Does FDA regulate laboratory-developed tests (LDTs) now that the LDT rule has been vacated?

Not as medical devices. On 31 March 2025 a federal court vacated FDA’s 2024 LDT rule, holding that laboratory-developed test services are professional medical services, not devices, and on 19 September 2025 FDA restored the earlier text of 21 CFR 809.3. LDTs are regulated by CMS under CLIA: the laboratory needs a CLIA certificate or exemption and must establish the test’s performance specifications before reporting results. A test sold to other laboratories as a kit is still an FDA-regulated device.Checked against its sources on 9 November 2026. How these answers are made

1What happened to FDA’s LDT rule.

On 6 May 2024 FDA published a final rule adding “including when the manufacturer of these products is a laboratory” to the definition of in vitro diagnostic products in 21 CFR 809.3(a), with a four-year phaseout of its enforcement discretion for LDTs, the first stage due in May 2025.1,2 On 31 March 2025 the US District Court for the Eastern District of Texas ruled in American Clinical Laboratory Association v. FDA: “The final rule is hereby SET ASIDE and VACATED.”3 The phaseout went with it.

The court accepted FDA’s authority over commercially distributed devices, “including devices used to perform standardized clinical tests (known as ‘test kits’)”, but held that “laboratory-developed test services are professional medical services that are qualitatively and categorically different from the tangible goods that FDA may regulate as ‘devices’”.3 On 19 September 2025 FDA removed the added words, a step it called “ministerial in nature”.2

The court chose vacatur over a remand that left the rule in place, seeing “no likelihood that FDA can justify its decision on remand”, and cited Fifth Circuit precedent that setting a rule aside “has nationwide effect, is not party-restricted”.3 What stays open is what the court did not decide, such as companion diagnostics (below), and what Congress, which it noted “has considered but declined to enact several bills”, may yet pass.3

2LDT vs IVD: what each must do.

The same assay carries different duties as a kit and as a laboratory’s own service.

Table 1. Either way, the laboratory running the test needs a CLIA certificate or exemption.8
DutyKit sold to laboratoriesLDT run by its own laboratory
Before marketingEstablishment registration and device listing; 510(k), De Novo or PMA unless the type is 510(k)-exempt4,5,6,7No FDA review; a CLIA certificate or exemption for the testing3,8
Quality systemThe QMSR, 21 CFR Part 8209CLIA’s quality system standards, 42 CFR 493 Subpart K10
PerformanceUnless waived, each lab using it as labelled verifies accuracy, precision, reportable range and reference intervals10,11The lab establishes every applicable specification, below10
LabellingLabelling under 21 CFR 809.1012The 809.30(e) statement on reports, if built from analyte specific reagents (ASRs)13
ProblemsMedical device reports and correction and removal reports14,15None for the test, which the court held is not a device; user facility reports still cover devices the lab uses3,14
Inspected byFDA16CMS or its agent, or an approved accreditation organisation17,18

3What counts as an LDT?

FDA described a test “designed, manufactured, and used within a single laboratory”,1 which CMS’s FAQ glosses as “a laboratory with a single CLIA certificate”.19 The court’s examples run from “routine tests such as pap smears and gram stains, to sophisticated molecular and genetic sequencing tests for cancer, heart disease, and rare and infectious diseases”.3 New York is more specific: assays not cleared or approved by FDA that the offering laboratory developed, including ones built from reagents or kits labelled ASR, research use only (RUO) or investigational use only (IUO), outside an FDA investigational device exemption (IDE).20 A test developed in one laboratory and run in a sister laboratory under a separate certificate fits none of the federal descriptions.

Under CLIA, any change to a cleared test that could affect its performance specifications makes it the laboratory’s own to establish. CMS adds: “A modified moderate complexity test (including modifications in its intended use) is considered uncategorized for CLIA and therefore becomes a high complexity test.”11

Table 2. Examples from CMS’s surveyor guidance and New York’s test approval page.11,20
TestBefore reporting, under CLIANew York
Cleared kit, as labelledVerify accuracy, precision and reportable range; check reference intervals10FDA-approved assay20
Plasma test run on urineEstablish: a new sample matrix changes the intended use11Modification, treated as an LDT20
Semi-quantitative cutoffEstablish11Modification, treated as an LDT20
PCR built from ASRsEstablish, and print the FDA statement on reports11,13LDT20
Kit labelled RUO or IUOEstablish10LDT, unless used under an FDA IDE20
Reagents on other analyserVerify, if FDA cleared that use: not a modification11FDA-approved assay20

4What CLIA requires before the first patient result.

