Skip to main content

21 CFR Part 11 and Annex 11 checklist.

Mark each control against eCFR Part 11 and EU GMP Annex 11 to see gaps, inspection focus and a Word assessment record.

The system and scope

Assessment scope

Access is controlled by the people responsible for the electronic records (21 CFR 11.3(b)(4)). Assesses technical controls alongside site SOPs and periodic review.

Controls checklist

Governance, risk and validation

No open gaps

  1. Risk assessment and GMP system inventory

    Part 11: FDA 2003 scope decision Annex 11: Sections 1, 4.3

    Documented risk assessment setting validation depth and data integrity controls, plus a system inventory entry recording GMP criticality and validation status.

    Record which predicate rules require each electronic record held by the system.

    Status for Risk assessment and GMP system inventory
  2. Validation for intended use and IT qualification

    Part 11: 11.10(a) Annex 11: Principle, Section 4

    Application validated for accuracy, reliability, consistent intended performance and detection of invalid or altered records; IT infrastructure qualified.

    Verify each configured workflow and calculation against your approved user requirements.

    Status for Validation for intended use and IT qualification
  3. Supplier agreements, documentation and change control

    Part 11: 11.10(k)(1), (k)(2) Annex 11: Sections 3, 10

    Controlled system documentation, formal change and configuration control, and written agreements defining responsibilities with suppliers and internal IT.

    Keep configuration changes and documentation revisions under one approved change control procedure.

    Status for Supplier agreements, documentation and change control
  4. Personnel qualification, roles and training

    Part 11: 11.10(i) Annex 11: Section 2

    Defined process owner, system owner, QA and IT responsibilities; users, developers and maintainers hold documented education, training and experience.

    Confirm training is complete before granting production write or approval privileges.

    Status for Personnel qualification, roles and training

Access control and security

2 gaps

  1. Authorised access and unique personal accounts

    Part 11: 11.10(d), 11.300(a) Annex 11: Section 12.1, 12.3, 12.4

    System access restricted to authorised individuals with unique user IDs; no shared writable accounts; grants, modifications and leaver removals recorded.

    Prohibit shared writable logins across instruments, workstations and application accounts.

    Status for Authorised access and unique personal accounts
  2. Role authority checks and administrator independence

    Part 11: 11.10(g) Annex 11: Section 12.1

    Technical authority checks restrict who can use the system, sign, alter a record or perform an operation; administrators are independent of the record content, as FDA’s 2018 Q&A asks.

    Separate IT administration from QA, laboratory and manufacturing users responsible for record content.

    Status for Role authority checks and administrator independence
  3. Operational sequence, device and accuracy checks

    Part 11: 11.10(f), (h) Annex 11: Sections 5, 6

    Operational checks enforce permitted event sequences and device checks validate input sources where appropriate; critical manual entries receive a second or validated check.

    Document why 11.10(f) or 11.10(h) is N/A if a workflow does not require enforced sequencing or terminal checks.

    Status for Operational sequence, device and accuracy checks
  4. Open-system determination and encryption controls

    Part 11: 11.30 Annex 11: Section 12

    Documented closed-vs-open determination; where external parties control access, additional safeguards such as document encryption and digital signature standards apply.

    Mark N/A only with a written rationale confirming your organisation controls all access to the system.

    Status for Open-system determination and encryption controls

Audit trails and data integrity

1 partial

  1. Secure, time-stamped audit trails with reason for change

    Part 11: 11.10(e) Annex 11: Section 9

    Computer-generated, time-stamped audit trails independently record creation, modification and deletion without obscuring previous entries, capturing reason for GMP changes.

    Verify audit trails are enabled, cannot be disabled by users and remain retained as long as the parent record.

    Status for Secure, time-stamped audit trails with reason for change
  2. Audit trail review and synchronised system clock

    Part 11: 11.10(e), 211.192 Annex 11: Sections 9, 12.4

    Audit trails are reviewed with the record, before release where the record must be; clocks are controlled and synchronised, as MHRA’s 2018 guidance (5.1) has them.

    Define which audit trails require batch-by-batch review versus periodic review based on risk.

