What system suitability criteria does an HPLC method need under USP <621>, and how is the maximum RSD worked out?

HPLC tests meet their monograph’s criteria plus two USP <621> defaults: symmetry (tailing) factor 0.8 to 1.8 for the quantified peak in organic impurities, related substances and assays, waived in impurity tests that prescribe an RSD; and S/N of at least 10 at a reporting threshold. In-house methods follow their approved criteria. Pure-substance assays without a stated limit allow RSD up to 0.349 × B√n / t(90%, n−1), B being the upper limit minus 100%; drug products follow their monograph.Checked against its sources on 25 November 2026. How these answers are made

Illustration of a seal beside a modular HPLC system and its chromatogram.

1The system suitability criteria in USP <621>, with their formulas.

System suitability is the check, run with each sequence, that the chromatographic system is fit for the test in hand.1 The harmonized chapter (USP <621>, Ph. Eur. 2.2.46 and JP <2.00>) is blunt: “Compliance with the system suitability criteria is required throughout the chromatographic procedure. No sample analysis is acceptable unless the suitability of the system has been demonstrated.”2,3

The chapter lists the parameters a method may use: “Column plate number, retention factor (mass distribution ratio), system repeatability, signal-to-noise, symmetry factor and resolution/peak-to-valley ratio”.3 The monograph chooses which apply and sets their limits, and the chapter adds defaults to them: “The following requirements are to be fulfilled, in addition to any other system suitability criteria stated in the monograph. When specific requirements are stated in the monograph, they supersede the requirements mentioned in this chapter.”2 An approved in-house method is held to its own criteria instead.4

Table 1. Parameters, formulas and defaults of the harmonized chromatography chapter.2,3
ParameterFormulaDefault in the chapterWhere the limit usually comes from
Repeatability, %RSD(100 / ȳ) × √(Σ(yᵢ − ȳ)² / (n − 1)), on at least 3 injections of a reference solution%RSDmax, for assays of a substance with no stated limitMonograph or method; see the next section
Symmetry (tailing) factor, AsW0.05 / 2d, measured at one-twentieth of peak height0.8 to 1.8 for the peak used for quantitationMonograph, which may set its own minimum or maximum
Resolution, Rs1.18 (tR2 − tR1) / (Wh1 + Wh2), widths at half-heightNoneMonograph, for a named pair of peaks
Plate number, N5.54 (tR / Wh)²NoneMonograph or method
Signal-to-noise, S/N2H / h, noise from a blank over 20 × Wh (at least 5 × Wh if needed)Limit of quantitation (S/N 10) at or below the reporting thresholdImpurity tests with a reporting threshold
Peak-to-valley, p/vHp / HvNoneMonograph, when two peaks are not baseline-separated
Retention factor, k(tR − tM) / tMNoneMonograph or method

The symmetry factor is the old USP tailing factor, “also known as the asymmetry factor or tailing factor”.2 And “It is the responsibility of the user to ensure that the calculation methods used in the software are equivalent to the requirements” of the pharmacopeia.3

2How the maximum permitted RSD is worked out.

For an assay of an active substance or excipient, “where the target value is 100% for a pure substance, and a system repeatability requirement is not specified”, the chapter derives the repeatability limit from the specification and the number of injections:2

%RSDmax = K × B × √n / t90%,n−1

  • K is 0.349, obtained as (0.6 / √2) × (t90%,5 / √6), where 0.6 / √2 (0.42%) is the RSD required on six injections when B is 1.0.
  • B is the upper assay limit in the monograph minus 100%. For 98.0% to 102.0%, B is 2.0.
  • n is the number of replicate injections of the reference solution, from 3 to 6.
  • t90%,n−1 is Student’s t at the 90% level (two-sided) with n − 1 degrees of freedom: 2.920, 2.353, 2.132 and 2.015 for n of 3, 4, 5 and 6.2,3
Table 2. Maximum permitted RSD of peak response for assays, from USP <621> Table 2 and JP <2.00> Table 2.00-⁠1.2,3
Bn = 3n = 4n = 5n = 6
2.0%0.41%0.59%0.73%0.85%
2.5%0.52%0.74%0.92%1.06%
3.0%0.62%0.89%1.10%1.27%

For other limits, compute it. A specification of 98.0% to 101.5% gives B = 1.5, so with six injections %RSDmax = 0.349 × 1.5 × 2.449 / 2.015 = 0.64%; with five, 0.349 × 1.5 × 2.236 / 2.132 = 0.55%; with three, 0.31%.