LDTs are high complexity: a test not on the published lists “is considered to be a test of high complexity” until reviewed.21 The laboratory establishes each applicable specification before reporting patient results:10

ItemRuleWhat the record shows
Accuracy493.1253(b)(2)(i)Comparison with reference materials, a reference method or a method shown to be accurate; for a qualitative test, presence or absence identified11
Precision(b)(2)(ii)Day-to-day, run-to-run and within-run variation, and operator variance11
Analytical sensitivity(b)(2)(iii)“the lowest concentration or amount of the analyte” that can be measured or distinguished from a blank11
Analytical specificity(b)(2)(iv)What the method measures, and interfering substances from product information, literature or own testing11
Reportable range(b)(2)(v)The upper and lower limits of the test system11
Reference intervals(b)(2)(vi)A range for your patient population, by specimen type and demographics11
Other characteristics(b)(2)(vii)Anything else the test needs; the example below adds specimen stability10
Calibration, controls(b)(3); 493.1256Type, number and frequency of controls set from the specifications, at least the 493.1256(d) minimums unless an Individualized Quality Control Plan applies11,22
Raw data(c)“The actual measurement(s) taken, reactions and/or observations must be recorded.”11
Who does the work493.1451(b)(2)The technical supervisor, responsible for establishing performance characteristics23
Who approves493.1251(d); 493.1445(e)(3)The director signs and dates the procedure before use, having ensured the method gives “the quality of results required for patient care”24,25
Ongoing accuracy493.1236(c)(1); Subpart HProficiency testing for analytes in Subpart I; for other tests, accuracy verified at least twice a year (tag D5217)26,11
Retention493.1105(a)(3)(i)While the test is in use, and at least 2 years27

A new analyte may have no reference method to compare against. For that case CMS says “research results may be used to document the accuracy of the test by correlation with the clinical presentation”; precision, analytical specificity, sensitivity and interfering substances still have to be determined.11

4.1Example: validation summary VS-⁠RX-⁠01.

Invented. The laboratory at Hospital H designs Test RX, a qualitative PCR test for a respiratory pathogen on nasopharyngeal swabs in transport medium, using primers and probes bought as ASRs, so 493.1253(b)(2) applies.10 Criteria were set in the protocol before testing.

Table 4. Example. 95% intervals are exact binomial. Section 493.1253 sets no specimen numbers; CMS says the study should show performance adequate for the clients’ needs, as determined by the director and clinical consultant.10,11 With 40 positives the lower bound is 86.8%, so the director’s sign-off records why the point estimate is adequate for the intended use.
CharacteristicStudyCriterionResult
Accuracy80 stored specimens, 40 positive and 40 negative by a cleared comparatorAt least 95% agreement each way, as a point estimatePositive 39 of 40 (97.5%; 95% CI 86.8 to 99.9%); negative 40 of 40 (100%; 91.2 to 100%)
Sensitivity (LoD)20 replicates at 2,000, 1,000, 500 and 250 copies/mLAt least 19 of 20 detected20, 20, 19 and 14 of 20: LoD 500 copies/mL
PrecisionLow positive at 1,000 copies/mL (twice the LoD) and a negative; 6 replicates a day for 3 days, 3 by each of two operatorsEvery replicate correct18 of 18 detected; 18 of 18 not detected
Specificity15 related organisms; blood and mucin added to positivesNo cross-reaction or interferenceNone seen
Reportable rangeQualitativeDefined outcomesDetected, Not detected, Invalid
Reference intervalThe 40 negative specimensNot detectedNot detected in all 40
Specimen stability5 negative swabs spiked to 1,000 copies/mL and 5 unspiked, held at 2 to 8 °C and tested at 0, 24, 48 and 72 hoursSpiked detected and unspiked not detected at every point20 of 20 detected; 20 of 20 not detected: stable 72 hours

Each run, and so each day of testing, includes a negative and a positive control; the positive is extracted with the specimens to catch extraction errors. An internal control in each specimen detects amplification inhibition, as 493.1256(d)(3) asks when inhibition is a significant source of false negatives, and its failure is reported as Invalid.22 The technical supervisor signed the study, and the director signed and dated the procedure on 9 March 2026, before the first report.23,24 Every report carries the 809.30(e) statement:13

“This test was developed and its performance characteristics determined by (Laboratory Name). It has not been cleared or approved by the U.S. Food and Drug Administration.”

