1One paragraph, and reports FDA could not see.
Section 820.22, “Quality audit”, was one paragraph:1
“Each manufacturer shall establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements and to determine the effectiveness of the quality system. Quality audits shall be conducted by individuals who do not have direct responsibility for the matters being audited. Corrective action(s), including a reaudit of deficient matters, shall be taken when necessary. A report of the results of each quality audit, and reaudit(s) where taken, shall be made and such reports shall be reviewed by management having responsibility for the matters audited. The dates and results of quality audits and reaudits shall be documented.” 21 CFR 820.22, eCFR as of 1 February 20261
Until the same date, 820.180(c) kept the audit reports themselves out of FDA’s reach: on request, a manager with executive responsibility certified in writing that the audits “have been performed and documented, the dates on which they were performed, and that any required corrective action has been undertaken.”2
2Fewer per 483 each year, more of them for the procedure.
| Cited as | FDA’s standard citation | FY2023 | FY2024 | FY2025 |
|---|---|---|---|---|
| 820.22, all paragraphs | All five standard citations | 108 | 93 | 107 |
| Cite 2327 | “Procedures for quality audits have not been [adequately] established.” | 41 | 50 | 57 |
| Cite 3415 | “Quality [audits] [reaudits] have not been performed.” | 29 | 15 | 25 |
| Cite 4059 | “Quality audits were not performed [at defined intervals] [at sufficient frequency] to determine whether the quality system activities and results comply with quality system procedures.” | 19 | 17 | 11 |
| Cite 2328 | “Individuals who conduct quality audits have direct responsibility for the matters being audited.” | 19 | 10 | 13 |
| Cite 2339 | “The dates of quality [audits] [reaudits] have not been documented.” | 0 | 1 | 1 |
| Device 483s | Total, for the rates | 720 | 672 | 791 |
Quality audits drew 308 citations in three years, falling from 15.0 to 13.8 to 13.5 per 100 device 483s. Because 820.22 has no paragraphs, it ranks seventh among device references in FY2025 but tenth among device sections.3 Within it, the procedure citation rose from 41 to 57 while the citation for audits not done at defined intervals or often enough fell from 19 to 11.3
3Three different failures.
FDA’s Data Dashboard holds 99 FY2025 citations of 820.22, and they split three ways.4 At least 24 found that quality audits “have not been performed”, which can mean areas left unaudited as well as no program. 31 kept “adequately” in the procedure citation, which may be the text left unedited, and 19 dropped it. And 12 found auditors with “direct responsibility for the matters being audited.”
The 99 are short of the yearly file’s 107 because the dashboard waits for each inspection’s final classification.4 Under 820.180(c), investigators who wrote these could ask for the procedures and a written certification, but not the reports themselves.2
4Clause 8.2.4, with the reports open to FDA.
Internal audit is clause 8.2.4 of ISO 13485:2016, part of the Quality Management System Regulation (QMSR) since 2 February 2026, and FDA’s compliance program lists the internal audits element against it.5,
| Cited as | Subject | FY2025 | FY2026 |
|---|---|---|---|
| 820.22 | Quality audit, the old section | 99 | 30 |
| ISO 13485 clause 8.2.4 | Internal audit | 0 | 36 |
Clause 8.2.4’s standard citations spell out more than 820.22’s did. At least 14 of the 36 say internal audits were not conducted, or not adequately, “at planned intervals to determine whether the QMS is effectively implemented and maintained”, at least 3 of them in a longer form that also names conformance “to planned and documented arrangements and applicable requirements”, which may be FDA’s text left unedited. At least 10 find no adequate procedure “to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results”.4
5What an audit procedure keeps and adds.
| 820.22 | Its successor | What to change |
|---|---|---|
| Procedures | Clause 8.2.4 | Cover each item in FDA’s citation: responsibilities, planning, conducting, recording and reporting |
| Compliance and effectiveness | Clause 8.2.4 | Little: 820.22 already asked whether the system complied and was effective. Audit against “planned and documented arrangements and applicable requirements”, the words FDA’s citation adds, and judge whether the system is “effectively implemented and maintained” |
| Defined intervals | Clause 8.2.4, “planned intervals” | Plan the program and record why each area gets its frequency: risk, past findings, changes |
| Independence | Clause 8.2.4, which asks that auditors be chosen, and audits run, for objectivity and impartiality | Keep the rule that no one audits work they are responsible for; use a colleague from another area or an outside auditor where the team is small |
| Corrective action and reaudit | Clause 8.2.4, with 8.5.2 | Link each finding to its correction or CAPA, and verify it before closing |
| Report reviewed by responsible management | Clause 8.2.4, and management review input | Keep the review; feed results to management review (clause 5.6.2) |
| Certification instead of reports | Withdrawn | Write reports on the assumption that FDA will read them: facts, evidence, and the finding’s status |
6A year’s audit program at a 60-person device maker.
