1The steps, and the record each produces.
Moving a GxP system to the cloud moves the work, not the responsibility. Annex 11 requires formal agreements whenever a supplier provides, configures, validates, maintains or retains a system, and says they should include “clear statements of the responsibilities of the third party”,1 and the 2025 draft revision adds that relying on a vendor does not change the requirements.2 Each step leaves a record an inspector can ask for.
| Step | What it settles | Record | Owner | Basis |
|---|---|---|---|---|
| 1. Plan | Scope, and which vendor evidence you will rely on, and why | Validation plan with the supplier-reliance rationale | Validation lead; QA approves | Draft 9.7;2 Annex 15 1.6;3 EMA clinical A2.54 |
| 2. Requirements | The functions you automate and rely on, each rated for GxP impact | URS and risk assessment | Process owner; QA approves | Annex 11 1 and 4.4;1 draft 6.1;2 FDA data integrity Q35 |
| 3. Vendor | Its quality system, development, testing, security and hosting, by questionnaire, review or audit | Supplier assessment report | QA | Annex 11 3.2 and 4.5;1 ICH Q10 2.7;6 FDA CSA7 |
| 4. Agreement | Who handles changes, releases, incidents, data access, backup, continuity and exit | Quality agreement and contract | QA and procurement | Annex 11 3.1;1 ICH Q10 2.7;6 MHRA 6.20;8 draft 7.52 |
| 5. Evidence map | Which requirements the vendor’s evidence covers, for which version, and the gaps | Requirement- | Validation lead | Annex 11 3.3;1 Annex 15 2.63 |
| 6. Configuration | The settings you chose, under control | Configuration specification | System owner | Draft 6.62 |
| 7. Testing | Your workflows, calculations, permissions, interfaces and a restore, including limits and error handling | Executed test evidence | Testers; QA reviews | FDA data integrity Q3;5 Annex 11 4.7 and 7.21 |
| 8. Migration | Values and meaning unchanged | Migration report | System owner | Annex 11 4.81 |
| 9. Release | Results against criteria; SOPs approved and users trained before GMP use | Validation summary report, SOPs, training records | Validation lead; QA approves | Annex 15 2.9 and 3.12;3 PI 006-4 4.1.49 |
| 10. New versions | Effect on your use, testing in proportion, approval before production | Release impact assessment | System owner; QA approves | FDA CSA Example 4;7 Annex 11 101 |
| 11. Periodic review | Whether it stays validated: incidents, deviations, release history, access reviews and the vendor’s service levels | Periodic review report | System owner; QA approves | Annex 11 11;1 draft 14.22 |
If the vendor offers its own requirements specification, start from it but don’t simply adopt it: the draft Annex 11 says the regulated user “should carefully review and approve the document” and “take ownership of the document covering the implemented version”.2 Strike the functions you won’t use and add what your process needs.
How far you can lean on the vendor depends on what you are buying, set out in GAMP 5 software categories; the vendor assessment itself follows supplier qualification, and the requirements are covered in how to write a URS.
2Who does what: a checklist to agree with the vendor.
Put each line in the contract, the quality agreement or an approved procedure.
