What did 21 CFR 820.20 require of management, and what replaced it?

21 CFR 820.20 made management with executive responsibility answerable for the quality system: set the quality policy and objectives, build an organization with the authority and resources to meet Part 820, appoint a management representative, review the system at defined intervals, and plan for quality. The QMSR puts it, from 2 February 2026, in ISO 13485 clauses 4.1 and 5, owned by top management. Management review drew 195 of its 315 citations in fiscal years 2023 to 2025.Checked against its sources on 26 November 2026. How these answers are made

1Management responsibility in the old Part 820.

Section 820.20, “Management responsibility”, put the quality system on the people who run the company. Its most cited paragraph, management review, read:1

“(c) Management review. Management with executive responsibility shall review the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency according to established procedures to ensure that the quality system satisfies the requirements of this part and the manufacturer’s established quality policy and objectives. The dates and results of quality system reviews shall be documented.” 21 CFR 820.20(c), eCFR as of 1 February 20261
Table 1. 21 CFR 820.20(a), (b), (d) and (e), from the eCFR as in effect on 1 February 2026.1
ParagraphWhat it required, in the regulation’s words
(a) Quality policyManagement “shall establish its policy and objectives for, and commitment to, quality” and ensure the policy “is understood, implemented, and maintained at all levels of the organization”
(b) Organization“an adequate organizational structure to ensure that devices are designed and produced in accordance with the requirements of this part”, with responsibility, authority, independence and “adequate resources, including the assignment of trained personnel”
(b)(3) Management representativeA documented appointment of “a member of management who, irrespective of other responsibilities” ensures the requirements are “effectively established and effectively maintained” and reports on performance
(d) Quality planning“a quality plan which defines the quality practices, resources, and activities relevant to devices that are designed and manufactured”
(e) Procedures“quality system procedures and instructions”, with “an outline of the structure of the documentation” where appropriate

2Management review drew most of the citations.

Table 2. 820.20 citations by fiscal year, ending 30 September. The other five standard citations, on the quality policy, quality plan and the representative’s duties, drew 11 to 16 a year between them. Handwritten 483s are not in these counts, and a citation is an observation, not FDA’s final finding.2
Cited asFDA’s standard citationFY2023FY2024FY2025
820.20, all paragraphsAll eleven standard citations94102119
Cite 3666“Procedures for management review have not been [adequately] established.”323149
Cite 3669“Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system [at defined intervals] [with sufficient frequency].”282618
Cite 3668“The results and/or dates of management reviews are not documented.”614
Cite 2302“Quality system procedures and instructions have not been established.”5814
Cite 419“The organizational structure has not been [adequately] established and maintained to ensure that devices are [designed] [produced] in accordance with 21 CFR 820.”6811
Cite 3665“No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.”61210
Device 483sAll of them, for scale720672791

Management responsibility rose from eleventh to eighth among device sections, from 13.1 to 15.0 citations per 100 device 483s.2 Management review, paragraph (c), drew 195 of its 315 citations over the three years, and 820.20(c) was the ninth most cited device reference in FY2025.2

3Review procedures, more than missed reviews.

Citations that management had not reviewed the system at defined intervals fell from 28 to 18, while citations of the procedure rose from 32 to 49.2 In FDA’s Data Dashboard, which counts an inspection once its classification is final, 42 of the 64 FY2025 citations of 820.20(c) were for procedures, 16 of them dropping “adequately”, and 18 for reviews not held at defined intervals or often enough, some keeping both options.3 Under 820.180(c) investigators could not read the review itself, only ask for the procedure and a written certification that reviews had been done, so the procedure was most of what they could examine.4

The citation that “Quality system procedures and instructions have not been established” rose from 5 to 14, from 0.7 to 1.8 per 100 device 483s. Its text has no options, so investigators use it for any gap under 820.20(e), and it doesn’t show how much was missing.2

The citation that “No management representative had been appointed” was used 28 times in three years.2

4Top management, and the new clause 4.1.2.

The Quality Management System Regulation (QMSR), which has incorporated ISO 13485:2016 into Part 820 since 2 February 2026, renamed the people.5 “Management with executive responsibility” became ISO 9000’s “top management”. FDA said the change “does not change that FDA expects medical device manufacturers, led by individuals with executive responsibilities, to embrace a culture of quality”, and that “quality cannot be inspected or tested into products or services.”6 FDA’s compliance program puts these requirements in its management oversight area, whose purpose includes that top management “Uses risk management and risk-based decision making effectively in the QMS”, and lists management review against clauses 5.6.1 to 5.6.3 and a risk-based approach against clause 4.1.2 b).7

Table 3. Citations in FDA’s Data Dashboard, inspections classified as final by 26 November 2026. Its FY2026 column, 1 October 2025 to 30 September 2026, covers both rules and is not yet complete.3
Cited asSubjectFY2025FY2026
820.20(c)Management review, the old paragraph6419
ISO 13485 clause 4.1.2Risk-based approach to the QMS processes044
ISO 13485 clause 5.6.1Management review026
ISO 13485 clause 4.1.1QMS documented and maintained022
ISO 13485 clause 5.1Management commitment05

The new citation is clause 4.1.2, the seventh most cited single clause in FY2026’s classified inspections: “A risk-based approach to the control of the appropriate processes needed for the quality management system has not been [adequately] applied”, 28 times with “adequately” and 16 without.3 Nothing in 820.20 asked for that. Clause 5.6.1’s citations keep the old pair, procedures not documented (13) and reviews not held “at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness” (10).3 The review’s report can now be read too: FDA did not keep 820.180(c), and what FDA can now see covers what that means for the record.6

5Updating the management procedures.

