What did 21 CFR 820.30 require for design controls, and what replaced it?

21 CFR 820.30 required makers of class II and III devices, and some class I devices, to control design: plans, inputs, outputs, reviews, verification, validation with risk analysis, transfer and changes, recorded in a design history file. Under the QMSR, from 2 February 2026, it is ISO 13485 clause 7.3, with risk management in clause 7.1. Design validation made all the rise in its 839 citations in fiscal years 2023 to 2025, risk analysis about half.Checked against its sources on 26 November 2026. How these answers are made

1Design controls as Part 820 wrote them.

Section 820.30, “Design controls”, opened with its scope:1

“Each manufacturer of any class III or class II device, and the class I devices listed in paragraph (a)(2) of this section, shall establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met.” 21 CFR 820.30(a)(1), eCFR as of 1 February 20261

The class I devices were those “automated with computer software” and five listed types, among them tracheobronchial suction catheters and surgeon’s gloves.1 The paragraph cited most, design validation, read in full:

“(g) Design validation. Each manufacturer shall establish and maintain procedures for validating the device design. Design validation shall be performed under defined operating conditions on initial production units, lots, or batches, or their equivalents. Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation, shall be documented in the DHF.” 21 CFR 820.30(g)1
Table 1. 21 CFR 820.30(b) to (j), from the eCFR as in effect on 1 February 2026.1
ParagraphWhat it required, in the regulation’s words
(b) PlanningPlans “that describe or reference the design and development activities and define responsibility for implementation”, “reviewed, updated, and approved as design and development evolves”
(c) InputRequirements that “address the intended use of the device, including the needs of the user and patient”, with “a mechanism for addressing incomplete, ambiguous, or conflicting requirements”, approved with date and signature
(d) OutputOutput defined “in terms that allow an adequate evaluation of conformance to design input requirements”, with the outputs “essential for the proper functioning of the device” identified, approved before release
(e) ReviewFormal reviews at appropriate stages, including “an individual(s) who does not have direct responsibility for the design stage being reviewed”
(f) VerificationConfirmation “that the design output meets the design input requirements”, recorded in the design history file (DHF)
(h) TransferProcedures to ensure “the device design is correctly translated into production specifications”
(i) Changes“identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation”
(j) DHFA DHF for each type of device, with “the records necessary to demonstrate that the design was developed in accordance with the approved design plan and the requirements of this part”

2All of the rise was in design validation.

Design controls were FDA’s third most cited device section in FY2023 and its second in FY2024 and FY2025:2

Table 2. 820.30 on FDA’s electronically generated device 483s, by the fiscal year the inspection ended. The counts leave out handwritten 483s, and each citation is an investigator’s observation rather than FDA’s finding.2
Cited asSubjectFY2023FY2024FY2025
820.30, all paragraphsAll 27 standard citations250274315
820.30(g)Design validation, including risk analysis6889115
820.30(i)Design change procedures464648
820.30(a)No design control procedures332840
820.30(j)Design history file263430
820.30(f)Design verification242624
820.30(c)Design input111218
820.30(e)Design review14813
820.30(h)Design transfer121112
820.30(d)Design output6138
820.30(b)Design and development planning1077
Device 483sIssued from FDA’s electronic tools720672791

Per 100 device 483s, 820.30 rose from 34.7 to 39.8 citations, and the whole of that rise was paragraph (g), from 9.4 to 14.5; the other nine paragraphs together stood at 25.3 per 100 in both FY2023 and FY2025.2

3Risk analysis, change procedures and missing design control.

Risk analysis is the largest part of design validation. Paragraph (g) has ten standard citations, and the most used is not about validation testing at all:2

Table 3. Standard citations of 820.30(g): the four most used, and the second citation for risk analysis, by fiscal year.2
Cited asFDA’s standard citationFY2023FY2024FY2025
Cite 3678“Risk analysis [was not performed] [is inadequate] [is incomplete].”264247
Cite 3101“Procedures for design validation have not been [adequately] established.”191732
Cite 3192“Design validation did not ensure the device conforms to defined user needs and intended uses.”4411
Cite 3677“Validation of device software [was not performed] [is inadequate] [is incomplete].”798
Cite 14718“Results of the design risk analysis were not [adequately] documented.”143

The two risk analysis citations rose from 27 to 50, from 3.8 to 6.3 per 100 device 483s, and made up 50 of the 115 citations of paragraph (g) in FY2025: about half of the paragraph’s rise per 100 483s. Design validation procedures (cite 3101) and user needs (cite 3192) made up most of the rest.2 The bracket shows the kind of failure. In FDA’s Data Dashboard, 30 FY2025 citations found the risk analysis inadequate, 6 of them also incomplete, and 3 found that “Risk analysis was not performed.”3 The usual finding is a risk analysis that exists and falls short, with the facts after “Specifically” on the 483 saying where.

