What are the ALCOA+ and ALCOA++ principles of data integrity?

ALCOA+ stands for attributable, legible, contemporaneous, original and accurate (ALCOA), plus complete, consistent, enduring and available: the attributes regulators expect of GxP data, on paper or electronic. ALCOA++ adds traceable: every change recorded, explained where needed and leaving the original visible. FDA and MHRA write ALCOA, PIC/S and WHO ALCOA+, and EMA’s 2023 clinical trials guideline and the draft EU GMP Chapter 4 ALCOA++. MHRA says the expectations are the same whichever acronym is used.Checked against its sources on 25 November 2026. How these answers are made

Illustration of a seal beside a balance and a spectrometer, each cabled to one box of raw data.

1ALCOA, ALCOA+ and ALCOA++: what each adds and where it comes from.

ALCOA names five attributes of a record; ALCOA+ adds four that should hold throughout the data life cycle for the retention period; ALCOA++ adds traceable.1,2

Table 1. The three acronyms. Meanings from EMA’s 2023 clinical trials guideline, section 4.5.2
TermStands forWhat it asks of data
ALCOAAttributable, legible, contemporaneous, original, accurateLinked to the person or system that generated it, readable in context, captured at the time, kept in its first form, a true representation of the observation
ALCOA+ALCOA plus complete, consistent, enduring, availableAlso complete with its metadata and audit trail, defined and managed consistently, intact for the retention period, available for review when needed
ALCOA++ALCOA+ plus traceableAlso every change to data or metadata traceable, explained if necessary, never obscuring the original

Regulators use all three, and two documents name no acronym at all:

Table 2. Which documents write which acronym.3,4,5,6,7,8,1,2,9,10,11
DocumentAcronymWhat it says
EMA GCP inspectors’ reflection paper on electronic source data, 2010, section 6.1NoneLists nine attributes “of universal importance to source data”: the ALCOA five plus complete, consistent, enduring and available when needed
FDA guidance, Electronic Source Data in Clinical Investigations, 2013ALCOA“Source data should be attributable, legible, contemporaneous, original, and accurate (ALCOA)”
MHRA GXP data integrity guidance, 2018, section 3.10ALCOAThe “+” was “subsequently added to emphasise the requirements”; there is “no difference in expectations regardless of which acronym is used”
FDA data integrity questions and answers, 2018ALCOATies each of the five attributes to sections of 21 CFR 211
EMA GMP questions and answers, data integrity question 13ALCOAMaps each ALCOA principle to EU GMP Parts I and II and Annex 11
PIC/S PI 041-⁠1, 2021, section 7.4ALCOA+Complete, consistent, enduring and available “can be added to the list”, citing EMA guidance for GCP inspections
WHO TRS 1033 Annex 4, 2021, section 1.2 and glossaryALCOA+The nine attributes are “commonly referred to as ‘ALCOA+’”, kept “throughout the data life cycle for the defined retention period”
EMA guideline on computerised systems and electronic data in clinical trials, 2023, section 4.5ALCOA++Adds traceable, and replaces the 2010 reflection paper
ICH E6(R3) good clinical practice, in effect in the EU from 23 July 2025, section 2.12.2NoneSource records “attributable, legible, contemporaneous, original, accurate and complete”, with changes traceable and the original entry not obscured
EU GMP Chapter 4, draft for consultation, July 2025, section 4.63 and glossaryALCOA++Sets the ten attributes out as a table for paper and electronic systems; not yet in force

None of these documents says who coined ALCOA; MHRA says only that it “was historically regarded as defining the attributes of data quality that are suitable for regulatory purposes”.5 Whichever acronym a document uses, audit against all ten attributes: MHRA sees no difference in expectations, and the draft Chapter 4 calls the added ones “implicit basic ALCOA principles”.5,10

2What each letter means, and how to audit it.