Most methods state their own limit, and then the formula does not apply. For a stated limit, USP’s national text sets the number of injections: “if the requirement is 2.0 or less the calculation is based on data from five replicate injections of the analyte, if the requirement is more than 2.0% data from six replicate injections are used.”2 The 2026 revision adds the missing % sign after the first 2.0.5 The Japanese Pharmacopoeia says that “In principle, total number of replicate injections should be 6”, and gives limits for fewer injections that keep the test as strict: a 2.0% limit at six injections becomes 1.76% at five, 1.43% at four and 0.95% at three.1,6

3Peak symmetry and system sensitivity, and where they apply.

The harmonized text adds two defaults to a monograph’s own criteria. For symmetry: “Unless otherwise stated, in a test or assay, the symmetry factor (tailing factor) of the peak used for quantitation is 0.8 to 1.8.”3 For sensitivity: “The limit of quantitation (corresponding to a signal-to-noise ratio of 10) is equal to or less than the reporting threshold.”2 Each pharmacopeia scopes them differently.

Table 3. Where the two defaults apply.4,5,7,8,9
PharmacopeiaSymmetry 0.8 to 1.8System sensitivity
Ph. Eur. 2.2.46, all monographs from 1 January 2023Peaks quantified by external standardization in LC or GC tests and assays; not area normalizationTests with a reporting threshold or disregard limit; not assays
USP <621>, official 1 June 2026Organic impurities, related substances and assay; waived in the two impurity tests when an RSD of peak response is prescribedImpurity tests whose monograph prescribes a reporting threshold; S/N not lower than 10
JP <2.00>, from 12 December 2022Only where a monograph written to <2.00> specifies itOnly where a monograph written to <2.00> specifies it

For Ph. Eur., the S/N at the reporting threshold “must be greater than 10.”4 USP’s 2026 revision measures sensitivity on a solution at the reporting threshold and keeps the symmetry waiver to impurity tests: USP refused to extend it to assays “because the risk of error is high since some impurities may be hidden in the tailing of a large main peak.”5,10 JP makes both conditional: “When the following criteria are specified in the system suitability tests, each requirement is to be fulfilled unless otherwise prescribed.”3 PMDA also said when it consulted on the text that “<2.00> is not intended to be applied retrospectively to the monographs listed in JP”, and the two liquid chromatography chapters are alternatives: “Liquid Chromatography <2.01> and Chromatography <2.00> cannot be applied in combination.”9,1

4System suitability record and review checklist.

A reviewer should be able to rebuild the decision from the record: the criteria and their source, every injection, and the calculation.

Table 4. System suitability record and review checklist, with the source of each line.
LineWhat to recordWhy
MethodMethod or monograph, version, effective date and pharmacopeia editionMonograph criteria apply, with the chapter defaults added2
Suitability solutionReference standard, lot, potency, expiry and preparation recordQualified primary or secondary standard, never the batch under test11
Injection planNumber of replicate injections, and bracketing or interspersed standards, fixed before the runn sets the RSD limit; JP allows standards before, after and among samples1
Criteria and sourceEach limit, marked as monograph, approved method or chapter defaultMonograph limits supersede the defaults2
RepeatabilityEach response, mean, SD, %RSD and n; for an unstated limit, B, t and %RSDmaxFormula and Table 22
SymmetryAs of the quantified peak, at one-twentieth of its heightDefault 0.8 to 1.83
Resolution, p/v, N and kThe named peaks, their tR and Wh, and each value specifiedMonograph or method criterion3
SensitivityBlank and reporting-threshold chromatograms, noise window, H, h and S/NImpurity tests with a reporting threshold5
DriftBracketing standard responses against their criterionData from a suspect period are not used12
OutcomePass or fail for each criterion; for a fail, the investigation reference and confirmation that no sample result was reportedNo sample analysis without demonstrated suitability2
Data and reviewSequence audit trail, integration events, every injection including failed or aborted ones; reviewer and dateFailing and suspect data stay in the record13

5How it runs in Seal.

