1What changed on 2 February 2026.
The Quality System (QS) regulation required device records to be “readily available for review and copying by FDA employee(s)”, but 820.180(c) said this did not apply to the reports of management reviews under 820.20(c), quality audits under 820.22 or supplier audits used to meet 820.50(a). Instead, on request, “an employee in management with executive responsibility shall certify in writing” that the reviews and audits had been performed and documented, on what dates, and that any required corrective action had been undertaken.1
The Quality Management System Regulation (QMSR) replaced the QS regulation on 2 February 2026. Part 820 now requires a quality management system that complies with ISO 13485:2016, plus FDA’s additions, and sections 820.20 to 820.30 are reserved.2,
| Record | QS regulation, to 1 February 2026 | QMSR, from 2 February 2026 |
|---|---|---|
| Management review reports | Exempt; the procedure was not | Open to inspection, as Clauses 5.6.1 to 5.6.3 |
| Internal audit reports | Exempt; the procedure was not | Open to inspection, as Clause 8.2.4 |
| Supplier audit reports | Exempt where used to evaluate suppliers | Open to inspection, under Clause 7.4.1 |
| Instead of the reports | Written certification by executive management | No equivalent |
| Inspection method | QSIT | Compliance Program 7382.850 |
Comments on the proposed rule asked FDA to keep the exception, arguing that it kept reports complete and useful. FDA refused. Manufacturers inspected by other regulators or audited by MDSAP auditing organisations already have no such exception, so making the records available adds no burden for them; the records “are maintained in the regular course of business and should be readily available”; and investigators had already seen what feeds them, such as nonconformances, complaints and the corrective actions audits produce.2,
2When investigators will read them.
Under Compliance Program 7382.850, each inspection type follows one of two models. Model 1 has the investigator pick at least one element from each of six QMS areas, guided by product risk. Model 2 sets minimum elements: Management Review and Internal Audits are among them, but in the Outsourcing and Purchasing area the only minimum is Outsourcing, Clause 4.1.5. Supplier evaluation sits in the Purchasing Process element, Clause 7.4.1.5 At its April 2026 town hall FDA said that for baseline surveillance and PMA preapproval inspections, management review, internal audit and supplier audit processes “are expected to be reviewed”.6
| Inspection type | Model | Management review, internal audits | Supplier audits |
|---|---|---|---|
| Baseline surveillance | 2 | Minimum elements | Outsourcing (4.1.5) minimum; Purchasing (7.4.1) if selected |
| PMA preapproval | 2 | Minimum elements | Outsourcing (4.1.5) minimum; Purchasing (7.4.1) if selected |
| Non-baseline surveillance | 1 | If the investigator selects them | If selected |
| Compliance follow-up | 1 | If selected | If selected |
| For-cause | 1 | If selected, often from the signal | If selected |
| Specific product risk | 1 | If selected | If selected |
| PMA postmarket | 1 | If selected | If selected |
FDA gave an example for the other types: if an inspection reveals design deficiencies that led to a significant recall, investigators may assess how the manufacturer has audited its design and development process.6 Sites actively enrolled in MDSAP receive no surveillance inspections, but remain subject to all non-surveillance inspections, including for-cause, compliance follow-up and specific product risk inspections.5
Reports written before the QMSR are included. FDA does not expect them to be revised or recreated, but a comparative analysis may help show earlier records meet the QMSR.6,
But they are still fair game for us to look at if the inspection direction leads us there.
FDA town hall on QMSR risk-based inspections, 1 April 2026.6
3How FDA has enforced it since February 2026.
By 7 November 2026, FDA’s warning letter index held seven letters that cite ISO 13485:2016 clauses after inspections starting on or after 2 February 2026: to Linemaster, Koven, Nipro, Pro-Optics, O.N. Diagnostics, Pensar Medical and Unilam. None cites Clause 5.6 or Clause 8.2.4. Four cite supplier evaluation under Clause 7.4.1, the work a supplier audit report records.7 Koven and Nipro lacked supplier evaluations their own procedures required; O.N. Diagnostics and Unilam had inadequate purchasing procedures. None faults what an audit or management review report said.7
Koven had 25 510(k) clearances for devices designed and made by a contract manufacturer:
There is no documentation that a site assessment (supplier audit) has been conducted for the contract manufacturer, a (b)(4) supplier which your firm’s procedure defines as “Highest/Critical Impact.”
