, which assumes all their residue goes into the first 1,000 tablets: 10 µg/day ÷ 8 tablets × 1,000 = 1,250 µg; ÷ 1,200 cm² × 25 cm² = 26 µg
Alert limit48 µg per swab at the 100 µg sites: mean plus three standard deviations (18 + 3 × 10) of 40 recovery-corrected 2025 verification results, same sites and method
Detergent limit10 µg/cm² of the site’s neutral detergent (24% carbon), or 2.4 µg C/cm²; its supplier’s 5 mg/day safe intake would allow 5,000 µg/cm² (5 mg/day × 100 kg ÷ 400 mg/day = 1.25 kg over 25 m²)
Detergent testA total organic carbon (TOC) swab beside each of the 12 sites, limit 60 µg C above the swab blank (10 µg/cm² × 25 cm² × 24%)
RinseA 2 L sampling rinse of the 2,000 cm² transfer line, which swabs cannot reach: Exampladine by HPLC, limit 4 µg/mL (2,000 cm² × 4 µg/cm² = 8,000 µg in 2 L); detergent by TOC, limit 2,400 µg C/L above the water blank (2,000 cm² × 2.4 µg C/cm² ÷ 2 L)
MicrobialAfter a 14-day clean hold: no more than 25 CFU per 25 cm² swab and no objectionable organisms, justified in RA-⁠26-⁠031
VisualVisually clean, inspected dry against SOP-⁠CL-⁠07’s list of areas, with torch and mirror, by trained, eye-tested inspectors assessed in practice. On spiked steel coupons under the line’s lighting and distances, Exampladine is readily visible from 2 µg/cm² (VR-⁠26-⁠009): half the 4 µg/cm² limit, but above the press parts’ 0.52 µg/cm²2,13
Sites12 swab sites: 7 hardest to clean (impeller shaft seal, chopper, discharge valve, dryer bowl rim, blender lid gasket, feed frame, punch tips) and 5 representative flat areas, plus the rinse
RecoverySpiked coupons, stainless steel and silicone gasket: Exampladine 85% and 62%, detergent by TOC 90% and 71%; the rinse, from spiked steel tubing, Exampladine 91% and detergent 94%. Results are divided by recovery
MethodHPLC method AM-⁠114, swab extracted into 10 mL. Quantitation limit 0.5 µg/mL, or 5 µg per swab: 5% of the general limit, 38% of the press-part limit, and 12.5% of the rinse limit. TOC method AM-⁠121: quantitation limit 10 µg C per swab, 17% of its limit
Runs and holdsThree consecutive cleans, set by RA-⁠26-⁠031, each after a full-scale 6-batch, 10-day campaign and a 72-hour dirty hold, at the procedure’s minimum detergent contact time and rinse volume, by a different operator because press parts are washed by hand
After validationFeed frame and punch tips, hand-washed with a limit below the visible level: swabbed at every changeover.2 Elsewhere: the visual check, with swabs as often as RA-⁠26-⁠031 justifies13
Table 5. Example. Four of run 2’s HPLC swab results, recovery-corrected.
SiteFoundRecoveryCorrectedLimitResult
Impeller shaft seal45 µg85%, steel52.9 µg100 µgPass; above alert, investigated13
Blender lid gasket21 µg62%, silicone33.9 µg100 µgPass
Feed frame6.8 µg85%, steel8.0 µg13 µgPass
Punch tips6.2 µg85%, steel7.3 µg13 µgPass

The calls worth copying:

  • The tighter historical limit stays. EMA’s Q&A says historically used limits “should be retained and can be considered alert limits” where process capability keeps results below the HBEL; this site keeps its as the acceptance limit.13
  • The press has its own limit. PI 006-⁠4 says not to assume a contaminant is “uniformly distributed”, and to consider unit doses in compression.2 EMA’s guideline is not meant to set limits “at the level of the calculated HBEL”, so the site halves the press figure.13
  • The gasket has its own recovery, shown “from all product contact materials”;1 the steel figure would report 24.7 µg, not 33.9 µg.
  • Each residue has its own method. Exampladine, rinse included, goes to specific HPLC: Annex 15 offers TOC for when “it is not feasible to test for specific product residues”.1 The detergent, a mixture, goes to TOC, with all carbon counted as detergent.5,2

5Which situation needs what.

