1How to calculate F0.
F0 is the lethality of a saturated steam process: “the equivalent time in minutes at a temperature of 121 °C delivered by the process to the load in its container with reference to micro-organisms possessing a theoretical Z-value of 10”.1 Each reading becomes a lethal rate, the minutes at 121 °C that one minute at that temperature is worth, and F0 sums them:
F0 = Δt × Σ 10^((T − 121)/10)
T is the temperature in °C at a load probe and Δt the reading interval in minutes; the cycle’s F0 is the one in the slowest-to-heat container, not at the drain or control sensor.2,
| Temperature | Lethal rate per minute |
|---|---|
| 110 °C | 0.079 |
| 115 °C | 0.251 |
| 118 °C | 0.501 |
| 120 °C | 0.794 |
| 121 °C | 1.000 |
| 122 °C | 1.259 |
In a spreadsheet, put the time of each reading in column A from row 2, the probe temperature in column B and the reading interval in minutes in G1 (0.5 for 30-second readings). In C2 enter =IF(B2>=110,10^((B2-
Kaye’s validation software averages each lethal rate with the one before it;3 to match its number, enter =(C2+C3)/2*$G$1 in D3, fill it down and sum column D.
Two more choices change the number, so the protocol fixes both. The Japanese Pharmacopoeia’s 121.1 °C reference gives a result 2.3% lower than EMA’s 121 °C.4 And EMA says: “If temperatures below 110 °C are included (during heat-up and cool-down) in the determination of F0, this should be justified.”1
The F0 a load needs follows from its bioburden: Health Canada’s minimum F0 is D121 × (log A − log B), where A is the count per container, B the accepted probability of survival (10⁻⁶) and D121 that of the most heat-resistant organism or, more conservatively, a resistant spore former.5
2Example: physical and biological F0 in a PQ run.
The site, product and data are invented. Run PQ-AC05-02 is a performance qualification (PQ) run of cycle FL-121 on autoclave AC-05, with the maximum load of 2,400 vials of a 10 ml aqueous solution. The chamber drain, where the control sensor sits, reached 121.0 °C at minute 5. Probe TC-07, in a vial at the cold spot found in the heat distribution studies, reached 121 °C at minute 8. The 15-minute exposure was timed from then, not from minute 5, because “the whole of the load should reach the required temperature before measurement of the sterilising time-period starts” (Annex 1, 8.52);6 the drain-controlled cycle was set to 18 minutes.
Those 3 minutes are the equilibration time: in HTM 01-01, the NHS steam sterilizer guidance, from when the reference point (here the drain) “first attains the sterilization temperature” until the slowest-to-heat part of the load does.7 For fluid loads, FDA investigators ask:8
“Has the firm determined lag times for all container sizes/mass, product viscosities, etc. and adjusted their cycles accordingly?”
| Minutes | TC-07 reading | Lethal rate | F0 added |
|---|---|---|---|
| 1–2 | 104.0, 108.5 °C | Below 110 °C | 0 |
| 3–7 | 111.2 to 120.3 °C | 0.105, 0.200, 0.363, 0.589, 0.851 | 2.11 |
| 8–22 | 121.4 °C each | 1.096 each | 16.45 |
| 23–25 | 119.6 to 112.4 °C | 0.724, 0.331, 0.138 | 1.19 |
| 26–27 | 108.1, 103.5 °C | Below 110 °C | 0 |
| Total | 19.75 |
- F0 at TC-07 is 19.75 minutes; at TC-02, the fastest-heating vial, 23.85. Both meet the protocol’s limits: at least 15 at every load probe, at most 30 at any, the ceiling set by the product’s heat degradation data.
- Kaye’s averaging also gives 19.75. Counting from 100 °C gives 19.89; a 121.1 °C reference gives 19.30.
- The bioburden limit is 100 CFU/100 ml, so a vial holds at most 10 CFU. Taking the indicator’s D121 of 1.8 minutes, confirmed in the product solution, the minimum F0 is 1.8 × (log 10 − log 10⁻⁶) = 12.6 minutes,5 so an F0 8 cycle would fall short unless the bioburden is shown to be less heat-resistant. TC-07’s 19.75 gives 19.75 ÷ 1.8 = 10.97 log reductions: a sterility assurance level (SAL) of 10 × 10^−10.97 = 1.1 × 10⁻¹⁰. EMA prefers this calculation, “from the maximum bioburden per container and the D-value of the biological indicator used in the validation”.1
- For every cycle but the Ph. Eur. reference, EMA asks for “Validation of F0Phys and F0Bio” and their relationship. F0Bio is the indicators’ log reduction times their D121: EMA’s overkill, “F0BIO > 12 minutes”, is for example “at least a 12 log reduction” of organisms “having a minimum D value of 1 minute”.1 The indicator beside TC-07, a product vial inoculated with 10⁶ spores, showed no growth, so F0Bio there is at least 6 × 1.8 = 10.8 minutes. The probe’s 10.97 log reductions exceed the indicator’s 6, so the two agree; growth there would contradict the probe.