Koven Technologies, Inc., warning letter CMS 734643, 21 July 2026.8
Nipro had a supplier management procedure, but:
your firm was unable to provide any records demonstrating that suppliers and services had ever been classified or evaluated in accordance with this procedure.
Nipro Renal Solutions USA Corporation, warning letter CMS 732874, 16 July 2026.9
Procedures were never exempt: a letter of 2 October 2026 to Chirp Innovations, inspected in November 2025, cited missing management review and quality audit documentation under 820.20 and 820.22.10
4What you may withhold, redact or mark.
With certification gone, the questions are about scope, and FDA’s guidance on limiting inspections answers most of them.
| Situation | What the sources say |
|---|---|
| Offer a certification instead | No longer provided for: 820.180(c) was not kept4 |
| Show some requested reports | Providing some, but not all, records FDA may inspect may be considered limiting an inspection11 |
| Redact budgets, sales or salaries | Generally reasonable: FDA’s inspection authority excludes financial, pricing, most sales and most personnel data11 |
| Redact findings or risks | Unreasonable redaction of information FDA may inspect may be considered limiting11 |
| Mark a report confidential | Allowed, to help FDA decide disclosure under part 20; it concerns disclosure, not access3, |
| Reports kept off-site | Reasonable extra time to collect them is a potentially reasonable explanation; other delay may be considered delaying11 |
| Export with fields removed | Omitting data from electronic records may be considered limiting11 |
The stakes are statutory. A device made at an establishment whose owner, operator or agent delays, denies or limits an inspection is deemed adulterated under section 501(j) of the FD&C Act.11
4.1Example: one request, logged.
Invented for illustration: how Halden Medical’s front room logs one request.
| Field | Entry | What it satisfies |
|---|---|---|
| Request | REQ-07, 11 March 2026, 10:12: “Supplier audit reports for your sterilisation contractor since 2023” | The investigator’s words, kept verbatim |
| Records in scope | SA-2023-04 (14 pages) and SA-2025-02 (17 pages), Kestrel Sterilisation | Both produced: “some, but not all” may be limiting11 |
| Redaction | SA-2025-02, page 6: unit prices only; reason recorded as pricing data under 704(a)(1) | Footnote 13: obscures only what FDA can’t inspect11 |
| Findings | Left unredacted, with links to CAPA-0311 | Redacting what FDA may inspect is unreasonable11 |
| Marking | Both stamped confidential before copying | 820.35(d)3 |
| Delay | SA-2023-04 fetched from the archive in 40 minutes; reason given | Reasonable extra time for off-site records11 |
| Delivery | 31 pages handed over at 11:05; list signed by the host | Shows later exactly what FDA took |
5A readiness checklist for these records.
Copy it into your inspection-
| Area | Check | Source |
|---|---|---|
| Retrieval | Every management review, internal audit and supplier audit report in the retention period can be produced on request, including pre-QMSR ones | FAQ questions 7 and 84 |
| Old records | Pre-QMSR reports have been compared with the QMSR, and gaps they reveal have a plan | FAQ question 7;4 town hall6 |
| Content | Management review records show the Clause 5.6.2 inputs and 5.6.3 outputs; audit records meet Clause 8.2.4 | Attachment A5 |
| Oversight | Reviews show top management using risk management and risk-based decisions, and providing resources | Management Oversight purpose5 |
| Audit scope | The audit programme covers the QMSR, not only ISO 13485: 820.10, 820.35, 820.45 and the FDA requirements they link | Town hall;6 820.10(b)3 |
| Audit method | Audits follow your own procedure for planning, conducting, recording and reporting | Town hall on Clause 8.2.46 |
| Risk | Audits test whether risk controls trace from patient risk through each affected process | Town hall6 |
| Follow-through | Each finding traces to the corrective action investigators could already see, and the two agree | Preamble, comment 552 |
| Suppliers | A supplier evaluation record exists for every supplier your procedure requires, at the depth it requires | Clause 7.4.1 letters8, |
| Redaction rule | Written in advance: only data outside FDA’s inspection authority | FDA guidance, footnote 1311 |
| Marking | Confidential reports are marked before they are shown | 820.35(d)3 |
| Requests | Someone owns requests made before an inspection and answers within FDA’s timeframe | FDA guidance on 704(a)(4)11 |
6Combination products and drug makers.
