What did 21 CFR 820.100 require for CAPA, and what replaced it under the QMSR?

21 CFR 820.100 required CAPA procedures: analyze quality data, investigate causes, identify and implement actions, verify they work without harming the device, share the information and report it to management review, all documented. Since 2 February 2026 the QMSR has replaced it with ISO 13485 clauses 8.5.2 and 8.5.3. It was FDA’s most cited device section, 991 times in fiscal years 2023 to 2025, and clause 8.5.2 107 times in FY2026 inspections classified so far.Checked against its sources on 26 November 2026. How these answers are made

1What 820.100 said before the QMSR.

Section 820.100 was the Quality System regulation’s requirement for corrective and preventive action (CAPA). It read in full:1

“(a) Each manufacturer shall establish and maintain procedures for implementing corrective and preventive action. The procedures shall include requirements for:” 21 CFR 820.100, eCFR as of 1 February 20261
Table 1. 21 CFR 820.100, removed with the rest of the Quality System regulation’s sections on 2 February 2026.1,2
ParagraphThe requirement, in the regulation’s words
(a)(1)“Analyzing processes, work operations, concessions, quality audit reports, quality records, service records, complaints, returned product, and other sources of quality data to identify existing and potential causes of nonconforming product, or other quality problems. Appropriate statistical methodology shall be employed where necessary to detect recurring quality problems”
(a)(2)“Investigating the cause of nonconformities relating to product, processes, and the quality system”
(a)(3)“Identifying the action(s) needed to correct and prevent recurrence of nonconforming product and other quality problems”
(a)(4)“Verifying or validating the corrective and preventive action to ensure that such action is effective and does not adversely affect the finished device”
(a)(5)“Implementing and recording changes in methods and procedures needed to correct and prevent identified quality problems”
(a)(6)“Ensuring that information related to quality problems or nonconforming product is disseminated to those directly responsible for assuring the quality of such product or the prevention of such problems”
(a)(7)“Submitting relevant information on identified quality problems, as well as corrective and preventive actions, for management review”
(b)“All activities required under this section, and their results, shall be documented.”

Since 2 February 2026 Part 820 is the Quality Management System Regulation (QMSR): a manufacturer must document a quality management system that complies with ISO 13485:2016, which it incorporates by reference.3,2 Corrective action is now clause 8.5.2 of ISO 13485 and preventive action clause 8.5.3, and FDA’s compliance program for device inspections lists both, with analysis of data (clause 8.4), under measurement, analysis and improvement.2,4 See the QMSR for the whole changeover.

2The most cited device section, three years running.

Section 820.100 was FDA’s most cited device section in each of its last three full fiscal years, at 43 to 47 citations per 100 device 483s, and 820.100(a) was the single most cited device reference each year:5

Table 2. Citations of 820.100 on device 483s from FDA’s electronic inspection tools, by fiscal year (inspections ending 1 October to 30 September). Manually prepared 483s are not counted, and a citation is an investigator’s observation, not FDA’s final finding.5
Cited asFDA’s standard citationFY2023FY2024FY2025
820.100, all paragraphsBoth standard citations336313342
820.100(a)“Procedures for corrective and preventive action have not been [adequately] established.”252254279
820.100(b)“Corrective and preventive action activities and/or results have not been [adequately] documented.”845963
Device 483sIssued from FDA’s electronic tools720672791

Paragraph (a) had one standard citation for everything, so the facts after “Specifically” carried the detail. The bracket shows what kind of failure it was. Of the 247 FY2025 citations of 820.100(a) in FDA’s Data Dashboard, 210 kept “adequately”: a procedure existed but was not adequate. The other 37 dropped the word, and under the old 820.3(k) “establish” meant define, document and implement, so those cover procedures missing and procedures never put into practice.6 The dashboard holds only inspections whose classification is final, so its counts run below the yearly file’s 279.6,5

3Clause 8.5.2 took its place in FY2026.