The attributes apply to paper, electronic and hybrid records alike.8 The acronym is guidance, but it rests on binding text: EU law requires a documentation system that “shall ensure data quality and integrity”, and FDA ties each of the five ALCOA attributes to sections of 21 CFR 211.12,6 Audit one record set at a time, such as a batch record or an analytical run, from capture to archive.7

Table 3. ALCOA+ audit checklist. Meanings from PIC/S PI 041-⁠1 section 7.5 and draft Chapter 4 section 4.63.8,10 Q numbers are questions in FDA’s 2018 guidance and EMA’s answers. Part 211 sections follow FDA’s footnote 5, and its questions 13, 1.e and 17 for complete, enduring and available; EU references draw on PIC/S section 11.1 and EMA question 13.6,13,7,14
AttributeWhat it asksAudit check21 CFR 211EU GMP
AttributableWho did it and when, and who changed it and whyEvery entry and change names one person or system; no shared account can change data211.101(d), 211.188(b)(11); FDA Q54.20, 4.21; Annex 11 12.4
LegibleReadable and unambiguous, every part of the recordCorrections leave the original readable, with reason, name and date211.180(e)4.7 to 4.10; Annex 11 9
ContemporaneousRecorded as it happensTimes run in order; nothing was noted on loose paper and copied in211.100(b), 211.160(a); FDA Q124.8; Annex 11 12.4
OriginalThe first capture, dynamic data kept dynamicOriginal or true copy kept; blank forms issued numbered and reconciled211.180(d), 211.194(a); FDA Q6, Q9, Q104.9, 4.27; Annex 11 8.2
AccurateA truthful record of what happenedCritical manual entries second-checked; spreadsheet formulas verified211.68(b), 211.1884.1, 6.17; Annex 11 6
CompleteEverything needed to recreate the event, metadata includedAborted, failed and invalidated runs are in the record; the review used electronic raw data211.188, 211.194; FDA Q2, Q134.8; EMA Q16
ConsistentCreated and stored in a defined, logical wayDate, unit and rounding conventions written and followed; clocks agreeNot mapped4.2; Annex 11 5
EnduringIntact for the whole retention periodBackups restore, archives retrieve and old files still open211.68(b); FDA Q1.eAnnex 11 7.2, 17
AvailableRetrievable whenever review or inspection needs itRecords can be produced for an inspector during the retention period211.180(c); FDA Q17Annex 11 7.1, 17
Traceable (ALCOA++)The history, changes and location of data can be tracedEach change and reprocessing has an entry with its reason; the first result is keptFDA Q1.c, Q14Annex 11 9; draft 4.63

The second-person reviewer runs these checks, as 211.194(a)(8) and FDA question 7 expect; QA verifies them at self-inspection, and the system owner supplies the evidence on accounts, clocks, backups and archives.13,6,8,14 Depth follows data criticality: for an oral tablet, PIC/S rates API assay data above friability data, and it expects self-inspection checks of audit trails to be “both random (selected based on chance) and targeted (selected based on criticality or risk)”.8

3What a failure looks like, letter by letter.

Each is a failure FDA names in its 2018 guidance (Q numbers), its preapproval inspection program or a warning letter.6,15

Table 4. Failures and where FDA names them.6,15,16
AttributeWhat a failure looks likeSource
AttributableAnalysts share one login; an instrument password is stuck to the laptopFDA Q5; GC America, 2026
LegibleA batch record entry covered with correction fluid and written overCP 7346.832
ContemporaneousWeights noted on scrap paper and copied in later; stability results backdatedFDA Q12; CP 7346.832
OriginalUncontrolled blank batch record forms, so a page can be rewritten and swappedFDA Q6
AccurateA result entered for a test never run, or copied from an earlier batchCP 7346.832
CompleteAn aborted injection missing from the reported sequence; samples run as “system suitability” until one passesFDA Q12, Q13
ConsistentEquipment named in the batch record differs from the applicationCP 7346.832
EnduringInstrument data deleted when the laptop shuts downGC America, 2026
AvailableAudit trails that cannot be produced for the investigatorGC America, 2026
TraceableOnly the final reprocessed chromatography result keptFDA Q14

In December 2025 FDA asked Unipack for a plan “to ensure you retain attributable, legible, complete, original, accurate, contemporaneous records throughout your operation”, having found under 211.194(a):17

“Furthermore, the sample weights were recorded on an adhesive note, and no standard weight was recorded. All raw data must be recorded in a controlled document at the time they are observed or measured to ensure data integrity.” Unipack LLC, warning letter 320-⁠26-⁠32, 19 December 202517

More patterns are in common data integrity violations.

