Agilent BioTek Gen5

Instrument Connection

Connect plate-reader results to assay, potency and process-development experiments.

Put this connection to work Set up Agilent BioTek Gen5 in the docs
Illustration of a seal following a sample vial to an analytical instrument and its result.

Microplate assays / Data integration

From instrument evidence to connected records.

Connect plate-reader results to assay, potency and process-development experiments. Seal connects the exported evidence to the experiments, samples and review workflows your team already uses.

  1. ExportThe vendor software saves the measurement or evaluated result.
  2. CollectSeal IoT on the site computer uploads the completed file.
  3. MapA Seal Script validates the data and writes your chosen fields.

Instrument acquisition, analysis and control stay in the vendor software. This connection uses the existing Seal file and entity APIs. Configure the collection and mapping for your installed software and export layout; selecting the instrument in the directory does not install a parser or start acquisition.

What evidence can Agilent BioTek Gen5 provide?

Source
Agilent / Agilent BioTek Gen5
Evidence to collect
Configured Excel or text-file exports from Gen5 experiments and protocols.

Export options depend on the installed software and licences. The Agilent BioTek Gen5 setup guide covers source preparation and collection in Seal.

Agree the export configuration before writing the Script. A report layout intended for printing may differ from the file export. Required column changes should produce a clear validation error before any result records are written.

Keep one representative export and its matching instrument report. Record the software version, licensed export options, chosen columns or sheets, and which results the file actually contains. Check the layout after a software or report-template change.

Vendor reference: Agilent BioTek support: Gen5 Excel and file export configuration. Product pages describe the source software; verify detailed export options against the manual for your installed version.

Choose the records you need.

One plate assay linked to sample-well results, with chosen endpoint or kinetic summaries and the original exports.

These are suggested fields, not a required schema. Reuse your existing sample, experiment, equipment and method references. Preserve complete exports as File content; store selected values as typed fields when people need to filter, compare, calculate or review them.

Suggested field mapping for Agilent BioTek Gen5
InformationSeal fieldsMapping rule
Plate and wellReference and textResolve sample identity within each experiment and plate.
Raw responseNumber and unitKeep read mode, channel or wavelength and timepoint explicit.
Calculated assay resultNumber and referenceRetain the curve, blank correction and dilution used.
Protocol evidenceFile and textLink protocol settings and the calculation revision to the result.

Context to retain: Experiment, plate, well and sample identity, read mode, wavelengths, time or kinetic cycle, blank correction, standard curve, dilution and calculation revision. A missing, failed or qualified result should stay distinguishable from a valid zero. Require explicit sample identities and reject unexpected layouts before the Script writes results.

Ready to configure the connection?

The platform docs contain the setup steps: prepare the source export, choose your System and template, configure collection, and test a mapping Script against your actual fields.

Follow the Agilent BioTek Gen5 setup guide ↗

Start with one representative export and its matching vendor report. The on-site equipment docs cover IoT installation, credentials and the first upload. The instrument mapping docs explain File content, typed records and ordinary Seal Scripts.

The site computer connects to Seal over outbound HTTPS. A cloud Script cannot read the instrument’s local drive. Collection frequency follows the export workflow: a folder watcher observes saved files and does not request new measurements.

Keep the source files until delivery and mapping are verified. For frequent or large datasets, choose useful summaries or intervals, retain coverage information and measure the intended volume before unattended operation.

Use the evidence in your existing workflow.

Imported records use normal Seal permissions and references. Compare typed results in views and formulas, link the analysis to a procedure, and use @ references to give neil the accessible samples, experiments and source evidence needed for a question.

Review these Gen5 assay results by sample. Show blanks, standards, replicate consistency and the calculation behind each concentration.

neil can help write the mapping Script from a representative file and the actual fields in your System. Ask for a proposed mapping and a dry run first. Review its sample matching and scientific assumptions before allowing writes. An instrument name alone does not define a file layout or an analysis method.

Define which fields the import owns. Preserve user-entered notes, review decisions and formula-controlled values. Keep source acquisition time, analysis revision and import time separate so later reanalysis remains understandable.

Prove the workflow with representative data.

A useful first acceptance check

Import a plate with standards, blanks, replicates and kinetic readings. Verify well mapping and raw-versus-calculated values, then test a changed export layout.

  1. Compare the result, not only the upload. Check the destination sample, selected values, units, qualifiers and linked File against the vendor report.
  2. Replay identical input. Verify the Script recognizes the source identity and revision before creating another result. File-upload calls alone do not prevent duplicate scientific records.
  3. Reprocess the analysis. Export a changed evaluation and check that it retains its prior evidence. Decide explicitly whether this creates a linked analysis or updates allowed fields under your normal review rules.
  4. Interrupt and recover. With disposable data, test a network interruption and restart. Also test failure after upload but before mapping completes. Confirm retained files, visible failures and correct retry behavior for the installed IoT app and Script.

Reconcile the source run or sample list with the imported analyses. A watcher cannot recover measurements the instrument never exported. Retain the checks and their evidence with the configured connection before relying on it for your intended use.

When the result is missing or looks wrong.

No file arrives
Confirm the instrument created a completed export in the watched folder. Check the IoT account’s file access, upload logs, destination System and template permissions.
The File exists, but result fields are empty
Check whether the mapping Script ran and recognized the export. Resolve missing sample identities or fields explicitly; uploading bytes does not populate an analysis automatically.
The values differ from the instrument report
Raw absorbance, fluorescence or luminescence is not a concentration. Preserve the blank, standard curve and dilution used for calculated results, and keep kinetic timepoints distinct from endpoint summaries. Compare the exact export, report and evaluation revision using the field rules above.
Results are repeated or an analysis is missing
Check export selection, source identities and revision matching. Reconcile at the analysis level; a filename or successful network response alone cannot establish complete coverage.