5Who inspects, and what they cite.

  • Who. With a certificate of compliance, CMS or a CMS agent may inspect “on a biennial basis or with such other frequency as CMS determines to be necessary”.17 Otherwise an approved state licensure programme or one of seven approved accreditation organisations, such as the College of American Pathologists, COLA or The Joint Commission, inspects,18,28 and CMS can still run validation and complaint inspections.29
  • What they use. Surveyors work from Appendix C of the State Operations Manual. Establishing specifications is tag D5423, verification D5421, calibration and controls D5425, documentation D5427.11
  • What they can do. Have staff test samples, observe testing, interview personnel and copy records.29 Appendix C reminds them that the laboratory “is responsible for establishing performance specifications for test systems using analyte specific reagents (ASR)”.11

6GMP methods, trial assays and other edge cases.

Table 5. CLIA covers examining “materials derived from the human body” for diagnosis, prevention or treatment, which is why a drug QC laboratory falls outside it.34 These states are examples: check each state your specimens come from. For GMP methods see method validation, verification and transfer.
Your situationRegulated byWhat applies
GMP QC methodFDA drug CGMP, not CLIAAccuracy, sensitivity, specificity and reproducibility established and documented30,31
Research labCLIA only if it reports patient resultsExempt if it does not report patient-specific results for diagnosis, prevention or treatment8
Research use onlyFDA labelling ruleLabelled “For Research Use Only. Not for use in diagnostic procedures.”12
New York specimensNew York as wellOnly laboratories with a state permit may test them; LDTs reviewed by the Clinical Laboratory Evaluation Program20
California specimensCalifornia as wellA state licence for moderate or high complexity testing, including for laboratories outside California32
A lab in WashingtonWashington, in place of CMSA Medical Test Site licence: CMS found the state programme equivalent to CLIA in 199333,18

6.1A trial assay or companion diagnostic: the sponsor’s checks.

The court’s opinion mentions neither companion diagnostics nor investigational devices,3 so a sponsor works from FDA’s device rules and its 2014 guidance.

  1. Will results select patients, assign treatment or choose the arm? Then the test will generally be considered investigational “unless employed for an intended use for which the device is already approved or cleared”, and “generally will be considered a significant risk device”. A sponsor may need an IND alone or an IND and an IDE, and should ask the drug and device centres which.35
  2. If not, is it exempt from Part 812? A 510(k)-cleared test used as labelled is. Otherwise, under 812.2(c)(3), a diagnostic device is exempt if the testing is noninvasive, needs no significant-risk sampling, introduces no energy and “Is not used as a diagnostic procedure without confirmation of the diagnosis by another, medically established diagnostic product or procedure”, and the device is labelled under 809.10(c).36
  3. Neither significant risk nor exempt? The abbreviated requirements of 812.2(b)(1) apply: IRB approval on the sponsor’s explanation of why the device is not significant risk, labelling under 812.5, and the rules on consent, monitoring, records, reports and promotion. The study then counts as having an approved IDE unless FDA says otherwise.36
  4. Do results reach the participant’s care? The testing laboratory needs a CLIA certificate or exemption unless it reports no patient-specific results.8
  5. Any participants in New York? If results drive enrolment, treatment arm, safety, toxicity or dosing, the laboratory needs a permit. The application includes the Risk Attestation Form, a narrative summary, the validation summary, sample reports, and an IRB or FDA no-significant-risk letter or the clinicaltrials.gov identifier. A test under an IDE counts as FDA-approved.20
  6. Will the drug’s labelling require the test? Then it is a companion diagnostic, which “in most circumstances” should be approved or cleared with the drug.37 How the ruling applies to a companion diagnostic offered as one laboratory’s service is untested.

The ruling covers the laboratory’s service, not what it ships. In October 2024, before the ruling, FDA warned a laboratory that mailed HIV dried blood spot kits to the public:

“your firm offers its HIV DBS card self-collection kit for delivery to individuals and intends for these individuals to use the kit to self-collect DBS samples and ship those samples to MTL for HIV testing.” Blackfly Investments, LLC dba Molecular Testing Labs, warning letter, 11 October 202438

FDA found the kit adulterated for lacking a PMA or an IDE, and misbranded for lacking a 510(k) notification.38

7Outside the US.