In Seal, an approved method becomes a controlled execution pattern: preparations, standards, system suitability, sequence rules, calculations and signatures in the order the work requires, and failed suitability follows the defined response. Configured checks can confirm analyst qualification and instrument state before work begins, standards keep their lot, potency and expiry, and the repeatability and maximum-⁠RSD calculations can be configured in the method with explicit inputs, units and reporting rules, then verified at boundary values and rounding whenever they change. Pharmaceutical QC shows the pattern, and Neil can bring samples, standards, method versions and instrument status together before a planned sequence, while the analyst still confirms suitability at execution.

Seal Edge uploads supported exports from the instrument workstation to the record they support, and the original file stays attached to the sample, batch or investigation. When a result goes to investigation, Seal reconciles planned against actual injections, so exploratory injections stay visible and cannot quietly become reportable. Across runs, analytical method lifecycle trends system suitability, invalid runs, OOS and OOT results by analyst, instrument, column and reagent to detect drift in the measurement system.

6Example: system suitability for an Exampladine assay.

Exampladine hydrochloride is an invented drug substance. Its specification sets assay limits of 98.0% to 101.5% and states no repeatability limit, so the chapter formula applies. The method fixes six replicate injections of the reference solution, so B = 101.5 − 100 = 1.5 and %RSDmax = 0.349 × 1.5 × √6 / 2.015 = 0.64%. Run 1 gives areas of 1,524,310, 1,529,870, 1,521,960, 1,531,440, 1,526,080 and 1,519,750: mean 1,525,568, SD 4,510, %RSD 0.30%.

The Exampladine peak elutes at 6.84 minutes, 0.105 minutes wide at half-height and 0.236 minutes wide at one-twentieth height, 0.101 minutes of that before the apex. Impurity B, the critical pair, elutes at 6.21 minutes with a half-height width of 0.098 minutes.

Table 5. Run 1 system suitability results.
CriterionLimitCalculationResult
Repeatability0.64% or less, computedSix areas, run 10.30%, passes
Symmetry factor0.8 to 1.8, chapter default0.236 / (2 × 0.101)1.17, passes
Resolution, impurity B and ExampladineNot less than 2.0, method1.18 × (6.84 − 6.21) / (0.098 + 0.105)3.66, passes
Plate numberNot less than 5,000, method5.54 × (6.84 / 0.105)²23,509, passes

In run 2 the sixth injection reads 1,499,800 after an autosampler fault. The six-injection RSD becomes 0.76%, above 0.64%, so the system fails and no sample from the run is reported. Dropping the sixth injection would give 0.26% on five, and five injections even have a limit of their own (0.55%), but repeatability is defined on “a consecutive series of measurements”, and the method set the number of injections before the run.3

The related substances test has a reporting threshold of 0.05%. At that concentration the Exampladine peak is 0.36 mAU high, and the blank’s noise range over 20 half-height widths is 0.058 mAU: S/N = 2 × 0.36 / 0.058 = 12.4, a pass. With 0.081 mAU of noise, S/N falls to 8.9 and the run fails.

7When system suitability fails, and the cases between.

A failed system produces no reportable result: FDA says that “systems not meeting these requirements should not be used”, and that when bracketing standards show drift, “all of the data collected during the suspect time period should be properly identified and should not be used.” The cause is identified and, if possible, corrected “before a decision is made whether to use any data prior to the suspect period.”12 The failed injections stay in the record: FDA expects every result, “including obvious errors and failing, passing, and suspect data”, to be retained and reviewed.13

When a system will not pass, the chapter allows these adjustments to a pharmacopeial liquid chromatography procedure without revalidation, provided suitability then passes; “Changes other than those indicated require revalidation of the procedure.”2

