What is orphan data in GMP, and how do you review for it?

Orphan data are injections, sequences or results a chromatography data system or other lab instrument acquired that no reported result accounts for. No rule uses the term, but 21 CFR 211.194(a) requires complete data, and FDA treats an unreported sample injection as possible testing into compliance. To find them, compare everything the system acquired with everything reported, by sequence, user and date, and explain each difference from the audit trail.Checked against its sources on 26 November 2026. How these answers are made

1What orphan data are.

Orphan data are records a laboratory system acquired that no reported result accounts for. In a chromatography data system (CDS) they are a single injection in a sequence named “test”, a sequence acquired and never processed, a run aborted partway, a sample injected twice with only the second reported, a reprocessed result left aside, or a folder no reviewer opens. Other instruments that store results make them too.

The term is the industry’s. None of the FDA, PIC/S, MHRA or WHO data integrity guidance uses it, and nor do any of the 80 warning letters counted below. The rules speak of complete data instead. The laboratory record must hold “complete data derived from all tests necessary to assure compliance with established specifications and standards” (21 CFR 211.194(a)), and FDA’s data integrity guidance says: “When generated to satisfy a CGMP requirement, all data become a CGMP record.”1,2

Not every orphan is a problem. A system acquires many injections that never feed a result: blanks to condition a column, standards to check a repaired pump, runs under a method validation protocol. They are accounted for when their purpose is documented and no sample was used. FDA’s test is transparency: all data, including obvious errors and failing, passing and suspect data, “must be in the CGMP records that are retained and subject to review and oversight.”2 Orphan data are a failure when the record cannot say what they were.

2Why FDA treats orphan data as a failure.

An unreported sample injection can hide a failing result. Analysts who inject a sample before the official sequence can see whether it will pass, and adjust or repeat until it does. FDA’s guidance calls this testing into compliance, and says: “FDA considers it a violative practice to use an actual sample in test, prep, or equilibration runs as a means of disguising testing into compliance.”2 Its OOS guidance calls repeated testing until a pass, with the failures disregarded, “unscientific and objectionable under CGMP”.3 FDA told Aspire, in 2022, that “The injection of trial samples is not acceptable, in part, because all data from analysis of product samples must be retained and reviewed”, and that “FDA considers it a violative practice to perform a trial injection.”4

Table 1. What the rules and guidance require of data that never reached a result.1,5,6,2,7,8
What the record must showBasis
Complete data from every test, failing and aborted runs includedRule: 211.194(a)
Any departure from the written test procedure, recorded and justifiedRule: 211.160(a)
Every unexplained discrepancy investigatedRule: 211.192
Aborted or incomplete injections in the audit trail, investigated and justifiedGuidance: FDA DI Q&A 12
No actual sample in test, prep or equilibration runsGuidance: FDA DI Q&A 13
Every reprocessed result kept, not only the lastGuidance: FDA DI Q&A 14
Excluded data kept with the data set, with the reasonGuidance: MHRA 6.10
Review that asks whether unreported data were generatedGuidance: PIC/S 9.8; MHRA 6.15

The guidance outside the US says the same. MHRA asks every routine data review to consider “is this the only data generated as part of this activity?”, and says excluded data must stay “available for review in a format that allows the validity of the decision to exclude the data to be confirmed.”7 WHO adds where to look: “The review of original electronic data should include checks of all locations where data may have been stored, including locations where voided, deleted, invalid or rejected data may have been stored.”9

3What FDA found in 17 of 80 letters.

