What is the difference between a deviation and a nonconformance?

A deviation is a departure from an approved instruction or established standard, such as a skipped step or an exceeded hold time; it may or may not affect quality. A nonconformance is a failure to meet a specification or requirement, such as a lot outside its limits, and it does affect quality. ICH says the terms are related, not synonyms. A deviated batch that meets its specifications may be released after investigation; failing drug product is rejected or reprocessed.Checked against its sources on 9 November 2026. How these answers are made

Illustration of a seal examining a circled change in a trace alongside run, sample and procedure records.

1Deviation vs nonconformance at a glance.

ICH answers the question directly: the terms are not synonyms, “However, they are related.” A deviation is a “departure from an approved instruction or established standard” that “may or may not have an impact on the quality of the material”. A non-conformance is “a status as a result of a failure of the material to meet specifications or appropriately established standards that impacts the quality of the material”.1

Table 1. Definitions from ICH Q7 and ISO 9000, which says nonconformity; forms usually say nonconformance.2,3,1
DeviationNonconformance
DefinitionDeparture from an approved instruction or established standardNon-fulfilment of a requirement
Effect on qualityMay or may not have oneHas one: the material fails a requirement
Usually measured againstHow the work is done: procedure, batch recordWhat the result must be: specification, drawing
Usual homeDrug and biologic GMPDevice rules and ISO-based systems
ExampleHold time exceeded; step done out of orderLot outside its moisture limit; part outside tolerance
First responseRecord, justify, assess the impactIdentify, segregate, evaluate
Usual outcomeInvestigation and a batch decisionDevices: rework, repair, concession or scrap. Failing drug product: rejection or reprocessing

Other definitions are broader. FDA’s 2006 drug guidance calls any deficiency in a “characteristic, product specification, process parameter, record, or procedure that renders the quality of a product unacceptable, indeterminate, or not according to specified requirements” a non-conformity, so a skipped step that leaves quality unknown is already one. ISO 9000 agrees, as a procedure states requirements.4,3 What differs in practice is what the record controls: a departure in the process, or an item that must not be used until someone decides its fate.

2Which record do I open?

Ask first whether the batch still meets its registered specifications. That is the line Annex 16 draws (below), and in the US a failing laboratory result goes to an OOS investigation, in-process tests included.5,6

Table 2. Which record to open first.
What you foundOpen firstBasis
Failing lab resultOOS investigation; if the result is confirmed, the batch or material fails and is dispositionedFDA OOS guidance;6 211.165(f), 211.84(e)7,8
Departure, results dueDeviation, material on hold; link an OOS and an NCR if a test then fails (example below)FDA 2006: “identify and/or segregate”;4 ICH Q7 2.172
Step not done as writtenDeviation, with specifications still met211.100(b);9 Annex 16 section 35
Wrong SOP or no entryDeviation; check the data it affects. See late entries211.100(b): “documented at the time of performance”;9 ICH Q7 2.152
Equipment or HVAC faultDeviation: out of calibration, utility loss or a monitoring excursion; assess every batch exposed since the last good check211.160(b)(4);10 211.19211
Defect seen, no testNonconformance: damage, wrong label, foreign matter found on receipt or in process. A drug site with no NCR record opens a deviation and quarantines the material211.82(a) receipt checks;12 FDA 2006: identify or segregate;4 ISO 13485 8.3.113
Departure planned aheadPlanned deviation, approved before it happensISO 9000: deviation permit3
A lasting changeChange controlFDA cites changes made outside change management14
Departure and bad lotOne investigation: linked deviation and NCR, or one deviation that holds the dispositionRecommendation; the QC unit decides either way15

Each record has its own page: deviations, nonconformance reports, OOS investigations, planned deviations and change control. A cause worth removing goes to a CAPA, linked to its source.16 At a contract manufacturer, the quality agreement says how deviations reach the owner.17

3Where the regulations draw the line.

  • 21 CFR 211.100(b): “Any deviation from the written procedures shall be recorded and justified.” 211.192 then applies one investigation duty to an unexplained discrepancy and to a batch’s failure “to meet any of its specifications”, as ICH Q7 6.53 does for APIs.9,11,2
  • For licensed biologics, a reportable deviation includes “a deviation from current good manufacturing practice, applicable regulations, applicable standards, or established specifications”, so there a failed specification is a deviation too.18
  • EU Annex 16: “Provided registered specifications for active substances, excipients, packaging materials and medicinal products are met, a QP may consider confirming compliance or certifying a batch where an unexpected deviation” has occurred. A deviation can end in certification; a failed specification cannot.5 PIC/S GMP Annex 16 sets the same rule for the Authorised Person,19 and in the UK the EU guidelines as they stood at the end of 2020, this Annex included, still apply unless the MHRA replaces them.20
  • 21 CFR 211.165(f): “Drug products failing to meet established standards or specifications and any other relevant quality control criteria shall be rejected. Reprocessing may be performed.” Reprocessing needs a written procedure and the quality control unit’s approval; a concession or use-as-is is not an option.7,21
  • ISO 13485 clause 8.3.1: nonconforming product must be “identified and controlled to prevent its unintended use or delivery”, as FDA reminded Nipro in 2026.13