  • EU. IVDR Article 5(5) exempts devices made and used only within an EU health institution from all but the relevant Annex I requirements, if all its conditions are met, among them no transfer to another legal entity, an EN ISO 15189 or nationally compliant laboratory and a public declaration. Points (b), (c) and (e) to (i) have applied since 26 May 2024; point (d), a documented justification that no equivalent device on the market meets the patients’ needs, applies from 31 December 2030.39
  • Great Britain. Regulation 33 of the UK MDR 2002 exempts a health institution that makes IVDs from certain requirements, if the devices are used on its premises or in the immediate vicinity and are not transferred to another legal entity.40

8What changed in 2026.

DateWhat happened
19 May 2026H.R. 8890, the Enhancing CLIA Act, introduced: an LDT “is not a medical device”, but labs would report certain harmful errors within 5 calendar days from two years after enactment. No further action recorded41,42
16 July 2026CMS and CDC ask how CLIA should change, including how labs establish performance specifications; comments closed 14 September 202643
8 October 2026CLIA proposed rule on virtual access, cytology proficiency testing and personnel; it does not mention 493.1253. Comments close 7 December 202644

9How it runs in Seal.

Seal does not decide which regime a test falls under. It keeps the laboratory’s own evidence, which is what CLIA asks for. In Seal’s LIMS, each result is captured as the work runs with its sample, preparation, method, specification and instrument file, and preparations record the reagent lots used. A method change carries its impact assessment and verification, and earlier runs keep the version they used.

The analytical method lifecycle keeps intended use, procedure versions, development and validation studies and routine performance connected, and assesses each laboratory’s readiness on its own evidence. For IVD manufacturers whose kits are devices, the medical device blueprint links each test method to the requirement or risk control it verifies.

AQuestions and answers

Is a GMP lab’s in-house test method an LDT?

No. LDT describes a clinical laboratory’s test on human specimens for patient care, which CLIA covers. A method a GMP quality control laboratory develops to test a drug falls under drug CGMP, which requires its accuracy, sensitivity, specificity and reproducibility to be established and documented.

Do laboratory-developed tests need FDA approval now?

No. With the rule vacated, the premarket review it planned for LDTs no longer applies. A laboratory offering an LDT still needs a CLIA certificate or exemption for the testing it performs, a state permit or licence where states such as New York and California require one, and must establish the test’s performance specifications before reporting results.

Can a laboratory sell its LDT to other laboratories?

Not as an LDT. A test distributed to other laboratories is a product rather than a laboratory’s own service, and the court accepted FDA’s authority over commercially distributed test kits. The kit would be an in vitro diagnostic device under FDA’s rules.

Does CLIA require clinical validation of an LDT?

Not in its validation rule. Section 493.1253 lists analytical characteristics such as accuracy, precision and analytical sensitivity, and CMS has said CLIA does not address the clinical validity of any test. Surveyors review a new test’s analytical validation at the routine survey, after the laboratory has started testing. The Enhancing CLIA Act, introduced in May 2026, would add a clinical validity standard.

Does an LDT report need an FDA disclaimer?

It does when the test was built with analyte specific reagents. FDA’s rule on those reagents requires the report to say that the laboratory developed the test and determined its performance characteristics, and that FDA has not cleared or approved it. The statement is not needed when the results come from the test FDA cleared or approved together with a class II or III reagent.

Does TRICARE cover laboratory-developed tests?

Usually not until FDA approves them, TRICARE says. Its demonstration, running to 18 July 2028, covers tests the Defense Health Agency has reviewed and included, with pre-authorisation from the regional contractor (in the East, on Humana Military’s LDT attestation form) for all but cystic fibrosis screening, and an accredited laboratory. Denials cannot be appealed.

