Skip to main content

How do you run a tablet friability test, how is friability calculated, and what is the limit?

Friability is the percentage of mass a dedusted tablet sample loses in 100 drum rotations at 24 to 26 rpm: (initial mass − final mass) ÷ initial mass × 100. Under USP <1216>, harmonized with Ph. Eur. and JP, a loss of not more than 1.0% from one test, or the mean of three, is acceptable for most products. Test whole tablets nearest 6.5 g, or 10 above 650 mg. Any obviously cracked, cleaved or broken tablet fails.Checked against its sources on 25 November 2026. How these answers are made

Illustration of a seal beside a rotary tablet press and a tray of tablets.

1Friability formula and acceptance limit.

Friability is the percentage of mass a dedusted sample of tablets loses when tumbled 100 times in a standard drum:

Friability (%) = (W1 − W2) ÷ W1 × 100

W1 is the mass of the dedusted sample before tumbling and W2 its mass after tumbling and dedusting again. The method is harmonized between the European, Japanese and United States pharmacopeias, and it measures abrasion and chipping, not breaking strength: friability “supplements other physical strength test, such as tablet crushing strength.”1

The harmonized text sets the decision rule as well as the procedure. “Generally, the test is run once.” “If obviously cracked, cleaved, or broken tablets are present in the tablet sample after tumbling, the sample fails the test.”1,2 “If the results are difficult to interpret or if the weight loss is greater than the target value, the test should be repeated twice and the mean of the three tests determined.”2 A loss of not more than 1.0% from a single test, or as the mean of three tests, “is considered acceptable for most products”, and effervescent and chewable tablets may have different specifications.1

Table 1. The harmonized decision rule.1,2 The 1.0% applies “unless otherwise specified in the dossier”.3
SituationDecision
Any test: an obviously cracked, cleaved or broken tablet after tumblingThe sample fails, whatever the mass loss
One test: loss at or below the target value, result clearReport the single result; it meets if not more than 1.0%
One test: loss above the target value, or result difficult to interpretRun two more tests and report the mean of three
Three tests: mean not more than 1.0%Meets
Three tests: mean above 1.0%Fails; investigate

USP publishes the method as <1216>, a “general information chapter”, so the 1.0% binds a batch only through its specification or master record.2 ICH Q6A says “It is normally appropriate to perform hardness and/or friability testing as an in-process control”, and puts acceptance criteria in the specification only where friability has a critical impact on quality, as for chewable tablets.4 In-process controls for tablets covers where friability sits in the compression plan.

2How to run the friability test.

The harmonized procedure has seven steps.1,2

  1. Check the apparatus. The drum has an internal diameter between 283.0 and 291.0 mm and a depth between 36.0 and 40.0 mm, with a curved projection that tumbles the tablets at each turn. It rotates at 24 to 26 rpm, so 100 rotations take 231 to 250 seconds. The texts set these requirements but not how to check them; a common practice is to time 100 rotations with a stopwatch, confirm the counter stops at 100, and measure the drum at qualification.
  2. Take the sample. For a unit mass of 650 mg or less, take whole tablets as near as possible to 6.5 g; above 650 mg, take 10 whole tablets.
  3. Dedust and weigh. Dedust the tablets carefully and weigh them accurately. The Japanese Pharmacopoeia defines weighing accurately as weighing to 0.1 mg or finer, chosen for the purpose of the test.5
  4. Tumble. Rotate the drum 100 times.
  5. Inspect. Remove the tablets and look for cracked, cleaved or broken tablets before anything else.
  6. Dedust and reweigh. Remove loose dust as before and weigh accurately.
  7. Calculate the loss and apply the decision rule.

If size or shape makes the tablets tumble irregularly, tilt the drum base to about 10° from the horizontal. Hygroscopic tablets need a suitable humidity environment. Drums with two scooping projections, and apparatus with several drums, are permitted.1

Table 2. Sample size by unit mass, computed from the 6.5 g and 650 mg rule.1 Use the actual mean tablet mass of the batch, not the label strength.
Unit massTablets in the sampleNominal sample mass
100 mg656.50 g
150 mg436.45 g
250 mg266.50 g
320 mg206.40 g
400 mg166.40 g
500 mg136.50 g
650 mg106.50 g
800 mg108.00 g
1,000 mg1010.00 g

3Friability test record template.

One record per sample point, with any repeat tests on the same record. Every field traces to a records rule or to the method.