Of 80 FDA drug CGMP warning letters from 2021 to 2026 with a computer or electronic data finding, 17 (21%) cite trial or unreported injections, or testing into compliance. Most concern chromatography, but not all:

Table 2. The 17 letters by the instrument that held the unreported data; 7 of them also cite aborted runs or tests.
Where the unreported data wereLettersExamples
Chromatography data systems (HPLC, UPLC, GC)11Shimoga, Curia, Grupo Quimico, Unipack, Aspire
Particle counters and air samplers4Fareva, Somerset, Toyobo, Genzyme
Filter integrity tester1Aspen
Spectrophotometer1Persōn & Covey

Method: we searched FDA’s warning letter full-text index, which holds letters posted from January 2021, on 26 November 2026 for 35 phrases, among them “211.68”, “data integrity”, “audit trail” and “trial injection”; we read every numbered violation of 175 of the letters found, and kept the 80 drug CGMP letters (finished drugs, including OTC drugs, medical gases, veterinary drugs, a licensed biologic and contract testing labs, and APIs, issued 27 January 2021 to 10 November 2026) in which at least one numbered violation faults a computerized system or electronic data. Each letter counts once per category, and assigning a finding to a category is a reader’s judgment, recorded with a quote from the letter; FDA’s requests to the firm and its standard remediation text are not counted, nor are Form FDA 483s. We then read each of the 17 to see which instrument held the data. A letter names what FDA chose to cite, not all that inspectors saw.

The letters show how orphan data stay hidden: nobody compared what the system held with what was reported.

  • Shimoga, 2026: “Multiple instances of unreported sample injections for related substances and assay testing by high performance liquid chromatography (HPLC) were not documented in laboratory batch records.” One batch “had an unreported injection with an out-of-specification (OOS) assay result”, and because the quality unit did not review the electronic data, the batch was released to the US.10
  • Grupo Quimico, 2024: the quality unit “only reviews printed analytical results.” Investigators found unprocessed sample injections, and “When your firm subsequently processed the drug product sample injections after their discovery by the investigator, the injections were determined to be OOS for ingredient (b)(4) assay.”11
  • Curia, 2026: analysts “performed thousands of unexplained injections using Empower over the last 30 months”, under a “Developer” role. Of four on one day, “Your analyst confirmed these trial injections could not be traced to any product, sample, or analysis.”12
  • Unipack, 2025: a stability sample “was initially injected on April 29, 2025, with no reported results, then re-injected later that same day, with passing results subsequently reported.”13
  • Outside the CDS: at Fareva, “only conforming retest results were retained and included in production documentation” after Grade A particle counts failed; at Aspen, after failed filter integrity tests, “Your operators did not record or print the failing results.”14,15 MHRA warns that standalone instruments are where “it may be possible to manipulate data or repeat testing to achieve the desired outcome with limited opportunity for detection”.7

4How to review for orphan data.

Compare everything the system acquired with everything that was reported, and explain the difference. Two reviews do this at different depths. The steps are practice; the template gives each check’s basis.

With each result. The reviewer who approves a result opens the sequence and its audit trail, not only the report, and asks MHRA’s question: is this the only data generated for this test? Aborted injections, re-injections and reprocessing in the sequence are visible here. Audit trail review covers the method.

Across the system, periodically. The per-result review cannot see what never reached a result. A periodic sweep, by someone independent of the laboratory, starts from the system’s side:

  1. Take the list of acquisitions from the system itself: every sequence and injection in the period, across all projects and folders, with instrument, user, time, sample type and processing status.
  2. Take the list of what was reported in the same period, from the LIMS or the batch record, with the sequence each result came from.
  3. Match by sequence. Each acquired sequence belongs to a reported result, an approved protocol, a qualification record or a logbook entry, or it is unexplained.
  4. Within reported sequences, match by injection. Each sample injection is in the result, or excluded with a documented reason and kept.
  5. Read the audit trail for aborts, deletions and changes to the sequence after acquisition started, and check that each aborted run with samples in it was assessed.
  6. Sort what is left by user, instrument and day. One short sequence is a question; a pattern by one user before official runs is a finding. FDA asked Indoco for trending “including, but not limited to, specific users, methods, and equipment.”16
  7. Raise a deviation for each unexplained item, process the data with the approved method, assess it against the specification, and follow the OOS procedure if it fails.

