What is chain of identity, and how does it differ from chain of custody in cell and gene therapy?

Chain of identity is the permanent link between a patient or donor and their cells, kept from collection to treatment by unique identifiers checked at every handoff. Chain of custody is the continuous record of who held the material, where, when and in what condition. Identity answers whose material it is; custody answers who had it. For cell products in the US, both rest on the 21 CFR 1271 tracking and labeling rules; in the EU, on ATMP traceability law.Checked against its sources on 26 November 2026. How these answers are made

Illustration of a seal beside a sample transport case and study records.

1What is the difference between chain of identity and chain of custody?

A personalized therapy needs both: a bag with a flawless custody record is still the wrong product if it was mislabeled at collection, and a correctly labeled bag can sit for hours in a transfer nobody recorded.

Table 1. The two chains side by side.1
AspectChain of identityChain of custody
QuestionWhose material is this?Who had it, where, when, in what state?
Recorded asThe same identifiers on every container, sample and recordA handoff event: sender, receiver, time, place, condition
Breaks whenA label, a keyed entry, a split or a relabel loses the linkA transfer, hold or excursion goes unrecorded
US ruleDistinct code and label controls, 1271.290(c) and 1271.250Receipt, shipment and disposition records, 1271.265(e)

The standard definitions come from a working group of the Standards Coordinating Body. Chain of identity is “the permanent and transparent association of a cell or gene therapy’s unique identifiers from procurement of tissue or cells throughout the full product(s) lifecycle including post treatment monitoring”. Chain of custody is “concurrent, permanent, auditable documentation illustrating the guardianship of a cell or gene therapy product from its origin through its final disposition”. FACT’s Common Standards for Cellular Therapies and ICCBBA use the same wording.2,3

Neither phrase appears in 21 CFR Part 1271, Regulation 1394/2007 or the EU GMP guidelines for ATMPs. FDA’s CAR T guidance uses chain of identity and ties it to the regulations: “To maintain the Chain of Identity (COI), labeling and tracking of material, from collection all the way through CAR T cell administration, must be documented (21 CFR Part 1271 Subpart D).” EU law calls the obligation traceability.1,4,5,6

2What US and EU rules require.

CAR T cells and other HCT/Ps regulated as drugs or biologics follow Part 1271 subparts C and D as well as drug CGMP.7,1 In the EU, Regulation 1394/2007 sets traceability for authorized ATMPs, Part IV applies it to manufacture, and EMA’s investigational ATMP guideline to trials.5,6,8

ControlUnited StatesEuropean Union
Two-way trackingDonor to consignee and back, 1271.290(b)1Product and materials to the hospital, Article 15(1)5; donation to recipient from the first clinical batch, 6.356
Distinct codeRelates the HCT/P to its donor and records; any new code linked to the old by procedure, 1271.290(c)1Internal code always linked to the donation code, 6.37(ii)6; donation and product codes on authorized ATMP packaging, Article 12(a)5
Label controlProcedures “to prevent mix-ups”; accuracy, legibility and integrity verified, 1271.2501Code readers checked; labels fit for ultra-low temperatures, 9.776
Autologous label“FOR AUTOLOGOUS USE ONLY”, 1271.90(c)(1)1Authorized ATMPs: “For autologous use only” and the unique patient identifier, Article 12(b)5
Trial labelInvestigational caution statement; for autologous CAR T, at least two unique identifiers recommended4Subject or treatment number and batch code, Annex VI9, which Part IV 9.81 applies to trial ATMPs6
Relabel or reprintLabels issued, used and returned reconciled; excess lot-numbered labels destroyed, 211.125(c) to (f)10Trial ATMPs: checked by a second person, 9.82; an expiry relabel repeats the batch number, 9.816
SeparationDefined areas or controls to prevent mix-ups, 211.42(c)10No two batches in the same area at once, closed systems aside, 4.186
ReleaseTracking records reviewed; release documented and dated, 1271.265(c)1Autologous or donor-matched: starting material matched to recipient, 11.27(a)(iii)6
At the hospitalTwo independent bedside checks, CAR T guidance4Hospital system links product to patient and back, Article 15(2)5
Retention10 years after administration, 1271.270(d)130 years after expiry, Article 15(4) and 6.37; can pass to the sponsor or holder by written agreement, 6.405,6

3Allogeneic and in vivo gene therapy.

