1What raw data means, text by text.
21 CFR 211 never uses the term “raw data”. It requires laboratory records to hold “complete data derived from all tests necessary to assure compliance with established specifications and standards”, including “all graphs, charts, and spectra from laboratory instrumentation”, and lets records be kept “either as original records or as true copies”.1 The one FDA regulation that defines raw data is the GLP rule for nonclinical studies, 21 CFR 58.3(k): records “that are the result of original observations and activities” and are “necessary for the reconstruction and evaluation of the report”, including “recorded data from automated instruments”.2
Guidance gives one definition. MHRA, PIC/S and WHO call raw data “the original record (data) which can be described as the first-capture of information, whether recorded on paper or electronically”, and MHRA and PIC/S add: “Information that is originally captured in a dynamic state should remain available in that state.”3,
| What the text says | Basis |
|---|---|
| Complete data from all tests, spectra included | Rule: 211.194(a) |
| A record of all calculations | Rule: 211.194(a)(5) |
| Originals or true copies kept | Rule: 211.180(d) |
| A backup of data entered | Rule: 211.68(b) |
| Raw data defined, for nonclinical studies | Rule: 21 CFR 58.3(k) |
| Dynamic records stay dynamic; simple printouts can be originals | Guidance: FDA DI Q&A 1d, 10 |
| Raw data is the first capture | Guidance: MHRA 6.2; PIC/S, WHO glossaries |
| Each firm defines its raw data, for electronic records | Guidance: EU GMP Chapter 4, Principle |
| The first-capture definition added | Draft guidance: Chapter 4 glossary |
EU GMP Chapter 4, guidance under Article 47 of Directive 2001/83/EC, leaves the line to the firm: “For electronic records regulated users should define which data are to be used as raw data. At least, all data on which quality decisions are based should be defined as raw data.”7 Its draft revision, whose consultation closed on 7 October 2025, adds the first-capture definition to its glossary.8,
2Raw data, instrument by instrument.
Ask two questions of each instrument. Does it store the record? If not, the printout or the written entry is the raw data; if it does, the stored record is. Can the stored record be reprocessed or recalculated? If so, it is dynamic, and only the file, or a true copy that keeps it dynamic, will do; if not, a verified printout can be its true copy.3,
| Instrument | The raw data | Basis |
|---|---|---|
| HPLC with a CDS | Each injection’s data file, method, sequence, audit trail | Guidance: FDA DI Q&A 10, 12 |
| Balance or pH meter, no storage | The printout: signed, dated, sample ID | Guidance: FDA DI Q&A 10; MHRA 6.2 |
| Balance or pH meter with storage | Stored readings, reconciled with printouts | Guidance: MHRA 6.2 |
| Karl Fischer titrator with software | Each stored determination, dynamic as it can be recalculated | Guidance: FDA DI Q&A 1d, 10; MHRA 6.2 |
| Spreadsheet calculation | The file: inputs and formulas | Rule: 211.194(a)(5); Guidance: FDA DI Q&A 1d |
| Paper observation | The entry made at the time | Rule: 211.160(a); Guidance: FDA DI Q&A 12 |
- Chromatography. FDA wants data saved “to durable media upon completion of each step or injection (e.g., peak integration or processing steps; finished, incomplete, or aborted injections)”.6 An injection nobody reports is still raw data; see orphan data. Which printouts and PDFs can stand in is set out in ALCOA+ and true copies.
- Balances and pH meters. For basic equipment that does not store data “or provides only a printed data output (e.g. balances or pH meters), then the printout constitutes the raw data.” Where it holds a limited volume before overwriting, MHRA wants it “periodically reviewed and where necessary reconciled against paper records”.3 Thermal printouts fade; keep a verified true copy too.
- Karl Fischer titrators. Many store each determination and can recalculate it: the stored determination is then the raw data, and dynamic, so a printout cannot replace it. At Intas, FDA found that “The KF instrument used for water content testing and assay testing is capable of storing electronic data; however, this capacity was not utilized, and you did not save this data electronically.”10
- Spreadsheets. FDA’s dynamic records include a spreadsheet whose formulas or entries can be changed.6 Keep the file. At Specialty Process Labs, the assay spreadsheet’s “formulas and outputs of the spreadsheet were printed at the time of the calculation and not saved.”11
- Paper. “It is not acceptable to record data on pieces of paper that will be discarded after the data are transcribed to a permanent laboratory notebook”.6 A colony count written at the plate is raw data.