SituationWhat to doSource
New product on the trainAssess it against the worst case, through change control, before it runsAnnex 15 10.10;1 FDA’s Jabil letter;15 ABBE letter below16
At a CDMOThe product owner supplies the HBEL assessment, or the data for oneEMA HBEL Q&A 513
Grouping similar itemsPIC/S: only like for like, or where differences are shown in practice not to affect cleaning; justify the one testedPI 006-⁠4 7.3.5;2 Annex 15 10.11
Manual cleaning, low PDEDedicated parts or automation; otherwise verify at each changeoverPI 006-⁠4 7.1.2 and 7.13.12
Limit below visible levelDo not rely on visual inspection; verify after every changeoverPI 006-⁠4 7.8.3 and 7.13.22
Proteins and peptidesIf cleaning degrades them, a toxicological evaluation may not apply; control bioburdenAnnex 15 10.6.1;1 PI 006-⁠4 7.1.32

6What inspectors cite.

“FDA documented an unidentified (b)(4) residue, apparent rust, and unidentified liquids inside or on product contact surfaces of various non-dedicated process equipment used in the production of (b)(4). Each was marked as cleaned and ready for use.” Flowchem Pharma Private Limited, API, warning letter 320-⁠26-⁠51, 11 March 202617
“You also failed to perform a new cleaning validation study when you began manufacturing a new drug product, End Zit Acne Drying Lotion, which you determined to be the most difficult to clean product due to the titanium dioxide active ingredient, on shared equipment.” ABBE Laboratories, Inc., warning letter 320-⁠25-⁠98, 6 August 202516
“The inspection resulted in a critical deficiency related to contamination risk in the Fluid Bed Coating (FBC) equipment due to an observed pool of liquid and residue in a section of the inlet air plenum of a fully cleaned FBC. The flaws were due to common design between FBCs (inclination of plenum not allowing gravity draining) and inadequate cleaning procedure.” Glukem Biocare Private Limited, statement of non-compliance MT/004NCR/2026, Malta Medicines Authority, 16 October 202618

7What changed from 2024 to 2026.

  • PIC/S PI 006-⁠4 came into force on 1 October 2026, replacing PI 006-⁠3 of 2007. It adds alert limits, unit-dose checks and continued verification for manually cleaned low-limit products.2,7
  • Annex 15 is being revised. EMA’s concept paper of 19 January 2026 proposes making it mandatory for active substance manufacturers; a draft is planned for consultation in December 2026 and publication in April 2028; the 2015 text applies meanwhile.19,20

8How it runs in Seal.

In Seal, the cleaning validation blueprint connects the product and equipment matrix to versioned limit calculations, worst-case rationale, cleaning instructions, hold clocks, swab and rinse samples and laboratory results. Each limit keeps its toxicological input, dose basis, batch sizes, surface area and safety factors. The production step checks the equipment’s clean state before use, an exceeded hold is flagged at use, and a change to a toxicological input, procedure or surface area shows the calculations, studies and campaigns in scope.

Neil can review cleaning validation evidence, mapping results to equipment locations, recovery assumptions and protocol limits; it does not derive a safety limit or release equipment from an incomplete package. In electronic logbooks, cleaning, inspection and planned use stay separate facts, so ‘cleaned’ is not read as ‘ready’.

AQuestions and answers

What is the difference between cleaning validation and cleaning verification?