Under the QMSR the same failures are cited as ISO 13485 clauses. FY2026, which began on 1 October 2025, mixes the two: inspections that ended before 2 February 2026 cite 820.100, and later ones, with few exceptions, clause 8.5.2.6,2

Table 3. Citations in FDA’s Data Dashboard, counting inspections whose classification was final on 26 November 2026. FY2026 ran from 1 October 2025 to 30 September 2026, and its counts grow as its inspections are classified.6
Cited asSubjectFY2025FY2026
820.100(a)CAPA procedures24769
820.100(b)CAPA documentation5813
ISO 13485 clause 8.5.2Corrective action0107
ISO 13485 clause 8.5.3Preventive action015
ISO 13485 clause 8.4Analysis of data014

Clause 8.5.2 is the second most cited single clause in FY2026’s classified inspections, after risk management in product realization (see the most common 483 observations).6 Its standard citations say more than the old one did. At least 34 of the 107 say the action itself failed, most often in these words: “Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.” Among the wordings FDA used three or more times, 32 more cite the procedure, and all of them name the requirement for “reviewing the effectiveness of corrective action taken” among what is missing; it is the element FY2026’s procedure citations name most. The fullest version lists each element FDA expects the procedure to define:6

“Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.” FDA Data Dashboard, standard citation of ISO 13485 clause 8.5.26

Three more FY2026 citations say corrective actions “were unduly delayed”.6

FDA’s compliance program lists two CAPA failures among its examples of serious deficiencies that should lead to an initial classification of Official Action Indicated: data analysis or risk information so poor that it “results in a decision not to proceed with formal investigations and/or corrective actions”, with harm or nonconformities left unmitigated, and “Failure to adequately correct the same or similar significant deficiencies from previous inspections(s).”4

4Rewriting a CAPA procedure for the QMSR.

Most of 820.100 has a counterpart in clause 8.5.2; FDA said that adopting ISO 13485’s definition of preventive action does not change what it expects, and that procedures must still cover “quality systems, processes, and products with actual or potential nonconformities”.2

Table 4. From the text of 820.100, the QMSR final rule’s responses to comments 12 and 20, and FDA’s standard citations of the clauses.1,2,6
820.100Where it is nowWhat to change in the procedure
(a)(1) Analyze quality data, with statistics where neededClause 8.4, analysis of data, cited 14 times in FY2026Keep the list of data sources and the trending method; reference clause 8.4, not 820.100(a)(1)
(a)(2) Investigate the causeClause 8.5.2, “determining the causes of nonconformities”Little; state how the depth of investigation follows the risk
(a)(3) Identify the actionClause 8.5.2, “evaluating the need for action to ensure that nonconformities do not recur”Record the decision not to act, and why, as well as the action
(a)(4) Verify or validate effectiveness and no adverse effectClause 8.5.2, verifying no adverse effect on “regulatory requirements or the safety and performance of the medical device” and “reviewing the effectiveness”Make the effectiveness review a defined step with criteria, owner and date; it is the element FDA’s FY2026 citations name most
(a)(5) Implement and record changesClause 8.5.2, “planning and documenting action needed and implementing such action”Route changes to documents and processes through change control
(a)(6) Share the informationNot listed in ISO 13485; FDA says clauses 8.2.2, 8.5.2 and 8.3.1 account for itKeep the step; cite those clauses
(a)(7) Feed management reviewClause 5.6.2 lists corrective action and preventive action among management review inputs; FDA’s comment calls (a)(7) not specifically listed, its substance coveredKeep CAPA status as a management review input under clause 5.6.2
Preventive actionIts own clause, 8.5.3Give preventive action its own trigger and record; one CAPA procedure can serve both clauses if it keeps them distinct
(b) Document everythingClause 8.5.2’s records of investigations and actionsLittle; FDA’s citation is that records of “the results of any investigations and actions taken” are not maintained

FDA also set the scale of the response: “The degree of corrective or preventive action taken to eliminate or minimize actual or potential nonconformities shall be appropriate to the magnitude of the problem and commensurate with the risks encountered.”2

5Example: one CAPA record, from trend to effectiveness review.

Invented for illustration. A maker of sterile single-use catheters trends pouch seal complaints quarterly against the previous four quarters.