Design changes are cited through their procedure. All 48 FY2025 citations of 820.30(i) use one standard citation, “Procedures for design change have not been [adequately] established”. In the dashboard 38 of 43 keep “adequately”, which may be the text left unedited, so the facts after “Specifically” say whether a procedure existed.2,3

Paragraph (a) has one standard citation. It reads “Procedures for design control have not been established”, with no “adequately” option, so its 40 FY2025 citations cover missing and inadequate design control procedures alike.2 Of the 27 dashboard citations of the DHF in FY2025, at least 12 found that it “was not established”, apart from 2 that keep every option.3

4Under the QMSR: clause 7.3, and risk in clause 7.1.

Since 2 February 2026 Part 820 is the Quality Management System Regulation (QMSR): it incorporates ISO 13485:2016 by reference, and 820.10(c) keeps the old scope, requiring makers of class II, class III and the same class I devices to comply with “Design and Development, Clause 7.3 and its Subclauses in ISO 13485”.4 FDA’s compliance program for device inspections lists the design and development elements against clauses 7.3.1 to 7.3.10, software validation and validation both under clause 7.3.7.5

Risk analysis did not move to clause 7.3. FDA wrote that the QS regulation “explicitly addressed risk management activities in the former” 820.30(g), and that the QMSR “incorporates risk management throughout its requirements”, naming clauses 4.1, 7.1, 7.3, 7.4, 7.5, 7.6 and 8.2.6 Investigators now cite it as clause 7.1, risk management in product realization, and it is the single ISO 13485 clause FDA cites most:3

Table 4. Citations in FDA’s Data Dashboard, inspections classified as final by 26 November 2026. FY2026 ran from 1 October 2025 to 30 September 2026, so it mixes inspections before and after the change, and its counts grow as its inspections are classified. Clauses 7.3.2, 7.3.4 and 7.3.5 were cited 5, 5 and 2 times.3
Cited asSubjectFY2025FY2026
820.30, all paragraphsDesign controls, the old section27778
ISO 13485 clause 7.1Planning of product realization, including risk management0128
ISO 13485 clause 7.3.9Control of design and development changes040
ISO 13485 clause 7.3.7Design and development validation023
ISO 13485 clause 7.3.10Design and development files020
ISO 13485 clause 7.3.1Design and development, general013
ISO 13485 clause 7.3.3Design and development inputs011
ISO 13485 clause 7.3.6Design and development verification011
ISO 13485 clause 7.3.8Design and development transfer08

FDA’s standard citations of clause 7.1 are for the process and for its records. At least 68 of the 128 say “One or more processes for risk management in product realization” have not been adequately documented (55) or not documented at all (13), and at least 37 that “Records of risk management activities in product realization” were not adequately maintained (20) or not maintained (17).3 Clause 7.3.9’s citations follow 820.30(i): procedures to control changes not adequately documented (13), and changes “not adequately reviewed, verified, validated, and approved before implementation” (9).3

Both failures are on FDA’s list of serious deficiencies that should lead to an initial classification of Official Action Indicated: “Failure to establish, implement, and/or maintain one or more processes for risk management in product realization”, and “Failure to control the design and development of product, including not adequately evaluating changes for risk and impact on product(s) prior to implementation.”5

5Re-citing a design control procedure.

FDA declined to map the old regulation to the new one, calling a direct comparison “cumbersome and not a useful tool”;6 the clause for each paragraph below is the one FDA’s compliance program lists for the same element.5

Table 5. From the text of 820.30 and 820.10(c), Compliance Program 7382.850 Attachment A, the QMSR final rule’s responses to comments 19 and 31, and FDA’s standard citations of the clauses.1,4,5,6,3
820.30Clause nowWhat to change
(a) Scope820.10(c), 7.3.1Cite 820.10(c) for which devices are covered; the list is unchanged
(b) to (h)7.3.2 to 7.3.8Re-cite each step to its clause. The steps stay, and FDA’s clause 7.3.7 citations add two conditions for validation: done to planned and documented arrangements, and completed before the product is released for use
(g) Risk analysis7.1, and risk throughoutTake risk management out of the validation step and make it a product realization process of its own, from design input through production and postmarket data, with its own records; FDA expects risk documentation reviewed and updated when a manufacturer learns of “previously unforeseen risks or new information”6
(g) Software validation7.3.7Keep it inside design validation
(i) Changes7.3.9Make the record show each of the four steps FDA’s citation names: reviewed, verified, validated, approved, before implementation
(j) DHF7.3.10Rename it the design and development file; FDA says it must “contain or reference all the records necessary to establish compliance with design and development requirements”6

FDA did not incorporate ISO 14971, the risk management standard, but “acknowledges that these other standards may be helpful in understanding application of ISO 13485”.6 DHF vs DMR vs DHR maps the old records to the new ones, with a template.

6Example: one design change, reviewed, verified, validated and approved.