Validation shows in advance, through a planned study of several cleans, that an approved procedure reliably cleans to the limit. FDA then generally does not expect each routine clean to be tested; in EU shared facilities, testing at each changeover is expected unless a documented risk assessment justifies less. Verification tests that specific equipment is clean after a particular clean. Verification is used during the validation study, for development and investigational products, and routinely where manual cleaning cannot be validated or residue cannot be seen at its limit.

Does dedicated equipment need cleaning validation?

In the US, cleaning only between batches of the same product needs no validation, only visibly clean equipment, under FDA’s 1993 guide. FDA’s Q&A still applies 21 CFR 211.67 to dedicated equipment and expects residues, cleaning agents included, to be adequately removed between campaigns. EU GMP Annex 15 asks for cleaning validation for all product contact equipment and makes no exception for dedicated equipment.

What recovery is acceptable in a swab recovery study?

None of FDA, Annex 15, PIC/S, ICH Q7 or Health Canada sets a minimum percentage; FDA’s 1993 guide mentions 50% and 90% only as examples of a level to establish. WHO’s cleaning validation appendix gives an example scale: above 80% good, above 50% reasonable and below 50% questionable. The texts ask for recovery to be shown for every product-contact material with each sampling method, at an appropriate and consistent level, and Health Canada warns that low or variable recovery may point to poor sampling or extraction. Health Canada has each material’s recovery used in the residue calculation; PIC/S says a correction factor may be required.

What do WHO’s cleaning validation guidelines require?

WHO’s cleaning validation appendix dates from 2006 and was carried, unrevised, into its 2019 validation guideline (TRS 1019). It asks for at least three consecutive successful cleans; allows bracketing of similar products, represented by the hardest to clean, and of the same equipment in different sizes; and uses rinse and placebo-batch samples only alongside other sampling. Its limit is the strictest of visually clean, 10 ppm and 0.1% of the therapeutic dose. WHO’s 2021 points to consider (TRS 1033) add health-based limits: calculate one, keep a historical limit only where it is stricter, and base alert limits on statistics. EMA, by contrast, lets historical limits serve as alert limits.

