MACO, and the swab and rinse limits it allows.

Enter the PDE and the changeover, and see each step from MACO to the swab and rinse limits, with a limit sheet for Word.

The changeover

µg/day
kg
mg/day
m²

Swab

cm²
%
mL

Rinse

m²
L
%

Limits for Quoravir into Pellastine

StepCalculationResult
MACO5 µg/day × 150 kg ÷ 2,000 mg/day0.375 g, or 375 mg
MSSR0.375 g × 1,000 ÷ 30 m²12.5 mg/m², or 1.25 µg/cm²
Acceptance limit1.25 µg/cm² ÷ 100.125 µg/cm²
Per swab, all recovered0.125 µg/cm² × 100 cm²12.5 µg
Swab limit, as reported12.5 µg × 80% recovery10 µg per swab, or 1 µg/mL in 10 mL
In the rinse, all recovered0.125 µg/cm² × 2 m² ÷ 25 L2,500 µg, or 0.1 µg/mL
Rinse limit, as reported0.1 µg/mL × 90% recovery0.09 µg/mL

Where each number comes from.

MACO is the PDE of A × the smallest batch of B ÷ the maximum daily dose of B. Spread over the surface the two share, it is the maximum safe surface residue (MSSR). WHO TRS 1033 Annex 2 names both, calling MACO the maximum safe carryover. The PDE comes from the HBEL report for A, set as EMA’s guideline on health-based exposure limits describes.

The acceptance limit is the MSSR divided by a safety factor you choose. The example keeps 10, which PIC/S PI 052-1 says “may be reasonable to provide confidence that permitted exposure is not breached”. Recovery is applied once, to the limit, so a result is compared as reported, and the limit sheet’s sampling basis says so, citing WHO’s section 5.8. How to calculate MACO works the same example, with the choices behind each figure.

Keep each limit with its inputs.

In Seal a cleaning limit is a governed calculation: its PDE, batch sizes, surface and safety factor stay with it, so a changed input shows what it reaches.