Table 3. Beyond autologous products. ST-⁠028’s label examples include related and unrelated donors, so its identifier fits matched products too.3
DesignIdentity linksUnited StatesEuropean Union
Allogeneic, matchedA donor to one named recipientEligibility statement and record summary travel with the product, 1271.55(a)(2) and (3)1Release checks the starting material against the recipient, 11.27(a)(iii)6
Allogeneic, off the shelfA donor or cell bank to many lots and patientsThe same records travel; tracking runs from the donor to every lot’s consignee, 1271.290(b)1Cell banks recommended, 8.10; every internal code stays linked to the donation code, 6.376
In vivo gene therapyA vector lot to each patientNot an HCT/P, so no Part 1271 tracking, 1271.3(d)1; distribution records reach the consignee, 211.19610Article 15 still links each product to its patient5

Autologous cells need no donor-eligibility determination, so no eligibility statement or summary has to travel with them (1271.90(a)(1)). The code may carry a name only for autologous, directed reproductive and first- or second-degree blood relative donations, so an unrelated donor’s link to the recipient runs through the code alone; the accompanying records must not identify the donor (1271.55(a)(1) and (c), 1271.290(c)).1 FDA’s example deficiencies for biological drug inspections include a name in the code and an HCT/P not accompanied by its eligibility statement or record summary.11

4Example: one autologous case from order to infusion.

An invented CAR T trial in 2026: the sponsor allocates an ISBT 128 identifier (format below), subject 104-⁠007 is treated at center T1, and site M1 makes Batch 26-⁠117. The sponsor’s validated limit from end of collection to receipt at M1 is 48 hours.

Table 4. Invented case. Bracketed numbers count recorded identity checks, 13 in all.
StepWhenIdentity checkCustody
Order acceptedThu 5 Mar, 14:20CHA999926000417, boxed P, allocatedCoordinator
CollectionMon 9 Mar, 09:40Donor identity verified; bag labeled with DIN A9997 26 300215, identifier, “FOR AUTOLOGOUS USE ONLY” (1)Apheresis unit
Pick-upMon 9 Mar, 11:05Bag and shipper scanned against the order (1)Unit to courier
Receipt at M1Tue 10 Mar, 08:15Keyed entry refused, then accepted (1)Courier to M1 goods-in
Manufacture10 to 21 MarLabel check before each of 6 processing steps (6)M1 suites and cryostore
ReleaseThu 26 MarRecords reviewed; origin matched to recipient (1)M1 quality unit
DeliveryTue 31 Mar, 10:30Identifier and destination checked (1)Courier to T1 pharmacy
InfusionThu 2 Apr, 09:00Two independent patient and label checks (2)T1 clinical team
  • Transit. Collection ended at 09:40 on Monday and M1 received the bag at 08:15 on Tuesday: 22 hours 35 minutes, inside the 48-hour limit.
  • The refused entry. The goods-in scanner failed and a technician keyed CHA999926000418. Its check character is N, not the boxed P, so the system refused it and asked again; the second entry matched.3
  • Retention. The Part 1271 records are kept until at least 2 April 2036.

5A checklist, custody record and system tests you can copy.

These arrangements make the rules work:

  1. An agreement naming who allocates the identifier (the collecting facility, sponsor or manufacturer) and when, one per given therapy.3
  2. Written chain of identity and custody procedures at the clinical program, collection facility and processing facility, with donor identity verified before collection.12
  3. Label checks built into the batch record before each processing step.4
  4. Every receipt, shipment and transfer recorded with the fields below.1
  5. The consignee told in writing, at or before distribution, of the tracking requirements and the tracking system.1
  6. Chain of identity and chain of custody audited at least once a year.2

5.1Custody record fields.