Table 5. Example. A CAPA record, invented for illustration.
StepRecord entryClause
TriggerQ2: 6 seal-leak complaints on 18,400 units shipped, 0.33 per 1,000, against a baseline of 7 on 70,000, 0.10 per 1,000; threshold for a CAPA is twice the baseline8.4, 8.2.2
Risk and scopeSterile barrier breach; severity high; all lots sealed on line 2 since its heater was replaced in March8.5.2, 7.1
CorrectionLine 2 lots on hold; 3 distributed lots assessed for field action under the nonconformity procedure, and for a correction or removal report under Part 8068.3.1, 8.3.3; 820.10(b)(4)
CauseSeal temperature drifting 6 °C below setpoint after the replacement; the replacement was not assessed for revalidation8.5.2
ActionHeater calibrated; sealing process revalidated; replacement parts added to the change control screen8.5.2, 7.5.6
No adverse effect30 pouches at the new settings: peel strength 1.9 to 2.4 N/15 mm against a minimum of 1.2; no change to labeling or registration8.5.2
Effectiveness reviewCriterion set at opening: no more than 0.10 per 1,000 over three months. Result: 1 complaint on 21,000 units, 0.05 per 1,000. Effective; closed by QA8.5.2
Management reviewCAPA status reported at the next review5.6

The rates: 6 ÷ 18,400 × 1,000 = 0.33; 7 ÷ 70,000 × 1,000 = 0.10; 1 ÷ 21,000 × 1,000 = 0.05. The trigger fired because 0.33 is more than twice 0.10. The review was planned when the CAPA opened, with its criterion and window: the element FY2026’s procedure citations name most.6

6What the CAPA procedure defines, and where to start.

The CAPA procedure should define, for each of the elements in FDA’s clause 8.5.2 citation above, who does it, by when and what is recorded: review of nonconformities and complaints, cause, the decision on action, the plan and its implementation, the check of no adverse effect, and the effectiveness review.6 These answers carry the pieces, each with its Word download:

To answer a 483 that cites clause 8.5.2, see responding to a 483.

7How it runs in Seal.

In Seal, the CAPA blueprint keeps the source deviation, investigation, actions, controlled change and follow-up evidence in one record, and the corrective action can be a controlled change to the workflow itself rather than only a revised document and a training task. The effectiveness review keeps its baseline, population, exposure and follow-up window, and closure can be configured to require the approved review.

Neil, Seal’s AI agent, can compare a CAPA’s event rates before and after the action against a defined population and exposure measure. The CAPA owner assesses effectiveness and QA approves the conclusion.

AQuestions and answers

Does 21 CFR 820.100 still apply?

No. The QMSR took effect on 2 February 2026 and removed the Quality System regulation’s sections, 820.100 among them. Inspections since then cite ISO 13485 clause 8.5.2 for corrective action and 8.5.3 for preventive action. An inspection that ended before that date could still cite 820.100, which is why FY2026’s data show both.

Is preventive action still required under the QMSR?

Yes. ISO 13485 gives preventive action its own clause, 8.5.3, and FDA said in the final rule that adopting ISO 13485’s definition does not change what it expects. In FY2026 inspections classified so far, FDA cited clause 8.5.3 15 times, against 107 for corrective action.

Should my SOPs still cite 820.100?

Replace the citation with ISO 13485 clauses 8.5.2 and 8.5.3. Keep sharing information with those responsible, which 820.100(a)(6) named and ISO 13485 does not list in so many words, and keep CAPA status as a management review input under clause 5.6.2. FDA said in the QMSR rule that their substance is covered by other clauses, so keeping them costs nothing and answers an auditor who asks.

Why was 820.100 cited so often?

CAPA is where other failures are meant to be caught, so a complaint trend not acted on or a nonconformance that recurred can be cited under CAPA as well as under its own section. Of FY2025’s 820.100(a) citations with final classifications, 210 of 247 said the procedure was not adequately established, not that it was missing.

Does every complaint or nonconformance need a CAPA?

No. The requirement is to evaluate the need for action and to make the action proportionate to the problem and its risk. Record the evaluation, including a decision not to open a CAPA and the reason, so the trend that later justifies one can be traced.

References6 sources

References

  1. 121 CFR 820.100, Corrective and preventive action, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.100 (read 26 November 2026)
  2. 2Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, responses to comments 12 and 20, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  3. 321 CFR 820.10, Requirements for a quality management system, and 820.7, Incorporation by reference, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  4. 4Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Parts III and V and Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)
  5. 5Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  6. 6FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)