Table 6. Liquid chromatography adjustments allowed without revalidation, unless the monograph says otherwise.2
ConditionIsocraticGradient
Stationary phaseSame substituent (no C18 to C8), similar characteristicsSame
Column length and particle sizeL/dp unchanged or within −25% to +50%Same
Flow rateRescaled for diameter and particle size, then ±50%; ±50% if the column is unchangedRescaled for diameter and particle size; each gradient segment time rescaled too
Mobile phaseMinor components ±30% relative, none beyond ±10% absoluteAllowed if suitability passes, principal peaks elute within ±15% of their retention time (unless the column changed), first peaks are retained and last peaks elute
pH and buffer salts±0.2 pH units; salts ±10%Same
Column temperature±10 °C±5 °C
Detector wavelengthNo changeNo change
Injection volumeScaled with column size, or varied if suitability passesSame
  • Using a sample to check the system. “FDA considers it a violative practice to use an actual sample in test, prep, or equilibration runs as a means of disguising testing into compliance.”13 An API or product may serve as the suitability material only if it is a properly characterized secondary standard from a different batch from the samples.11 See trial injections and working standards.
  • Adjusting until it passes. “Multiple adjustments can have a cumulative effect on the performance of the system”, and a procedure that passes only “with repeated adjustments to the operating conditions” should be re-evaluated, FDA says.2,14
  • Integrating until it passes. A suitability result rescued by reintegration needs the same justification as any other (manual integration).
  • The software measures something else. FDA’s 1994 guidance measured tailing width “at either 5% (0.05) or 10% (0.10)” of peak height.15 The pharmacopeial factor uses 5%, so check which the processing method reports.
  • Area normalization. The Ph. Eur. symmetry default “does not apply to analytical procedures using area normalisation as the quantitation procedure.”4

8What inspectors cite.

  • Taizhou Kangping Medical Science and Technology, warning letter 320-⁠26-⁠01, 9 October 2025. “Your high-performance liquid chromatography (HPLC) analyses for (b)(4) Assay Content lacked an appropriate determination of system suitability, as outlined in USP <621>, before sample injections.” FDA found that the method’s standard-curve approach “has the practical effect of masking system drift and performance issues that affect the quantitative accuracy of batch-specific test results”.16
  • Ava Inc., warning letter 320-⁠26-⁠64, 14 April 2026. An analyst ran a trial injection, repeated it, and the firm reported the more favorable result. FDA wrote: “The injection of trial samples is not acceptable, in part, because all data from analysis of product samples must be retained and reviewed.” Of the response it added: “Furthermore, you do not adequately address the capability of your system suitability to determine the readiness of the equipment for analysis.”17
  • Curia New York, warning letter 320-⁠26-⁠127, 18 September 2026. Analysts made “over 70 trial injections using TotalChrom in October 2025 alone”, and FDA asked for an independent investigation of all trial injections, “with supporting documentation showing that the trial injections were not sample injections.”18

9Older limits, in-house methods and what is changing.

Many methods still carry the values in FDA’s 1994 reviewer guidance: capacity factor above 2, RSD of 1% for at least five injections, resolution above 2, tailing factor of no more than 2 and more than 2,000 plates. “For low level impurities, higher variations may be acceptable.”15 FDA’s 2015 methods guidance still refers readers to it and to <621>.14 Those values guide method writing; the pharmacopeial defaults apply where a monograph is used.

For an in-house method, Ph. Eur. 2.2.46 “is not mandatory”, and relying on it is “subject to approval by the competent authority” in the marketing authorization.4 ICH defines suitability tests, set during development, as checks that “verify that the measurement system and the analytical operations associated with the analytical procedure are fit for the intended purpose”.19 See ICH Q2(R2) and validation, verification and transfer.

The dates have moved several times:

  • USP. The harmonized <621> became official on 1 December 2022, but USP delayed its symmetry and sensitivity subsections by a year, then to 1 May 2025, then to 1 June 2026.20,21,22,23 The revision that narrows them was approved in USP-⁠NF 2026 Issue 3 and became official on 1 June 2026.7
  • Ph. Eur. From the 13th Edition (implementation January 2027), new and revised monographs will state under Assay which suitability criteria from the purity test apply, including its selectivity check.4
  • JP. The 19th edition took effect on 10 April 2026, but MHLW’s English pages end at the 18th edition, so this page cites JP XVIII Supplement I.24,25

AQuestions and answers

Does the maximum permitted RSD formula apply to tablets and other drug products?

Not directly. The harmonized formula covers an assay of an active substance or excipient whose target is 100% and whose monograph states no repeatability limit. Drug product monographs and approved methods normally state their own limit. When setting one, the Japanese Pharmacopoeia advises taking account of the width of the product’s content specification and of the limit used for the drug substance assay.