- A weaker indicator flatters the result: “the SAL demonstrated becomes numerically higher, but does not provide as high a safety margin”. So EMA expects a D121 of at least 1.5 minutes, or additional data, and “The SAL should always be established in relation to a D-value that is higher than that of the normal bioburden at routine production.”1 Annex 1 normally accepts the certificate’s D-value (8.43);6 Health Canada says it “must be assessed along with the material to be sterilized and its formulation”, since resistance can change with pH and other parameters.9
3Heat penetration summary: template.
FDA accepts a data summary of each loading study;10 this one adds what investigators ask about and the signatures.
| Field | Run PQ-AC05-02 | Source |
|---|---|---|
| Autoclave | AC-05, with manufacturer and model | FDA II.A.3, II.B.3 |
| Cycle and load | FL-121; maximum load, 2,400 × 10 ml vials | FDA II.B.3; CP question 20 |
| Protocol | VP-AC05-01, approved by QA before the first run | PI 006-4 4.1.6, 4.3.3.2 |
| Run | 14 October 2026, 09:40; 2 of 3 consecutive | FDA II.B.1, II.B.3 |
| Executed by | Site validation engineer or qualified contractor, signed | PI 006-4 4.3.3.1 |
| Equilibration time | 3 minutes: drain at minute 5, TC-07 at minute 8 | HTM 2.85, 2.86; CP question 24 |
| Dwell time | 15 minutes, from TC-07 at 121 °C | FDA II.B.3; Annex 1 8.52 |
| Dwell range | 121.4 to 121.9 °C, average 121.6 °C; spread 0.5 °C, limit 2.0 °C | FDA II.B.3; Annex 1 8.59 |
| Minimum F0 | 19.75 at TC-07; limit 15 | FDA II.A.3, II.B.3 |
| Maximum F0 | 23.85 at TC-02; limit 30 | FDA II.A.3, II.B.3 |
| Probes | 12, as on the load diagram; calibration passed before and after the run | FDA II.B.2; CP question 22; GUI-0074 |
| Indicator lot | Geobacillus stearothermophilus lot 2611, 10⁶ spores per inoculated vial; D121 1.8 min and z-value from the certificate, D121 confirmed in the product solution; count, purity and identity verified before first use | FDA II.C.3, II.C.4; Annex 1 8.43; GUI-0074 phase 2 |
| Indicator results | In product vials beside each probe, TC-07 included; all negative, positive control grew; F0Bio at least 10.8 | FDA II.C.5; Annex 1 8.42; EMA Table 1; GUI-0074 phase 2 |
| Deviations | None; any would also enter the deviation system | PI 006-4 3.4.3(l) |
| Data integrity | Confirmed by a QA reviewer who did not run it | PI 006-4 3.4.3(m) |
| Report approval | QA, once the third run’s indicators are read | FDA PV IV.C.1; PI 006-4 3.4.3(b) |
3.1When a PQ run fails.
A failed run cannot be one of FDA’s “minimum of three consecutive, successful cycles”, so the series starts again after the investigation that Annex 1 requires for any cycle that deviated, even one with a longer heating phase (8.41).10,
| Failure | What follows | Source |
|---|---|---|
| Low F0 at a probe | Investigate and correct the cycle or load; negative BIs don’t override it | Annex 1 8.41, 8.42; CP question 23 |
| Positive BI | Sterilization not achieved: look for a processing error, and with none, the process is unacceptable | GUI-0074 phase 3; CP question 23 |
| Probe out of calibration | A deviation; the report says whether that probe’s data still stand | PI 006-4 3.4.3(l); CP question 22 |
4How long autoclave PQ takes, and who approves it.
