Skip to main content

How do you choose an electronic lab notebook (ELN) for life sciences?

Choose an ELN with your own experiments, because every product looks good on its own. List the kinds of work it will hold and the rule each comes under, then run one real experiment through each: type a wrong value, sign it, correct it, find what used it, abandon an entry and export the lot. Then run a four-week pilot that measures whether one scientist can repeat another’s work from the record alone, and settle the exit before you sign.Checked against its sources on 24 November 2026. How these answers are made

Illustration of a seal reading a laboratory notebook ruled as a table, with a red thread from one entry to a beaker.

1Decide what the notebook has to prove.

Most labs do three or four of these at once, and each asks something different of the same page. What an ELN is explains which rules reach which work; this is what to look for because of it.

Table 1. Five uses of one notebook.
WorkWhat the record is forBasisWhat to look for
Exploratory researchAnother scientist repeating it without asking you. NIH sets that test for its own labs: a scientifically literate person with no knowledge of the project can follow it “entirely from the ELN”NIH policy; good practice elsewhere1Quick free entries, linked files, search
Work that may support a patentDates and authorship that stand for decades; NIH keeps such notebooks for 30 years after filingNIH policy, as an example of a term1Export that outlives the product
GLP studyReconstructing the study report. Changes keep the first entry, with reason, date and person21 CFR 58.130(e), 58.190, 58.1952,3,4Study archive, index, one person responsible for it
Phase 1 materialEnough to replicate the process; exempt from part 211, still under the statute210.2(c); FDA phase 1 guidance5,6Process records that keep why a step changed
GMP tests and validationComplete data from every test, reviewed against the original by a second person211.194(a)7Controlled templates, review of the original, audit trail

Count your entries for a month by those rows before you see a product. A notebook bought for the last row alone makes research slow enough that people go back to paper. One bought for the first row alone has to be replaced when the first study starts. The thing to look for is one notebook that can hold free pages in one project and locked templates in another, with the same audit trail under both.

The steps, the team and the records of a selection are the same as for a LIMS and are in how to choose a LIMS. LIMS or ELN settles which of the two a piece of work belongs in.

2An ELN requirements checklist to copy.

Table 2. ELN requirements checklist. Basis numbers are 21 CFR sections, FDA’s data integrity Q&A, MHRA’s guidance and Annex 11.2,3,8,9,10,11,7,12,13,14
RequirementThe system mustBasisTest it by
Who and whenTake the person from the login and the time from a server58.130(e); 11.10(e)Setting the laptop clock back a day
CorrectionsKeep the first value readable, with reason, person and time58.130(e); Annex 11, 9Step 5 of the script
SignaturesShow name, time and meaning on screen and in the export, tied to what was signed11.50; 11.70Changing a value after signing
TemplatesIssue approved versions; leave started entries on their versionFDA Q&A 6; 211.160(a)Changing one halfway through a run
Free pagesLet a scientist record what was not planned, at once211.160(a): departures recorded and justifiedStep 8, with a stopwatch
CalculationsShow formula and units, recalculate, keep both results211.194(a)(5)Step 5
Original filesKeep the instrument’s file in the form that can still be reprocessed, linked to the entryFDA Q&A 10; MHRA 6.7Opening it from the entry, in its own software
MaterialsLink each lot of reagent, standard and cell bank, with its expiry211.194(c)Choosing an expired lot
Nothing lostKeep abandoned and failed entries in the numbering211.194(a): complete data; MHRA 6.7 on gapsAbandoning one, then listing all numbers
ReviewShow a second person the original data and the changes, and stop an author reviewing their own211.194(a)(8)Trying to review your own
AccessSeparate projects, guests and administrators; record changes of access11.10(d) and (g); Annex 11, 12.3Logging in as a contract lab
RetrievalFind by sample, lot, material, instrument and text11.10(c); 58.190(e)Step 9
CopiesGive complete copies, readable and electronic11.10(b)Step 10
MigrationBring old entries across with nothing altered in value or meaningAnnex 11, 4.8Ten old entries, compared by eye

Mark a line as a must only if some of your own work needs it. A research group with no GLP or GMP work has four musts: corrections, original files, nothing lost and copies.

3Questions to put to any vendor, and what a good answer contains.