4Example: one HPLC assay reviewed against ALCOA+.

Example, invented for illustration: on 4 November 2026 the Quillby QC laboratory’s second-person reviewer checks the 3 November assay run for Zelvatrin 40 mg tablets, Batch 26-⁠041, before the result goes to QA.

Table 5. Example. An invented review; each line traces to the checklist above.
AttributeWhat the reviewer foundOutcome
AttributableAnalyst’s own login, named in the audit trail for acquisition and processingMeets
LegibleOne worksheet correction, original readable, with reason, initials and dateMeets
ContemporaneousBalance printout 09:12, worksheet entry 09:13, sequence start 10:05Meets
OriginalWeights of 50.21 mg and 50.08 mg on the signed balance printout, attachedMeets
AccurateWeights typed into the LIMS second-checked against the printoutMeets
CompleteSequence of 14 injections, report of 13: injection 6, the first of sample preparation 1, aborted on a pressure alarm and was re-injected as 7Gap: no investigation
ConsistentInstrument and server clocks set from the site time serverMeets
EnduringServer in the nightly backup; last restore test passed on 9 September 2026Meets
AvailableRaw data, audit trail and alarm log retrieved from the server for the reviewMeets
TraceableIntegration reprocessed once at 14:40, reason “baseline drift at 4.1 min”; both results keptMeets

Injection 6 is the finding: FDA expects aborted injections to be “investigated and justified”, and excluded data need a documented, scientifically sound justification.6 The reviewer returned the run; investigation LI-⁠26-⁠117 attaches the pump’s alarm log, shows the bracketing standards met system suitability, and concludes injection 6 gave no reportable data. QA approved it on 5 November; only then was the result reported: 99.1% and 99.7%, mean 99.4% of label claim, against 95.0% to 105.0%. A printed report alone would have hidden the gap, which is why EMA expects reviewers to see electronic raw data.7

5Is this record acceptable?

A printout, PDF or copy stands in for the original only if it keeps its content and meaning, metadata included.6,5

RecordAcceptable as the record?Source
Balance or pH printoutYes, as the original where the instrument stores no data: signed and dated by its maker, with the sample or batch ID, and attached to the recordFDA Q10;6 PIC/S 8.98
HPLC or FT-⁠IR printoutNo. The file can be reprocessed, so it is the originalFDA Q10;6 MHRA 6.11.15
PDF of a chromatogramFDA: no, a static copy loses the dynamic record. PIC/S: paper or PDF only if justified, with all raw data, metadata, audit trails and processing runs kept and a documented check, which it calls “likely to be onerous”FDA Q10;6 PIC/S 7.7.28
Electronic file copyYes, as a true copy keeping metadata and dynamic form; PIC/S would prohibit PDFs that lose metadataFDA Q9;6 PIC/S 7.78
Scan of a paper recordYes, once verified by a dated signature or a validated processMHRA 6.11.25
Contract lab summaryNot a true copy; rely on it only after assessing the labPIC/S 7.88
Hybrid recordYes, if a procedure defines the whole data set and its review; PIC/S discourages hybridsMHRA 4.3;5 PIC/S 9.108
Entry minutes laterFDA, MHRA and PIC/S set no tolerance in minutes: save at the time of performance, without bundling critical stepsFDA Q12;6 MHRA 6.125
Clocks out of stepNo. Synchronize clocks, restrict access to them and record the time zone across sitesPIC/S 9.5, 9.8;8 MHRA 5.15

6How FDA tests ALCOA at a pre-approval inspection.

