What did 21 CFR 820.198 require for complaint files, and what replaced it under the QMSR?

21 CFR 820.198 required complaint files and procedures, run by a formally designated unit, to receive, review and evaluate every complaint, decide whether to investigate, investigate possible failures to meet specifications, assess MDR reportability and record set details. The QMSR moved it on 2 February 2026 to ISO 13485 clause 8.2.2, with record contents in 21 CFR 820.35(a). Its 833 citations in fiscal years 2023 to 2025 put it second or third among device sections every year.Checked against its sources on 26 November 2026. How these answers are made

1Complaint files under the old Part 820.

Section 820.198, “Complaint files”, opened: “Each manufacturer shall maintain complaint files. Each manufacturer shall establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit.”1 Its seven paragraphs, in the regulation’s words:

Table 1. 21 CFR 820.198, from the eCFR as in effect on 1 February 2026 (full text).1
ParagraphWhat it required
(a)Procedures ensuring that “All complaints are processed in a uniform and timely manner”, “Oral complaints are documented upon receipt”, and complaints are evaluated for reporting under Part 803, Medical Device Reporting (MDR)
(b)Review of all complaints “to determine whether an investigation is necessary”; with no investigation, a record of “the reason no investigation was made and the name of the individual responsible for the decision not to investigate”
(c)Investigation of “Any complaint involving the possible failure of a device, labeling, or packaging to meet any of its specifications … unless such investigation has already been performed for a similar complaint and another investigation is not necessary”
(d)MDR-reportable complaints “promptly reviewed, evaluated, and investigated by a designated individual(s)”, kept separately or “clearly identified”, with a determination of whether the device failed its specifications, whether it was used for treatment or diagnosis, and its relationship to the event
(e)A record of each investigation: device name, date received, “Any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s) and control number(s) used”, complainant’s name, address and phone, nature and details, “The dates and results of the investigation”, any corrective action and any reply
(f), (g)Records “reasonably accessible” to the manufacturing site, and in the United States where the complaint unit is abroad

2Second to CAPA among device references, three years running.

Complaint files were FDA’s second or third most cited device section in each of the last three full fiscal years, and 820.198(a) the second most cited device reference each year, behind CAPA’s 820.100(a):2

Table 2. 820.198 on device 483s, by the fiscal year the inspection ended (1 October to 30 September). FDA’s file leaves out handwritten 483s, and a citation records what the investigator observed, not a final agency finding.2
Cited asFDA’s standard citationFY2023FY2024FY2025
820.198, all paragraphsAll ten standard citations276271286
Cite 14713“Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been [adequately] established.”181191211
Cite 3168“Complaint files are not [adequately] maintained.”213932
820.198(c)Possible failures to meet specifications not reviewed, evaluated or investigated321716
820.198(e)Investigation records missing required information231316
820.198(b)Decision whether to investigate1454
820.198(d)MDR-reportable complaints567
Device 483sEvery device 483 in the file, for scale720672791

The citations moved up to paragraph (a) over the three years: (a) rose from 202 to 243 while (b) to (e) together fell from 74 to 43.2 An uninvestigated failure or a record without its UDI can be cited under the paragraph missed or under the procedure, and more of the section’s citations now go under the procedure; the counts don’t say why. FDA’s Data Dashboard shows what kind of procedure failure: of its 218 FY2025 citations of 820.198(a), 146 keep “adequately” in the procedure citation, which may be the text left unedited, 42 drop it, and 30 that the complaint files were not maintained, adequately or at all.3 The 218 fall short of the yearly file’s 243 because the dashboard counts an inspection only once its classification is final.3

3Clause 8.2.2, Part 803 and 820.35 since February.