Table 3. Ten questions.
AskA good answer containsBe wary of
What can a user delete, and what can an administrator?Nothing signed; drafts leave a numbered trace; the administrator’s actions are in a trail too“Only admins can delete”
Who are the administrators at your other customers?Someone with no interest in the data, as MHRA expects13The lab head
What happens to a signature when something under it changes?It stays with the version it was given to and the new version is unsignedIt moves to the new version
A value was wrong. How do I find what used it?A list of entries and reports that read the value, from the entry itselfText search for the number
Where does our file live, and what checks that it has not changed?The place, the check, and what you see when it failsA copy of the chart as a picture
Who writes templates?Your scientists, with your approval route, in hoursA day rate
How does a colleague with gloves on use it?Scanner, scale and photograph at the bench, shown on your benchA laptop on a desk
What was your longest outage last year, and what did labs do?A date, a duration, and how late entries were markedA percentage
Is our content used to train models?No, or a switch, written in the contractA link to a policy that can change
How do we leave, and what will we hold on that day?Entries, files, signatures and audit trail in open formats, rehearsed in the pilotA PDF for each entry

NIH tells its scientists that data in the notebooks it supplies “are under the full control of NIH and will not be deleted by the vendor”.15 Ask for a sentence like it in your contract. For EU GMP work the 2025 draft of Annex 11 would make it an expectation: an exit strategy “by which the regulated user may retain control of system data”, and a chance to test new versions before release.16

4A demonstration script on your own experiment.

Choose an experiment in which someone once made a mistake that mattered. This one is invented; replace every value.

The case. Wenmoor grows three clones and takes forward any with a growth rate of at least 0.050 per hour. For clone 2 the counter’s file says 2.1 million cells per mL at 72 hours and 8.4 million at 96 hours. The scientist types 4.8.

Table 4. Example: one transposed pair of digits, two decisions.
ValueAs typedFrom the counter’s file
Count at 72 h, million per mL2.12.1
Count at 96 h, million per mL4.88.4
Growth rate, ln(second ÷ first) ÷ 240.0344 per hour0.0578 per hour
Doubling time, ln 2 ÷ rate20.1 h12.0 h
Decision at 0.050DroppedTaken forward
  1. Make the template from your own method, with units and the formula above. Let one of your scientists do it, and time it.
  2. Enter the counts and attach the counter’s file. If the notebook can read the file, see whether typing is still possible, and whether it says the two disagree.
  3. Sign as author, then as reviewer. Watch what the reviewer is shown. If the file is not in front of them, the error passes.
  4. Use the result elsewhere. In a second entry, rank the three clones by reading the rate from the first.
  5. Correct 4.8 to 8.4. You should see both values, the reason, the new rate, both rates kept, and two signatures that now belong to the earlier version.
  6. Look at the ranking. It should say that something under it changed. If it shows 0.0344 with no mark, the dropped clone stays dropped.
  7. Change the template to add viability while a run is open. The open run keeps the old fields.
  8. Record something unplanned: a flask fell over. Photograph, sentence, time, at the bench, in under a minute.
  9. Find every entry that used medium lot M-⁠0417. You know the answer from the paper. Here it is six.
  10. Export the project and open it without the product. It should contain the numbers in the next table.
  11. Log in as a contract lab and look for a project you were not given.
Table 5. Example: what step 10 has to give back.
In the exportCountWhy
Entry numbers1514 finished and one abandoned
Signatures30Author and reviewer on 14, and both again on the corrected one
Instrument files22Opened in the instrument’s own software, not as pictures
Versions of the template29 finished entries on the first, 5 on the second

5Test it for four weeks before you sign.

A script shows what the product can do. A pilot shows what people do with it. Take one team, all its new work, and no paper beside it. Measure the same things on the last 90 paper entries first.

Table 6. Example, invented: six scientists, four weeks.
MeasurePaper, 90 entriesPilot, 84 entries
Signed and reviewed within five working days38 (42%)71 (85%)
Original file found from the entry52 (58%)80 (95%)
Of 20 picked at random, repeatable by a colleague without asking11 (55%)17 (85%)
Find every entry that used one lot3 h4 min
Median time to write one up35 min41 min

The last line got worse, and it is the line the scientists will quote. Six more minutes on 84 entries is 504 minutes, or 8.4 hours between six people in four weeks. The team found the cause: three fields typed twice, once at the scale and once in the template. It asked for the scale to fill them, and agreed to buy when a repeat week showed 36 minutes. The third line is the one to defend. It is the reason to have a notebook at all.