Invented for illustration. A 60-person maker of class II devices has 12 QMS processes on its audit plan.
| Plan item | Entry |
|---|---|
| Frequency | Each process once a year; CAPA, complaints, production controls and purchasing twice, for their risk and last year’s findings: 12 + 4 = 16 audits |
| Auditors | Quality engineer audits production and purchasing; a design engineer audits quality’s own processes; an outside auditor audits management review and the audit program itself |
| Done | 15 of 16; the second purchasing audit moved to January, with the reason and the new date approved |
| Findings | 11: 2 major (complaint trending not done for two quarters; a validated sealer run outside its window), 9 minor |
| Follow-up | Both majors to CAPA; reaudited within 90 days and closed on evidence |
| Reporting | Each report reviewed and signed by the manager of the area; the year’s results to management review |
15 of 16 is 94% of the plan done in the year, and the one deferral has a reason and an approved date. An investigator reading it against the clause 8.2.4 citations would find the intervals planned, auditors away from their own work, a procedure behind it, and results reported and acted on.4
7The audit procedure, item by item.
The procedure needs: the program and how risk sets each area’s frequency, auditor qualification and independence, how an audit is planned, conducted and recorded, how findings are graded and tracked to correction or CAPA, reaudit, who reviews the report, and the hand-off to management review. The checklist and templates are in these answers:
- The GMP internal audit, with a self-inspection checklist, written for medicines; its program, grading and report sections carry over.
- Management review and audit reports under the QMSR, with a readiness checklist for these records.
- Management review, with a template, where audit results go.
- 21 CFR 820.100 and clause 8.5.2, for findings that need corrective action.
8How it runs in Seal.
In Seal’s audit management workflow, the program records what is audited, when and by whom, with the reason for each choice and for any deferral. Each engagement keeps its focus, window, lead and scope, and the assigned auditor’s role, qualification and independence are recorded with its approval. A finding links its assessment, the change it leads to and the verification as separate records, each with its own owner and state.
Neil can assemble the evidence for an audit request: it reads the question, the procedures and the linked records, drafts an indexed evidence brief and makes missing evidence explicit, and your audit owner checks scope and completeness before anything is shared.
AQuestions and answers
Has 21 CFR 820.22 been replaced?
Yes. Since 2 February 2026 the QMSR has required internal audits under ISO 13485 clause 8.2.4, and FDA may read the reports. 820.22 still appears in FY2026’s citations, a year that began under the old rule.
Can FDA read our internal audit reports now?
Yes. The old 21 CFR 820.180(c) exception, which let a manager certify that audits were done instead of showing the reports, did not carry into the QMSR. FDA may now read internal audit reports, supplier audit reports and management reviews.
How often must internal audits be done?
At planned intervals that you set and justify. Neither the old rule nor FDA’s clause 8.2.4 citation names a number; plan by risk, past findings and changes, and audit every process within a cycle you can defend.
Who may audit in a small company?
Anyone competent who has no direct responsibility for the matter audited. A colleague from another area can audit quality’s own processes, or an outside auditor can, so the quality manager never audits their own work.
References7 sources
References
- 121 CFR 820.22, Quality audit, eCFR, as in effect on 1 February 2026. https:/
/ www. ecfr. gov/ on/ 2026- 02- 01/ title- 21/ section- 820. 22 (read 26 November 2026) - 221 CFR 820.180, General requirements (records), paragraph (c), eCFR, as in effect on 1 February 2026. https:/
/ www. ecfr. gov/ on/ 2026- 02- 01/ title- 21/ section- 820. 180 (read 26 November 2026) - 3Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ inspection- references/ inspection- observations (read 26 November 2026) - 4FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https:/
/ datadashboard. fda. gov/ oii/ cd/ inspections. htm (read 26 November 2026) - 521 CFR 820.10, Requirements for a quality management system, eCFR, current to 24 November 2026. https:/
/ www. ecfr. gov/ current/ title- 21/ chapter- I/ subchapter- H/ part- 820/ subpart- A/ section- 820. 10 (read 26 November 2026) - 6Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Attachment A. https:/
/ www. fda. gov/ media/ 80195/ download (read 26 November 2026) - 7Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, response to comment 55, FDA, 2 February 2024. https:/
/ www. federalregister. gov/ documents/ 2024/ 02/ 02/ 2024- 01709/ medical- devices- quality- system- regulation- amendments (read 26 November 2026)