FDA’s process validation guidance says “The quality control unit must review and approve the qualification plan and report”.12
| Days | Step | What it involves | Record |
|---|---|---|---|
| 1–5 | IQ and OQ | Control and recording instruments calibrated | IQ and OQ reports |
| 6–7 | Empty chamber | Heat distribution within the protocol’s uniformity limits | Chamber profile |
| 8–10 | Loaded chamber | Maximum and minimum loads, probes between units, repeated to see if the cold spot moves | Certified profile per load |
| 11–16 | Heat penetration | Probes in containers filled to maximum with the slowest-heating solution, BIs alongside; three consecutive runs per load, one a day | A summary per run, as above |
| 17–23 | Incubation | The last run’s BIs read after 7 days | BI results, positive controls |
| 24–30 | Report | Data integrity check, deviations, QA approval | Routine parameters and loading patterns |
Indicators set the end date: FDA’s guidance for indicator makers says “Generally, FDA recommends 7 days as the conventional incubation time”, and a shorter time has to be validated.14 Each further loading pattern repeats days 8 to 30, a little over three weeks.
5What autoclave validation must show.
21 CFR 211.113(b) requires “validation of all aseptic and sterilization processes”,15 and Annex 1 asks for sterilizing conditions “in all parts of each type of load to be processed” (8.36).6 Validation also covers these checks.
| Check | What to show | When |
|---|---|---|
| Steam quality | Non-condensable gases, dryness, superheat; condensate meets the WFI monograph | PQ at every user point, then periodically |
| Leak test | Chamber holds vacuum | Normally weekly for vacuum cycles |
| Air removal | Air removal test cycle or an air detector | Normally daily, for cycles that purge air |
| Porous loads | Items dry on removal | Every cycle |
| Routine cycle | Unique cycle ID; record reviewed for batch release | Every cycle |
For limits, HTM 01-01 Part C gives the European standard values: non-condensable gases at most 3.5%, superheat at most 25 °C, and dryness at least 0.9, or 0.95 for metal loads.7
5.1Where FDA, the EU and Health Canada differ.
- FDA: its 1994 guidance sets no minimum F0 and no requalification interval: the application states its own minimum and maximum F0 and requalification “frequency”.10
- EU: Annex 1, in operation since 25 August 2023, requires revalidation “at least annually for load patterns that are considered worst case”, and other patterns at a frequency justified in the contamination control strategy (8.39).6
- Health Canada sets the number of runs by risk management and makes physical and microbial requalification at least once a year a “must”.2,
13
6Which steam cycle to choose.
EMA lists its cycles from most to least lethal, and “the first feasible process should be selected”: the first the product survives.1
| Cycle | Lethality | Hold | Bioburden | Dossier |
|---|---|---|---|---|
| Ph. Eur. reference | ≥ 15 min | ≥ 121 °C | 100 CFU/100 ml, not routine | Confirmation of validation only |
| Overkill | F0 > 12 | ≥ 121 °C | 100 CFU/100 ml, not routine | Validation summary |
| F0 > 8, above 115 °C | F0 > 8 | > 115 °C | Routine; characterized if BI D121 < 1.5 | Summary, plus more data with a low-D BI |
| F0 > 8, above 110 °C | F0 > 8 | > 110 °C | Routine and characterized | As above |
| Post-aseptic heat | F0 < 8 | > 110 °C | 0 CFU/100 ml after aseptic filtration | Summary plus additional data |
7What inspectors ask and cite.
FDA investigators also ask about routine cycles:8
“Is the slowest to heat spot (‘cold spot’) in each load monitored during each autoclave cycle?”
FDA’s warning letter of 29 October 2019 to Cadila Healthcare, Moraiya, shows where a failed requalification leads. In June 2018 a periodic requalification did not reach the required F0; the firm named a root cause, judged other products unaffected and did not extend the CAPA. In March 2019 the same equipment failed again:16
“Again, several sensors did not achieve the required F0, and one did not reach the (b)(4) temperature. In addition, at the end of the incubation, the biological indicators at multiple locations in the (b)(4) showed microbiological growth.”
A batch was recalled. FDA found the investigation “did not sufficiently address why your originally validated cycle parameters were not met and why the process fell out of a state of control” and asked for a review of every affected US batch still in expiry, including the F-value, z-value and indicator D-value data behind it.16
8What changed from 2024 to 2026.
- ISO 17665:2024, published in March 2024, replaced ISO 17665-1:2006 and ISO/TS 17665-2 and 17665-3.17 EMA’s guideline still cites Parts 1 and 2.1
- Health Canada’s GUI-0074, published on 12 November 2024, replaced its separate moist heat, gaseous and irradiation guides.18
- PIC/S PI 006-4 came into force on 1 October 2026, the first revision since PI 006-3 of September 2007.11
- USP’s 2026 Pharmacopeial Forum proposes renaming <1229> “Sterilization” and omitting <1229.5>, <1229.9> and <1229.14>, folded into it. <1229.1> Steam Sterilization by Direct Contact would become “Moist Heat Sterilization”, absorbing text from <1229.2> Moist Heat Sterilization of Aqueous Liquids, which turns into “Sterilizing Filtration of Liquids”.19,
20, 21
9How it runs in Seal.
