What is a quality manual, and is one required for pharma and medical devices?

A quality manual is the top-level document describing a company’s quality system: its policy, scope, processes and how they link, responsibilities and documentation structure. EU GMP, PIC/S, WHO and UK GMP ask medicine makers for one “or equivalent documentation”. US drug rules never name one, though FDA’s non-binding ICH Q10 guidance does. Device makers under the QMSR need one: ISO 13485, which it incorporates, requires it, and so do the EU MDR’s quality system annexes.Checked against its sources on 26 November 2026. How these answers are made

1Quality manual template.

GMP texts set contents, not a layout. ICH Q10 names four: the quality policy, the scope, the processes and their links, and management responsibilities.1 In each section, say how the site meets the requirement and point to the procedure that does it.

For devices, ISO 13485 clause 4.2.2 sets the minimum, the four rows marked 4.2.2, and FDA inspects it under management oversight.2 MDSAP says the manual “is to outline the structure of the documentation and to describe the interaction of processes”.3 For US device gaps, see ISO 13485 vs 21 CFR 820.

SectionWhat to writeRequirement it serves
Document controlNumber, version, effective and review dates, approvers’ signatures, and what each version changed and whyEU GMP 4.3, and 4.5: “regularly reviewed and kept up-to-date”4
Standards followedThe regulations and standards the system meets, by marketFDA: “The quality standard that will be followed”5
ScopeSites, products, lifecycle stages and activities coveredICH Q10 1.8(b);1 4.2.2
Exclusions (devices)Each exclusion or non-application, and why. US class II, III and listed class I devices can’t exclude design4.2.2; 21 CFR 820.10(c);6 FDA on ISO 13485 clause 17
Outsourced workWhat contractors do, and how you select, agree with and monitor themFDA: scope “including any outsourcing”;5 ICH Q10 2.71
Policy and objectivesSenior management’s statement, committing to compliance and continual improvement; how objectives are set and measuredICH Q10 1.8(a), 2.2 and 2.3;1 EU GMP 2.48
OrganizationChart showing the heads of production, QC and QA and the qualified person (QP); where job descriptions sitEU GMP 2.2 and 2.38
ResponsibilitiesWhat senior management, the quality unit and each department ownICH Q10 1.8(d) and 2.1;1 21 CFR 211.22(d)9
Process mapEach process, its owner, its procedure and the processes it feeds4.2.2; ICH Q10 1.8(c)1
Document structureThe document levels, how they reference each other, and the quality system procedures or their index4.2.2; MDSAP Task 13
Management reviewWho reviews the system, how often and with what inputsEU GMP 1.6;10 ICH Q10 2.61

2Is a quality manual required?

Products and marketRuleIs a manual required?
Medicines: EU and PIC/SEU GMP and PE 009-⁠18 Part I, 1.7; WHO TRS 986, 1.7; ICH Q10, 1.8A quality manual “or equivalent documentation”10,11,12,1
Medicines, UKHuman Medicines Regulations 2012, regulation C17(3)The same: EU GMP guidelines as they applied at the end of 2020 continue unless MHRA replaces them13,14
Active substancesICH Q7, 2.11No manual named; each manufacturer should “establish, document, and implement an effective system for managing quality”15
Distributors, EUEU GDP guidelines, 1.2“A quality manual or equivalent documentation approach should be established”16
Medicines, US21 CFR Parts 210 and 211Not by name.9,17 FDA’s non-binding ICH Q10 guidance names one;18 its quality systems guidance lists what the documented system should cover, including the template’s Standards followed and Outsourced work rows5
Devices, USQMSR, 21 CFR 820.10Yes, through ISO 13485 clause 4.2.2,6,2 except class I devices whose classification exempts them from Part 820 other than its records and complaint file requirements (820.35), such as non-sterile tongue depressors19
Devices, EUMDR and IVDR, Annex IX 2.2Yes, where the notified body assesses the QMS: “in the form of a quality manual”20,21

“Or equivalent documentation” means a medicine maker needs the description, not a document with that title.10 ISO 9001 certification and ISO/IEC 17025 accreditation are outside this page. The QMSR replaced the device Quality System regulation on 2 February 2026, so a manual that still cites 21 CFR 820.20 or “the QSR” is out of date.7,6 The Commission’s draft revision of EU GMP Chapter 1, consulted on until 3 December 2025, keeps the 1.7 wording.22

2.1US drug makers and pre-approval inspections.