Part 211 never uses “nonconformity” or “nonconformance”, though ICH Q10 lists “non-conformances” and “deviations” among CAPA sources.22,16 ISO 9000:2026 has replaced the 2015 edition that 21 CFR 820.7 still names, keeping the definitions this page uses under new numbers: nonconformity is now 3.5.13.23,24

4What each record must hold.

Build both forms from these fields. Your procedure sets the classes and their criteria: the rules speak of “critical” (ICH Q7) or “significant” (EU GMP) deviations without a classification scheme.2,25

Table 3. Fields for a deviation or nonconformance form, with what each satisfies.
FieldWhat to enterSource
Number and linksUnique number; the related OOS, NCR, CAPA or change recordICH Q7 6.522
FoundDate, time, place, who found it, and the batch, lot or item211.100(b);9 ICH Q7 2.152
Requirement and factsThe instruction or specification departed from, its version, what happened; for a deviation, the justification211.100(b): “recorded and justified”;9 ISO 9000:2015 3.6.93
ContainmentWhat was held or segregated, and whenFDA 2006;4 ISO 13485 8.3.113
ClassThe grade under your procedureICH Q7 2.16;2 EU 1.8(vii)25
Impact and scopeEffect on quality; other batches and products that may be associated211.192;11 Annex 16 3.15
Cause and CAPARoot or most likely cause, any human error justified; each CAPA raisedEU 1.4(xiv);25 ICH Q10 3.2.216
DecisionBatch decision or disposition, approved by the quality unit; for a concession, the justification211.22(a);15 QMSR preamble26
ClosureConclusions, follow-up and date211.19211

Neither Part 211, ICH Q7 nor EU GMP Chapter 1 sets a deadline for closing either record: FDA asks for “timely” investigations, and your procedure sets the target and how an overrun is approved; see deviation investigation reports.22,2,25,6 For US drugs and biologics, either record becomes a field alert or BPDR only when distributed product may be affected.18,27 Both records then feed the yearly review: “investigations conducted under § 211.192” in the US, “all significant deviations or non-conformances” in the EU, and “all critical deviations or non-conformances and related investigations” for APIs.28,25,2

5Example: one event, linked records.

Example, invented for illustration. The operator stops the final blend of batch 26-⁠058 of Examplex 5 mg tablets at 4 minutes instead of the 6 in the batch record. Blend uniformity testing then puts 2 of 10 samples outside the 90.0 to 110.0% acceptance range, at 88.6% and 111.4%.

Table 4. Example. One invented event and the records it opens.
RecordWhat it holdsCloses when
DEV-⁠26-⁠0140The 4-minute blend against the required 6, who found it, the impact on batch 26-⁠058The investigation is approved and a CAPA opened
OOS-⁠26-⁠0213Phase I laboratory investigation: no assignable cause, results confirmedThe result stands and the investigation moves to production
NCR-⁠26-⁠0077The blend on hold, the 2 failing results, the dispositionThe blend is rejected and its destruction recorded
InvestigationCause: the electronic batch record (version 3) does not enforce the 6-minute blend timeApproved once, linked to all three
CAPA-⁠26-⁠0044Version 4 enforces the 6-minute blend, under change control CC-⁠26-⁠0091The next 10 batches each blend for 6 minutes and pass every uniformity sample

The OOS record confirms the failure before the quality control unit rejects the blend.6,15 The NCR closes there, but the investigation extends to every batch made from version 3.11

6What inspectors cite.

A class set too low, so defects below a rate limit go uninvestigated. FDA told SCA Pharmaceuticals, an outsourcing facility:

“It appears that your visual inspection system is inadequate because you classify particulate defects as major and therefore only characterize and investigate the observed particles when the defect rate is above the major defect threshold of (b)(4)%.” SCA Pharmaceuticals, LLC, warning letter, 14 September 202629

FDA added that extrinsic particulates “should be classified as critical”.29 Base class criteria on risk, and say in the procedure who may downgrade or cancel an event, and on what evidence; deviation investigation reports covers Genzyme’s cancelled deviations.