6Example: scoring three suppliers.

Each criterion scores 0 to 3 and is multiplied by its weight, so the most is 300. All three met every must.

Table 7. Example, invented: a tie that the points cannot break.
CriterionWeightSupplier ASupplier BSupplier C
Writing at the bench253 (75)2 (50)1 (25)
Corrections, signatures, review202 (40)3 (60)3 (60)
Templates and calculations152 (30)3 (45)3 (45)
Files and instruments152 (30)2 (30)3 (45)
Search and reuse103 (30)2 (20)2 (20)
Export and exit102 (20)3 (30)2 (20)
Price over five years53 (15)1 (5)2 (10)
Points100240240225

A is better where scientists write and B where reviewers read. So Wenmoor went back to its count: 320 entries a month, 70 of them GLP or GMP, which is 22%. It chose A, on condition that its five regulated methods became locked templates and that the correction script passed on each. A lab with the proportions the other way round should read the same table and choose B. Write that reason down; in three years it will be the only part anyone needs.

7The validation and data integrity work it brings.

Table 8. Three depths of work for one product. Plan and price the column you need.14,8,12,17,18,4,3,19
WorkResearch onlyGLP studyGMP tests
Requirements and supplier assessmentShort, keptFullFull; Annex 11, 3 and 4.4
ValidationTry the musts, keep the note11.10(a), for the records you rely onEach workflow you use, FDA Q&A 3
TemplatesOwner and versionApproved with the protocolControlled by quality, FDA Q&A 6
CalculationsChecked once with known valuesChecked and lockedChecked, locked and changed under control
ReviewWitness, if your patent advice asksStudy director answers for data being recorded and verified, 58.33(b); quality assurance checks the report against it, 58.35(b)(6)Second person, original records, audit trail
RestoreOnceBefore use, then periodicallyBefore use, then periodically; Annex 11, 7.2
Keep forYour policyThe term in 58.195(b), in an indexed archiveAt least a year past the batch’s expiry, 211.180(a)

If one product serves two columns, show that the lighter one cannot reach the records of the heavier. FDA says the risk from research functions on a shared system should be assessed and mitigated.12 Spreadsheet validation covers the calculations you bring with you, and GAMP 5 software categories how much to test.

8What an inspector will look for.

Pages from the associated laboratory notebook 830-⁠7 contained no reference to the four trial injections.

Curia New York, warning letter 320-⁠26-⁠127, 18 September 2026. The injections were on an HPLC, under a role called “Developer”.20 The notebook did not have to be wrong to be cited; it only had to be silent. Add it to the script: ask for every injection in a week that no entry mentions. If nobody can answer, the link between the notebook and the instrument’s software is the requirement you are missing.

9Before you sign.

  • The account belongs to the organisation, and leavers’ entries stay with it.
  • The export from step 10 is in the contract, with its formats, a price of nothing, and how long after the end you can take it.
  • For GLP work, who holds the archive. If a testing facility closes, 58.195(h) has its records go to the sponsor’s archives, so the sponsor needs a copy it can read.4
  • Five years of price: scientists, readers, guests, instruments, storage, test environments, validation documents.
  • How releases are announced, where you try them, and how long you may wait.
  • Two customers your size to ring. Ask what is still on paper, and whether they have ever exported everything.

10Scorecard template.

Table 9. Scorecard: Supplier ____ Script run by ____ Date ____
CriterionWeightScore 0 to 3What we sawPoints
Writing at the bench, with seconds timed25
Corrections, signatures, review (must)20
Templates and calculations15
Files and instruments (must)15
Search and reuse10
Export and exit (must)10
Price over five years5

11How it runs in Seal.

Seal ELN runs structured, versioned protocols with linked samples, reagents, calculations and instrument files on the same platform as process development, LIMS and GMP batch records. A later revision of a protocol doesn’t rewrite earlier runs.

The electronic lab notebook comparison lists what to check, Seal included.

AQuestions and answers

Do early-stage R&D labs need a 21 CFR Part 11 ready ELN?

Part 11 applies to electronic records that FDA’s rules require and to electronic records submitted to FDA, so most discovery notes are outside it. The parts worth having from the first day are the ones that cannot be added later: a correction that keeps the first value, a server’s time, and entries nobody can delete.

How long should an ELN trial run?