FDA’s closest text is ICH Q10, which it published as non-binding guidance in April 2009: “A Quality Manual or equivalent documentation approach should be established”.18 Its 2006 quality systems guidance says what the documented system should cover.5 The binding rule is 211.22(d): the quality unit’s responsibilities and procedures must be in writing and followed.9 Part 211 has no internal audit or management review section; the guidance maps system review to record reviews such as the annual review in 211.180(e).5

CP 7346.832, FDA’s compliance program for pre-approval inspections (issued 29 June 2026), doesn’t mention a quality manual. Investigators look for “adequate documents describing the roles and responsibilities of the relevant disciplines in the development process”, and QA oversight of development, scale-up and technology transfer.23 See pre-approval inspections.

3Example: a quality manual extract.

Invented for illustration, with every name, number and date: Kilvarra Aseptics, a contract manufacturer in Ireland. It adds a field for client products, because each contract should “describe clearly who undertakes each step of the outsourced activity”.24

Table 3. Example. A quality manual extract, invented for illustration.
FieldEntry
DocumentQM-⁠001, version 6, effective 2 March 2026, review due by 2 March 2027. Approved by the managing director, the head of quality and a qualified person.
ScopeAseptic filling, inspection, labeling and packaging of liquid vials, clinical and commercial, for 14 clients at Building 2, Kilvarra Park. Clients supply the drug substance.
Client productsEach client’s quality agreement sets who approves deviations, changes and batch release for that client’s product. Section 9 describes how those decisions reach the client.
Quality policyWe make every batch as our clients’ authorizations require, meet GMP in each market we supply, and improve the system from what deviations, complaints, audits and trends show us.
Management reviewTwice a year, chaired by the managing director, with the inputs listed in SOP-⁠QA-⁠001.
Table 4. Example. The same manual’s process map, invented for illustration.
ProcessOwnerProcedureFeeds
DeviationQASOP-⁠QA-⁠010CAPA, change control, product quality review
CAPAQASOP-⁠QA-⁠011Change control, management review
Change controlQASOP-⁠QA-⁠012Documents, training, validation
Management reviewManaging directorSOP-⁠QA-⁠001Objectives, resources, quality policy

3.1A sponsor with no plant.

A virtual company’s manual describes oversight, not operations. Also invented: Corrow Therapeutics, a 40-person biotech, outsources drug substance to one contract development and manufacturing organization (CDMO), fill-finish and EU certification to another, and release testing to a laboratory.

Table 5. Example. A virtual sponsor’s manual, invented for illustration.
SectionEntryRequirement
ContractorsAudited before first use, then every two years; agreed measures reviewed quarterly, with improvement actionsEU GMP 7.5 and 7.7;24 ICH Q10 2.7(c)1
Quality agreementsOne per contractor, naming who approves deviations, changes and out-of-specification investigationsEU GMP 7.14 and 7.15;24 FDA quality agreements guidance25
US dispositionCorrow QA approves or rejects each batch after reviewing the contract sites’ recordsThe owner’s quality unit approves or rejects, “including for final release”25
EU releaseThe fill-finish CDMO’s QP certifies each EU batch, with the drug substance and test records the agreements route to themEU GMP Chapter 7 Principle and 7.824

The last row answers this EU GMP sentence: “The Quality Management System of the Contract Giver must clearly state the way that the Qualified Person certifying each batch of product for release exercises his full responsibility.”24 See quality agreements and the qualified person.

4How to write and approve a quality manual.