A deviation closed on passing results. Meeting the specification does not end the investigation of a deviation, and FDA cited Fresenius Kabi under 211.192:

“famotidine injection manufacturing had experienced multiple (b)(4) bioburden action limit failures. However, your firm failed to sufficiently investigate this critical deviation and permitted batch release following passing finished drug product test results (i.e., sterility, endotoxin).” Fresenius Kabi USA LLC, warning letter, 22 September 202630

Testing of retained samples, prompted by adverse event complaints, later found an endotoxin failure, and three batches were recalled.30

7How it runs in Seal.

Seal keeps the two as separate records that can link. In the nonconformance blueprint, a nonconformance begins with an item, material, product or output that does not meet a requirement and needs control and disposition, and a deviation begins with a departure from an approved instruction or process and needs investigation. One event can create or relate to the other, and their containment, evaluation and decision responsibilities stay separate. When a nonconformance is opened, an inventory hold is placed on the material, and an issue to a batch is flagged.

The deviation blueprint links each deviation to the batch, sample and procedure version behind it. Approved trace rules find the affected population, keeping potentially affected, assessed unaffected, confirmed affected and unknown states distinct.

AQuestions and answers

Is a nonconformance the same as a nonconformity?

In practice, yes. ISO 9000 defines nonconformity as the non-fulfilment of a requirement, and nonconformance is the everyday word for the same thing. ISO treats conformance as a deprecated synonym of conformity, so standards say nonconformity while procedures and forms often say nonconformance.

What is the difference between a deviation and a discrepancy?

Discrepancy is the broader word. FDA’s quality systems guidance defines it as a result outside the expected range or an unfulfilled requirement, and says it may be called a non-conformity, defect, deviation or out-of-specification result. Under 21 CFR 211.192, any unexplained discrepancy must be thoroughly investigated, whatever your procedures call it.

Is an incident the same as a deviation?

Not as a GMP term. 21 CFR Part 211 and EU GMP Chapter 1 do not use the word. ICH Q7 and EU GMP Annex 11 use it for computerised system events, which are recorded and assessed, and Annex 11 lists deviation records and incidents as separate inputs to a system’s periodic review. In EU device law an incident is a malfunction or deterioration of a device on the market, handled through vigilance. If your procedures use incident for events with no product impact, define it and say when one becomes a deviation.

Does every deviation or nonconformance need a CAPA?

No. Both need recording and an assessment of their effect. A CAPA follows when the investigation finds a cause worth removing, when the event is significant, or when trending shows it recurring.

Do medical device companies use the word deviation?

The device regulation and ISO 13485 are written in terms of nonconformities and nonconforming product. A departure from a procedure is still a nonconformity under ISO 9000, because the procedure states a requirement, so a device quality system can handle it as a process nonconformity whatever its procedures call it. The record still has to exist: FDA’s 2026 warning letter to Nipro Renal Solutions cited at least thirteen endotoxin failures for which the firm could show no documented investigation or nonconformance record.