    Status for Audit trail review and synchronised system clock

Electronic signatures

1 gap

  1. Signature manifestation: name, date/time and meaning

    Part 11: 11.50 Annex 11: Section 14

    Signed records show the signer’s printed name, the date and time of signing, and its meaning, on screen and on any printout. Annex 11 asks for the time and date.

    Verify both on-screen displays and exported PDFs/printouts carry all three 11.50 elements.

    Status for Signature manifestation: name, date/time and meaning
  2. Permanent signature-to-record linking

    Part 11: 11.70 Annex 11: Section 14(b)

    Electronic and handwritten signatures executed to electronic records are permanently linked so they cannot be excised, copied or otherwise transferred to falsify a record by ordinary means.

    Test that editing a signed record either blocks changes or creates a new version requiring fresh approval.

    Status for Permanent signature-to-record linking
  3. Unique attribution, accountability policy and FDA letter

    Part 11: 11.10(j), 11.100(a)-(c) Annex 11: Sections 14(a), 15

    Each electronic signature is unique to one verified individual and never reassigned; signers acknowledge accountability in writing; FDA receives a non-repudiation letter; EU batch release is restricted to Qualified Persons.

    Verify disabled user IDs cannot be reassigned to a new employee with the same name.

    Status for Unique attribution, accountability policy and FDA letter
  4. Two-component signing and credential controls

    Part 11: 11.200(a)-(b), 11.300(a)-(e) Annex 11: Sections 12, 14

    Non-biometric signatures use at least two distinct components (such as ID and password) with session re-entry rules, anti-collusion password resets, periodic review, lockout safeguards and urgent misuse reporting.

    Test both continuous-session repeat signing and post-timeout full re-authentication.

    Status for Two-component signing and credential controls

Copies, backup and archiving

1 partial

  1. Accurate inspection copies and release printouts

    Part 11: 11.10(b) Annex 11: Section 8

    Ability to generate accurate, complete human-readable and electronic copies (including metadata and audit trails) for inspection; batch-release printouts show whether data were changed since entry.

    Test exporting a complete record with its signatures, attachments and audit trail in a standard readable format.

    Status for Accurate inspection copies and release printouts
  2. Backup, tested restore, retention and archiving

    Part 11: 11.10(c) Annex 11: Sections 7.1, 7.2, 17

    Regular backups with periodic restore verification; archived records protected for accessibility, readability and integrity throughout retention and across system changes.

    Test restoring actual production-like data and confirm timestamps, signatures and audit trails remain intact.

    Status for Backup, tested restore, retention and archiving

Periodic evaluation and continuity

No open gaps

  1. Periodic evaluation of validated state

    Part 11: Not stated Annex 11: Section 11

    Scheduled periodic evaluation confirming the system remains valid and compliant, reviewing functionality, changes, deviations, incidents, upgrades, performance, reliability and security.

    Include user-access reconciliation, leaver checks, restore tests and open CAPAs in each review cycle.

    Status for Periodic evaluation of validated state
  2. Incident management and business continuity

    Part 11: 11.300(c), (d) (credentials) Annex 11: Sections 13, 16

    All incidents reported and assessed, with root cause analysis of critical incidents feeding CAPA; documented and tested continuity fallback for systems supporting critical processes.

    Test the fallback procedure and define how data recorded during an outage are entered and second-checked afterwards.

    Status for Incident management and business continuity

Readiness summary for CDS-⁠02 Chromatography Data System v5.4

18 of 18 controls assessed: 12 in place, 2 partial, 3 gaps, 1 N/A.

Control groupIn placePartialGapN/AOpen
Governance, risk and validation40000
Access control and security10210
Audit trails and data integrity11000
Electronic signatures30100
Copies, backup and archiving11000
Periodic evaluation and continuity20000

