Skip to main content

What were the most common FDA 483 observations in fiscal year 2025?

In fiscal year 2025, FDA’s most cited drug observation was 21 CFR 211.22(d), quality unit procedures not written or not followed (243 citations), then 211.192, discrepancies and failures not thoroughly investigated (236). For devices it was 820.100(a), CAPA procedures (279), then 820.198(a), complaint procedures (243). The counts are FDA’s, from 483s its electronic tools produced. Under the QMSR, the single device clause cited most in FY2026 inspections classified so far is ISO 13485 clause 7.1, risk management.Checked against its sources on 26 November 2026. Latest change: The device rows and the ISO 13485 clause table now link the new answers on 820.20, 820.22, 820.30, 820.50, 820.70, 820.75, 820.90, 820.100 and 820.198. How these answers are made

1The ten most cited drug observations in fiscal year 2025.

FDA counts how often each regulation was cited on the Form 483s its electronic inspection tools produce. In fiscal year 2025, inspections ending 1 October 2024 to 30 September 2025, those tools produced 713 drug 483s, which cited Part 211 2,608 times.1 These are the ten references cited most, each count summed over that reference’s standard citations:

Table 1. Part 211 citations on drug 483s from FDA’s electronic inspection tools, by fiscal year. FDA’s standard text offers the bracketed options; the investigator picks one or more and adds “Specifically” and the facts.1
Cited asFDA’s standard wording, main citationFY2023FY2024FY2025The fix
211.22(d)“The responsibilities and procedures applicable to the quality control unit are not [in writing] [fully followed].”152201243Quality unit
211.192“There is a failure to thoroughly review [any unexplained discrepancy] [the failure of a batch or any of its components to meet any of its specifications] …”175173236Investigations
211.100(a)“Your firm failed to establish [adequate] written procedures for production and process controls …”98134169Production procedures
211.160(b)“Laboratory controls do not include the establishment of scientifically sound and appropriate [specifications] [standards] [sampling plans] [test procedures] …”84119121Laboratory controls
211.68(b)“Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.”8180110Computer controls
211.113(b)“Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not [established] [written] [followed].”94108105Sterile processes
211.67(a)“Equipment and utensils are not [cleaned] [maintained] [sanitized] at appropriate intervals …”687695Equipment cleaning
211.84(d)(2)“Reports of analysis from component suppliers are accepted in lieu of testing each component …”486990Component testing
211.63“Equipment used in the manufacture, processing, packing or holding of drug products is not [of appropriate design] [of adequate size] [suitably located] …”748581IQ, OQ and PQ
211.67(b)“Written procedures are not [established] [followed] for the cleaning and maintenance of equipment …”447078Equipment cleaning
Drug 483sIssued from FDA’s electronic tools510561713

The top two have held for three years, at 34.1 and 33.1 citations per 100 drug 483s in FY2025.1 Under 211.22(d), procedures not followed are cited as often as procedures not written. The Data Dashboard keeps the option each investigator chose, and of its 222 FY2025 citations of 211.22(d), 178 say the procedures were not “fully followed”, 97 of them also that they were “not in writing”, and 41 only that they were not in writing.2 Keeping both options is also what an unedited citation says, so the 97 may not all mean both were found.

The FDA inspection readiness checklist turns the 15 most cited into checks to run on a site’s own records before an inspection.

2The ten most cited device observations in fiscal year 2025.

FY2025 is the last full year of the Quality System regulation. Its 791 device 483s cited Part 820 2,318 times:1

Table 2. Part 820 citations on device 483s from FDA’s electronic inspection tools, by fiscal year. 820.30(g) and 820.50 tie for third in FY2025; 820.30(g) is summed over ten standard citations, of which risk analysis is the largest.1
Cited asFDA’s standard wording, main citationFY2023FY2024FY2025The fix
820.100(a)“Procedures for corrective and preventive action have not been [adequately] established.”25225427921 CFR 820.100, CAPA
820.198(a)“Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been [adequately] established.”20223024321 CFR 820.198, complaints
820.30(g)“Risk analysis [was not performed] [is inadequate] [is incomplete].”688911521 CFR 820.30, design and risk
820.50“Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been [adequately] established.”848211521 CFR 820.50, purchasing
820.75(a)“A process whose results cannot be fully verified by subsequent inspection and test has not been [adequately] validated according to established procedures.”9010111421 CFR 820.75, process validation
820.90(a)“Procedures have not been [adequately] established to control product that does not conform to specified requirements.”11010811321 CFR 820.90, nonconforming product
820.22“Procedures for quality audits have not been [adequately] established.”1089310721 CFR 820.22, audits
820.184“Procedures for device history records have not been [adequately] established.”1129588DHF, DMR and DHR
820.20(c)“Procedures for management review have not been [adequately] established.”66587121 CFR 820.20, management
820.25(b)“Procedures for training and identifying training needs have not been [adequately] established.”666068Training
Device 483sIssued from FDA’s electronic tools720672791

Since 2 February 2026 none of these sections exists: Part 820 is now the Quality Management System Regulation (QMSR), which incorporates ISO 13485:2016 and adds a few requirements of its own, such as the content of complaint records in 820.35(a).3,4 See the QMSR and ISO 13485 vs 21 CFR 820 for the changeover.