Table 3. Friability test record. Round only the reportable result; the Japanese Pharmacopoeia obtains an n-figure value by rounding a value of n + 1 figures.5
FieldWhat to recordSource
Product and batchName, dosage form, batch number, stage (cores before coating, end of compression, release) and sample point211.194(a)(1);6 EU GMP 6.17(i), (ii)7
Method and limitMethod and version, the limit, and the target value above which two more tests run211.194(a)(2);6 EU GMP 6.17(iii);7 USP <1216>2
ApparatusFriabilator and drum identity, calibration status; speed check, such as 100 rotations timed at 231 to 250 seconds (practice)211.160(b)(4);6 EU GMP 6.17(ix);7 USP <1216>2
BalanceBalance identity, readability, calibration status211.160(b)(4);6 JP General Notice 245
SampleNumber of tablets and the unit mass that set it; humidity if the tablets are hygroscopicJP G6-⁠5-⁠1811
Initial mass W1Mass after dedusting, with the balance printout or captured value211.194(a)(3), (4)6
Tumbling100 rotations, run time, drum tilt if usedJP G6-⁠5-⁠1811
ObservationCracked, cleaved or broken tablets after tumbling: none, or how many and whatJP G6-⁠5-⁠181;1 EU GMP 6.17(iv)7
Final mass W2Mass after dedusting again211.194(a)(4)6
Calculation(W1 − W2) ÷ W1 × 100, calculated from unrounded masses211.194(a)(5)6
Tests 2 and 3Why they ran, with W1, W2 and loss for eachUSP <1216>;2 FDA OOS guidance8
ResultEvery individual loss, the reportable result, the limit and meets or fails211.194(a)(6);6 EU GMP 6.17(viii)7
SignaturesPerformed by and date; reviewed by and date211.194(a)(7), (8);6 EU GMP 6.17(v) to (vii)7

4How it runs in Seal.

In Seal, friability is one of the checks the compression step expects: weight, hardness and friability run during compression, and the batch step knows whether a result permits continuation, requires adjustment within an approved range, or triggers a hold and investigation. The specification holds the test, acceptance criteria, calculation, units and rounding as approved data, and the calculation uses an approved formula that records its inputs.

People should not retype a value from the balance: Seal captures it and keeps the instrument, acquisition time and extracted values. Configured checks can confirm analyst qualification and instrument state before the test starts, a result outside limits is flagged for assessment, and retests need documented authorization and stay distinct from the original data. See the oral solid dose and pharmaceutical QC blueprints.

5Example: a first result above 1.0%.

The product and data are invented. Exampladine 320 mg uncoated tablets, batch 26-⁠207, are sampled at the end of compression. At 320 mg, 20 tablets weigh 6.40 g and 21 would weigh 6.72 g, so the sample is 20 tablets. The method sets the target value at the 1.0% limit.

Table 4. Batch 26-⁠207, three friability tests. Invented data, calculated by script.
TestW1W2Mass lostLoss
16.4125 g6.3433 g69.2 mg1.079%
26.4087 g6.3540 g54.7 mg0.854%
36.4210 g6.3660 g55.0 mg0.857%
Mean of three0.930%, reported 0.9%

Test 1 lost 1.079%, above the target value, and no tablet was cracked, so the analyst ran tests 2 and 3. The mean of three is 0.930%, reported as 0.9%, which meets the limit of not more than 1.0%. The record keeps all three results. The repeat is part of the method, not an out-of-specification retest. FDA’s OOS guidance “applies to chemistry-based laboratory testing”; applied here by analogy, it accepts that “Where averaging of separate tests is appropriately specified by the test method, a single averaged result can be reported as the final test result.”8

Batch 26-⁠208 lost 6.4302 g to 6.3903 g in a single test, 0.621%. One tablet came out of the drum with its top crown split away. That tablet is obviously cleaved, so the sample fails although the mass loss is well inside the limit.

6Failures and edge cases.

  • A mean above 1.0%, or a broken tablet. The result fails, so treat it as an out-of-specification result and investigate. Further testing cannot be averaged in: FDA says “averaging the result(s) of the original test that prompted the investigation with additional retest or resample results obtained during the OOS investigation is not appropriate because it hides variability among the individual results.”8 The steps are in the OOS investigation page. During compression, a failure also means segregating tablets made since the last good check.
  • The target value. The text leaves it to the user. Fix it in the method before testing; deciding afterwards whether to run tests 2 and 3 invites selective testing.
  • Chewable and effervescent tablets. Their limits may differ.1 FDA’s chewable tablet guidance says “careful attention should be given to tablet size, thickness, and friability, as well as taste”, and Q6A puts their acceptance criteria in the specification.9,4 Set the limit from development and stability data.
  • Coated tablets. “This chapter provides guidelines for the friability determination of compressed, uncoated tablets.”2 Test cores before coating; Q6A treats core friability as an in-process test outside the specification.4 Cores held before coating are retested in a hold time study.
  • Scored tablets. FDA’s scoring guidance asks for data “Confirming that the split tablet portions meet the USP Friability requirement”, alongside a loss-of-mass test on 15 split tablets.10
  • Irregular shapes and hygroscopic tablets. Use the 10° tilt and a controlled humidity; record both.1
  • Process design. Friability depends on formulation and compression settings; quality by design shows a compression force floor set from friability data.