Priority remediation

PriorityControlStatusInspection focus and evidence
1Authorised access and unique personal accountsPart 11: 11.10(d), 11.300(a) Annex 11: Section 12.1, 12.3, 12.4GapShared LAB login still enabled on two HPLC workstations.FDA: with shared credentials “a unique individual cannot be identified”. A shared read-only account may view data, never act on it (2018 Q&A 5). Expected evidence: Current user access list, approval records for grants and removals, leaver revocation check and periodic access review.
2Role authority checks and administrator independencePart 11: 11.10(g) Annex 11: Section 12.1GapThree analysts hold administrator rights and can edit methods.FDA: administrator rights should go “to personnel independent from those responsible for the record content” (2018 Q&A 4). Expected evidence: Role permission matrix, negative testing showing refused actions, and separation between IT administrators and lab or QA users.
3Signature manifestation: name, date/time and meaningPart 11: 11.50 Annex 11: Section 14GapPrinted report shows user ID instead of the printed name.A user ID isn’t a printed name, and 11.50(b) wants all three on “any human readable form”, so check the printout as well as the screen. Expected evidence: On-screen view and PDF or paper printout of a signed record showing the name, timestamp, meaning (such as review or approval) and, across sites, the time zone.
4Secure, time-stamped audit trails with reason for changePart 11: 11.10(e) Annex 11: Section 9PartialAudit trail locked, but reason for change is optional.Annex 11 section 9 weighs the audit trail by risk but wants a reason for every change or deletion; in EU law “audit trails shall be maintained” (Directive 2017/1572, Art. 9(2)). Expected evidence: Locked audit-trail configuration screenshot/export and sample entries showing user, timestamp, old value, new value and reason.
5Backup, tested restore, retention and archivingPart 11: 11.10(c) Annex 11: Sections 7.1, 7.2, 17PartialNightly backups running; annual restore test overdue by 2 months.One EU statement of non-compliance lists a restore check “not performed in 2025, although it was SOP requirement” (EudraGMDP NNGYK/11120-1/2026). Expected evidence: Backup schedule, dated restore test report matching record counts/checksums, retention schedule and archive retrieval check.

Where each control comes from.

Each row maps to 21 CFR Part 11 (11.10(a) to (k), 11.30, 11.50, 11.70, 11.100, 11.200 and 11.300, as amended 2 March 2023) and EU GMP Annex 11 (revision 1, Principle and sections 1 to 17).

FDA’s 2003 guidance doesn’t intend to enforce Part 11’s own validation, audit trail, retention and copying requirements, but enforces the rest, and every rule that requires the record. Part 11 requirements marks each clause; the 2018 Q&A says what FDA expects under Part 211.

Worked: one gap, from clause to closure.

Example, all invented: the shared login in the opening assessment. Closing it takes more than a new account, because the records made under it stay.

StepEntry
Control and clauseAuthorised access: 11.10(d) and 11.300(a); Annex 11, 12.1; 21 CFR 211.68(b)
What was foundLogin LAB shared by six analysts on HPLC-⁠07 and HPLC-⁠11, with rights to process and delete
Records it touches14 months of release and stability work: 412 sequences on HPLC-⁠07 and 388 on HPLC-⁠11, 800 in all
The same dayLAB disabled; six personal accounts with the analyst role, none able to delete
Who did whatLAB’s entries matched to the instrument logbook: 791 sequences to one analyst, 9 to none, taken to a deviation
Whether results changedAudit trails of all 800 read: 37 reprocessed, 33 with a reason that stands and 4 without, in the same deviation
ProductThe 4 and the 9 fall in three released lots; QA assesses each under 21 CFR 211.192
Why it happenedThe instrument’s set-up script made the account, and the access review read the directory, not the instruments
Proof it is closedScript corrected under change control; local accounts on all 14 instruments listed and reviewed; a LAB sign-in refused on each
ApprovedLaboratory head and QA, with the next access review’s date

Where a checklist stops.

It assesses one system in one use. A supplier’s features answer some rows. Procedures, training and the certification letter answer others, and only you can.

Scope comes first. Part 11 reaches records another FDA rule requires. Record which those are, and whether people rely on the screen or the printout, before marking a row.

“N/A” is a finding too. 11.10(f) and (h) apply “as appropriate” and 11.30 to open systems; each N/A needs its reason.

It is Annex 11 as in force. The Annex 11 revision says what the 2025 draft would add, and Part 11 and Annex 11 compared where the two differ today.

Sources.

Each was read on 24 November 2026.

Keep the evidence as the work runs.

In Seal every record carries version history, locked audit trails, role-based access and signed approvals, with supplier evidence ready for your review.