Who and how often. No rule sets either. Practice is a QA reviewer independent of the laboratory, with read-only access, at an interval set by risk: monthly or quarterly for a CDS or instrument whose data decide release, and with the periodic review for systems that hold less. Include standalone instruments, whose data never reach the CDS, and record the review with “a positive statement regarding whether issues were found or not, the date that review was performed and the signature of the reviewer” (MHRA 6.15).7

5How it runs in Seal.

In Seal, Seal Edge watches a chosen folder on the instrument workstation and uploads supported exports to a configured Seal record, and each acquisition is a record with relationships: it links to a sample, the sample links to a batch, and an investigation points to the evidence it assesses. Reprocessing, reintegration and report generation create derived records, so an earlier result is not overwritten by a later one. You can ask Neil, Seal’s AI agent, which acquisitions support a batch or where an investigation is missing evidence; Neil follows the records the requester is permitted to see, and your team inspects the sources and reviews the conclusion.

To test this on your own data, the data integrity blueprint starts with one chromatographic assay, followed from sample and sequence through acquisition, processing, OOS assessment and batch disposition, and includes a deleted injection and a trial sequence.

6Orphan data review template.

The first table is the checklist for one review; the second decides each sequence or injection that no reported result accounts for.

Table 3. Orphan data review checklist, one system and period.9,8,1,7,2,6
CheckLook forBasis
ScopeEvery project, folder and instrument, for the whole periodGuidance: WHO Appendix 1, example 4
Reviewer and intervalIndependent of the laboratory, read-only; interval set by riskPractice
AcquiredA list from the system itself, with user, time and sample typeGuidance: PIC/S 9.8
ReportedEvery result from the period, with its sequenceRule: 211.194(a)
SequencesEach one linked to a result, protocol, qualification or logbookRule: 211.194(a)
Sample injectionsEach in the result, or excluded with a reason and keptGuidance: MHRA 6.10
Aborted runsA reason in the audit trail; samples in them assessedGuidance: FDA DI Q&A 12
ReprocessingEvery result kept, not only the lastGuidance: FDA DI Q&A 14
DeletionsNone, or each with its recordGuidance: PIC/S 9.8
PatternsShort sequences and aborts by user, instrument and dayPractice
Unexplained itemsA deviation each; data processed and assessedRule: 211.192
Sign-offIssues found or not, date, reviewer’s signatureGuidance: MHRA 6.15
Table 4. The decision for each sequence or injection the review finds outside a reported result.1,7,2,5,6,3
FoundAccounted for whenBasis
Standard, blank or system suitability injection in a reported sequenceThe result’s own review covers itRule: 211.194(a)
Sample injection in a reported sequence, not in the resultThe exclusion has a documented scientific reason, and the data are keptGuidance: MHRA 6.10
Sequence with no sample, for a documented purpose: equilibration, qualification, an approved protocolIts record is namedGuidance: FDA DI Q&A 13
Aborted or incomplete runThe reason is documented, and any sample injected before the abort is assessedGuidance: FDA DI Q&A 12
A sample injection outside a reported sequence, or any run with no documented purposeNever by itself: open a deviation, process the data, assess against the specification, and follow the OOS procedure if it failsRule: 211.160(a), 211.192; Guidance: FDA OOS

System suitability belongs to the sequence it qualifies, and uses standards, not the batch under test; HPLC system suitability covers the rules.

7Example: Exampla’s October 2026 review of ExamCDS.

Invented for illustration. Exampla Pharma Services LLC, a contract manufacturer and testing laboratory in Durham, North Carolina, runs chromatography on ExamCDS 3.4: 14 HPLC and 4 GC systems. Its data integrity procedure, QA-⁠DI-⁠001, requires a monthly orphan data review by QA of ExamCDS and of each instrument PC that stores results. On 4 November 2026 a QA reviewer, who has read-only access to ExamCDS, did the ninth, QA-⁠ODR-⁠009, for October.