3Raw data vs processed data.
Processed data are what software or a person derives from raw data. MHRA’s example of processing is the “conversion of a raw electronic signal to a chromatogram and subsequently a calculated numerical result”.3
| Raw data | Processed from it |
|---|---|
| Detector signal for each injection | Integrated peaks, then assay % |
| FT-IR spectral file | Library match, identity result |
| Titrant volume and sample weight | Water content % |
| Balance printout | Potency-adjusted quantity |
| Colonies counted on a plate | CFU per gram |
Both belong to the record. Part 211 asks for “A record of all calculations performed in connection with the test”;1 FDA wants each reprocessed result “retained for review”;6 and MHRA wants “adequate traceability of any user-defined parameters used within data processing activities to the raw data, including attribution to who performed the activity.”3 Natural Essentials kept only one side: during the inspection, the firm admitted “preserving raw chromatographic data but failing to retain the processed chromatograms.”12 Reprocessing adds a new version of the processed data and leaves the raw data as captured; how integration may change a result is on manual integration.
4How it runs in Seal.
In Seal’s scientific data management records, each instrument acquisition is a package whose capture rule names the files it must hold; for chromatography, the sequence, method, raw, audit, result and report files. A missing file is kept distinct from a corrupt transfer or an unsupported format. The original file is preserved with its source, acquisition identity, timestamps and integrity evidence, and extracted values are stored as related records, so the source remains available for review.
Reviewers reach the preserved acquisition from the result, and reprocessing creates a derived record rather than overwriting the earlier one.
5What FDA cites when raw data are missing.
| Finding in a warning letter | Of the 80 letters |
|---|---|
| Raw data not retained | 13 (16%) |
| Data not available when FDA asked | 6 (8%) |
| Either | 17 (21%) |
Method: we searched FDA’s warning letter full-text index, which holds letters posted from January 2021, on 26 November 2026 for 35 phrases, among them “211.68”, “211.194”, “data integrity”, “audit trail” and “electronic data”. Of the 258 letters found, we set aside 83 whose numbered violations hold no computer or electronic data term or use one in another sense, and read every numbered violation of the other 175. We kept the 80 drug CGMP letters (finished drugs and APIs, issued 27 January 2021 to 10 November 2026) in which at least one numbered violation faults a computerized system or electronic data. Each letter counts once per category; FDA’s requests to the firm and its standard remediation text are not counted, nor are Form FDA 483s or findings on paper alone, such as discarded balance printouts. We then read each of the 13 to see what was lost. Placing a finding in a category is our judgment, and a letter names what FDA chose to cite, not all that inspectors saw.
| What was lost | Letters |
|---|---|
| Instrument files not saved, overwritten or kept only on paper | 5 |
| Calculations or processed results not kept | 3 |
| Monitoring data shown but not stored, or overwritten | 3 |
| Raw data or records missing, no system named | 2 |
Only 4 of the 13 cite 211.68 or ICH Q7 5.4 or 6.6; the rest cite laboratory records and controls, facilities, investigations or the quality unit, because missing raw data is a records failure first.
- Monitoring with no memory. Pharmathen, 2026: “Your monitoring devices displayed real-time data only and lack data storage capability.”13 Somerset, 2025: the review of aborted air samples was limited by an “unacceptable data retention capability of only 115 entries.”14
- Kept, but not produced. The Center for Instrumental Analysis of China Pharmaceutical University, 2025, kept NMR data on archived hard drives and “did not provide any CGMP data stored on the external hard drives for review during the inspection.”15 FDA may inspect and copy electronic records, so raw data must be retrievable, not only kept.6
6Raw data register template.
One row for each system or instrument type that creates GxP data, plus paper worksheets. It answers EU GMP’s request to define raw data, and shows a US inspector where the complete data for each test are.
| Field | What it records | Basis |
|---|---|---|
| System or instrument | Name, ID, version, inventory row | Guidance: Annex 11, 4.3 |
| Activities | Tests or steps it records | Practice |
| Raw data | The first capture: file, printout or entry | Guidance: EU GMP Chapter 4; MHRA 6.2 |
| Static or dynamic | Which, and so what must be kept | Guidance: FDA DI Q&A 1d, 10 |
| Processed data | What is derived, and where it is kept | Guidance: MHRA 6.9 |
| Metadata | IDs, times, method, audit trail | Guidance: FDA DI Q&A 1b; PIC/S 9.1.5 |
| Where kept | Location; backup or true copy | Rule: 211.68(b), 211.180(d) |
| Review | Who reviews it, before which decision | Rule: 211.194(a)(8) |
| Retention | Period and its basis | Rule: 211.180(a) |
7Example: Exampla’s QC raw data register.
