What did 21 CFR 820.22 require for quality audits, and what replaced it?

21 CFR 820.22 required procedures for quality audits, and audits by people without direct responsibility for what they audit, to check that the quality system complies and is effective, with reports reviewed by the management responsible, corrective action and reaudit where needed, and dates and results recorded. ISO 13485 clause 8.2.4 replaced it on 2 February 2026; FDA may now read the reports. It drew 308 citations in fiscal years 2023 to 2025, fewer per 483 each year.Checked against its sources on 26 November 2026. How these answers are made

1One paragraph, and reports FDA could not see.

Section 820.22, “Quality audit”, was one paragraph:1

“Each manufacturer shall establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements and to determine the effectiveness of the quality system. Quality audits shall be conducted by individuals who do not have direct responsibility for the matters being audited. Corrective action(s), including a reaudit of deficient matters, shall be taken when necessary. A report of the results of each quality audit, and reaudit(s) where taken, shall be made and such reports shall be reviewed by management having responsibility for the matters audited. The dates and results of quality audits and reaudits shall be documented.” 21 CFR 820.22, eCFR as of 1 February 20261

Until the same date, 820.180(c) kept the audit reports themselves out of FDA’s reach: on request, a manager with executive responsibility certified in writing that the audits “have been performed and documented, the dates on which they were performed, and that any required corrective action has been undertaken.”2

2Fewer per 483 each year, more of them for the procedure.

Table 1. Citations of 820.22 by fiscal year, from the device 483s FDA’s inspection software produced. Handwritten 483s are not counted, and a citation is not FDA’s final word.3
Cited asFDA’s standard citationFY2023FY2024FY2025
820.22, all paragraphsAll five standard citations10893107
Cite 2327“Procedures for quality audits have not been [adequately] established.”415057
Cite 3415“Quality [audits] [reaudits] have not been performed.”291525
Cite 4059“Quality audits were not performed [at defined intervals] [at sufficient frequency] to determine whether the quality system activities and results comply with quality system procedures.”191711
Cite 2328“Individuals who conduct quality audits have direct responsibility for the matters being audited.”191013
Cite 2339“The dates of quality [audits] [reaudits] have not been documented.”011
Device 483sTotal, for the rates720672791

Quality audits drew 308 citations in three years, falling from 15.0 to 13.8 to 13.5 per 100 device 483s. Because 820.22 has no paragraphs, it ranks seventh among device references in FY2025 but tenth among device sections.3 Within it, the procedure citation rose from 41 to 57 while the citation for audits not done at defined intervals or often enough fell from 19 to 11.3

3Three different failures.

FDA’s Data Dashboard holds 99 FY2025 citations of 820.22, and they split three ways.4 At least 24 found that quality audits “have not been performed”, which can mean areas left unaudited as well as no program. 31 kept “adequately” in the procedure citation, which may be the text left unedited, and 19 dropped it. And 12 found auditors with “direct responsibility for the matters being audited.”

The 99 are short of the yearly file’s 107 because the dashboard waits for each inspection’s final classification.4 Under 820.180(c), investigators who wrote these could ask for the procedures and a written certification, but not the reports themselves.2

4Clause 8.2.4, with the reports open to FDA.

Internal audit is clause 8.2.4 of ISO 13485:2016, part of the Quality Management System Regulation (QMSR) since 2 February 2026, and FDA’s compliance program lists the internal audits element against it.5,6 FDA did not keep the 820.180(c) exception, so it may now read internal audit reports, supplier audit reports and management reviews. It said such records “are maintained in the regular course of business and should be readily available”, and that “robust management review, as well as internal and supplier audit programs, are fundamental to the culture of quality”.7 What FDA can now see covers how investigators use that access.

Table 2. Dashboard citations, inspections classified as final by 26 November 2026. FY2026 began under the old rule and ended under the new one, and will grow as classifications come in.4
Cited asSubjectFY2025FY2026
820.22Quality audit, the old section9930
ISO 13485 clause 8.2.4Internal audit036

Clause 8.2.4’s standard citations spell out more than 820.22’s did. At least 14 of the 36 say internal audits were not conducted, or not adequately, “at planned intervals to determine whether the QMS is effectively implemented and maintained”, at least 3 of them in a longer form that also names conformance “to planned and documented arrangements and applicable requirements”, which may be FDA’s text left unedited. At least 10 find no adequate procedure “to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results”.4

5What an audit procedure keeps and adds.