Table 5. One row per handoff, written when it happens.
FieldWhat to recordRule
MaterialThe distinct code, and who supplied the material1271.265(e)(1)1
ActivityReceipt, shipment, transfer or return, and its result1271.265(e)(2)1
DateThe date of each activity; the time too where a transit limit applies1271.265(e)(3)1
QuantityBags or volume handled1271.265(e)(4)1
DispositionThe next holder or consignee1271.265(e)(5)1
ConditionsThe shipping conditions set for the product, and whether they held1271.265(d)1
PersonWho did the work and when each entry was made, written at the time1271.270(a)1
DepartureAny departure from procedure, recorded and justified when it happens1271.265(c)(3)1

5.2System validation tests.

Part 1271 requires custom or customized software relied on for core good tissue practice, labeling and distribution among them, to be validated and the results approved before use (1271.160(d)), which FDA investigators check.1,11 Test with the reserved FINs A9990 to A9999.3

Table 6. Run before go-live and again after any change to the software.1
TestExpected resultRule
Wrong-patient scanThe step is blocked and the mismatch recorded1271.250(a)1
Mistyped identifierRefused by its check character; entry asked againST-⁠028 4.43
New or split codeThe new code stays linked to the old one1271.290(c)1
System down mid-stepA manual or alternative system carries the recordPart IV 10.136
Unauthorized editRefused; only authorized people change records211.68(b)10
Trace reportDonor to consignee or final disposition and back from any one code1271.290(b)1

6How the ISBT 128 chain of identity identifier works.

ICCBBA’s standard ST-⁠028, version 1.1.0 of June 2025, gives the field a shared format: one code for a patient’s given therapy, allocated before or at the first collection and never reassigned. It does not replace the donation identification number (DIN) on each collection; it is added to every donation for that therapy. FACT-⁠JACIE’s Immune Effector Cell Standards recommend it for collections sent for further manufacture.3,12

Table 7. FINs A9990 to A9999 are reserved for validation testing and example labels.3
ElementLengthExample
Fixed text2CH
Facility number (FIN)5A9999
Year of issue226
Sequence number6000417
Check character1, boxedP

Bar codes encode the 15 characters as ISBT 128 Data Structure 040. The boxed check character is printed only beside the eye-readable form, to catch keying errors. It follows ISO/IEC 7064 Mod 37-⁠2: multiply each character’s value (digits as themselves, A to Z as 10 to 35) by 2 to the power of its position from the right, add them up, divide by 37, take the remainder from 38, and subtract 37 if the result is 37 or more. For CHA999926000417 the total is 824,114, the remainder 13, and 38 − 13 = 25, the letter P. The example’s mistyped CHA999926000418 totals 824,116 and gives N, so the entry is refused and repeated.3

7What inspectors check, and what they have cited.

CAR T and other cell and gene therapy sites are inspected under FDA’s compliance program for biological drugs, not the one for HCT/Ps, and the deficiencies it lists include “Failure to establish a tracking system that enables the tracking of HCT/Ps back and forth from the donor to the consignee or final disposition.”13,11 Part 1271 sets no time limit for a trace, but the records “must permit the prompt identification of the consignee” (1271.290(e)).1

At the February 2007 pre-license inspection of Dendreon’s Morris Plains, New Jersey, site for sipuleucel-⁠T, an autologous cell therapy, FDA observed:14

“The quality control laboratory did not demonstrate adequate ability to maintain the chain of identity for the autologous product.”

The commercial system used a bar code to maintain identity, but, the same observation notes, “The QC laboratory does not have the capacity to read the barcode”.14 In July 2025, at Celgene’s Summit, New Jersey, site, which makes the approved CAR T products Breyanzi and Abecma, FDA observed:15,16

“The transfer of Breyanzi and Abecma (b) (4) between (b) (4) due to a (b) (4) event that occurred on October 14, 2023, was not consistently documented in production batch records for all affected lots.”

8What changed in 2024 to 2026, and what is pending.

DateChangeStatus
January 2024FDA CAR T guidance with chain of identity controlsFinal, nonbinding4
1 July 2025EMA investigational ATMP guideline, two-way traceability by coding systemIn effect8
By October 2025FACT-⁠JACIE Immune Effector Cell Standards, third edition: chain of identity and custody procedures at every facilityCurrent12,17
13 March 2026Draft fourth edition of the FACT Common Standards, with “Greater emphasis on traceability, product identity, and labeling standardization”Draft; the third edition stays current18,17
7 August 2027SoHO Regulation covers donor cells up to distribution to the ATMP maker, with EU-unique codes and 30-year traceability; for autologous ATMPs, only its testing and collection rulesAdopted, applies from that date19

9How it runs in Seal.