References20 sources

References

  1. 1EudraLex Volume 4, Annex 15: Qualification and Validation, European Commission, in operation since 1 October 2015, cover note, sections 10 and 11, and glossary. https://health.ec.europa.eu/system/files/2016-11/2015-10_annex15_0.pdf (read 26 November 2026)
  2. 2PI 006-⁠4, Recommendations on Qualification and Validation, PIC/S, in force from 1 October 2026, section 7 and revision history. https://picscheme.org/docview/11277 (read 26 November 2026)
  3. 3Guideline on setting health based exposure limits for use in risk identification in the manufacture of different medicinal products in shared facilities, EMA/CHMP/CVMP/SWP/169430/2012, EMA, 20 November 2014, in effect from 1 June 2015, section 4.1. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-setting-health-based-exposure-limits-use-risk-identification-manufacture-different-medicinal-products-shared-facilities_en.pdf (read 26 November 2026)
  4. 4Guide to Inspections: Validation of Cleaning Processes, FDA, July 1993. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-guides/validation-cleaning-processes-793 (read 26 November 2026)
  5. 5Questions and Answers on Current Good Manufacturing Practice Requirements: Equipment, FDA, content current as of 16 November 2022, questions 3, 5, 7, 8, 10 and 12. https://www.fda.gov/drugs/guidances-drugs/questions-and-answers-current-good-manufacturing-practice-requirements-equipment (read 26 November 2026)
  6. 6Cleaning validation guide (GUI-⁠0028), Health Canada, 29 June 2021, section 7.2. https://www.canada.ca/en/health-canada/services/drugs-health-products/compliance-enforcement/good-manufacturing-practices/validation/cleaning-validation-guidelines-guide-0028/document.html (read 26 November 2026)
  7. 7PI 006-⁠3, Recommendations on Validation Master Plan, Installation and Operational Qualification, Non-⁠Sterile Process Validation, Cleaning Validation, PIC/S, 25 September 2007, section 7, in ECA Academy’s copy because PIC/S’s own link now goes to a members’ login. https://www.gmp-compliance.org/files/guidemgr/PI%20006-3%20Recommendation%20on%20Validation%20Master%20Plan.pdf (read 26 November 2026)
  8. 8Good manufacturing practices: guidelines on validation, Annex 3, Appendix 3, Cleaning validation, WHO Technical Report Series No. 1019, WHO, 2019, section 3.3. https://cdn.who.int/media/docs/default-source/medicines/norms-and-standards/guidelines/production/trs1019-annex3-gmp-validation.pdf (read 26 November 2026)
  9. 921 CFR 211.67, Equipment cleaning and maintenance, eCFR. https://www.ecfr.gov/current/title-21/section-211.67 (read 26 November 2026)
  10. 1021 CFR 211.182, Equipment cleaning and use log, eCFR. https://www.ecfr.gov/current/title-21/section-211.182 (read 26 November 2026)
  11. 11Good manufacturing practice and good distribution practice, MHRA and Department of Health and Social Care, last updated 13 May 2024. https://www.gov.uk/guidance/good-manufacturing-practice-and-good-distribution-practice (read 26 November 2026)
  12. 12EudraLex Volume 4, Part I, Chapter 4: Documentation, European Commission, in operation since 30 June 2011, sections 4.18, 4.29 and 4.31. https://health.ec.europa.eu/system/files/2016-11/chapter4_01-2011_en_0.pdf (read 26 November 2026)
  13. 13Questions and answers on implementation of risk-based prevention of cross-contamination in production and the guideline on setting health-based exposure limits, EMA/CHMP/CVMP/SWP/246844/2018, EMA, 19 April 2018, questions 5 to 8. https://www.ema.europa.eu/en/documents/other/questions-and-answers-implementation-risk-based-prevention-cross-contamination-production-and-guideline-setting-health-based-exposure-limits-use-risk-identification-manufacture-different_en.pdf (read 26 November 2026)
  14. 14ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, ICH, 10 November 2000, section 12.7. https://database.ich.org/sites/default/files/Q7%20Guideline.pdf (read 26 November 2026)
  15. 15Jabil Inc. (Pharmaceutics International, Inc.), warning letter 320-⁠26-⁠120, MARCS-⁠CMS 731037, FDA, 27 August 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/jabil-inc-731037-08272026 (read 26 November 2026)
  16. 16ABBE Laboratories, Inc., warning letter 320-⁠25-⁠98, MARCS-⁠CMS 708309, FDA, 6 August 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/abbe-laboratories-inc-708309-08062025 (read 26 November 2026)
  17. 17Flowchem Pharma Private Limited, warning letter 320-⁠26-⁠51, MARCS-⁠CMS 720719, FDA, 11 March 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/flowchem-pharma-private-limited-720719-03112026 (read 26 November 2026)
  18. 18Glukem Biocare Private Limited, statement of non-compliance with GMP MT/004NCR/2026, Malta Medicines Authority, inspection ending 15 September 2026, issued 16 October 2026, Part 3, EudraGMDP document 188112. https://eudragmdp.ema.europa.eu/inspections/gmpc/prepReviewSubmittedGMPC.do?key=188112 (read 26 November 2026)
  19. 19Concept paper on the revision of the guidelines on Good Manufacturing Practice for medicinal products, Annex 15, Qualification and Validation, EMA/INS/GMP/163466/2026, EMA GMP/GDP Inspectors Working Group, 19 January 2026. https://www.ema.europa.eu/system/files/documents/scientific-guideline/concept-paper-annex-15_revised-timetable-en.pdf (read 26 November 2026)
  20. 20EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https://health.ec.europa.eu/medicinal-products/eudralex/eudralex-volume-4_en (read 26 November 2026)