| Area | The vendor provides | You do | Source |
|---|---|---|---|
| Quality system | Its QMS, development and test records, certifications such as SOC reports or ISO | Assess them; audit where the risk calls for it | Annex 11 3.2, 4.5;1 FDA CSA7 |
| Hosting | Data centres, security, patching, availability against agreed service levels | Know where the data sits and which laws apply; monitor the service levels | MHRA 6.20;8 draft 7.32 |
| Subcontractors | The names of its hosting and other providers | Approve them before work passes to them | Chapter 7.11;10 WHO 4.311 |
| Configuration | The options and a test environment | Specify, verify and control the configuration you choose | Draft 6.6;2 FDA CSA Example 47 |
| Your workflows | Nothing it can provide alone | Validate each GMP workflow and calculation as configured | FDA data integrity Q3;5 MHRA 6.198 |
| Releases | Notice, release notes, its test results | Assess the impact, test what it touches, keep the record | FDA CSA Example 4;7 draft 7.5;2 Annex 11 101 |
| Data access | Timely access to data, metadata and audit trails, for you and inspectors | Check you can retrieve and read them | MHRA 6.20;8 Annex 11 7.11 |
| Vendor staff access | Named support roles; no change or deletion of your GxP data without your approval; their actions in the audit trail | Approve that access and review it regularly; check back-end changes can’t bypass the audit trail | Annex 11 12.1 to 12.4;1 draft 11.10 and 11.11;2 EMA clinical A4.194 |
| Incidents | Notice, within agreed times, of failures, data errors and security incidents affecting your data; root cause and actions for critical ones | Assess each for GxP impact; for a critical one, check the root cause and follow the CAPA through | Annex 11 13;1 draft 7.5 iii and vii2 |
| Backup and continuity | Backups, restore, tested continuity arrangements | Check a restore during validation; keep your own continuity plan | Annex 11 7.2, 16;1 MHRA 6.208 |
| Inspections | Support during inspections and audits | Make supplier quality and audit information available to inspectors | Annex 11 3.4;1 draft 7.52 |
| Exit | Export of all records in a readable form | Keep control of the data for its whole retention period | Draft 7.5;2 MHRA 6.20;8 FDA clinical Q1712 |
MHRA adds that arrangements must exist to restore the system to its original validated state, including the validation and change control information needed to do so.8 What a quality agreement can and can’t do is covered in what a quality agreement should include.
3Example: assessing a vendor release.
Example, invented: Velloran Biologics and Kestrane LIMS are not real. Velloran runs Kestrane LIMS 7.2 for QC release testing at Site 2. Its agreement requires 30 days’ notice of each release. Kestrane notifies release 7.3 on 2 October 2026 for production on 9 November 2026, 38 days later, with release notes listing 14 changes and its own test summary. On 12 October Kestrane upgrades Velloran’s validation tenant, its one test environment, to 7.3, and Velloran confirms that tenant’s configuration matches the production baseline, record CV-73-00.
| Changes | Count | In Velloran’s use? | Action | Evidence |
|---|---|---|---|---|
| Unlicensed modules | 6 | No, disabled for the tenant | Confirm they stay disabled after upgrade | Configuration check CV-73-01, pass |
| Screens and speed | 4 | Yes, low risk | Rely on Kestrane’s tests, reviewed | Supplier assessment SA-07, addendum 3 |
| Approval workflow | 1 | Yes, URS-012, high | Rerun OQ-07 | OQ-07 run 3, pass, 19 Oct |
| Audit trail export | 1 | Yes, URS-021, high | Rerun OQ-15, add OQ-15b for the new format | Both pass, 20 Oct |
| Calculation library | 1 | Yes, URS-018, high | Rerun OQ-12 with boundary values | Pass, 21 Oct |
| Security patch | 1 | Yes, infrastructure | Review Kestrane’s evidence; recheck role access | Review and access check, pass, 21 Oct |
The 14 changes add up: 6, 4, 1, 1, 1 and 1. Four test runs and two checks in the validation tenant, and two reviews of Kestrane’s evidence, let QA approve the change on 28 October, twelve days before production. The record holds what FDA’s SaaS example describes: the vendor’s description of the changes and its testing, the assessment of their effect on intended use, testing in proportion to the impact, and a summary of the risk assessment and activities performed.7 CV-73-00 answers Annex 11 4.7, which asks for test environments to have documented assessments of their adequacy; EMA’s clinical guideline asks the cloud provider for a test environment identical to production when the user does its own validation.1,
4When a release doesn’t go to plan.
| Situation | What to do | Record | Basis |
|---|---|---|---|
| Scheduled, with notice | Assess against your requirements, retest what it touches, approve before production | Release impact assessment | FDA CSA Example 4;7 Annex 11 10;1 draft 7.5 ix2 |
| Emergency security patch | Accept it under your emergency change procedure, then assess it: recheck access and the audit trail | Emergency change record, assessed after the fact | PI 011-3 18.2;13 draft 15.132 |
| No chance to test first | Assess the vendor’s release testing, leave new functions unused until assessed, and tighten periodic review | Assessment of the vendor’s release evidence | EMA clinical A2.1;4 Annex 11 111 |
| Region or subcontractor | Approve it before data moves; check where data sits and which laws apply | Change control and amended agreement | Chapter 7.11;10 MHRA 6.208 |
| Vendor closes or is sold | Run the exit plan: export data, metadata and audit trails readably, and test retrieval | Exit or migration report | Annex 11 4.8 and 17;1 draft 7.5 viii;2 MHRA 6.208 |
Agree the emergency route in the contract before you need it: the draft Annex 11 says critical vulnerabilities may need patching immediately.2
5What the vendor’s evidence can and can’t cover.
