Your FMEA, scored and ranked.
Score each failure mode for severity, occurrence and detection, and see its RPN, its rank and what the rules say to do, ready for Word.
The assessment
Failure modes
- F1
- F2
- F3
- F4
Ranking for Line 3, Exemplar 10 mg/mL injection
| Rank | Failure mode | S × O × D | Decision |
|---|---|---|---|
| 1 | F1 RABS glove pinhole: ingress into Grade A | 5 × 4 × 3 = 60 | ActRPN 40 or more |
| 2 | F4 Vial neck crack: closure integrity lost | 5 × 1 × 4 = 20 | Actseverity 5 with detection 4 |
| 3 | F3 Low fill volume: under-filled vials | 4 × 3 × 3 = 36 | Act, or accept with a rationale QA approves |
| 4 | F2 Stopper not seated: closure integrity lost | 5 × 2 × 2 = 20 | Act, or accept with a rationale QA approves |
Scales and rules
| Score | Severity | Occurrence | Detection |
|---|---|---|---|
| 5 | Could harm a patient, such as loss of sterility | More than once in 10 batches | Not detectable |
| 4 | Batch fails a specification | Once in 10 to 100 batches | Found only by chance, or by a check not proven for it |
| 3 | Within specification, a quality attribute affected | Once in 100 to 1,000 batches | Likely found by a periodic or sampled check |
| 2 | GMP non-conformance, no product effect | Rarer than once in 1,000 batches | Found by a validated check of every unit or batch |
| 1 | No noticeable effect | None in over 5,000 batches of relevant data | Always found: a validated automatic check stops the line |
- Controls a regulation or GMP guideline requires are in place before scoring and are never accepted away by an RPN.
- Act when the RPN is 40 or more, or when severity is 5 and detection 4 or 5.
- From 20 to 39, act or record a rationale for accepting, approved by QA.
- Below 20, accept and review at the next periodic review.
Where each number comes from.
The RPN is severity × occurrence × detection, as WHO’s quality risk management guideline (TRS 981 Annex 2) defines it. Rows are ranked by what the rules say to act on, then by RPN, then by severity and detection.
Neither ICH Q9(R1) nor WHO sets the scales or the thresholds. These are the example ones of FMEA and the RPN, QRM-SC-02 v3, which scores the same line: replace them with your own approved scales and rules.
Keep the FMEA with its evidence.
In Seal each failure mode keeps its causes, controls and the basis for each score, so a failed verification or a complaint links back to the row it reopens.