References30 sources

References

  1. 1ICH Q7 Guideline: Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, Questions and Answers, ICH, Step 4, 10 June 2015, question 20.1. https://database.ich.org/sites/default/files/Q7%20Q%26As%20Questions%20%26%20Answers.pdf (read 8 November 2026)
  2. 2ICH Q7, Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients, ICH, 10 November 2000, sections 2.15 to 2.17, 2.50, 6.52, 6.53 and glossary. https://database.ich.org/sites/default/files/Q7%20Guideline.pdf (read 8 November 2026)
  3. 3ISO 9000:2015, Quality management systems: Fundamentals and vocabulary, fourth edition, entries 3.6.4, 3.6.9, 3.6.11, 3.12.5 and 3.12.6, ISO Online Browsing Platform, edition marked withdrawn. https://www.iso.org/obp/ui/en/#iso:std:iso:9000:ed-4:v1:en (read 8 November 2026)
  4. 4Guidance for Industry: Quality Systems Approach to Pharmaceutical CGMP Regulations, FDA, September 2006, section IV.C.4 and glossary. https://www.fda.gov/media/71023/download (read 8 November 2026)
  5. 5EudraLex Volume 4, Annex 16: Certification by a Qualified Person and Batch Release, European Commission, 12 October 2015, sections 3 and 3.1. https://health.ec.europa.eu/system/files/2016-11/v4_an16_201510_en_0.pdf (read 8 November 2026)
  6. 6Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, Revision 1, FDA, May 2022, sections I, III and IV. https://www.fda.gov/media/158416/download (read 8 November 2026)
  7. 721 CFR 211.165, Testing and release for distribution, paragraph (f), eCFR. https://www.ecfr.gov/current/title-21/section-211.165 (read 8 November 2026)
  8. 821 CFR 211.84, Testing and approval or rejection of components, drug product containers, and closures, paragraph (e), eCFR. https://www.ecfr.gov/current/title-21/section-211.84 (read 8 November 2026)
  9. 921 CFR 211.100, Written procedures; deviations, paragraph (b), eCFR. https://www.ecfr.gov/current/title-21/section-211.100 (read 8 November 2026)
  10. 1021 CFR 211.160, General requirements, paragraph (b)(4), eCFR. https://www.ecfr.gov/current/title-21/section-211.160 (read 8 November 2026)
  11. 1121 CFR 211.192, Production record review, eCFR. https://www.ecfr.gov/current/title-21/section-211.192 (read 8 November 2026)
  12. 1221 CFR 211.82, Receipt and storage of untested components, drug product containers, and closures, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-211.82 (read 8 November 2026)
  13. 13Nipro Renal Solutions USA, Corporation, warning letter, MARCS-⁠CMS 732874, FDA, letter dated 16 July 2026 and listed 24 July 2026, violation 1. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/nipro-renal-solutions-usa-corporation-732874-07242026 (read 8 November 2026)
  14. 14TCI Co., Ltd BioCosme PABP BRANCH, warning letter 320-⁠26-⁠122, MARCS-⁠CMS 736888, FDA, 2 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/tci-co-ltd-biocosme-pabp-branch-736888-09022026 (read 8 November 2026)
  15. 1521 CFR 211.22, Responsibilities of quality control unit, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-211.22 (read 8 November 2026)
  16. 16ICH Q10, Pharmaceutical Quality System, ICH, Step 4 version of 4 June 2008, section 3.2.2. https://database.ich.org/sites/default/files/Q10%20Guideline.pdf (read 8 November 2026)
  17. 17Contract Manufacturing Arrangements for Drugs: Quality Agreements, guidance for industry, FDA, November 2016, sections IV.B, IV.B.1.a and IV.B.1.e. https://www.fda.gov/media/86193/download (read 8 November 2026)
  18. 1821 CFR 600.14, Reporting of biological product deviations by licensed manufacturers, paragraph (b), eCFR. https://www.ecfr.gov/current/title-21/section-600.14 (read 8 November 2026)
  19. 19PIC/S Guide to Good Manufacturing Practice for Medicinal Products, PE 009-⁠18 (Annexes), Annex 16, section 3, PIC/S, 24 September 2026. https://picscheme.org/docview/11333 (read 8 November 2026)
  20. 20The Human Medicines Regulations 2012 (SI 2012/1916), regulation C17(3), legislation.gov.uk. https://www.legislation.gov.uk/uksi/2012/1916/regulation/C17 (read 8 November 2026)
  21. 2121 CFR 211.115, Reprocessing, eCFR. https://www.ecfr.gov/current/title-21/section-211.115 (read 8 November 2026)
  22. 2221 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals, full text, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211 (read 8 November 2026)
  23. 2321 CFR Part 820, Quality Management System Regulation, sections 820.3 and 820.7, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 8 November 2026)
  24. 24ISO 9000:2026, Quality management: Fundamentals and vocabulary, fifth edition, Foreword and entries 3.5.1, 3.5.9, 3.5.13, 3.6.9 and 3.6.10, ISO Online Browsing Platform. https://www.iso.org/obp/ui/en/#iso:std:iso:9000:ed-5:v1:en (read 8 November 2026)
  25. 25EudraLex Volume 4, Part I, Chapter 1: Pharmaceutical Quality System, European Commission, in operation since 31 January 2013, sections 1.4(xiv), 1.8(vii) and 1.10(iv). https://health.ec.europa.eu/system/files/2016-11/vol4-chap1_2013-01_en_0.pdf (read 8 November 2026)
  26. 26Medical Devices; Quality System Regulation Amendments, final rule, FDA, 89 FR 7496, 2 February 2024, effective 2 February 2026, response to comment 12. https://www.govinfo.gov/content/pkg/FR-2024-02-02/html/2024-01709.htm (read 8 November 2026)
  27. 2721 CFR 314.81, Other postmarketing reports, paragraph (b)(1), NDA field alert report, eCFR. https://www.ecfr.gov/current/title-21/section-314.81 (read 8 November 2026)
  28. 2821 CFR 211.180, General requirements, paragraph (e)(2), eCFR. https://www.ecfr.gov/current/title-21/section-211.180 (read 8 November 2026)
  29. 29SCA Pharmaceuticals, LLC, warning letter 724202, FDA, 14 September 2026, section C, deficient corrective action 2. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/sca-pharmaceuticals-llc-724202-09142026 (read 8 November 2026)
  30. 30Fresenius Kabi USA LLC, warning letter 320-⁠26-⁠129, MARCS-⁠CMS 733107, FDA, 22 September 2026, violation 1. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/fresenius-kabi-usa-llc-733107-09222026 (read 8 November 2026)