Under the Quality Management System Regulation (QMSR), in force from 2 February 2026, complaint handling is clause 8.2.2 of ISO 13485:2016, which the regulation incorporates; to comply with clause 8.2.3, 820.10(b)(3) requires the manufacturer to “notify FDA of complaints that meet the reporting criteria of part 803”; and 820.35(a) sets what complaint records must hold.4 FDA’s compliance program for device inspections lists the complaint handling element against 820.10(b)(3) and (4), 820.35 and clauses 8.2.2 and 8.2.3, where 820.10(b)(4) sends advisory notices to Part 806.5,4 FDA withdrew paragraph (a)(3), the MDR evaluation, saying 820.10(b) makes “explicit that compliance with other parts of Title 21 is central to a comprehensive QMS system”.6

Table 3. FDA’s Data Dashboard, counting inspections classified as final by 26 November 2026. FY2026 (1 October 2025 to 30 September 2026) holds inspections from both sides of the change, and its counts rise as more are classified.3
Cited asSubjectFY2025FY2026
820.198(a)Complaint procedures and files, old section21861
ISO 13485 clause 8.2.2Complaint handling099
820.35(a)QMSR complaint records011
820.35(c)UDI recorded for each device or batch06

Clause 8.2.2 is the third most cited single clause in FY2026’s classified inspections, after risk management in product realization and corrective action (see the most common 483 observations).3 Its standard citations keep the old shape. At least 37 of the 99 say procedures for timely complaint handling “have not been adequately documented” (28) or “have not been documented” (9), at least 25 that complaint handling records were not maintained, adequately or at all, at least 7 that the procedures lack “requirements and responsibilities for receiving and recording information”, and at least 5 that the “Justification for not investigating a complaint(s) was not adequately documented.”3

Complaints also feed risk. Among FDA’s examples of serious deficiencies that should lead to an initial classification of Official Action Indicated is “Information gathered through the feedback process and/or postmarket surveillance is not used as potential input(s) into risk management for monitoring and maintaining the product realization or improvement processes.”5

4What stays, and what moves, in a complaint procedure.

Table 4. From the text of 820.198, 820.10 and 820.35, and the QMSR final rule’s response to comment 54.1,4,6
820.198NowWhat to do
(a) A formally designated unitNot in 820.35; FDA still expects procedures to name “which one group or unit is ultimately responsible for coordinating all complaint handling functions”Keep the designated unit; where several sites handle complaints, one corporate procedure, as FDA calls multiple interpretations “unacceptable”
(a) Uniform and timelyClause 8.2.2; FDA’s citation is for procedures for “timely complaint handling”State the time limits for each step
(a)(3), (d) MDR evaluation, kept separate820.10(b)(3): notify FDA of complaints meeting Part 803’s criteriaKeep the reportability decision in every complaint record; cite 820.10(b)(3) and Part 803
(b) Reason for not investigating, and who decidedClause 8.2.2 and 820.35(a); FDA says that “if any complaint is not investigated, the firm shall document the reason it has not investigated that complaint”Keep the reason and who decided; when the information can’t be obtained, the record shows “the efforts it made to ascertain the information”
(c) Investigate possible failures820.35(a): records of “the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications”Little; the similar-complaint exception is kept
(e) Record contents820.35(a)(1) to (7)Device name, date received, UDI or UPC, complainant’s name, address and phone, nature and details, any correction or corrective action, any reply; and 820.35(c)’s UDI for each device or batch
(f), (g) Accessible recordsClause 4.2.5; FDA reads its “readily identifiable and retrievable” as substantially similar to “reasonably accessible”: records shown during the inspection, or from a remote site “by the next working day or two, at the latest”Keep the records retrievable for an inspection wherever the unit sits

Two items of the old record list are not in 820.35(a)’s: “control number(s) used” and “The dates and results of the investigation”. Nor are the determinations 820.198(d) required for an MDR-reportable complaint: whether the device failed its specifications, whether it was used for treatment or diagnosis, and its relationship to the event. Keep all of them in the record: 820.35(a) still requires records of the investigation itself, and FDA said the information Part 803 requires “must appear on certain records of complaints”.1,4,6

5Twelve similar complaints, one investigation.

Invented for illustration. In March a maker of a home blood pressure monitor receives 12 complaints that the monitor inflates the cuff, then shows error E3 and no reading. Each is logged on receipt, 3 of them from phone calls written up by the support desk the same day.