In Seal, autologous cell therapy manufacturing runs each patient’s therapy as one governed case created by the treatment order. Chain of identity relates the patient, collection, starting material, every intermediate and sample, the final product, the shipment and the treatment-center handoff; chain of custody records who held each physical object, where, when and in what condition. At every critical handoff the expected identifiers and the physical object are verified, and a mismatch is flagged on the record and opens containment. Reprinting, relabeling and manual recovery keep the prior value and the reason for the change.

Patient orchestration runs the clinic workflow on the same platform. A delivery with an unexpected identifier enters a visible exception path: the observed and expected values are recorded, the mismatch is flagged, and the decision is signed in sequence, before the next step.

AQuestions and answers

What should happen when a chain of identity check fails?

Stop the step and hold the material. A keyed identifier that fails its check character is simply entered again. Any other mismatch is a deviation: record it, investigate it and release nothing until the link to the patient is proved. If a distributed licensed product may be affected, the license holder reports a biological product deviation to FDA within 45 calendar days of learning of it.

Who is responsible for chain of identity, the manufacturer or the treatment center?

Both. In the US, the manufacturer must tell each consignee in writing, at or before distribution, about the tracking requirements and its tracking system, so the treatment center can keep its part of the chain. In the EU, the hospital keeps its own record linking each product to the patient who received it.

Can the patient’s name and date of birth go on a cell therapy bag?

In the US, for autologous material, yes. Part 1271 bars names, social security numbers and medical record numbers from the identification code except for autologous donations, directed reproductive donations and donations from first- or second-degree blood relatives; it says nothing about dates of birth. EU rules ask for the unique patient identifier on an authorized autologous ATMP and the subject identification or treatment number on a trial product, and investigational ATMP traceability runs through a coding system that respects data protection law.

Does the ISBT 128 chain of identity identifier replace the donation identification number?

No. Each collection keeps its own donation identification number and gains the identifier alongside it. Pooling is the case that changes the number: donations pooled at an ICCBBA-licensed facility get a new donation identification number.

Does HLA typing prove chain of identity?

No. FDA’s CAR T guidance says HLA typing does not detect the genetic modification, so it is not a sufficient identity test, and it does not replace the requirements for maintaining chain of identity.

Can a contract manufacturer use the sponsor’s tracking system?

Yes. Under Part 1271, an establishment that performs only some manufacturing steps may join a tracking system set up by another establishment for the same product, as long as that system meets every tracking requirement. In the EU, a written agreement can also move retention of the 30-year traceability data to the sponsor or marketing authorization holder.

