Skip to main content

2 software routes

METTLER TOLEDO instrument integration

METTLER TOLEDO integrations connect measurements managed in LabX and thermal analysis from STARe, two applications with different result structures.

Illustration of a seal beside a METTLER TOLEDO XPR205 analytical balance.

METTLER TOLEDO instruments to SealInstrument export

Seal reads
Exports from LabX and STARe
Writes back
No
Source of truth
Instrument software
Set up
Seal Edge and an upload Script

1Choose the source software and evidence.

Each route starts from a documented output of the vendor software. Check the installed version, licensed options and export settings against the linked docs. A listed route is not an acquisition driver or a universal parser.

METTLER TOLEDO LabX connection overview →

CSV or XML through the LabX Auto Import / Export option.

One weighing or analytical task with typed results and source evidence; relate it to the sample or material record.

METTLER TOLEDO STARe connection overview →

STARe exported curves and evaluation reports; verify the configured data-export option and layout.

One sample thermal evaluation with selected transition or mass-change results and original curve evidence.

2Verify the installed LabX or STARe interface.

LabX export options depend on the configured application and licensed capabilities. Confirm the source instrument, result type and export layout rather than assuming every instrument connected to LabX produces an identical table. Keep a representative export with its matching report and the software configuration needed to reproduce it.

STARe workflows involve curves and evaluated thermal-analysis results. Confirm the installed data-export option and the chosen report layout. A native project, a plotted curve and a selected evaluation table serve different purposes; identify which is suitable for mapping and which should be retained as source evidence.

3Keep units, method context and the source result together.

Laboratory results need the sample or material identity, the measurement unit and the relevant method or task context. Preserve the source result identifier so repeated collection can be recognised. Where the source reports several values for one task, map their distinct meanings explicitly instead of selecting a numeric column by position.

For thermal analysis, preserve the curve axes, method program and evaluation context available in the source output. A selected transition or peak result depends on how the measurement and evaluation were performed. Keep the original curve linked to the evaluated value so interpretation does not depend on a number stripped of its experimental setting.

4Review calculations and source revisions before using results.

Decide which calculations belong in the vendor application and which are explicitly performed by the receiving Script. Preserve reported values and units before any transformation. A unit conversion or normalisation should have a defined rule and a test example, especially where a result is later compared with a specification.

Check an identical export, a corrected source result and a changed report layout. The first should not duplicate the measurement; the second needs traceable revision handling; the third may require an updated mapping. Keep those outcomes separate from the review or release decision made in your operational workflow.

5Example: connect a laboratory task and thermal characterisation

The same material can have several measurements without losing the source and method behind each one.

Material
Resolve the existing sample or material reference for the laboratory work.
LabX result
Retain the source task, result identity, reported unit and original export.
Thermal result
Link the STARe curve and evaluation with its method context and selected typed values.

Use these relationships as a mapping example for your own templates. The receiving fields and record grain are configured through Seal Scripts and APIs.

6Configure the collection and mapping for your site.

In the laboratory

METTLER TOLEDO instruments

  • LabXSaves CSV or XML through the LabX Auto Import / Export option
  • STAReSaves STARe exported curves and evaluation reports

Export folder

On the workstation or a network share

Seal Edge

Waits until each file is complete, then sends it

Outbound HTTPSNo inbound connection

Nothing is written back to the instrument or its software

In Seal

Seal Script

Checks the file’s layout and matches it to its record

Record

Weighing

  • Net mass
  • Tare
  • Balance
  • Original file, kept
How a result from METTLER TOLEDO software reaches Seal.

For local exports, use a completed-file handoff on the instrument workstation or another approved site computer. Seal Edge collects the file and uploads it to the chosen Seal destination over outbound HTTPS. Keep source files until delivery and mapping have been verified.

A Seal Script can validate the expected layout, resolve source identities and write selected typed fields. Keep the original export as File content. Acquisition and vendor analysis remain in the source application; Seal collects the completed results and does not send instructions to the instrument.

Follow the on-site equipment setup instructions, then use the source-specific guide above to test the actual export and mapping. For a result exchange with an existing LIMS, also define that system’s supported receiving operation and sample identifiers.

7Test the configured workflow with known evidence.

  1. Compare a known result, its reported unit and any configured conversion against the source report.
  2. Verify the axes and evaluation context associated with a selected thermal-analysis result.
  3. Confirm that corrected results and changed export layouts are handled explicitly and preserve source evidence.

Keep the test outcome with the configuration and repeat the relevant checks when software, reports, analysis settings or destination fields change.

Discuss your METTLER TOLEDO workflow with a representative export, the installed software version and the records you want to connect.

8The METTLER TOLEDO models.

These are METTLER TOLEDO instruments laboratories connect to Seal, listed from the manufacturer’s own documents. A model not listed works the same way if its software saves a file.

Balances

Weighings: the net mass with its unit, and the tare where one was used.