Is the HPLC tailing factor limit 2.0 or 1.8?

Both appear. A tailing factor of no more than 2 comes from FDA’s 1994 reviewer guidance and survives in many older methods. The harmonized pharmacopeial default is a symmetry factor of 0.8 to 1.8 for the peak used for quantitation, unless the monograph says otherwise. In USP it has applied since 1 June 2026 to organic impurities, related substances and assays, and is waived in the two impurity tests when the monograph prescribes an RSD of peak response. An approved method’s own limit governs that method.

How many replicate injections does system suitability need?

As many as the method states. In USP, when a monograph sets an RSD limit of 2.0% or less, five injections are used, and six when the limit is above 2.0%. The harmonized formula works for three to six injections, with a tighter limit for fewer. The Japanese Pharmacopoeia normally expects six and lists equivalent, tighter limits for three to five.

Can a sample from the batch be used as the system suitability solution?

No. FDA expects suitability to be checked with qualified primary or secondary reference standards. An API or finished product may serve only if it has been qualified as a reference standard, is used under a written procedure, and comes from a different batch from the samples being tested.

Does a failed system suitability test need an investigation?

It needs a documented cause before any sample result from the run is used. Samples are not reported from a failed system, the failed injections stay in the record, and where bracketing standards show drift, data from the suspect period are set aside. FDA says the cause should be “identified and, if possible, corrected” before deciding whether earlier data can be used. A method that passes only after repeated adjustments needs re-evaluation.