FDA’s CSA guidance, reissued on 3 February 2026 for device production and quality system software, lets manufacturers leverage validation done by “developers, suppliers, cloud service providers”. It accepts that an onsite audit of a software vendor may not be feasible, and lists other evidence: SOC reports and ISO certificates, development, quality assurance and cybersecurity practices, and data integrity controls such as audit trails, access controls and accurate copies of records.7
EU GMP already has a rule in force. Annex 15 says that where a third party providing validation services supplies validation protocols and other documentation, “appropriate personnel at the manufacturing site should confirm suitability and compliance with internal procedures before approval”, and vendor protocols may be supplemented before use.3 The draft Annex 11 would let validation documentation come from the vendor “in parts or in whole”, if the regulated user authorises it and checks it covers the implemented version.2
The limit is your use. FDA’s example: qualifying an MES platform “does not demonstrate” that a master record built in it has the correct calculations.5 MHRA says vendor testing in isolation from the intended process “is likely to be limited to functional verification only and may not fulfil the requirements for performance qualification”.8
Infrastructure is a judgement call. FDA’s example is cloud storage: when it holds quality records, focus assurance on the features that protect those records and on Part 11; when it holds data that is not a quality record, the device validation clauses would not apply.7 How much of your own testing each function needs is covered in risk-based validation.
6How inspectors enforce it.
FDA holds the user responsible for validation the vendor performed, and for data the vendor holds:
You are responsible for the validation of your HPLC system despite your vendor having completed it on your behalf.
Westwood Laboratories LLC, warning letter of 17 April 2024, on the software of its own laboratory HPLC system, which its vendor had validated. The firm could not find the validation documents during the inspection and needed the vendor’s help to retrieve them.14
a third-party vendor contracted by Applied Therapeutics deleted electronic data in Q-global®, including associated audit trails, for the (b)(4) for all 47 subjects enrolled in the study at all (b)(4) clinical sites.
Applied Therapeutics, Inc., warning letter of 27 November 2024, on Pearson’s web-based Q-global® system in a clinical study. The deletion came two days after FDA preannounced a site inspection. The firm’s planned fixes included sponsor-held backups and removing vendors’ ability to delete files; FDA still found the response inadequate for lack of detail.15
Both point at the same clauses: documentation you can produce and explain from your own site, which the draft Annex 11 asks for in 7.4, and access, backup and exit terms that keep the data under your control.2,
7How FDA, EU and MHRA differ.
| Question | EU GMP and PIC/S | MHRA | FDA |
|---|---|---|---|
| What requires it | “The application should be validated; IT infrastructure should be qualified”, Annex 11 principle1 | Validated “for their intended purpose”, 6.19, citing Annexes 11 and 158 | Drugs: 211.68(b), each CGMP workflow validated, Q3;16, |
| Written agreement | “formal agreements must exist”, Annex 11 3.11 | A technical agreement or contract defining responsibilities, 6.208 | Drugs: for outsourced activities, responsibilities “should be included in a written agreement”, ICH Q10 2.7;6 devices: a service agreement on security, data integrity, availability, change management and continuity, CSA Example 4;7 clinical: recommended, question 1712 |
| Vendor audit | Need based on risk; audit information available to inspectors on request, Annex 11 3.2, 3.41 | Need based on risk, 6.208 | Drugs: the provider’s suitability assessed before outsourcing, audits among the means, ICH Q10 2.7;6 devices: onsite audit may not be feasible, CSA;7 clinical: FDA will generally not review your audit reports of IT providers, question 1012 |
| Vendor documents | Site personnel confirm suitability before approval, Annex 15 2.63 | Vendor data alone, without your configuration and use, “is not acceptable”, 6.198 | Devices: vendor work leveraged by risk, CSA; drugs: each workflow still validated, data integrity Q37, |
| Cloud named | Not in Annex 11 of 2011; the draft names SaaS, 6.3;2 WHO 4.1 names cloud-based systems11 | SaaS, PaaS and IaaS named, 6.208 | Cloud infrastructure among “computer or related systems” under 211.68;5, |
| Clinical trials | EMA guideline of 2023: cloud provider qualified, explicit contracts, 6.74 | The same data integrity guidance covers GCP, 2.48 | FDA may inspect IT providers that took on sponsor duties, and others when trial data integrity is in doubt, clinical question 1912 |
| Current texts | Annex 11 of 2011; the July 2025 draft is not adopted on 8 November 202620 | Data integrity guidance, March 20188 | Drug data integrity Q&A, 2018; device CSA, February 20265, |
The draft Annex 11 would expand today’s section 3 on suppliers into a section on supplier and service management: oversight through agreed service levels and indicators, and contracts covering audits, inspection support, an exit strategy and testing of new versions.2 The full set of changes is on the Annex 11 revision page.