In Seal, a validated load pattern becomes the template each autoclave load is scanned against, and the cycle’s phases, lethality, probe traces and indicators stay attached to that load and its items. A wet pack is bounded to the positions and items it affects, and load release is a quality decision, not the controller’s cycle-complete flag.
Released items carry their sterile hold into the aseptic batches that use them, so a later failed indicator or calibration finding is bounded to the affected items and the batches that consumed them.
AQuestions and answers
What is the difference between overkill and bioburden-based sterilization?
An overkill cycle delivers so much lethality that the bioburden hardly matters: EMA lists it as a cycle with F0 above 12 minutes, and the Japanese Pharmacopoeia as a 12 log reduction, also referred to as F0 12, regardless of bioburden. A bioburden-based cycle delivers less heat, so it depends on knowing how many organisms each container holds and how heat-resistant they are. FDA says more bioburden information may be needed for those cycles, and EMA requires routine bioburden testing for them.
How many temperature probes does autoclave validation need?
FDA’s 1994 guidance asks for the number of thermal monitors and their location in the chamber to be described, with a diagram if helpful. Health Canada asks for multiple probes in each run, each calibrated before and after use, placed to show the temperature distribution and find the slowest-heating units. The protocol sets the number from the chamber, the load and the heat distribution results, and justifies it.
Does each fill volume need its own heat penetration study?
Not automatically, but it needs to be covered. FDA says more studies may be needed when different fill volumes run in the same container line, and its investigators ask whether lag times were determined for every container size, mass and viscosity. A worst-case approach, such as validating the largest fill of the slowest-heating solution, has to be shown to cover the rest.
Is F0 used for dry heat sterilization?
No. EMA and the Japanese Pharmacopoeia define F0 for moist heat, with a z-value of 10 °C. The Japanese Pharmacopoeia defines a separate FH for dry heat: equivalent minutes at 160 °C for organisms with a z-value of 20 °C. For dry heat, EMA asks for the cycle’s time and temperature and a bioburden limit to be stated, with the Ph. Eur. reference of at least 160 °C for at least 2 hours, and cycles below it validated physically and biologically to an SAL of 10⁻⁶.
Does a liquid load need a Bowie-Dick test?
Only if its cycle purges air. The Bowie-Dick test shows whether air removal from, and steam penetration into, a standard test pack is even and rapid; it does not show that the load reached sterilizing conditions. Annex 1 asks for an air removal test cycle, normally daily, or an air detector when the process includes air purging, as porous loads do. For fluid cycles it asks for temperature, time and/or F0 instead, and an autoclave that also runs porous cycles still needs the test for them.
References21 sources
References
- 1Guideline on the sterilisation of the medicinal product, active substance, excipient and primary container, EMA/
CHMP/ CVMP/ QWP/ 850374/ 2015, European Medicines Agency, effective 1 October 2019, section 4.1.1, Table 1 and definitions. https:/ / www. ema. europa. eu/ en/ documents/ scientific- guideline/ guideline- sterilisation- medicinal- product- active- substance- excipient- and- primary- container_ en. pdf (read 9 November 2026) - 2Guide to validation of terminal sterilization process of drugs (GUI-0074), Process validation, sterilization by moist heat, Health Canada, 12 November 2024. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ guide- validation- terminal- sterilization- process- drugs- gui- 0074/ process- validation- sterilization- moist- heat. html (read 8 November 2026) - 3Understanding Lethality Calculation: Minutes or Seconds?, Kaye (Amphenol), 23 June 2023, the validation system maker’s description of its own calculation. https:/
/ www. kayeinstruments. com/ en/ news/ blog- post/ understanding- lethality- calculation- minutes- or- seconds (read 8 November 2026) - 4The Japanese Pharmacopoeia, 18th edition, English version, General Information G4-10-162, Sterilization and Sterilization Indicators, Ministry of Health, Labour and Welfare. https:/