Invented for illustration. Exampla Pharma Services LLC, a contract manufacturer and testing laboratory in Durham, North Carolina, names each system’s raw data in one line in section 2 of QA-DI-001, its data integrity procedure. Its QC laboratory keeps the detail each line cannot hold in register QA-RDR-001: 6 rows covering ExamCDS 3.4 with its 14 HPLC and 4 GC, ExamLIMS 6.1, 12 standalone instruments and paper worksheets. Revision 1 was issued with QA-DI-001 revision 3 on 12 January 2026. Revision 2, prepared by the lab manager as system owner and approved by QA’s computer systems lead on 6 October 2026, added Karl Fischer titrators KF-01 and KF-02, found outside the inventory under deviation DEV-26-136.
| System | Raw data | Kept and reviewed |
|---|---|---|
| ExamCDS 3.4, 14 HPLC and 4 GC | Every injection’s file, method, sequence, audit trail | ExamCDS, nightly backup; result review with audit trail; monthly orphan data review |
| ExamLIMS 6.1 | Only what is first entered there, such as sample login | ExamLIMS; each result checked against its raw data |
| 3 FTIR, 1 UV-Vis | Spectral files on each PC | PC, nightly backup; reviewer opens the file |
| KF-01, KF-02 | Stored determinations; printout a copy | PC, backup since 5 October 2026 |
| 6 balances | The printout; they store no data | With the worksheet; thermal paper copied |
| Worksheets | Observations at the time, such as appearance | Paper; second-person review |
The ExamLIMS row is the one people get wrong. Results reach ExamLIMS from ExamCDS or are typed from instruments and worksheets, so in ExamLIMS they are processed or copied data; the raw data stay where they were captured, and the reviewer checks each result against them. In October 2026, ExamCDS acquired 10,512 injections, and 828 of them were in sequences that reported no result, such as method validation, qualification, conditioning or aborted runs; all are raw data, which is why the orphan data review accounts for each.
Each row also names the metadata kept: for ExamCDS, user, instrument, date, time and audit trail; for a balance printout, balance ID, date, time and sample ID, written on by the analyst. Revision 2’s KF row followed DEV-26-136: the titrators store each determination and their software can recalculate it, so the stored determination is the raw data and the printout cannot replace it; all 2,156 stored determinations were traced to reported results or the titer log. QC raw data are kept to the end of the 8th year after the year of the test, so a 2026 result is kept to 31 December 2034; stability data supporting a client’s EU marketing authorization are kept while it is in force, as EU GMP 4.12 asks.7
AQuestions and answers
Is a printout ever raw data?
Yes, from basic equipment that stores no data, such as many balances and pH meters: FDA and MHRA both treat the printout as the original. Sign and date it, mark it with the sample or batch, and keep it, or a verified true copy if it is thermal paper. A printout from an instrument that stores a dynamic record, such as an HPLC or FT-IR, is not the raw data.
Are metadata part of the raw data?
They are kept with it. FDA: “Data should be maintained throughout the record’s retention period with all associated metadata required to reconstruct the CGMP activity.” For a chromatogram that means the user, instrument, date and time, method, sequence and audit trail; for a balance printout, the balance ID, date, time and sample.
What is the difference between raw data and source data?
None in practice. MHRA calls raw data “synonymous with ‘source data’ which is defined in ICH GCP”, and WHO says the same. Source data is the clinical term; raw data the GMP and GLP one.
Is raw data the same as complete data?
No. Raw data are the first capture. Complete data, which 21 CFR 211.194(a) requires, are everything the test produced: the raw data, every calculation and processed result, the method, the second-person review and the record of any deviation or reprocessing.
Can raw data be deleted once the report is approved?
No. Raw data are kept for the retention period of the record they support: in the US, for records tied to a batch, at least 1 year after its expiration date under 211.180(a). EU GMP 4.12 keeps raw data supporting the marketing authorization, such as validation or stability data, while the authorization is in force.