  1. Map before you write. List each process, its owner, its procedure and what it feeds: the manual identifies their “sequences, linkages and interdependencies”.1
  2. Draft by reference. Have the head of quality draft it and each process owner confirm their row of the process map. Point to procedures rather than copying them, so the manual changes only when the system does.
  3. Approve it. Under EU GMP 4.3, documents containing instructions “should be approved, signed and dated by appropriate and authorised persons”, with a defined effective date.4 Senior management has “the ultimate responsibility” for the system and sets the policy, so its signature belongs beside the head of quality’s.1
  4. Issue and review it. The policy should be “communicated to and understood by personnel at all levels” and “reviewed periodically for continuing effectiveness”.1 WHO expects management review “at least annually” unless otherwise justified: a natural point to confirm the manual still fits.12
  5. Revise it when the system changes. Changes to the policy, objectives, QMS or regulatory requirements, such as a new site, product, market or reorganization, prompt quality system planning, whose outputs can include “the quality manual and associated procedures”.3 Revise through document control.

5Quality manual vs quality plan, site master file and SOPs.

DocumentWritten forWhat it holdsSource
Quality manualThe people who run the system, and auditorsPolicy, scope, processes and links, responsibilitiesICH Q10 1.81
Quality planOne product or projectIts objectives, processes, resources, acceptance criteria and checksISO 13485 7.1, as MDSAP audits it;3 FDA: “The documented result of quality planning”5
Site master fileA regulator planning an inspectionOne site’s GMP activities, as far as possible within 25 to 30 pages plus appendicesEU explanatory notes 1.2 and 1.326
Quality policyEveryone in the companySenior management’s “overall intentions and direction” on qualityICH Q10 2.21
SOPThe people doing a task“directions for performing certain operations”EU GMP Chapter 44

A site master file carries only a “Brief description of the quality management systems run by the company”, so keep it in step with the manual.26 For procedures, see how to write a GMP SOP.

6What inspectors do with a quality manual.

MDSAP auditors are told to “Verify that a quality manual has been documented”, and that the link between nonconforming product and corrective action should be “evident in the quality manual”.3 FDA investigators read the manual against the records. Its August 2026 warning letter to Peter Schiff Enterprises, a device maker inspected in December 2025 under the old Quality System regulation, set three manual sections against what the firm did:27

Table 7. Example from enforcement. The audit and review findings were each “a repeat observation from the 2017, 2013, and 2011 FDA inspections”.27 ISO 13485 clauses as FDA inspects them now.2
Manual saidFDA foundRecord to have ready
Section 9: a Design History File for three devicesNo design control proceduresDesign and development file for each device (ISO 13485 7.3.10)
Section 15: internal audits “periodically to verify compliance with the QMS and FDA requirements”None since 1966; no frequency or criteria setAudit schedule and reports (8.2.4)
Section 16: management reviews at planned intervalsNone since 1966Review records with inputs and actions (5.6)

Do the same with your own manual: name the record behind each promise, such as Kilvarra’s twice-yearly review.

Say how you meet each rule, not just its number. A contract laboratory, whose manual cited the nonclinical laboratory rules:28

“while your quality manual … and SOP PRO42-⁠1 … referenced 21 CFR 58.35, your procedures did not detail how your operations would meet 21 CFR parts 210 and 211 requirements.” Analytical Food Laboratories, warning letter, 10 October 202429

Fix the quality unit, not just the document. Safrel released OTC drugs made by others:

“In your response, you state that a QU was formally established, and that a quality manual and several SOPs have been created or revised. Your response is inadequate.” Safrel Pharmaceuticals, warning letter, 7 August 202630

7How it runs in Seal.

Seal’s QMS runs the processes a quality manual describes and links: deviations, CAPA, change control, documents, training, audits, suppliers, complaints and risk. The manual itself is a controlled document. In document control, each version moves from draft through review and approval to effective, and signature records show who approved it, when and with what meaning. Where policy requires periodic review, the interval, reminders and escalation sit on the document record.

A revision can identify the roles it affects, and a configured automation assigns the training linked to the new version. The periodic review draws on deviations, audit findings, CAPAs and changes, and ends in a recorded decision to continue, revise, consolidate, replace or retire the document.