Long enough for real work to go wrong, which is about four weeks. Put all of one team’s new work in it with no paper beside it, include a correction, an abandoned entry and a departure, and end with an export that someone opens without the product.

Can an ELN replace Excel spreadsheets in analytical labs?

For fixed calculations, yes, if the formula is versioned, locked and checked against known answers, and each result shows which version made it. The checking does not go away with the spreadsheet; it moves to the template, where it has to be done once for each version.

How long should ELN records be retained?

By the work. US GMP records tied to a batch are kept at least a year past its expiry. GLP records follow the terms in 21 CFR 58.195, which run from two to five years from different events. Research follows your policy and patent advice. Where an entry serves two, keep it for the longer.

How do you move old notebooks into a new ELN?

Move what people still use. Leave finished paper notebooks as they are, scan and index the ones in use, and migrate electronic entries with their files, signatures and history where the product can hold them. Compare a sample by eye for value and meaning, and keep the old system readable until that is signed.

References20 sources

References

  1. 1Intramural Electronic Lab Notebook Policy, NIH Office of Intramural Research. https://oir.nih.gov/sourcebook/intramural-program-oversight/electronic-lab-notebooks/intramural-electronic-lab-notebook-policy (read 24 November 2026)
  2. 221 CFR 58.130, Conduct of a nonclinical laboratory study, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58/subpart-G/section-58.130 (read 24 November 2026)
  3. 321 CFR 58.190, Storage and retrieval of records and data, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58/subpart-J/section-58.190 (read 24 November 2026)
  4. 421 CFR 58.195, Retention of records, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58/subpart-J/section-58.195 (read 24 November 2026)
  5. 521 CFR 210.2, Applicability of current good manufacturing practice regulations, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-210/section-210.2 (read 24 November 2026)
  6. 6CGMP for Phase 1 Investigational Drugs, Guidance for Industry, FDA, July 2008, section V.E. https://www.fda.gov/media/70975/download (read 24 November 2026)
  7. 721 CFR 211.194, Laboratory records, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-J/section-211.194 (read 24 November 2026)
  8. 821 CFR 11.10, Controls for closed systems, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-11/subpart-B/section-11.10 (read 24 November 2026)
  9. 921 CFR 11.50, Signature manifestations, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-11/subpart-B/section-11.50 (read 24 November 2026)
  10. 1021 CFR 11.70, Signature/record linking, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-11/subpart-B/section-11.70 (read 24 November 2026)
  11. 1121 CFR 211.160, General requirements (laboratory controls), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-I/section-211.160 (read 24 November 2026)
  12. 12Data Integrity and Compliance With Drug CGMP: Questions and Answers, Guidance for Industry, FDA, December 2018, questions 3, 6 and 10. https://www.fda.gov/media/119267/download (read 24 November 2026)
  13. 13‘GXP’ Data Integrity Guidance and Definitions, revision 1, MHRA, March 2018, sections 6.7 and 6.16. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 24 November 2026)
  14. 14EudraLex Volume 4, Annex 11: Computerised Systems, revision 1, European Commission, in operation from 30 June 2011. https://health.ec.europa.eu/system/files/2016-11/annex11_01-2011_en_0.pdf (read 24 November 2026)
  15. 15Frequently asked questions about the use of ELNs at NIH, NIH Office of Intramural Research. https://oir.nih.gov/sourcebook/intramural-program-oversight/electronic-lab-notebooks/frequently-asked-questions-about-use-elns-nih (read 24 November 2026)
  16. 16Annex 11: Computerised Systems, draft for consultation, European Commission and PIC/S, July 2025, section 7.5. https://health.ec.europa.eu/document/download/40231f18-e564-4043-94de-c031f813d38b_en?filename=mp_vol4_chap4_annex11_consultation_guideline_en.pdf (read 24 November 2026)
  17. 1721 CFR 58.33, Study director, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58/subpart-B/section-58.33 (read 24 November 2026)
  18. 1821 CFR 58.35, Quality assurance unit, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58/subpart-B/section-58.35 (read 24 November 2026)
  19. 1921 CFR 211.180, General requirements (records and reports), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-J/section-211.180 (read 24 November 2026)
  20. 20Warning letter 320-⁠26-⁠127 to Curia New York, Inc., MARCS-⁠CMS 728790, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 24 November 2026)

See your process in Seal.

Bring a procedure or a recurring problem. See how your team configures it with Neil and improves the next run.