On 2 November, IT’s system administrator exported every October acquisition from all 9 ExamCDS projects, and the reviewer exported every result ExamLIMS had received from ExamCDS, which carries the sequence number with each result. The September review, QA-⁠ODR-⁠008, had left nothing open.

Table 5. Example. QA-⁠ODR-⁠009, ExamCDS, 1 to 31 October 2026. The firm, systems and records are invented.
LineSequencesInjections
Acquired in October43810,512
In reported sequences: 5,102 sample injections and 4,582 standards, system suitability and blanks3919,684
Method validation or transfer, each under an approved protocol12486
Instrument qualification after maintenance6108
Column conditioning and equilibration, mobile phase only, in the logbook17153
Aborted, each with a reason in the audit trail1180
No documented purpose11

Within the reported sequences, 5,093 of the 5,102 sample injections were in reported results. The other 9 were excluded after a vial or air bubble error, each with the reason in the analyst’s worksheet, kept with the sequence and followed by a re-injection that was reported. Of the 11 aborted sequences, 8 stopped before any sample was injected. The other 3 stopped after 14 sample injections, two for pump pressure faults and one for a leak; each had a laboratory investigation, and the samples were re-run in reported sequences. The audit trail showed no deleted injection or sequence, and no sequence changed after its acquisition began.

The same review took the 6 instrument PCs that store results outside ExamCDS: 3 FTIR, 1 UV-⁠Vis and the Karl Fischer titrators KF-⁠01 and KF-⁠02. The two titrators joined the review in October, once the inventory check of 1 October had found them and DEV-⁠26-⁠136 had added them to the raw data register. Their 480 stored measurements were 431 in reported results and 49 checks in the instrument logbooks and the titer log; none was unexplained.

That left one ExamCDS injection. At 07:42 on Monday 19 October, an analyst had made a single injection on HPLC-⁠09, in a sequence named “chk” in a client’s project, from the vial of a 12-month stability sample for assay. It was never processed. The reported sequence for the same sample started on HPLC-⁠09 at 08:15, 33 minutes later, passed system suitability and gave an assay of 99.1% against a specification of 95.0% to 105.0%.

QA opened deviation DEV-⁠26-⁠149 on 4 November. The analyst explained that the column had been replaced that morning, and that the sample was injected to check its retention time instead of the system suitability standard. QA judged it a trial injection: an actual sample from the batch under test, injected outside the reported sequence with no written procedure allowing it, where FDA expects a sample used for system suitability to be “a properly characterized secondary standard” from a different batch.2 It found no sign of testing into compliance: nothing in the method, column or sample preparation changed between it and the reported sequence, and none of the analyst’s other short sequences lacked a purpose. QA processed the injection with the approved processing method, as a new result that leaves the acquisition unchanged: 98.6%, within specification. It was not OOS, but the injection and its result were added to the test record in ExamLIMS, so the record is complete. The reported result stands: its sequence passed system suitability, and the trial result agrees with it. The client was told under its quality agreement.

To set the scope, QA checked every sequence of 1 to 3 injections the analyst had acquired since February 2026: 23, each with a documented purpose. The 8 earlier monthly reviews had found no unexplained injection. The cause was a gap in the lab’s column change instruction, which said nothing about what to inject first. The CAPA, recorded under DEV-⁠26-⁠149, added the system suitability standard to it, the analyst was retrained, and the next 3 reviews will list every short sequence. QA closed DEV-⁠26-⁠149 on 20 November 2026, 16 days after opening it; the November review is due in the first week of December.

AQuestions and answers

Is orphan data always a data integrity violation?

No. Blanks, column conditioning, instrument qualification and runs under a validation protocol never feed a batch result, and they are acceptable when their purpose is documented and no sample was used. Orphan data become a violation when no record explains them, and especially when a sample was injected outside the reported sequence.

What if an orphan injection shows an out-of-specification result?

Treat it as an OOS result for that sample. Open a deviation, start the laboratory investigation under your OOS procedure, and consider the result in the batch decision. Without an assignable laboratory error it cannot be invalidated in favor of the reported passing result, and released batches it affects need an impact assessment.