References44 sources

References

  1. 1Medical Devices; Laboratory Developed Tests, final rule, 89 FR 37286, Federal Register, 6 May 2024. https://www.federalregister.gov/documents/2024/05/06/2024-08935/medical-devices-laboratory-developed-tests (read 9 November 2026)
  2. 2Medical Devices; Laboratory Developed Tests; Implementation of Vacatur, final rule, 90 FR 45134, Federal Register, 19 September 2025. https://www.federalregister.gov/documents/2025/09/19/2025-18239/regulation-identification-number-0910-aj05-medical-devices-laboratory-developed-tests-implementation (read 9 November 2026)
  3. 3American Clinical Laboratory Association v. FDA, No. 4:24-⁠CV-⁠479-⁠SDJ, memorandum opinion and order, US District Court for the Eastern District of Texas, 31 March 2025, govinfo. https://www.govinfo.gov/content/pkg/USCOURTS-txed-4_24-cv-00479/pdf/USCOURTS-txed-4_24-cv-00479-0.pdf (read 9 November 2026)
  4. 421 CFR Part 807, Establishment Registration and Device Listing, sections 807.20 and 807.81, eCFR. https://www.ecfr.gov/current/title-21/part-807 (read 9 November 2026)
  5. 521 CFR 860.200, De Novo classification, purpose and applicability, eCFR. https://www.ecfr.gov/current/title-21/section-860.200 (read 9 November 2026)
  6. 621 CFR 814.1, Premarket approval of medical devices, scope, eCFR. https://www.ecfr.gov/current/title-21/section-814.1 (read 9 November 2026)
  7. 721 CFR 862.9, Limitations of exemptions from section 510(k), eCFR. https://www.ecfr.gov/current/title-21/section-862.9 (read 9 November 2026)
  8. 842 CFR 493.3, Applicability, paragraphs (a) and (b)(2), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-A/section-493.3 (read 9 November 2026)
  9. 921 CFR 820.1, Scope, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.1 (read 9 November 2026)
  10. 1042 CFR 493.1253, Standard: Establishment and verification of performance specifications, eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-K/section-493.1253 (read 9 November 2026)
  11. 11State Operations Manual, Appendix C, Survey Procedures and Interpretive Guidelines for Laboratories and Laboratory Services, Rev. 236, CMS, 23 January 2026, D5217, D5421 to D5427 and IQCP under 493.1256(d). https://www.cms.gov/regulations-and-guidance/guidance/manuals/downloads/som107ap_c_lab.pdf (read 9 November 2026)
  12. 1221 CFR 809.10, Labeling for in vitro diagnostic products, paragraphs (a) and (c), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-809/subpart-B/section-809.10 (read 9 November 2026)
  13. 1321 CFR 809.30, Restrictions on the sale, distribution and use of analyte specific reagents, eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-809/subpart-C/section-809.30 (read 9 November 2026)
  14. 1421 CFR 803.1, Medical device reporting, scope, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-803.1 (read 9 November 2026)
  15. 1521 CFR 806.1, Medical devices; reports of corrections and removals, scope, eCFR. https://www.ecfr.gov/current/title-21/section-806.1 (read 9 November 2026)
  16. 16Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026. https://www.fda.gov/media/80195/download (read 9 November 2026)
  17. 1742 CFR 493.1777, Standard: Inspection of laboratories that have requested or have been issued a certificate of compliance, paragraph (b), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-Q/section-493.1777 (read 9 November 2026)
  18. 1842 CFR 493.551, General requirements for laboratories, paragraph (a), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-E/section-493.551 (read 9 November 2026)
  19. 19CLIA Overview and Laboratory Developed Tests (LDTs) Frequently Asked Questions, Centers for Medicare & Medicaid Services, version of 22 October 2013. https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/ldt-and-clia_faqs.pdf (read 9 November 2026)
  20. 20Test Approval, Clinical Laboratory Evaluation Program, New York State Department of Health, Wadsworth Center. https://www.wadsworth.org/regulatory/clep/clinical-labs/obtain-permit/test-approval (read 9 November 2026)
  21. 2142 CFR 493.17, Test categorization, paragraph (c)(4), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-A/section-493.17 (read 9 November 2026)
  22. 2242 CFR 493.1256, Standard: Control procedures, paragraph (d), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-K/section-493.1256 (read 9 November 2026)
  23. 2342 CFR 493.1451, Standard: Technical supervisor responsibilities, paragraph (b)(2), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-M/section-493.1451 (read 9 November 2026)