AQuestions and answers

Is a quality manual the same as a QMS?

No. The quality management system is the set of processes, people, records and controls that run quality. The quality manual is the document that describes that system and points to its procedures. A good manual with a system that doesn’t work as described is a finding, not a defense.

How long should a quality manual be?

EU GMP, PIC/S, WHO and ICH Q10 set no length. ICH Q10 names the policy, scope, process links and responsibilities; ISO 13485 adds the document structure and the procedures or references to them. Keep detail that changes often in the procedures, so the manual changes only when the system does.

Does a CDMO need a different quality manual for each client?

Usually not. One manual describes the site’s quality system. Each client’s contract or quality agreement sets who does what for that client’s product, and the manual should say how those agreements work with the site’s own deviation, change and release processes.

Can a quality manual be electronic?

Yes. EU GMP applies its documentation rules whatever the medium, paper or electronic. The manual must be controlled like any other quality system document: one effective version, approved and dated, with its history kept, and available to the people who need it.

References30 sources

References

  1. 1ICH Q10, Pharmaceutical Quality System, sections 1.8, 2.1 to 2.3, 2.6 and 2.7, ICH, 4 June 2008. https://database.ich.org/sites/default/files/Q10%20Guideline.pdf (read 26 November 2026)
  2. 2Compliance Program 7382.850, Inspection of Medical Device Manufacturers, Attachment A, QMS area tables, FDA, implemented 2 February 2026. https://www.fda.gov/media/80195/download (read 26 November 2026)
  3. 3MDSAP AU P0002.009, MDSAP Audit Approach, Chapter 1, Task 1: QMS Planning, Implementation, Changes and Quality Manual; Chapter 6, Task 1: planning of production and service processes, and Task 22: nonconforming product, Medical Device Single Audit Program, revised 6 August 2024. https://www.fda.gov/media/166672/download (read 26 November 2026)
  4. 4EudraLex Volume 4, EU GMP Guide Part I, Chapter 4: Documentation, sections 4.3 and 4.5, European Commission, in operation from 30 June 2011. https://health.ec.europa.eu/system/files/2016-11/chapter4_01-2011_en_0.pdf (read 26 November 2026)
  5. 5Quality Systems Approach to Pharmaceutical CGMP Regulations, guidance for industry, sections IV.A.3 and IV.A.5, the table of CGMP citations for management responsibilities, and the glossary, FDA, September 2006. https://www.fda.gov/media/71023/download (read 26 November 2026)
  6. 621 CFR 820.10, Requirements for a quality management system, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.10 (read 26 November 2026)
  7. 7Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, FDA, 2 February 2024, effective 2 February 2026. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  8. 8EudraLex Volume 4, EU GMP Guide Part I, Chapter 2: Personnel, sections 2.2 to 2.4, European Commission, in operation since 16 February 2014. https://health.ec.europa.eu/system/files/2016-11/2014-03_chapter_2_0.pdf (read 26 November 2026)
  9. 921 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, including 211.22, eCFR, up to date as of 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 26 November 2026)
  10. 10EudraLex Volume 4, EU GMP Guide Part I, Chapter 1: Pharmaceutical Quality System, sections 1.6 and 1.7, European Commission, in operation from 31 January 2013. https://health.ec.europa.eu/system/files/2016-11/vol4-chap1_2013-01_en_0.pdf (read 26 November 2026)
  11. 11PE 009-⁠18, Guide to Good Manufacturing Practice for Medicinal Products, Part I, Chapter 1, section 1.7, PIC/S, 24 September 2026. https://picscheme.org/docview/11332 (read 26 November 2026)
  12. 12WHO good manufacturing practices for pharmaceutical products: main principles, WHO Technical Report Series No. 986, Annex 2, sections 1.6 and 1.7, World Health Organization, 2014. https://www.who.int/publications/m/item/trs986-annex2 (read 26 November 2026)