  • LA
  • LA84E
  • LA104E
  • LA203E
  • LA204E
  • LA403E
  • LA2002E
  • LA4002E
  • MA
  • MA54
  • MA55
  • MA95
  • MA103
  • MA104
  • MA203
  • MA303
  • MA503
  • MA602
  • MA602P
  • MA1002
  • MA2002
  • MA2002P
  • MA3002
  • MA4002
  • MA5001
  • MA6001P
  • MA6002
  • MA12001L
  • MA16001L
  • MA32000L
  • MA32001L
  • ME
  • ME-T
  • ML
  • ML-T
  • MR
  • MR104
  • MR203
  • MR204
  • MR303
  • MR304
  • MR503
  • MR603
  • MR1002
  • MR2002
  • MR3002
  • MR4002
  • MR6001
  • MR6002
  • MS Semi-Micro
  • MS-L
  • MS-S
  • MS-TS
  • MX
  • MX104
  • MX105
  • MX105DU
  • MX204
  • MX205DU
  • MX303
  • MX303N
  • MX304
  • MX603
  • MX603N
  • MX1203
  • MX1203N
  • MX2002
  • MX4002
  • MX6001
  • MX6002
  • MX6002DR
  • MX8001
  • MX12001L
  • MX12002
  • MX16001L
  • MX32000L
  • MX32001L
  • PL-E
  • XP
  • XPE
  • XPR2U
  • XPR3
  • XPR6U
  • XPR6UD5
  • XPR10
  • XPR10U
  • XPR26PC
  • XPR36
  • XPR36DR
  • XPR56
  • XPR56DR
  • XPR104DUE
  • XPR105
  • XPR105DUE
  • XPR106DUH
  • XPR106DUHQ
  • XPR204
  • XPR204DUE
  • XPR204E
  • XPR204S
  • XPR205
  • XPR205DUE
  • XPR225DR
  • XPR225DUE
  • XPR226DR
  • XPR226DRQ
  • XPR303S
  • XPR303SN
  • XPR304E
  • XPR305DR
  • XPR305DRQ
  • XPR404S
  • XPR504S
  • XPR603S
  • XPR603SN
  • XPR1202S
  • XPR1203S
  • XPR1203SN
  • XPR2002S
  • XPR2003S
  • XPR3003S
  • XPR3004SD5
  • XPR4001S
  • XPR4002S
  • XPR5003S
  • XPR6001S
  • XPR6002S
  • XPR6002SDR
  • XPR6003SD5
  • XPR8001S
  • XPR8002S
  • XPR12001L
  • XPR12001S
  • XPR12002S
  • XPR12003SD5
  • XPR15002L
  • XPR16001L
  • XPR20002LDR
  • XPR32000L
  • XPR32001L
  • XPR64000L
  • XPR64001L
  • XS
  • XSE

Titrators

The titrant volume to each endpoint and the content calculated from it, with the titration curve.

  • Compact G10S
  • Compact G20S
  • EasyPlus Easy Cl
  • EasyPlus Easy EPM
  • EasyPlus Easy Ox
  • EasyPlus Easy pH
  • EasyPlus Easy Pro
  • Excellence T5
  • Excellence T7
  • Excellence T9

pH meters

pH with temperature, and the calibration slope and offset it was measured against.

  • FiveEasy F20
  • FiveEasy Plus FP20
  • FiveGo F2
  • NineFocus NF2000
  • NineFocus NF5000
  • Seven2Go S2
  • Seven2Go S8
  • SevenDirect SD20
  • SevenDirect SD50

Conductivity meters

Conductivity, corrected to a reference temperature.

  • FiveEasy F30
  • FiveEasy Plus FP30
  • FiveGo F3
  • NineFocus NF3000
  • Seven2Go S3
  • Seven2Go S7
  • SevenDirect SD30

Density meters

Density at the measuring temperature, and values derived from it such as specific gravity or concentration.

  • Densito
  • DensitoPro
  • Easy D30
  • Easy D40
  • Excellence D4
  • Excellence D5
  • Excellence D6

Multiparameter meters

pH, conductivity, dissolved oxygen or other readings from the probes connected.

  • NineFocus NF2300
  • NineFocus NF2340
  • SevenDirect SD23
  • SevenGo Duo SG23
  • SevenGo Duo SG68
  • SevenGo Duo SG78
  • SevenGo Duo SG98

Karl Fischer titrators

Water content for each sample, from the titrant or charge needed to reach the endpoint.

  • EasyPlus Easy KFV
  • EVA C1
  • EVA C3
  • EVA V1
  • EVA V3

Melting point apparatus

The melting point or range of each sample.

  • MP30
  • MP55
  • MP70
  • MP80
  • MP90

Moisture analyzers

Moisture or dry content as a percentage, from the mass lost on heating.

  • HC103
  • HE53
  • HE73
  • HS153
  • HX204

Refractometers

Refractive index at the measuring temperature, and values derived from it such as Brix.

  • Easy Brix
  • Easy R40
  • Excellence R4
  • Excellence R5
  • MyBrix

Thermal analyzers

Heat flow, mass or dimension against temperature, evaluated into transitions, mass loss or expansion.

  • DMA 1
  • TGA 2
  • TGA/DSC 3+
  • TMA/SDTA 2+

DSC

Heat flow against temperature, evaluated into transitions such as melting temperature and enthalpy.

  • DSC 1
  • DSC 3
  • DSC 5+

Other

  • DP70
  • DP90
  • FiveGo F4
  • NineFocus NF4000
  • Seven2Go S4
  • Seven2Go S9