  24. 2442 CFR 493.1251, Standard: Procedure manual, paragraph (d), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-K/section-493.1251 (read 9 November 2026)
  25. 2542 CFR 493.1445, Standard: Laboratory director responsibilities, paragraph (e)(3), eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-M/section-493.1445 (read 9 November 2026)
  26. 2642 CFR 493.1236, Standard: Evaluation of proficiency testing performance, paragraphs (a) and (c), eCFR. https://www.ecfr.gov/current/title-42/section-493.1236 (read 9 November 2026)
  27. 2742 CFR 493.1105, Standard: Retention requirements, eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-J/section-493.1105 (read 9 November 2026)
  28. 28List of Approved Accreditation Organizations under the Clinical Laboratory Improvement Amendments (CLIA), CMS, version 8 August 2025. https://www.cms.gov/files/document/list-accreditation-organizations-august-2025.pdf (read 9 November 2026)
  29. 2942 CFR 493.1773, Standard: Basic inspection requirements, eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-Q/section-493.1773 (read 9 November 2026)
  30. 3021 CFR 211.165, Testing and release for distribution, paragraph (e), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-I/section-211.165 (read 9 November 2026)
  31. 3121 CFR 211.194, Laboratory records, paragraph (a)(2), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-C/part-211/subpart-J/section-211.194 (read 9 November 2026)
  32. 32New Application, Out-of-⁠State Clinical Laboratory License, Laboratory Field Services, California Department of Public Health, citing Business and Professions Code section 1241(a). https://www.cdph.ca.gov/Programs/OSPHLD/LFS/Pages/New-OOS-ClinicalLaboratoryLicense.aspx (read 9 November 2026)
  33. 33Medical Test Sites (MTS), Washington State Department of Health. https://doh.wa.gov/licenses-permits-and-certificates/facilities-z/medical-test-sites-mts (read 9 November 2026)
  34. 3442 CFR 493.2, Definitions, eCFR. https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-A/section-493.2 (read 9 November 2026)
  35. 35In Vitro Companion Diagnostic Devices, guidance for industry and FDA staff, FDA, August 2014, section VI, Investigational use. https://www.fda.gov/media/81309/download (read 9 November 2026)
  36. 3621 CFR 812.2, Applicability, paragraphs (b)(1), (c)(2) and (c)(3), eCFR. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-812/subpart-A/section-812.2 (read 9 November 2026)
  37. 37In Vitro Companion Diagnostic Devices, final guidance, FDA, August 2014, docket FDA-⁠2011-⁠D-⁠0215. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/in-vitro-companion-diagnostic-devices (read 9 November 2026)
  38. 38Blackfly Investments, LLC dba Molecular Testing Labs, warning letter, MARCS-⁠CMS 680324, CBER, FDA, 11 October 2024. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/blackfly-investments-llc-dba-molecularr-testing-labs-680324-10112024 (read 9 November 2026)
  39. 39Regulation (EU) 2017/746 on in vitro diagnostic medical devices, consolidated text of 10 January 2025, Articles 5(5) and 113(3), EUR-⁠Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:02017R0746-20250110 (read 9 November 2026)
  40. 40General information on the Health Institution Exemption, MHRA, GOV.UK, updated 20 January 2026. https://www.gov.uk/government/publications/health-institution-exemption-for-general-medical-devices/general-information-on-the-health-institution-exemption (read 9 November 2026)
  41. 41H.R. 8890, Enhancing Clinical Laboratory Innovation and Access Act of 2026, as introduced in the House, 119th Congress, govinfo, 19 May 2026. https://www.govinfo.gov/content/pkg/BILLS-119hr8890ih/html/BILLS-119hr8890ih.htm (read 9 November 2026)
  42. 42Bill status, H.R. 8890, 119th Congress, govinfo bulk data, updated 19 September 2026. https://www.govinfo.gov/bulkdata/BILLSTATUS/119/hr/BILLSTATUS-119hr8890.xml (read 9 November 2026)
  43. 43Request for Information; Clinical Laboratory Improvement Amendments of 1988 (CLIA) Regulations, CMS-⁠3485-⁠NC, 91 FR 43586, Federal Register, 16 July 2026. https://www.federalregister.gov/documents/2026/07/16/2026-14358/request-for-information-clinical-laboratory-improvement-amendments-of-1988-clia-regulations (read 9 November 2026)
  44. 44Clinical Laboratory Improvement Amendments of 1988 (CLIA); Virtual Access, Gynecologic Cytology Proficiency Testing (PT), Personnel Qualifications, and Other Changes, proposed rule, CMS-⁠3478-⁠P, 91 FR 64480, Federal Register, 8 October 2026. https://www.federalregister.gov/documents/2026/10/08/2026-20613/clinical-laboratory-improvement-amendments-of-1988-clia-virtual-access-gynecologic-cytology (read 9 November 2026)