Table 5. Example. An investigation record, invented for illustration.
RecordEntryRule
Each complaintDevice name, date received, UDI, complainant, nature and details, reply820.35(a)(1) to (5), (7)
ReportabilityEach assessed against the malfunction test of 803.50(a)(2): no injury reported, and the monitor shows an error and no reading rather than a wrong one, so a recurrence would not be “likely to cause or contribute to a death or serious injury”; not reportable, with rationale and reviewer in each record820.10(b)(3), 803.50(a)7
InvestigationOne, on complaint 1’s returned unit: pressure sensor drift that firmware 3.2 flags as a fault; covers the other 11820.35(a), clause 8.2.2
Justification on the other 11“Same error code, firmware and fault log as complaint 1; investigated under it”820.35(a)
CorrectionEach of the 12 complainants sent a replacement monitor820.35(a)(6)
Corrective actionA CAPA on the sensor, opened because 12 complaints in a month on 24,000 monitors sold in the past year is 0.50 per 1,000, against a trended 0.10Clause 8.5.2

The rate: 12 ÷ 24,000 × 1,000 = 0.50 per 1,000 monitors in a month, five times the trend. Each of the 11 records carries its own justification and the investigation it relies on, which is what the 5 FY2026 justification citations found missing.3

6Writing the complaint procedure.

The complaint procedure needs, in writing: the unit responsible, how oral complaints are documented on receipt, the time limit for each step, the reportability decision, when an investigation is required and how a similar complaint is linked to an earlier one, the record fields of 820.35(a), and the hand-off to CAPA.1,4 Its parts are covered in these answers:

7How it runs in Seal.

In Seal, complaint handling keeps the report as received, and its technical investigation, safety assessment and reportability assessment against 21 CFR 803.50 are separate linked records, each with its own owner or reviewer. The medical device blueprint enters complaints as structured data and lists the fields the QMSR asks of an investigated or reported complaint.

Neil, Seal’s AI agent, can compare recurring complaints against lot history and distribution exposure to define a focused investigation, and flags a lot with no distribution denominator before recurrence rates are compared. Complaint and quality owners assess significance and the reporting obligations.

AQuestions and answers

Is 21 CFR 820.198 still in force?

Not since 2 February 2026, when the QMSR replaced it. Complaint handling is ISO 13485 clause 8.2.2, reports to FDA go through 21 CFR 820.10(b)(3) and Part 803, and 21 CFR 820.35(a) sets what a complaint record holds. FY2026 inspections that ended before February could still cite 820.198, so that year’s data hold both.

Do I still need a formally designated complaint unit?

Part 820 no longer uses the phrase, but FDA said in the final rule that each manufacturer should establish in its procedures which one group or unit is ultimately responsible for coordinating all complaint handling functions, and that a system allowing multiple interpretations of handling would be unacceptable. Keep the designation.

Can one investigation cover several complaints?

Yes, if an investigation has already been performed for a similar complaint and another is not necessary. Each complaint’s record must then document the justification for not investigating it, which FDA cites under clause 8.2.2 when it is missing.

What must a device complaint record contain now?

For complaints that are MDR reportable, that the manufacturer decides to investigate, or that it investigates anyway: the device name, date received, any UDI or UPC and other identification, the complainant’s name, address and phone, the nature and details, any correction or corrective action, and any reply. Records of the review, evaluation and investigation are kept as well.

Is every complaint an MDR?

No. Every complaint is evaluated for reportability, and only events that meet Part 803’s criteria are reported to FDA. The evaluation and its rationale belong in the complaint record either way.

References7 sources

References

  1. 121 CFR 820.198, Complaint files, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.198 (read 26 November 2026)
  2. 2Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  3. 3FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)
  4. 421 CFR 820.10, Requirements for a quality management system, and 820.35, Control of records, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820 (read 26 November 2026)
  5. 5Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Part V and Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)
  6. 6Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, responses to comments 43, 54 and 58, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)
  7. 721 CFR 803.50, If I am a manufacturer, what reporting requirements apply to me?, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-803/subpart-E/section-803.50 (read 26 November 2026)