8How it runs in Seal.
Seal’s validation and AI assurance guide sets out the same split for Seal itself. Seal provides the platform controls and the supplier verification and release evidence for the version in scope; your team owns the intended use, process requirements, risk acceptance and production approval, including integrations and migrated data. Your reviewers can request Seal’s supplier evidence for their assessment. The hosting region is agreed at onboarding, and on the platform every change runs its tests before your team releases it, with failures kept on the record.
For the other cloud systems you rely on, the computer system assurance blueprint records each system’s intended use, the supplier evidence assessed against the claims it supports, and its configuration baseline. Each supplier release becomes a change control, and it cannot be published until every approved test has a passing run on the version being released and every sign-off is given.
AQuestions and answers
Do we have to audit our SaaS vendor?
Not always. Annex 11 and MHRA base the need for an audit on risk. For drugs, FDA’s adoption of ICH Q10 names audits as one way to assess a provider before outsourcing, and its software assurance guidance for device makers accepts that an onsite audit of a software vendor may not be feasible. A documented assessment of the vendor’s certifications, development and testing practices, security and data integrity controls can stand in its place when the risk allows. If the risk calls for an audit and the vendor refuses one, treat that as a finding in your selection.
Is a SOC 2 report or ISO certificate enough to validate a cloud system?
No. FDA lists them as inputs to a vendor assessment, and they speak to the vendor’s controls. They don’t show that your configured workflows, calculations and records do what your process needs, which is the intended use you have to validate.
Do we need to qualify AWS, Azure or Google Cloud ourselves?
Responsibility is shared. If you build on them directly, you qualify the layers you configure and run, such as your servers, databases and their settings, and assess the provider for the layers it runs, in proportion to what the records stored there are used for; Annex 11 expects IT infrastructure to be qualified. If a SaaS vendor runs on them, the vendor controls its hosting, and you assess how it does so, including its control of that subcontractor, as part of the vendor assessment.
Does it matter where a cloud vendor stores our data?
Yes. MHRA asks regulated companies to consider the physical location of the data and the laws that apply there, as well as ownership, retrieval, retention and security. Ask for the hosting region, backup locations and any support access from other countries.
What happens to our GxP records if the SaaS vendor goes out of business?
You still have to keep them readable for the whole retention period. Annex 11 expects access to data to be ensured throughout that period, and MHRA and the draft revision of Annex 11 ask contracts to cover archiving, continued readability and an exit strategy. Agree an export format and test it before you need it.