Table 3. From the text of 820.22 and 820.180(c), Compliance Program 7382.850 Attachment A, the QMSR final rule’s response to comment 55, and FDA’s standard citations of the clause.1,2,6,7,4
820.22Its successorWhat to change
ProceduresClause 8.2.4Cover each item in FDA’s citation: responsibilities, planning, conducting, recording and reporting
Compliance and effectivenessClause 8.2.4Little: 820.22 already asked whether the system complied and was effective. Audit against “planned and documented arrangements and applicable requirements”, the words FDA’s citation adds, and judge whether the system is “effectively implemented and maintained”
Defined intervalsClause 8.2.4, “planned intervals”Plan the program and record why each area gets its frequency: risk, past findings, changes
IndependenceClause 8.2.4, which asks that auditors be chosen, and audits run, for objectivity and impartialityKeep the rule that no one audits work they are responsible for; use a colleague from another area or an outside auditor where the team is small
Corrective action and reauditClause 8.2.4, with 8.5.2Link each finding to its correction or CAPA, and verify it before closing
Report reviewed by responsible managementClause 8.2.4, and management review inputKeep the review; feed results to management review (clause 5.6.2)
Certification instead of reportsWithdrawnWrite reports on the assumption that FDA will read them: facts, evidence, and the finding’s status

6A year’s audit program at a 60-person device maker.

Invented for illustration. A 60-person maker of class II devices has 12 QMS processes on its audit plan.

Table 4. Example. An internal audit program, invented for illustration.
Plan itemEntry
FrequencyEach process once a year; CAPA, complaints, production controls and purchasing twice, for their risk and last year’s findings: 12 + 4 = 16 audits
AuditorsQuality engineer audits production and purchasing; a design engineer audits quality’s own processes; an outside auditor audits management review and the audit program itself
Done15 of 16; the second purchasing audit moved to January, with the reason and the new date approved
Findings11: 2 major (complaint trending not done for two quarters; a validated sealer run outside its window), 9 minor
Follow-upBoth majors to CAPA; reaudited within 90 days and closed on evidence
ReportingEach report reviewed and signed by the manager of the area; the year’s results to management review

15 of 16 is 94% of the plan done in the year, and the one deferral has a reason and an approved date. An investigator reading it against the clause 8.2.4 citations would find the intervals planned, auditors away from their own work, a procedure behind it, and results reported and acted on.4

7The audit procedure, item by item.

The procedure needs: the program and how risk sets each area’s frequency, auditor qualification and independence, how an audit is planned, conducted and recorded, how findings are graded and tracked to correction or CAPA, reaudit, who reviews the report, and the hand-off to management review. The checklist and templates are in these answers:

8How it runs in Seal.

In Seal’s audit management workflow, the program records what is audited, when and by whom, with the reason for each choice and for any deferral. Each engagement keeps its focus, window, lead and scope, and the assigned auditor’s role, qualification and independence are recorded with its approval. A finding links its assessment, the change it leads to and the verification as separate records, each with its own owner and state.

Neil can assemble the evidence for an audit request: it reads the question, the procedures and the linked records, drafts an indexed evidence brief and makes missing evidence explicit, and your audit owner checks scope and completeness before anything is shared.

AQuestions and answers

Has 21 CFR 820.22 been replaced?

Yes. Since 2 February 2026 the QMSR has required internal audits under ISO 13485 clause 8.2.4, and FDA may read the reports. 820.22 still appears in FY2026’s citations, a year that began under the old rule.

Can FDA read our internal audit reports now?

Yes. The old 21 CFR 820.180(c) exception, which let a manager certify that audits were done instead of showing the reports, did not carry into the QMSR. FDA may now read internal audit reports, supplier audit reports and management reviews.

How often must internal audits be done?

At planned intervals that you set and justify. Neither the old rule nor FDA’s clause 8.2.4 citation names a number; plan by risk, past findings and changes, and audit every process within a cycle you can defend.

Who may audit in a small company?

Anyone competent who has no direct responsibility for the matter audited. A colleague from another area can audit quality’s own processes, or an outside auditor can, so the quality manager never audits their own work.

References7 sources

References

  1. 121 CFR 820.22, Quality audit, eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.22 (read 26 November 2026)
  2. 221 CFR 820.180, General requirements (records), paragraph (c), eCFR, as in effect on 1 February 2026. https://www.ecfr.gov/on/2026-02-01/title-21/section-820.180 (read 26 November 2026)
  3. 3Inspection Observations, FY2023, FY2024 and FY2025 Excel files, Summary and Devices sheets, inspections ending 1 October 2022 to 30 September 2025, FDA, page current as of 16 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/inspection-references/inspection-observations (read 26 November 2026)
  4. 4FDA Data Dashboard, Inspections, Citations dataset (entire), downloaded 26 November 2026, inspections ending through 17 November 2026, final classifications only. https://datadashboard.fda.gov/oii/cd/inspections.htm (read 26 November 2026)
  5. 521 CFR 820.10, Requirements for a quality management system, eCFR, current to 24 November 2026. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-820/subpart-A/section-820.10 (read 26 November 2026)
  6. 6Compliance Program 7382.850, Inspection of Medical Device Manufacturers, FDA, implemented 2 February 2026, Attachment A. https://www.fda.gov/media/80195/download (read 26 November 2026)
  7. 7Medical Devices; Quality System Regulation Amendments, final rule, 89 FR 7496, effective 2 February 2026, response to comment 55, FDA, 2 February 2024. https://www.federalregister.gov/documents/2024/02/02/2024-01709/medical-devices-quality-system-regulation-amendments (read 26 November 2026)