7What inspectors cite.

We found no recent FDA warning letter that cites a friability test by name. Letters do cite the consequence: brittle or crumbling tablets reaching patients, and investigations that stop short.

Cohance Lifesciences Limited, 30 January 2026: FDA described complaints about tablets “reported to be crumbling, disintegrating, pitted, and with dust in the bottle”. The firm later attributed the cause to a raw material, and FDA replied: “Despite identifying a root cause, your response lacks information on chemical testing of reserve samples to confirm that affected tablets meet specifications”.11

Izeen Pharma Inc., 16 May 2019: “Several complaints regarding these lots reported tablet size variation, splitting, brittleness, and disintegration upon touching.” The firm’s investigation traced it to “the physical properties of the blend and an unsuitable high-speed compression machine”, but “you failed to evaluate other lots produced using the same manufacturing process”.12

Zhejiang Huahai Pharmaceutical, 6 June 2025: in a compression reject-limit study comparing force with weight, “there is no rationale provided for why other parameters, including hardness, were not measured or considered.”13

The common thread is a missing link between complaints and the batch’s own mechanical-strength data. When a complaint reports chipped or crumbling tablets, pull the friability and hardness results for that batch and its neighbors, as complaint handling describes.

8USP, Ph. Eur. and JP: one text, and what changed.

The Pharmacopoeial Discussion Group signed Revision 1 of the friability text on 7 June 2022.14 Each pharmacopeia then implemented it on its own date, and the sign-off records no local requirements for any of them.15

Table 5. The harmonized friability text by pharmacopeia. This page quotes the JP18 Supplement I English text and the USP Stage 4 text.
PharmacopeiaTextIn force or official from
Japanese PharmacopoeiaGeneral Information G6-⁠5-⁠181, Tablet Friability Test (JP18 Supplement I)12 December 202214
European Pharmacopoeia2.9.7 Friability of uncoated tablets, revised for Supplement 11.31 January 202416
United States PharmacopeiaGeneral chapter <1216> Tablet Friability, Stage 4 text1 August 20232
Indian PharmacopoeiaAdded to the Revision 1 sign-off as a new PDG memberSign-off May 202415

Revision 1 made three changes that matter. The speed became “between 24 and 26 rpm”, replacing “rotates at 25 ± 1 rpm”.1,17 The acceptance sentence, which read “A maximum mean mass loss from the three samples”, now accepts a single test or the mean of three.17,1 And the drum dimensions gained a decimal place. EDQM summarized the revision as made “to unify the presentation of the apparatus dimensional requirements and to clarify the test criteria”.16 ICH Q4B Annex 9 had already accepted a single determination when it reached Step 4 in October 2009; its R1 revision of 2010 added Health Canada.3

The JP says its General Information “should not be taken as indicating standards for conformity of drugs” except where an approval specifies it, so in Japan, as under USP <1216>, the limit binds through the dossier and the specification.17,2 Interchangeability under Q4B Annex 9 continues to spread: China’s NMPA implemented it on 1 October 2025 and Egypt’s EDA on 1 May 2026.18 JP19 took effect on 10 April 2026; MHLW’s English pages still carry JP18 and its supplements.19,20

AQuestions and answers

What is the friability limit for chewable tablets?

There is no harmonized number. The pharmacopeial text says chewable and effervescent tablets may have different friability specifications, and ICH Q6A puts acceptance criteria for chewable tablets in the specification because friability affects their quality. Set the limit from development and stability data, and justify it in the dossier.

Can the friability test be run on film-coated tablets?

The harmonized chapter is written for compressed, uncoated tablets. For coated products, run it on the cores before coating, as an in-process control. The finished coated tablet is controlled through the other attributes in its specification.

What is the difference between tablet hardness and friability?

Hardness, or breaking force, is the force needed to break one tablet in a set plane. Friability is the mass a sample of tablets loses by abrasion and chipping when tumbled. The pharmacopeia treats friability as a supplement to breaking strength, so most compression plans check both.

How many tablets are needed for a friability test?

For tablets of 650 mg or less, take whole tablets as near as possible to 6.5 g: 20 tablets at 320 mg, for example. For tablets above 650 mg, take 10 whole tablets. Base the count on the batch’s actual mean tablet mass.

Does the Indian Pharmacopoeia use the same friability test?

The Indian Pharmacopoeia Commission joined the sign-off of the harmonized Revision 1 text in May 2024, as a new member of the Pharmacopoeial Discussion Group, with no local requirements recorded. Check that the IP edition you work to carries the revised text.