References20 sources
References
- 1EudraLex Volume 4, Annex 11: Computerised Systems, European Commission, revision 1, in operation from 30 June 2011, principle and sections 1, 3, 4, 7, 10, 11, 12, 13, 16 and 17. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ annex11_ 01- 2011_ en_ 0. pdf (read 8 November 2026) - 2Annex 11: Computerised Systems, draft for stakeholders’ consultation, European Commission, July 2025, sections 6, 7, 9.7, 9.9, 11.10, 11.11, 14.2 and 15.13. https:/
/ health. ec. europa. eu/ document/ download/ 40231f18- e564- 4043- 94de- c031f813d38b_ en?filename=mp_ vol4_ chap4_ annex11_ consultation_ guideline_ en. pdf (read 8 November 2026) - 3EudraLex Volume 4, Annex 15: Qualification and Validation, European Commission, in operation from 1 October 2015, sections 1.6, 2.6, 2.9 and 3.12. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ 2015- 10_ annex15_ 0. pdf (read 8 November 2026) - 4Guideline on computerised systems and electronic data in clinical trials, EMA/
INS/ GCP/ 112288/ 2023, European Medicines Agency, 9 March 2023, section 6.7, Annex 2, sections A2.1 and A2.5, and Annex 4, section A4.19. https:/ / www. ema. europa. eu/ en/ documents/ regulatory- procedural- guideline/ guideline- computerised- systems- and- electronic- data- clinical- trials_ en. pdf (read 8 November 2026) - 5Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, FDA, December 2018, questions 1.f and 3. https:/
/ www. fda. gov/ media/ 119267/ download (read 8 November 2026) - 6Q10 Pharmaceutical Quality System, guidance for industry adopting ICH Q10, FDA, April 2009, section III.G, Management of Outsourced Activities and Purchased Materials (ICH 2.7). https:/
/ www. fda. gov/ media/ 71553/ download (read 8 November 2026) - 7Computer Software Assurance for Production and Quality Management System Software, guidance for industry and FDA staff, FDA, 3 February 2026, sections III and V and Appendix A, Example 4. https:/
/ www. fda. gov/ media/ 188844/ download (read 8 November 2026) - 8‘GXP’ Data Integrity Guidance and Definitions, Revision 1, MHRA, March 2018, sections 2.4, 6.19 and 6.20. https:/
/ assets. publishing. service. gov. uk/ media/ 5aa2b9ede5274a3e391e37f3/ MHRA_ GxP_ data_ integrity_ guide_ March_ edited_ Final. pdf (read 8 November 2026) - 9PI 006-4, Recommendations on Qualification and Validation, PIC/S, in force from 1 October 2026, section 4.1.4. https:/
/ picscheme. org/ docview/ 11277 (read 8 November 2026) - 10EudraLex Volume 4, Chapter 7: Outsourced Activities, European Commission, revision 1, in operation from 31 January 2013, section 7.11. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ vol4- chap7_ 2012- 06_ en_ 0. pdf (read 8 November 2026) - 11WHO Technical Report Series No. 1019, Annex 3, Good manufacturing practices: guidelines on validation, Appendix 5, Validation of computerized systems, section 4, WHO, 2019. https:/
/ www. who. int/ docs/ default- source/ medicines/ norms- and- standards/ guidelines/ production/ trs1019- annex3- gmp- validation. pdf (read 8 November 2026) - 12Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers, guidance for industry, FDA, October 2024, section III.C and questions 10, 17 and 19. https:/
/ www. fda. gov/ media/ 166215/ download (read 8 November 2026) - 13PI 011-3, Good Practices for Computerised Systems in Regulated GxP Environments, PIC/S, 25 September 2007, section 18.2. https:/
/ picscheme. org/ docview/ 3444 (read 8 November 2026) - 14Warning letter to Westwood Laboratories LLC, MARCS-CMS 674728, FDA, 17 April 2024. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ westwood- laboratories- llc- 674728- 04172024 (read 8 November 2026) - 15Warning letter to Applied Therapeutics, Inc., MARCS-CMS 696833, FDA, dated 27 November 2024, posted 3 December 2024. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ applied- therapeutics- inc- 696833- 12032024 (read 8 November 2026) - 1621 CFR 211.68, Automatic, mechanical, and electronic equipment, eCFR, paragraph (b). https:/
/ www. ecfr. gov/ current/ title- 21/ section- 211. 68 (read 8 November 2026) - 1721 CFR Part 11, Electronic Records; Electronic Signatures, eCFR, section 11.10(a). https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- A/ part- 11 (read 8 November 2026) - 18Part 11, Electronic Records; Electronic Signatures: Scope and Application, guidance for industry, FDA, August 2003, section III.C.1. https:/
/ www. fda. gov/ media/ 75414/ download (read 8 November 2026) - 1921 CFR 820.10, Requirements for a quality management system, eCFR, paragraph (a). https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- B/ section- 820. 10 (read 8 November 2026) - 20EudraLex Volume 4, Good Manufacturing Practice guidelines, European Commission, listing Annex 11 as the January 2011 revision. https:/
/ health. ec. europa. eu/ medicinal- products/ eudralex/ eudralex- volume- 4_ en (read 8 November 2026)