Table 4. From the text of 820.20, Compliance Program 7382.850 Attachment A, the QMSR final rule’s responses to comments 27 and 55, and FDA’s standard citations of the clauses.1,7,6,3
820.20Under the QMSRWhat to change
Management with executive responsibilityTop managementRename it; name who that is at your company
(a) Policy and objectivesClauses 5.3, 5.4.1Little; make the objectives measurable and reviewed
(b) Organization, responsibility, resourcesClauses 5.5.1, 6.1, 6.2Keep the organization chart, and the independence and authority the old text required
(b)(3) Management representativeClause 5.5.2Keep the documented appointment
(c) Management reviewClauses 5.6.1 to 5.6.3Keep the defined intervals, and add what 820.20(c) never listed: the inputs clause 5.6.2 requires and the outputs clause 5.6.3 says to record. Write the record knowing FDA may read it
(d) Quality planClauses 5.4.2, 7.1Plan the QMS and each product’s realization
(e) Procedures and their structureClauses 4.1.1, 4.2Keep them; add the quality manual (clause 4.2.2) if you had none
Not in 820.20Clause 4.1.2 b)Add a documented risk-based approach to controlling each QMS process, and show where it set a control

6Example: a risk-based approach to the QMS processes.

Invented for illustration. A device maker scores each QMS process for its effect on device safety and its chance of failure, each from 1 to 3, and lets the product set the controls.

Table 5. Example. A risk-based approach to QMS processes, invented for illustration; a score of 6 or more sets the highest controls.
QMS processEffectChanceScoreControls the score sets
Outsourced sterilization326Monthly metrics at management review; audited twice a year; quality approves every change
Complaint handling326Monthly metrics; audited twice a year; quality approves every change
Document control224Quarterly metrics; audited once a year
Training224Quarterly metrics; audited once a year
Warehouse and shipping212Metrics at the annual review; audited once a year

The scores: 3 × 2 = 6, 2 × 2 = 4 and 2 × 1 = 2. The record shows the approach was applied, which is what FDA’s 44 clause 4.1.2 citations found missing or inadequate: each score sets the audit frequency, the review cadence and who approves changes.3

7What top management keeps on record.

Top management needs, on record: the quality policy and measurable objectives, the organization with responsibilities and authorities, the management representative’s appointment, a management review procedure with its interval, inputs and outputs, the QMS plan, the quality manual and procedure structure, and the risk-based approach to the QMS processes. These answers hold templates for most of it:

8How it runs in Seal.

In Seal the records a management review draws on are the quality records themselves. The QMS calculates quality KPIs such as deviation rates, CAPA aging and training status from the underlying records. The audit program supplies audit and inspection results and follow-up still open from earlier reviews, so the review reads them from the program rather than from a summary assembled for the meeting.

Automated reporting runs each recurring report against a defined population, time window and data cut, so a reviewer can trace a figure to the records behind it. A signed report stays fixed as later data arrives, and agreed findings become assigned follow-up linked to the issued report.

AQuestions and answers

Is 21 CFR 820.20 still in effect?

No. Since 2 February 2026 management responsibility has been ISO 13485 clause 5 under the QMSR, with the general QMS requirements, the risk-based approach among them, in clause 4.1. 820.20 still appears in FY2026’s citations because many of that year’s inspections ended before the change.

Who is top management?

The person or group who directs and controls the manufacturer at the highest level, as ISO 9000 defines it. FDA adopted that definition in place of management with executive responsibility and said it expects these executives to lead a culture of quality.

Do we still need a management representative?

Yes. ISO 13485 clause 5.5.2 keeps a management representative appointed by top management, with authority over the QMS and its reporting. Keep the documented appointment the old 820.20(b)(3) asked for.

Can FDA read our management review minutes now?

Yes. The old 21 CFR 820.180(c) exception, which let a manager certify that reviews were held instead of showing them, did not carry into the QMSR, so FDA may read management review records.

What is the risk-based approach FDA now cites?

ISO 13485 clause 4.1.2 asks for a risk-based approach to controlling the processes of the quality system. FDA cited it 44 times in FY2026 inspections classified so far, finding that a risk-based approach to controlling the processes needed for the QMS had not been applied, or not adequately.

References7 sources

References

  1. 121 CFR 820.20, Management responsibility, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.20 (read 26 November 2026)
  2. 2Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  3. 3FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)
  4. 421 CFR 820.180, General requirements (records), paragraph (c), eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.180 (read 26 November 2026)
  5. 521 CFR 820.10, Requirements for a quality management system, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.10 (read 26 November 2026)
  6. 6Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, responses to comments 27 and 55, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  7. 7Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)