3How the lists moved from FY2023 to FY2025.

Drug 483s from FDA’s electronic tools rose 40%, from 510 to 713, while their Part 211 citations rose 32%, from 1,983 to 2,608: a 483 carried 3.9 Part 211 citations on average in FY2023 and 3.7 in FY2025.1 So compare rates per 100 483s, not raw counts:

Table 3. Drug citations that moved most, and the steady second, per 100 drug 483s from FDA’s electronic tools (510 in FY2023, 713 in FY2025).1
Cited asSubjectFY2023FY2025Per 100 drug 483s, FY2023 to FY2025
211.84(d)(2)Components accepted on suppliers’ reports48909.4 to 12.6
211.22(d)Quality unit procedures15224329.8 to 34.1
211.100(a)Production and process control procedures9816919.2 to 23.7
211.67(b)Cleaning and maintenance procedures44788.6 to 10.9
211.192Investigations17523634.3 to 33.1
211.113(b)Sterile product procedures and validation9410518.4 to 14.7
211.63Equipment design, size and location748114.5 to 11.4
211.110(a)In-process control procedures524810.2 to 6.7
211.160(a)Laboratory controls approved, followed and documented41278.0 to 3.8

Device 483s rose from 720 to 791. The largest rise is in design validation, 820.30(g), from 9.4 to 14.5 citations per 100, and risk analysis is about half of it: its two standard citations went from 27 to 50 of the paragraph’s 68 and 115.1 Purchasing went the other way. Citations of 820.50 for missing purchasing procedures rose from 11.7 to 14.5 per 100, but the section as a whole fell from 27.2 to 24.0, because supplier evaluation under 820.50(a)(1) went from 34 citations to 6. Device history records, 820.184, fell from 17.1 to 11.5.1

Table 4. Device citations that moved most, with purchasing as a whole, per 100 device 483s from FDA’s electronic tools (720 in FY2023, 791 in FY2025).1
Cited asSubjectFY2023FY2025Per 100 device 483s, FY2023 to FY2025
820.30(g)Design validation, with risk analysis681159.4 to 14.5
Cite 3678Risk analysis not performed, inadequate or incomplete26473.6 to 5.9
820.50Purchasing procedures, cited to the section8411511.7 to 14.5
820.50, all paragraphsPurchasing controls as a whole19619027.2 to 24.0
820.50(a)(1)Supplier evaluation and its records3464.7 to 0.8
820.198(a)Complaint procedures20224328.1 to 30.7
820.184, all paragraphsDevice history records, UDI included1239117.1 to 11.5
820.100(b)CAPA documentation846311.7 to 8.0
820.72(a)Inspection, measuring and test equipment60398.3 to 4.9

4FY2026 in the dashboard: the first device citations under the QMSR.

FDA has not yet posted the FY2026 spreadsheet. Its Data Dashboard already holds FY2026’s citations for inspections whose classification is final, and under the QMSR investigators cite the clause of ISO 13485 that was not met.2,4 As downloaded on 26 November 2026, it holds 596 FY2026 device inspections with citations, whose 483s cite ISO 13485 clauses 1,178 times and Part 820 sections 546 times.2 FY2026 ran from 1 October 2025 to 30 September 2026, beginning four months before the change, and most of those Part 820 citations come from inspections that ended before it.2,3

Table 5. ISO 13485:2016 clauses cited on device 483s in FY2026 (inspections ending 1 October 2025 to 30 September 2026) whose classification was final when the Data Dashboard was downloaded on 26 November 2026. The last column is the old section closest in subject, from the text of Part 820 as it stood on 1 February 2026 and the QMSR rule.2,5,3
Cited asFDA’s short description, main citationFY2026Nearest Quality System regulation section
ISO 13485 clause 7.1“Product realization risk management processes not documented”128Risk analysis in design validation, 820.30(g), the only explicit risk requirement
ISO 13485 clause 8.5.2“Corrective action not taken”107820.100, CAPA
ISO 13485 clause 8.2.2“Complaint handling procedures not documented”99820.198, complaint files
ISO 13485 clause 7.4.1“Procedures for purchasing process not documented”78820.50, purchasing controls
ISO 13485 clause 7.5.6“Process for production and service provision not validated”64820.75, process validation
ISO 13485 clause 8.3.1“Procedure for nonconforming product not documented”51820.90, nonconforming product
ISO 13485 clause 4.1.2“Risk-based approach not applied to QMS processes”44No explicit counterpart
ISO 13485 clause 7.3.9“Procedures to control design and development changes not documented”40820.30(i), design changes
ISO 13485 clause 4.1.5“Inadequate monitoring and control over outsourced processes”38820.50, which covered purchased services
ISO 13485 clause 7.5.1“Production and service provision not adequate”37820.70, production and process controls