Objective 3 of FDA’s preapproval inspection program, CP 7346.832, is a data integrity audit, covered on every PAI: investigators “Audit and verify raw data at the facility that are associated with the product”.15 For biologics license applications the program points to CP 7346.832M instead.15 They compare chromatograms, spectra and notebooks with the application’s summary data; finished product stability, dissolution, content uniformity and API impurity data are named as good candidates.15 Its own example of a concern is the gap in the example above:

“Examples of data integrity concerns include failure to scientifically justify not reporting relevant data, such as aberrant test results or absences in a submitted chromatographic sequence.” Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 202615

A pattern of data reliability problems can widen the inspection to marketed products under CP 7356.002.15 Preparation is in pre-approval inspection.

7How it runs in Seal.

Seal keeps with each record the user, API key or trigger behind each entry, the capture time, the original file and its source reference, and each change as a new version linked to the earlier ones, as the data integrity blueprint sets out. Where a source system such as a chromatography data system keeps its own audit trail, that trail stays authoritative there; its export is attached to the Seal record and reviewed beside it, and a required review can hold the result or decision until it is done.

Each record keeps its audit history and electronic signatures, with controls to support 21 CFR Part 11 and EU GMP Annex 11, as the regulatory overview describes, and lab, manufacturing and quality work run on one platform of linked records.

AQuestions and answers

Are ALCOA-C and ALCOA-CCEA regulatory terms?

No. FDA’s 2018 questions and answers, MHRA’s 2018 guidance, PIC/S PI 041-⁠1, WHO TRS 1033 and EMA’s GMP answers and 2023 clinical trials guideline do not use them. ALCOA-⁠CCEA spells out the four attributes ALCOA+ adds: complete, consistent, enduring and available. MHRA says the expectations are the same whichever acronym is used, so audit against the attributes.

What is the difference between ALCOA+ and ALCOA++?

ALCOA++ adds traceable: the history, changes or location of data can be traced through recorded identifications. EMA’s 2023 clinical trials guideline and the 2025 draft of EU GMP Chapter 4 use it. The draft’s glossary describes the added emphasis as implicit in the basic ALCOA principles, and MHRA says expectations do not differ whichever acronym is used.

Do ALCOA principles apply in clinical trials?

Yes. EMA’s 2023 guideline on computerised systems and electronic data in clinical trials says data integrity is achieved when data fulfill the ALCOA++ principles, and that all data collected or generated in a clinical trial should fulfill them. ICH E6(R3), in effect in the EU since 23 July 2025, asks the same of source records without the acronym: attributable, legible, contemporaneous, original, accurate and complete, with every change traceable.

Can someone record data on another person’s behalf?

Only exceptionally. PIC/S PI 041-⁠1, in its footnote 8, allows a scribe in two situations: where the act of recording would put the product or activity at risk, such as line interventions by sterile operators, and to accommodate cultural, literacy or language limitations. MHRA’s 2018 guidance also names necropsy in GLP. The scribe records at the time of the task, the record names both people, the operator countersigns where possible, and an approved procedure lists the activities it covers.

Do the ALCOA+ attributes apply to metadata?

Yes. FDA’s 2018 guidance says data should be kept for the whole retention period with the metadata needed to reconstruct the activity, such as time stamps, user and instrument IDs and audit trails, and that the link between data and metadata should be preserved securely. PIC/S counts metadata as part of a complete electronic record.

Who is responsible for ALCOA+ in a company?

The manufacturer. PIC/S PI 041-⁠1 puts full responsibility on the company being inspected to assess its data management for vulnerabilities and to design good data governance. FDA’s 2018 guidance says management with executive responsibility should create a quality culture in which staff identify and promptly report data integrity issues.