References20 sources

References

  1. 1The Japanese Pharmacopoeia, 18th edition, Supplement I, English version, General Information, Tablet Friability Test G6-⁠5-⁠181 (PDG G-⁠06 Rev. 1), MHLW, in effect from 12 December 2022. https://www.mhlw.go.jp/content/11120000/001372370.pdf (read 25 November 2026)
  2. 2<1216> Tablet Friability, Stage 4 harmonisation text, official 1 August 2023, United States Pharmacopeial Convention, 2022, Internet Archive copy of 31 October 2026. https://web.archive.org/web/20261031025328/https://www.usp.org/sites/default/files/usp/document/harmonization/excipients/m99935.pdf (read 25 November 2026)
  3. 3ICH Q4B Annex 9(R1), Evaluation and Recommendation of Pharmacopoeial Texts for Use in the ICH Regions on Tablet Friability General Chapter, sections 2.1 and 2.2, ICH, 27 September 2010. https://database.ich.org/sites/default/files/Q4B%20Annex%209%28R1%29%20Guideline.pdf (read 25 November 2026)
  4. 4ICH Q6A Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and New Drug Products: Chemical Substances, sections 2.3 and 3.3.2.1(c), ICH, 6 October 1999. https://database.ich.org/sites/default/files/Q6A%20Guideline.pdf (read 25 November 2026)
  5. 5The Japanese Pharmacopoeia, 18th edition, English version, General Notices 24 and 25, MHLW, 2021. https://www.mhlw.go.jp/content/11120000/001372337.pdf (read 25 November 2026)
  6. 621 CFR Part 211, sections 211.160(b)(4) and 211.194(a), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 25 November 2026)
  7. 7EudraLex Volume 4, Part I, Chapter 6, Quality Control, paragraph 6.17, European Commission, in operation from 1 October 2014. https://health.ec.europa.eu/system/files/2016-11/2014-11_vol4_chapter_6_0.pdf (read 25 November 2026)
  8. 8Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, Revision 1, section IV.C.1, FDA, May 2022. https://www.fda.gov/media/158416/download (read 25 November 2026)
  9. 9Quality Attribute Considerations for Chewable Tablets, guidance for industry, section III, FDA, August 2018. https://www.fda.gov/media/98598/download (read 25 November 2026)
  10. 10Tablet Scoring: Nomenclature, Labeling, and Data for Evaluation, guidance for industry, section III.A, FDA, March 2013. https://www.fda.gov/media/81626/download (read 25 November 2026)
  11. 11Cohance Lifesciences Limited, warning letter 320-⁠26-⁠40, MARCS-⁠CMS 718812, FDA, 30 January 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/cohance-lifesciences-limited-718812-01302026 (read 25 November 2026)
  12. 12Izeen Pharma Inc., warning letter, MARCS-⁠CMS 565311, FDA, 16 May 2019. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/izeen-pharma-inc-565311-05162019 (read 25 November 2026)
  13. 13Zhejiang Huahai Pharmaceutical Co., Ltd., warning letter 320-⁠25-⁠78, MARCS-⁠CMS 707145, FDA, 6 June 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/zhejiang-huahai-pharmaceutical-co-ltd-707145-06062025 (read 25 November 2026)
  14. 14Pharmacopoeial Discussion Group harmonised general chapters, G-⁠06 Tablet Friability, sign-off and implementation dates, PMDA, as of 7 October 2026. https://www.pmda.go.jp/english/rs-sb-std/standards-development/jp/0027.html (read 25 November 2026)
  15. 15G-⁠06 Tablet Friability, Rev. 1, sign-off cover sheet version 2, Pharmacopoeial Discussion Group, May 2024. https://www.pmda.go.jp/files/000269520.pdf (read 25 November 2026)
  16. 16European Pharmacopoeia Commission, 174th session, 22 and 23 November 2022, summary of adopted texts, EDQM, as published by RIVM. https://rivm.nl/sites/default/files/2024-07/besluitenlijst%20EFC-vergadering%20174_november%202022.pdf (read 25 November 2026)
  17. 17The Japanese Pharmacopoeia, 18th edition, English version, General Information, preface and Tablet Friability Test G6-⁠5-⁠150, MHLW, 2021. https://www.mhlw.go.jp/content/11120000/001372365.pdf (read 25 November 2026)
  18. 18Quality Guidelines, Q4B Annex 9(R1) implementation status, ICH. https://www.ich.org/page/quality-guidelines (read 25 November 2026)
  19. 19Implementation date of JP editions and supplements, PMDA. https://www.pmda.go.jp/files/000250430.pdf (read 25 November 2026)
  20. 20The Japanese Pharmacopoeia (English version), MHLW. https://www.mhlw.go.jp/stf/seisakunitsuite/bunya/0000066597.html (read 25 November 2026)

See your process in Seal.

Bring a procedure or a recurring problem. See how your team configures it with Neil and improves the next run.