Invented for illustration. A maker of a class II powered breast pump must replace its vacuum motor, which the supplier has discontinued.7

Table 6. Example. A design change record, invented for illustration.
StepRecord entryClause
IdentifyDC-⁠26-⁠014: motor M-⁠1 replaced by M-⁠2; affected outputs: motor specification, firmware drive table, bill of materials, assembly instruction7.3.9
ReviewDesign review with engineering, quality, manufacturing, regulatory and a reviewer outside the project; effect on pumps already in the field assessed: none, as M-⁠2 is fitted only to new builds7.3.9, 7.3.5
510(k)Assessed under 21 CFR 807.81(a)(3):8 no new 510(k), as the change does not significantly affect safety or effectiveness and the specifications are verified unchanged; rationale filed with the change7.3.9
RiskRisk file updated: M-⁠2’s higher stall torque could raise vacuum above the 250 mmHg limit if the firmware limit failed; the relief valve, a separate control, is unchanged; both to be verified7.1
Verify30 pumps on M-⁠2: maximum vacuum 238 to 246 mmHg against the 250 limit; relief valve opens at 252 to 258 against a 260 limit; noise 44 to 47 dBA against 50; 500-hour life test on 6 pumps with no failure7.3.9, 7.3.6
ValidateNot repeated: the user interface, settings and instructions are unchanged, and the verification covers every changed output; rationale recorded and approved7.3.9, 7.3.7
ApproveApproved by the design owner and quality before the first M-⁠2 build; effective from serial 26-⁠041007.3.9
FileChange record, test reports and updated risk file added to the design and development file7.3.10

The margins: 250 − 246 = 4 mmHg below the vacuum limit at worst, and 260 − 258 = 2 mmHg below the relief valve’s limit. A decision to verify without repeating validation was allowed by the old text’s “validation or where appropriate verification”, and clause 7.3.9 likewise asks for changes to be verified or validated as appropriate. Either way the reason is written down; without it, the record invites the clause 7.3.9 citation that changes were “not adequately reviewed, verified, validated, and approved before implementation”.1,3

7The two procedures, and where to start them.

Two procedures now: design and development, covering the plan, inputs, outputs, reviews, verification, validation including software, transfer, changes and the file; and risk management, covering product realization from design input to postmarket data, with its records. Start with these answers; most carry a Word download:

8How it runs in Seal.

For design control in Seal, each design input links to its outputs, verification, validation and reviews, and the traceability matrix is a view of those links. A requirement with no linked test is listed as a gap, and the design file is organized from records kept throughout development rather than assembled before an audit.

The medical device blueprint links requirements to design outputs, verification and risk controls. Its own advice holds here: map the configured quality system to the QMSR, ISO 13485 and the other requirements that apply to your device, because the presence of a workflow does not itself demonstrate conformity.

AQuestions and answers

Does 21 CFR 820.30 still apply?

Only in inspections that ended before 2 February 2026, which is why FY2026’s data still show it. The QMSR replaced it with ISO 13485 clause 7.3 and its subclauses, which 21 CFR 820.10(c) requires of the same devices.

Which class I devices need design controls?

Class I devices automated with computer software, and five listed types: tracheobronchial suction catheters, non-powdered surgeon’s gloves, protective restraints, manual radionuclide applicator systems and radionuclide teletherapy sources. The list in 21 CFR 820.10(c) matches the old one in 820.30(a)(2), except that the gloves are now named as non-powdered.

Is ISO 14971 required under the QMSR?

Not by name. FDA incorporated ISO 13485 and Clause 3 of ISO 9000, not ISO 14971, though it said such standards may help in applying ISO 13485. Risk management itself is required, through clause 7.1 and the other clauses that call for it, and FDA cites clause 7.1 more than any other single clause.

Do designs finished before June 1997 need design controls?

Not retrospectively. FDA’s compliance program says a manufacturer need not apply design controls to stages completed before 1 June 1997, unless the design has changed since. If it has made no design changes since then, or has no design project FDA can review, investigators review only its design change procedures, which it must still have.

What if a contractor designed our device?

You remain responsible. FDA’s compliance program says you must control the contractor under ISO 13485 clauses 4.1.5 and 7.4, and maintain, or have reasonable access to, copies of a design and development file for each device type or family.

References8 sources

References

  1. 121 CFR 820.30, Design controls, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.30 (read 26 November 2026)
  2. 2Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  3. 3FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)
  4. 421 CFR 820.10, Requirements for a quality management system, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.10 (read 26 November 2026)
  5. 5Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Parts III and V and Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)
  6. 6Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, responses to comments 9, 18, 19 and 31, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  7. 721 CFR 884.5160, Powered breast pump, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-884/subpart-F/section-884.5160 (read 26 November 2026)
  8. 821 CFR 807.81(a)(3), When a premarket notification submission is required, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-807/subpart-E/section-807.81 (read 26 November 2026)