References17 sources

References

  1. 1WHO Technical Report Series No. 1033, Annex 4, Guideline on data integrity, World Health Organization, 2021. https://cdn.who.int/media/docs/default-source/medicines/norms-and-standards/guidelines/inspections/trs1033-annex4-guideline-on-data-integrity.pdf (read 25 November 2026)
  2. 2Guideline on computerised systems and electronic data in clinical trials, EMA/INS/GCP/112288/2023, EMA GCP Inspectors Working Group, 9 March 2023, section 4.5. https://www.ema.europa.eu/en/documents/regulatory-procedural-guideline/guideline-computerised-systems-and-electronic-data-clinical-trials_en.pdf (read 25 November 2026)
  3. 3Reflection paper on expectations for electronic source data and data transcribed to electronic data collection tools in clinical trials, EMA/INS/GCP/454280/2010, EMA GCP Inspectors Working Group, 9 June 2010, superseded, section 6.1. https://www.ema.europa.eu/en/documents/regulatory-procedural-guideline/reflection-paper-expectations-electronic-source-data-and-data-transcribed-electronic-data-collection-tools-clinical-trials-superseded_en.pdf (read 25 November 2026)
  4. 4Guidance for Industry: Electronic Source Data in Clinical Investigations, FDA, September 2013, section II, docket FDA-⁠2010-⁠D-⁠0643. https://downloads.regulations.gov/FDA-2010-D-0643-0092/content.pdf (read 25 November 2026)
  5. 5‘GXP’ Data Integrity Guidance and Definitions, revision 1, MHRA, March 2018, sections 3.10, 4, 5 and 6.11 to 6.12. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 25 November 2026)
  6. 6Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, FDA, December 2018. https://www.fda.gov/media/119267/download (read 25 November 2026)
  7. 7Guidance on good manufacturing practice and good distribution practice: Questions and answers, Annex 11 questions 2 and 3 and data integrity questions 3 to 5, 13, 14 and 16, European Medicines Agency. https://www.ema.europa.eu/en/human-regulatory-overview/research-development/compliance-research-development/good-manufacturing-practice/guidance-good-manufacturing-practice-good-distribution-practice-questions-answers (read 25 November 2026)
  8. 8PI 041-⁠1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, PIC/S, 1 July 2021, sections 5.4, 7.5, 7.7, 7.8, 8.9, 9.5, 9.6, 9.8, 9.10 and 11.1 and footnote 8. https://picscheme.org/docview/4234 (read 25 November 2026)
  9. 9ICH E6(R3) Guideline for good clinical practice, Step 5, EMA/CHMP/ICH/135/1995, EMA, 23 January 2025, in effect from 23 July 2025, section 2.12.2. https://www.ema.europa.eu/en/documents/scientific-guideline/ich-e6-r3-guideline-good-clinical-practice-gcp-step-5_en.pdf (read 25 November 2026)
  10. 10EudraLex Volume 4, Chapter 4, Documentation, draft for consultation, European Commission, July 2025, section 4.63 and glossary. https://health.ec.europa.eu/document/download/fa336a6e-753b-46fc-b53b-9178bacd8878_en?filename=mp_vol4_chap4_consultation_guideline_en.pdf (read 25 November 2026)
  11. 11EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission. https://health.ec.europa.eu/medicinal-products/eudralex/eudralex-volume-4_en (read 25 November 2026)
  12. 12Commission Directive (EU) 2017/1572 of 15 September 2017, principles and guidelines of good manufacturing practice for medicinal products for human use, Article 9, EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:32017L1572 (read 25 November 2026)
  13. 1321 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 25 November 2026)
  14. 14EudraLex Volume 4, Annex 11: Computerised Systems, revision 1, European Commission, in operation from 30 June 2011. https://health.ec.europa.eu/system/files/2016-11/annex11_01-2011_en_0.pdf (read 25 November 2026)
  15. 15Compliance Program 7346.832, Preapproval Inspections, FDA, issued 29 June 2026, Part III, Objective 3, Data Integrity Audit, and its coverage table. https://www.fda.gov/media/193382/download (read 25 November 2026)
  16. 16Warning Letter 320-⁠26-⁠76 to GC America, Inc., MARCS-⁠CMS 727602, FDA, 14 May 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/gc-america-inc-727602-05142026 (read 25 November 2026)
  17. 17Warning Letter 320-⁠26-⁠32 to Unipack LLC, MARCS-⁠CMS 716621, FDA, 19 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/unipack-llc-716621-12192025 (read 25 November 2026)