References16 sources
References
- 121 CFR Part 211, sections 211.68(b), 211.160(a), 211.180(a) and (d), and 211.194(a), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ part- 211 (read 26 November 2026) - 221 CFR 58.3, Definitions, paragraph (k), eCFR. https:/
/ www. ecfr. gov/ current/ title- 21/ section- 58. 3 (read 26 November 2026) - 3‘GXP’ Data Integrity Guidance and Definitions, revision 1, sections 6.2 and 6.9, MHRA, March 2018. https:/
/ assets. publishing. service. gov. uk/ media/ 5aa2b9ede5274a3e391e37f3/ MHRA_ GxP_ data_ integrity_ guide_ March_ edited_ Final. pdf (read 26 November 2026) - 4PI 041-1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, section 9.1.5 and glossary, PIC/S, 1 July 2021. https:/
/ picscheme. org/ docview/ 4234 (read 26 November 2026) - 5Guideline on data integrity, WHO Technical Report Series No. 1033, Annex 4, glossary, World Health Organization, 2021. https:/
/ cdn. who. int/ media/ docs/ default- source/ medicines/ norms- and- standards/ guidelines/ inspections/ trs1033- annex4- guideline- on- data- integrity. pdf (read 26 November 2026) - 6Data Integrity and Compliance With Drug CGMP: Questions and Answers, questions 1b, 1d, 10, 12, 14 and 17, FDA, December 2018. https:/
/ www. fda. gov/ media/ 119267/ download (read 26 November 2026) - 7EudraLex Volume 4, Chapter 4: Documentation, Principle and section 4.12, European Commission, in operation from 30 June 2011. https:/
/ health. ec. europa. eu/ system/ files/ 2016- 11/ chapter4_ 01- 2011_ en_ 0. pdf (read 26 November 2026) - 8EudraLex Volume 4, Chapter 4: Documentation, draft for consultation, glossary, European Commission, July 2025. https:/
/ health. ec. europa. eu/ document/ download/ fa336a6e- 753b- 46fc- b53b- 9178bacd8878_ en?filename=mp_ vol4_ chap4_ consultation_ guideline_ en. pdf (read 26 November 2026) - 9Stakeholders’ Consultation on EudraLex Volume 4, Chapter 4, Annex 11 and New Annex 22, 7 July to 7 October 2025, closed, European Commission. https:/
/ health. ec. europa. eu/ consultations/ stakeholders- consultation- eudralex- volume- 4- good- manufacturing- practice- guidelines- chapter- 4- annex_ en (read 26 November 2026) - 10Warning letter, Intas Pharmaceuticals Limited, MARCS-CMS 652067, violation 1, FDA, 28 July 2023. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ intas- pharmaceuticals- limited- 652067- 07282023 (read 26 November 2026) - 11Warning letter, Specialty Process Labs LLC, MARCS-CMS 624281, violation 3, FDA, 3 May 2022. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ specialty- process- labs- llc- 624281- 05032022 (read 26 November 2026) - 12Warning letter, Natural Essentials, Inc., MARCS-CMS 738543, violation 4, FDA, 10 November 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ natural- essentials- inc- 738543- 11102026 (read 26 November 2026) - 13Warning letter, Pharmathen International S.A., MARCS-CMS 723680, violation 3, FDA, 27 May 2026. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ pharmathen- international- sa- 723680- 05272026 (read 26 November 2026) - 14Warning letter, Somerset Therapeutics Private Limited, MARCS-CMS 711340, violation 1, FDA, 4 September 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ somerset- therapeutics- private- limited- 711340- 09042025 (read 26 November 2026) - 15Warning letter, Center for Instrumental Analysis of China Pharmaceutical University, MARCS-CMS 696906, violation 1, FDA, 23 January 2025. https:/
/ www. fda. gov/ inspections- compliance- enforcement- and- criminal- investigations/ warning- letters/ center- instrumental- analysis- china- pharmaceutical- university- 696906- 01232025 (read 26 November 2026) - 16EudraLex Volume 4, Annex 11: Computerised Systems, revision 1, section 4.3, European Commission, in operation from 30 June 2011. https:/
/ health. ec. europa. eu/ document/ download/ 8d305550- dd22- 4dad- 8463- 2ddb4a1345f1_ en?filename=annex11_ 01- 2011_ en. pdf (read 26 November 2026)