References19 sources

References

  1. 121 CFR Part 1271, Human Cells, Tissues, and Cellular and Tissue-⁠Based Products, sections 1271.1, 1271.3(d), 1271.55, 1271.90, 1271.160, 1271.250, 1271.265, 1271.270 and 1271.290, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-L/part-1271 (read 26 November 2026)
  2. 2Summary of Changes, Third Edition FACT Common Standards for Cellular Therapies, STS.RES.5.016 R3, FACT, 6 April 2023. https://www.factglobal.org/media/eb2jokew/sts-res-5-016_summary-of-changes-third-edition-fact-common-standards-for-cellular-therapies_r3_07apr2023.pdf (read 26 November 2026)
  3. 3ISBT 128 Standard, Chain of Identity (CoI) Identifier, version 1.1.0, ICCBBA ST-⁠028, sections 1.4, 1.6, 2, 4.1 to 4.5 and 4.7, ICCBBA, June 2025. https://iccbba.org/wp-content/uploads/2025/08/1a7593_0aa7a5f7202c4b43be4cc11bcd2048ef.pdf (read 26 November 2026)
  4. 4Considerations for the Development of Chimeric Antigen Receptor (CAR) T Cell Products, Guidance for Industry, sections IV.B, IV.C.2.c and IV.C.3, FDA, January 2024. https://www.fda.gov/media/156896/download (read 26 November 2026)
  5. 5Regulation (EC) No 1394/2007 on advanced therapy medicinal products, Articles 12 and 15, EUR-⁠Lex, 13 November 2007. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32007R1394 (read 26 November 2026)
  6. 6Guidelines on Good Manufacturing Practice specific to Advanced Therapy Medicinal Products, C(2017) 7694 final, sections 4.18, 6.35 to 6.40, 8.10, 9.77, 9.81, 9.82, 10.13 and 11.27, European Commission, 22 November 2017. https://health.ec.europa.eu/document/download/ad33d9dd-03f0-4bef-af53-21308ce2187d_en?filename=2017_11_22_guidelines_gmp_for_atmps.pdf (read 26 November 2026)
  7. 721 CFR 210.1, Status of current good manufacturing practice regulations, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-210/section-210.1 (read 26 November 2026)
  8. 8Guideline on quality, non-clinical and clinical requirements for investigational advanced therapy medicinal products in clinical trials, EMA/CAT/22473/2025, EMA, 20 January 2025. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-quality-non-clinical-clinical-requirements-investigational-advanced-therapy-medicinal-products-clinical-trials_en.pdf (read 26 November 2026)
  9. 9Regulation (EU) No 536/2014 on clinical trials on medicinal products for human use, Annex VI, section A.1, EUR-⁠Lex, 16 April 2014. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32014R0536 (read 26 November 2026)
  10. 1021 CFR Part 211, sections 211.42(c), 211.68, 211.125 and 211.196, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 26 November 2026)
  11. 11Compliance Program 7345.848, Inspection of Biological Drug Products (CBER), parts III and V, packaging and labeling system and computer validation, FDA, implemented 1 October 2010. https://www.fda.gov/media/73834/download (read 26 November 2026)
  12. 12Summary of Changes, Third Edition FACT-⁠JACIE International Standards for Immune Effector Cells, STS.RES.5.026 R1, standards B5.1.18, B5.1.19, C5.1.22, C5.1.23, C7.1.3, C7.3.2, C9.7, D5.1.20 and D5.1.21, FACT, 27 October 2025. https://factglobal.org/media/2qhj13x2/sts-res-5-026_summary-of-changes-3rd-ed-fact-jacie-standards-for-iecs_r1-28oct2025.pdf (read 26 November 2026)
  13. 13Compliance Program 7341.002, Inspection of Human Cells, Tissues, and Cellular and Tissue-⁠Based Products (HCT/Ps), part II, FDA. https://www.fda.gov/media/73949/download (read 26 November 2026)
  14. 14Form FDA 483, Dendreon Corporation, Morris Plains, New Jersey, pre-license inspection for sipuleucel-⁠T, observation 3, FDA, 16 February 2007. https://www.fda.gov/media/167053/download (read 26 November 2026)
  15. 15Form FDA 483, Celgene Corporation, Summit, New Jersey, observation 2, FDA, 18 July 2025. https://www.fda.gov/media/192234/download (read 26 November 2026)
  16. 16Approved Cellular and Gene Therapy Products, FDA, content current as of 30 October 2026. https://www.fda.gov/vaccines-blood-biologics/cellular-gene-therapy-products/approved-cellular-and-gene-therapy-products (read 26 November 2026)
  17. 17FACT Standards, FACT Common Standards for Cellular Therapies, FACT. https://factglobal.org/standards/ (read 26 November 2026)
  18. 18Summary of Changes, Draft Fourth Edition FACT Common Standards for Cellular Therapies, STS.RES.5.029 R1, FACT, 13 March 2026. https://factglobal.org/media/brrnnupo/sts-res-5-029_sum-of-changes-draft-4th-ed-fact-common-standards-for-ct_r1_13mar2026.pdf (read 26 November 2026)
  19. 19Regulation (EU) 2024/1938 on standards of quality and safety for substances of human origin intended for human application, Articles 2(6), 2(7), 42, 43 and 87, EUR-⁠Lex, 13 June 2024. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32024R1938 (read 26 November 2026)

See your process in Seal.

Bring a procedure or a recurring problem. See how your team configures it with Neil and improves the next run.