/ www. mhlw. go. jp/ content/ 11120000/ 000904454. pdf (read 8 November 2026) - 5Guide to validation of terminal sterilization process of drugs (GUI-0074), Phase 1, process design, Health Canada, 12 November 2024. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ guide- validation- terminal- sterilization- process- drugs- gui- 0074/ phase- 1- process- design. html (read 8 November 2026) - 6EudraLex Volume 4, Annex 1, Manufacture of Sterile Medicinal Products, European Commission, 22 August 2022, document history and sections 6.17 and 8.36 to 8.61. https:/
/ health. ec. europa. eu/ system/ files/ 2022- 08/ 20220825_ gmp- an1_ en_ 0. pdf (read 9 November 2026) - 7Health Technical Memorandum 01-01, Management and decontamination of surgical instruments, Part C, Steam sterilization, Department of Health, 2016, paragraphs 1.41, 2.85, 2.86, 3.20 and 3.30 and Table 5. https:/
/ www. england. nhs. uk/ wp- content/ uploads/ 2021/ 05/ HTM0101PartC. pdf (read 8 November 2026) - 8Compliance Program 7356.002A, Sterile Drug Process Inspections, Attachment A, moist heat questions 8 to 24, FDA, implemented 11 September 2015. https:/
/ www. fda. gov/ media/ 75174/ download (read 9 November 2026) - 9Guide to validation of terminal sterilization process of drugs (GUI-0074), Phase 2, process performance qualification, Health Canada, 12 November 2024. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ guide- validation- terminal- sterilization- process- drugs- gui- 0074/ phase- 2- process- performance- qualification. html (read 9 November 2026) - 10Guidance for Industry for the Submission Documentation for Sterilization Process Validation in Applications for Human and Veterinary Drug Products, FDA, November 1994, sections II.A to II.C. https:/
/ www. fda. gov/ media/ 71442/ download (read 9 November 2026) - 11PI 006-4, Recommendations on Qualification and Validation, PIC/S, in force 1 October 2026, document history and sections 3.4.3, 4.1.6, 4.3.3, 6.3.9 and 6.3.11. https:/
/ picscheme. org/ docview/ 11277 (read 8 November 2026) - 12Process Validation: General Principles and Practices, guidance for industry, FDA, January 2011, section IV.C.1. https:/
/ www. fda. gov/ media/ 71021/ download (read 8 November 2026) - 13Guide to validation of terminal sterilization process of drugs (GUI-0074), Phase 3, ongoing process verification, Health Canada, 12 November 2024. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ guide- validation- terminal- sterilization- process- drugs- gui- 0074/ phase- 3- ongoing- process- verification. html (read 8 November 2026) - 14Biological Indicator (BI) Premarket Notification [510(k)] Submissions, guidance for industry and FDA staff, FDA, 4 October 2007, section E. https:/
/ www. fda. gov/ media/ 71134/ download (read 8 November 2026) - 1521 CFR 211.113, Control of microbiological contamination, eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- C/ part- 211/ subpart- F/ section- 211. 113 (read 8 November 2026) - 16Warning letter 320-20-05 to Cadila Healthcare Limited, Moraiya, FDA, 29 October 2019, observation 2. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ cadila- healthcare- limited- 584856- 10292019 (read 8 November 2026) - 17ISO 17665:2024, Sterilization of health care products, Moist heat, ISO, March 2024. https:/
/ www. iso. org/ standard/ 80271. html (read 8 November 2026) - 18Overview: Guide to validation of terminal sterilization process of drugs (GUI-0074), Health Canada, published 12 November 2024. https:/
/ www. canada. ca/ en/ health- canada/ services/ drugs- health- products/ compliance- enforcement/ good- manufacturing- practices/ guidance- documents/ guide- validation- terminal- sterilization- process- drugs- gui- 0074. html (read 8 November 2026) - 19<1229> Sterilization of Compendial Articles, Pharmacopeial Forum briefing, case SUB-2643, United States Pharmacopeia, 2026. https:/
/ doi. usp. org/ USPNF/ USPNF_ M7408_ 30101_ 01. html (read 8 November 2026) - 20<1229.1> Steam Sterilization by Direct Contact, Pharmacopeial Forum briefing, case SUB-2644, United States Pharmacopeia, 2026. https:/
/ doi. usp. org/ USPNF/ USPNF_ M7409_ 20101_ 01. html (read 9 November 2026) - 21<1229.2> Moist Heat Sterilization of Aqueous Liquids, Pharmacopeial Forum briefing, case SUB-2645, United States Pharmacopeia, 2026. https:/
/ doi. usp. org/ USPNF/ USPNF_ M7411_ 30101_ 01. html (read 9 November 2026)