Drug-led combination products. A site that shows compliance with drug CGMP and uses the streamlined route in 21 CFR 4.4(b)(1) must also satisfy, among others, ISO 13485 Clause 4.1, Clause 5 and its subclauses, Clause 6.1, 820.10 and Clause 7.4, purchasing. Management review is in Clause 5; internal audits, Clause 8.2.4, are not on that list.12 Such a site keeps a QMSR management review, whose report has no exception.
Drug manufacturers. The QMSR governs finished devices.3 Parts 210 and 211 “do not specifically address the requirement to conduct, or to keep records of, internal quality assurance audits”, but records of corrective action and of any investigation into affected batches are open to inspection, as are complaint investigations, vendor qualification and failure investigations.13 FDA’s Compliance Policy Guide 130.300 still says that in routine inspections of any entity with a written quality assurance programme, FDA “will not review or copy reports and records that result from audits and inspections of the written quality assurance program”, and may seek a written certification instead. It still names 820.22, now reserved; its page notes that it predates the QMSR rule and is not binding.14,
7What is still draft.
FDA’s October 2025 draft guidance on QMS information in PMA and HDE submissions recommends including “a summary of the management review inputs and outputs for the subject device”, and the internal audit procedure. It is not final.15,
8How it runs in Seal.
Now that FDA can read these reports, it helps if each one is built from the records it summarises. In Seal’s audit management, supplier audits run from the same programme as internal audits, an evidence package references each source record at the version selected for it, and a finding links its assessment, proposed change and verification, each with its own owner and state. Management review reads audit and inspection results from those records rather than from a summary assembled for the meeting.
During an inspection, inspection readiness keeps each request’s verbatim wording, owner, evidence, redactions, response and delivery. A delivered package retains its source record versions, reviewers, approvers and recipient, so you can show later exactly what FDA was given. Your team decides what is disclosed.
AQuestions and answers
Does FDA still accept a management certification instead of the reports?
Not for devices. The written certification came from 21 CFR 820.180(c), which the QMSR does not keep, and FDA says it may now inspect the reports themselves. A certification option still appears in FDA’s separate policy for quality assurance audits, Compliance Policy Guide 130.300, which applies to drug makers and other regulated entities; its page notes that it predates the QMSR.
Can FDA ask for management review reports before an inspection starts?
Yes. Since the Food and Drug Omnibus Reform Act of 2022, section 704(a)(4) of the FD&C Act lets FDA require device establishments to provide records it may inspect in advance of or in lieu of an inspection. FDA’s guidance says failing to provide them in a timely manner without a reasonable explanation may be treated as delaying or limiting the inspection.
Can an investigator take copies of our internal audit reports?
Yes. Section 301(e) of the FD&C Act makes refusing access to, or copying of, certain required records a prohibited act, as FDA’s inspection guidance notes, and the same guidance says not allowing access to or copying of records FDA is entitled to inspect may be considered limiting an inspection. Compliance Program 7382.850 tells investigators to collect the records that support an observation. Agree beforehand who logs what was copied, so you can show later exactly what FDA took.
Can a notified body or MDSAP audit replace our internal audit?
No. FDA has said a notified body audit focused on ISO 13485 conformity may not replace an internal audit of the QMSR, which also includes Part 820’s own provisions, and that third-party audits typically do not follow your internal audit procedure. Their findings can still inform your audit plan. MDSAP enrolment removes FDA surveillance inspections, not your internal audit obligation.
Should management review minutes be written differently now that FDA can read them?