References25 sources

References

  1. 1Japanese Pharmacopoeia XVIII, Supplement I, General Tests 2.01 Liquid Chromatography, section 6 System suitability, MHLW Ministerial Notification No. 355, 12 December 2022. https://www.mhlw.go.jp/content/11120000/001372370.pdf (read 25 November 2026)
  2. 2USP General Chapter <621> Chromatography, Stage 4 harmonisation text, sections Definitions, System suitability (System repeatability, Table 2, System sensitivity, Peak symmetry) and Adjustment of chromatographic conditions, United States Pharmacopeia, official 1 December 2022, archived copy. https://web.archive.org/web/20261012151035/https://www.usp.org/sites/default/files/usp/document/harmonization/gen-chapter/harmonization-november-2021-m99380.pdf (read 25 November 2026)
  3. 3Japanese Pharmacopoeia XVIII, Supplement I, General Tests 2.00 Chromatography, sections 2 Definitions and 3 System suitability, Table 2.00-⁠1, MHLW Ministerial Notification No. 355, 12 December 2022. https://www.mhlw.go.jp/content/11120000/001372370.pdf (read 25 November 2026)
  4. 4EDQM FAQs, General chapter 2.2.46 (11.0): system sensitivity, peak symmetry and implementation for existing monographs and in-house procedures; Chromatography: impurities and assay, EDQM, Council of Europe, updated 2021 to 2024. https://faq.edqm.eu/pages/viewpage.action?pageId=48201789 (read 25 November 2026)
  5. 5Commentary, USP-⁠NF 2026 Issue 3, General Chapter <621> Chromatography, comment summaries 1 to 12, United States Pharmacopeia, 21 November 2025, archived copy. https://web.archive.org/web/20251204230107/https://www.uspnf.com/sites/default/files/usp_pdf/EN/USPNF/usp-nf-commentary/usp-nf-2026-issue-3-commentary-20251121.pdf (read 25 November 2026)
  6. 6Japanese Pharmacopoeia XVIII, Supplement I, General Information, System Suitability <G1-⁠2-⁠181>, section 2.1, MHLW Ministerial Notification No. 355, 12 December 2022. https://www.mhlw.go.jp/content/11120000/001372370.pdf (read 25 November 2026)
  7. 7USP-⁠NF 2026 Issue 3, compendial approvals, entry for <621> Chromatography, United States Pharmacopeia, posted 21 November 2025, archived copy. https://web.archive.org/web/20251204233005/https://www.uspnf.com/sites/default/files/usp_pdf/EN/USPNF/usp-nf-commentary/usp-nf-2026-issue-3-approvals.pdf (read 25 November 2026)
  8. 8Implementation of PDG harmonized texts into Japanese Pharmacopoeia (General Chapters), entry G-⁠20 Chromatography, Pharmaceuticals and Medical Devices Agency. https://www.pmda.go.jp/english/rs-sb-std/standards-development/jp/0027.html (read 25 November 2026)
  9. 9Briefing on drafts of General Tests and General Information related to Chromatography, Pharmaceuticals and Medical Devices Agency, September 2021. https://www.pmda.go.jp/files/000242711.pdf (read 25 November 2026)
  10. 10<621> Chromatography, Notice of Intent to Revise, United States Pharmacopeia, posted 25 September 2024, archived copy. https://web.archive.org/web/20250612193955/https://www.uspnf.com/notices/621-nitr-20240927 (read 25 November 2026)
  11. 11Questions and Answers on Current Good Manufacturing Practice Requirements, Laboratory Controls, question 16, FDA, content current as of 16 November 2022. https://www.fda.gov/drugs/guidances-drugs/questions-and-answers-current-good-manufacturing-practice-requirements-laboratory-controls (read 25 November 2026)
  12. 12Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production, Guidance for Industry, Revision 1, section III.A, FDA, May 2022. https://www.fda.gov/media/158416/download (read 25 November 2026)
  13. 13Data Integrity and Compliance With Drug CGMP, Questions and Answers, Guidance for Industry, question 13, FDA, December 2018. https://www.fda.gov/media/119267/download (read 25 November 2026)
  14. 14Analytical Procedures and Methods Validation for Drugs and Biologics, Guidance for Industry, sections IV.H and VIII, FDA, July 2015. https://www.fda.gov/media/87801/download (read 25 November 2026)
  15. 15Reviewer Guidance, Validation of Chromatographic Methods (CMC 3), sections IV.D.1 and IV.J, Center for Drug Evaluation and Research, FDA, November 1994, archived copy. https://web.archive.org/web/20090511235535/http://www.fda.gov/cder/guidance/cmc3.pdf (read 25 November 2026)
  16. 16Warning Letter 320-⁠26-⁠01, Taizhou Kangping Medical Science and Technology Co., Ltd., MARCS-⁠CMS 711081, FDA, 9 October 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/taizhou-kangping-medical-science-and-technology-co-ltd-711081-10092025 (read 25 November 2026)
  17. 17Warning Letter 320-⁠26-⁠64, Ava Inc., MARCS-⁠CMS 721180, FDA, 14 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/ava-inc-721180-04142026 (read 25 November 2026)
  18. 18Warning Letter 320-⁠26-⁠127, Curia New York, Inc., MARCS-⁠CMS 728790, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 25 November 2026)
  19. 19ICH Q2(R2) Validation of Analytical Procedures, introduction and glossary, ICH, 1 November 2023, error correction 2025. https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130_ErrorCorrection_2025.pdf (read 25 November 2026)
  20. 20<621> Chromatography, Notice of Adoption of Harmonized Standard, United States Pharmacopeia, posted 19 November 2021, archived copy. https://web.archive.org/web/20251115184012/https://www.usp.org/harmonization-standards/pdg/excipients/chromatography (read 25 November 2026)
  21. 21<621> Chromatography, Revision Bulletin, United States Pharmacopeia, posted 28 October 2022, archived copy. https://web.archive.org/web/20221204161250/https://www.uspnf.com/sites/default/files/usp_pdf/EN/USPNF/revisions/gc-621-rb-notice-20221028.pdf (read 25 November 2026)
  22. 22<621> Chromatography, Revision Bulletin, United States Pharmacopeia, posted 29 September 2023 and updated 17 November 2023, archived copy. https://web.archive.org/web/20231205013510/https://www.uspnf.com/sites/default/files/usp_pdf/EN/USPNF/revisions/20230929_621_Chromatography_RB_Notice_20231117_update.pdf (read 25 November 2026)
  23. 23<621> Chromatography, Revision Bulletin, United States Pharmacopeia, posted 22 November 2024, archived copy. https://web.archive.org/web/20250219085037/https://www.uspnf.com/sites/default/files/usp_pdf/EN/USPNF/revisions/gc-621-rb-notice-20241122.pdf (read 25 November 2026)