  13. 13The Human Medicines Regulations 2012, regulation C17, legislation.gov.uk. https://www.legislation.gov.uk/uksi/2012/1916/regulation/C17 (read 26 November 2026)
  14. 14Medicines: good manufacturing practice and good distribution practice, MHRA, GOV.UK, updated 13 May 2024. https://www.gov.uk/guidance/good-manufacturing-practice-and-good-distribution-practice (read 26 November 2026)
  15. 15ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, section 2.1, paragraph 2.11, ICH, 10 November 2000. https://database.ich.org/sites/default/files/Q7%20Guideline.pdf (read 26 November 2026)
  16. 16Guidelines of 5 November 2013 on Good Distribution Practice of medicinal products for human use, 2013/C 343/01, Chapter 1, section 1.2, Official Journal of the European Union, EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:52013XC1123(01) (read 26 November 2026)
  17. 1721 CFR Part 210, Current Good Manufacturing Practice in Manufacturing, Processing, Packing, or Holding of Drugs; General, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-210 (read 26 November 2026)
  18. 18Q10 Pharmaceutical Quality System, guidance for industry, section II.G (ICH 1.8), FDA, April 2009. https://www.fda.gov/media/71553/download (read 26 November 2026)
  19. 1921 CFR 880.6230, Tongue depressor, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-880/subpart-G/section-880.6230 (read 26 November 2026)
  20. 20Regulation (EU) 2017/745 on medical devices, Annex IX section 2.2 and Annex XI section 6.2, EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:32017R0745 (read 26 November 2026)
  21. 21Regulation (EU) 2017/746 on in vitro diagnostic medical devices, Annex IX section 2.2 and Annex XI section 3.2, EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:32017R0746 (read 26 November 2026)
  22. 22Stakeholders’ Consultation on EudraLex Volume 4, Good Manufacturing Practice Guidelines: Chapter 1, with the draft guideline, European Commission, open 3 September to 3 December 2025. https://health.ec.europa.eu/consultations/stakeholders-consultation-eudralex-volume-4-good-manufacturing-practice-guidelines-chapter-1_en (read 26 November 2026)
  23. 23Compliance Program 7346.832, Preapproval Inspections, Parts II and III, FDA, issued 29 June 2026, implemented 10 August 2026. https://www.fda.gov/media/193382/download (read 26 November 2026)
  24. 24EudraLex Volume 4, EU GMP Guide Part I, Chapter 7: Outsourced Activities, Principle and sections 7.5, 7.7, 7.8, 7.14 and 7.15, European Commission, in operation since 31 January 2013. https://health.ec.europa.eu/system/files/2016-11/vol4-chap7_2012-06_en_0.pdf (read 26 November 2026)
  25. 25Contract Manufacturing Arrangements for Drugs: Quality Agreements, guidance for industry, sections III and IV, FDA, November 2016. https://www.fda.gov/media/86193/download (read 26 November 2026)
  26. 26EudraLex Volume 4, Explanatory Notes on the preparation of a Site Master File, sections 1.2, 1.3 and 2.1, European Commission. https://health.ec.europa.eu/system/files/2016-11/2011_site_master_file_en_0.pdf (read 26 November 2026)
  27. 27Peter Schiff Enterprises, warning letter, CMS 727888, FDA, 26 August 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/peter-schiff-enterprises-727888-08262026 (read 26 November 2026)
  28. 2821 CFR 58.35, Quality assurance unit, and 58.1, Scope, Good Laboratory Practice for Nonclinical Laboratory Studies, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-A/part-58 (read 26 November 2026)
  29. 29Analytical Food Laboratories, Inc., warning letter 320-⁠25-⁠04, MARCS-⁠CMS 687513, FDA, 10 October 2024. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/analytical-food-laboratories-inc-687513-10102024 (read 26 November 2026)
  30. 30Safrel Pharmaceuticals LLC, warning letter 320-⁠26-⁠111, MARCS-⁠CMS 730520, FDA, 7 August 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/safrel-pharmaceuticals-llc-730520-08072026 (read 26 November 2026)