Who should do a periodic orphan data review?

Someone independent of the laboratory whose data are reviewed, usually QA, with read-only access to the system and its audit trail. Analysts and their supervisors review their own sequences with each result, but the periodic sweep looks for what those reviews could miss, so it should not depend on them.

Does orphan data apply outside chromatography?

Yes. Any instrument that stores results can hold some that were never reported: particle counters, air samplers, filter integrity testers, spectrophotometers, Karl Fischer titrators and balances with memory. In FDA’s letters counted on this page, 6 of the 17 that cite unreported data concern instruments other than chromatography.

Can a CDS be configured to prevent orphan data?

It can make them harder to create and easier to find. Remove roles that let analysts run outside the approved methods, require each sequence to start from an approved template with system suitability first, save each injection on completion, and keep audit trails on. Configuration does not replace the review, because an injection can still be made for a reason no one recorded.

References16 sources

References

  1. 121 CFR 211.194, Laboratory records, paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-211.194 (read 26 November 2026)
  2. 2Data Integrity and Compliance With Drug CGMP: Questions and Answers, guidance for industry, questions 1, 12, 13 and 14, FDA, December 2018. https://www.fda.gov/media/119267/download (read 26 November 2026)
  3. 3Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production, guidance for industry, revision 1, section IV.B.1, FDA, May 2022. https://www.fda.gov/media/158416/download (read 26 November 2026)
  4. 4Warning letter, Aspire Pharmaceuticals, Inc., MARCS-⁠CMS 630328, CGMP violation 3, FDA, 22 November 2022. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/aspire-pharmaceuticals-inc-630328-11222022 (read 26 November 2026)
  5. 521 CFR 211.160, General requirements (laboratory controls), paragraph (a), eCFR. https://www.ecfr.gov/current/title-21/section-211.160 (read 26 November 2026)
  6. 621 CFR 211.192, Production record review, eCFR. https://www.ecfr.gov/current/title-21/section-211.192 (read 26 November 2026)
  7. 7‘GXP’ Data Integrity Guidance and Definitions, revision 1, sections 4.3, 6.10 and 6.15, MHRA, March 2018. https://assets.publishing.service.gov.uk/media/5aa2b9ede5274a3e391e37f3/MHRA_GxP_data_integrity_guide_March_edited_Final.pdf (read 26 November 2026)
  8. 8PI 041-⁠1, Good Practices for Data Management and Integrity in Regulated GMP/GDP Environments, section 9.8, PIC/S, 1 July 2021. https://picscheme.org/docview/4234 (read 26 November 2026)
  9. 9Guideline on data integrity, WHO Technical Report Series No. 1033, Annex 4, Appendix 1, example 4, World Health Organization, 2021. https://cdn.who.int/media/docs/default-source/medicines/norms-and-standards/guidelines/inspections/trs1033-annex4-guideline-on-data-integrity.pdf (read 26 November 2026)
  10. 10Warning letter, Shimoga Chemicals, MARCS-⁠CMS 727904, data integrity concerns, FDA, 13 July 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/shimoga-chemicals-727904-07132026 (read 26 November 2026)
  11. 11Warning letter, Grupo Quimico SRL, MARCS-⁠CMS 674010, violation 1, FDA, 30 April 2024. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/grupo-quimico-srl-674010-04302024 (read 26 November 2026)
  12. 12Warning letter, Curia New York, Inc., MARCS-⁠CMS 728790, violation 1, FDA, 18 September 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/curia-new-york-inc-728790-09182026 (read 26 November 2026)
  13. 13Warning letter, Unipack LLC, MARCS-⁠CMS 716621, violation 2, FDA, 19 December 2025. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/unipack-llc-716621-12192025 (read 26 November 2026)
  14. 14Warning letter, Fareva Amboise, MARCS-⁠CMS 723502, violation 3, FDA, 10 April 2026. https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/fareva-amboise-723502-04102026 (read 26 November 2026)