The single clause cited most is risk management in product realization, ahead of corrective action and complaints. Summed to two-level clauses the order changes: monitoring and measurement, 8.2, has 167 citations, design and development, 7.3, has 138 and production, 7.5, has 131, against 128 for 7.1, which has no subclauses.2 In the QMSR rule FDA wrote that the QS regulation “explicitly addressed risk management activities in the former” 820.30(g), and that the QMSR “incorporates risk management throughout its requirements”, naming clauses 4.1 and 7.1 among others.3 So risk analysis, about half of the largest rise among device citations before the change, is the single clause cited most after it.1,2

For drugs, investigations lead in FY2026, and sterile processes, 211.113(b), have climbed from eighth in the dashboard’s FY2025 to fifth:

Table 6. Part 211 citations in the Data Dashboard, final classifications only, from 639 drug inspections with citations in FY2025 and 415 in FY2026 to date.2
Cited asSubjectFY2025FY2026
211.192Investigations213146
211.22(d)Quality unit procedures222130
211.160(b)Laboratory controls11281
211.100(a)Production and process control procedures15579
211.113(b)Sterile product procedures and validation7974

Read the order, not the totals. The dashboard leaves out inspections still waiting for a final enforcement action, so the inspections that end worst may be missing longest, and a year’s counts keep growing as classifications are made final.2

5What the counts include and leave out.

  • Only 483s from FDA’s electronic tools. FDA says the spreadsheets “are not a comprehensive listing of all inspectional observations”, and that “Not all Form FDA 483s are generated by these tools as some 483s are manually prepared.” Citations on a manually prepared 483, a handwritten one for example, are in neither the spreadsheets nor the dashboard.1,2
  • Citations, not final findings. Each count is the number of times a reference was cited as an observation, which an investigator lists when, in their judgment, a product may be in violation. FDA notes that the citations “may not necessarily represent citations on final classification letters” (see Form 483 and NAI, VAI and OAI).1,2
  • By the inspection’s end date. FY2025 is inspections ending 1 October 2024 to 30 September 2025.1
  • Summed by reference. A reference such as 211.192 has several standard citations; this page adds them up. Lists that rank single citations show lower numbers: 211.192’s main citation was cited 164 times in FY2025, and 820.198(a)’s 211.1
  • Programs counted apart. A 483 that cites two program areas counts in both, which is why FY2025’s areas add up to 4,912 483s against 4,862 issued from the system.1
  • The dashboard is narrower. It posts citations only when every classification of the inspection is final, leaves out state, pre-approval, mammography and nonclinical laboratory inspections and those awaiting final enforcement action, and is updated weekly. Its FY2025 count for 211.22(d) is 222, against the spreadsheet’s 243.2
  • No firm is named here. To see one company’s 483s, see looking up a firm’s 483s and warning letters; to answer one, responding to a 483.

6How it runs in Seal.

The inspection readiness blueprint turns each Form 483 observation into commitments: the cited requirement, the response, interim controls, CAPA, owners and verification stay connected, and an action is not treated as closed because a response letter was sent. For the corrective actions themselves, the CAPA blueprint keeps the source deviation, investigation, actions, controlled change and follow-up evidence in one record, and closure can be configured to require the approved effectiveness review.

Neil can assemble the evidence behind a specific question, such as whether every investigation closed this year extended to other batches, and make missing evidence explicit; your audit owner checks scope and completeness before anything is shared.

AQuestions and answers

Why do other FY2025 lists give 211.192 as 164?

They rank FDA’s single standard citations. 211.192 has five, and the main one, the failure to thoroughly review a discrepancy or failure, was cited 164 times. This page adds the five together, 236, because each is a citation of the same section. The same applies to 820.198(a): 211 for the main citation, 243 for the reference.

When will FDA publish the FY2026 numbers?

FDA had not posted the FY2026 spreadsheet on 26 November 2026. Its Inspection Observations page, which lists one file for each fiscal year from 2006 to 2025, was last updated on 16 December 2025. When FY2026’s file appears, this page moves to FY2026 at the same address. Until then, the Data Dashboard’s FY2026 citations, final classifications only, are the earliest view.

Is a 483 citation the same as a violation?

No. A 483 lists conditions an investigator judged may violate the requirements, and FDA decides afterwards, weighing the firm’s response, how to classify the inspection and whether to act. FDA notes that 483 citations may not match the citations on a final classification letter.

Do these lists include warning letters?

No. They count citations on Form 483s. A warning letter is a later, separate step that FDA takes for significant violations, and it can cite sections the 483 did not.

Can I see which companies were cited for a section?

The Data Dashboard’s citations dataset names the firm for each inspection, and you can search a firm by name or FEI number there. This page counts sections only. For checking one supplier or contract manufacturer, Seal’s answer on looking up FDA 483s and warning letters walks through the sources.

References5 sources

References

  1. 1Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary, Drugs and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  2. 2FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)
  3. 3Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, response on risk management, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  4. 421 CFR Part 820, Quality Management System Regulation, sections 820.10 and 820.35, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  5. 521 CFR Part 820, Quality System Regulation, sections 820.30, 820.50, 820.70, 820.75, 820.90, 820.100 and 820.198, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)

See your process in Seal.

Bring a procedure or a recurring problem. See how your team configures it with Neil and improves the next run.