Write them as any quality record: the Clause 5.6.2 inputs actually reviewed, and the decisions and actions Clause 5.6.3 requires, each with an owner and a due date. FDA heard the argument that inspector access would make reports less complete or useful and dropped the exception anyway, and it already sees the nonconformances, complaints and corrective actions they summarise, so minutes that leave problems out will not match the records an investigator opens next.
References16 sources
References
- 121 CFR 820.180, General requirements, eCFR as in force on 1 January 2026. https:/
/ www. ecfr. gov/ on/ 2026- 01- 01/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- M/ section- 820. 180 (read 7 November 2026) - 2Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, comment 55 and the discussion of 820.35, Federal Register, 2 February 2024. https:/
/ www. federalregister. gov/ documents/ 2024/ 02/ 02/ 2024- 01709/ medical- devices- quality- system- regulation- amendments (read 7 November 2026) - 321 CFR Part 820, Quality Management System Regulation, sections 820.1, 820.10 and 820.35, eCFR, current to 5 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820 (read 7 November 2026) - 4Quality Management System Regulation: Frequently Asked Questions, questions 7 to 9, FDA, content current as of 5 November 2026. https:/
/ www. fda. gov/ medical- devices/ quality- management- system- regulation- qmsr/ quality- management- system- regulation- frequently- asked- questions (read 7 November 2026) - 5Compliance Program 7382.850, Inspection of Medical Device Manufacturers, Part III and Attachment A, FDA, 2 February 2026. https:/
/ www. fda. gov/ media/ 80195/ download (read 7 November 2026) - 6FDA’s Quality Management System Regulation (QMSR): Medical Device Risk-Based Inspections, town hall transcript, FDA, 1 April 2026. https:/
/ www. fda. gov/ media/ 191960/ download (read 7 November 2026) - 7Warning Letters, searchable index, searched for “ISO 13485:2016”, with each letter opened, FDA. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ compliance- actions- and- activities/ warning- letters (read 7 November 2026) - 8Warning letter to Koven Technologies, Inc., CMS 734643, FDA, 21 July 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ koven- technologies- inc- 734643- 07212026 (read 7 November 2026) - 9Warning letter to Nipro Renal Solutions USA Corporation, CMS 732874, FDA, dated 16 July 2026, indexed 24 July 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ nipro- renal- solutions- usa- corporation- 732874- 07242026 (read 7 November 2026) - 10Warning letter to Chirp Innovations, LLC, CMS 731847, FDA, 2 October 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ chirp- innovations- llc- 731847- 10022026 (read 7 November 2026) - 11Circumstances that Constitute Delaying, Denying, Limiting, or Refusing a Drug or Device Inspection, guidance for industry, revision 1, FDA, June 2024. https:/
/ www. fda. gov/ media/ 86328/ download (read 7 November 2026) - 1221 CFR 4.4, current good manufacturing practice for co-packaged or single-entity combination products, eCFR, current to 5 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 4/ subpart- A/ section- 4. 4 (read 7 November 2026) - 13Questions and Answers on Current Good Manufacturing Practice Requirements: Records and Reports, FDA. https:/
/ www. fda. gov/ drugs/ guidances- drugs/ questions- and- answers- current- good- manufacturing- practice- requirements- records- and- reports (read 7 November 2026) - 14CPG Sec. 130.300, FDA Access to Results of Quality Assurance Program Audits and Inspections, FDA, June 2007, page current as of 3 February 2026. https:/
/ www. fda. gov/ regulatory- information/ search- fda- guidance- documents/ cpg- sec- 130300- fda- access- results- quality- assurance- program- audits- and- inspections (read 7 November 2026) - 15Quality Management System Information for Certain Premarket Submission Reviews, draft guidance, sections on management responsibility and on monitoring and measurement, FDA, October 2025. https:/
/ www. fda. gov/ media/ 189345/ download (read 7 November 2026) - 16Quality Management System Information for Certain Premarket Submission Reviews, guidance page, FDA, draft, docket FDA-2025-D-4051. https:/
/ www. fda. gov/ regulatory- information/ search- fda- guidance- documents/ quality- management- system- information- certain- premarket- submission- reviews (read 7 November 2026)