How do you get ISO 13485 certification, how long does it take, and what does it cost?

ISO 13485 certification comes from a certification body, not ISO: a Stage 1 readiness audit, a Stage 2 audit of the whole system, then a three-year certificate with yearly surveillance. Vendor planning guides put the whole project at four to 18 months and, for a small company, €15,000 to €100,000 all-in. In a worked example, a 45-person manufacturer takes nine months from procedures taking effect to certificate; its certification fees at NSAI’s rates are about €27,800 over three years.

Illustration of a seal examining a circled change in a trace alongside run, sample and procedure records.

1ISO 13485 certification, step by step.

ISO writes the standard but “does not perform certification or issue certificates”. Certification bodies do, and accreditation bodies check them, under rules now published by Global ACI, which has taken over the roles of IAF and ILAC.¹ Certify against ISO 13485:2016, confirmed in 2025.²

Table 1. TECH-3-004 and TECH-3-007 are the Global ACI documents on audit time and on ISO 13485 certification.
StepWhat happensRule
1. ImplementBuild and run the quality system to ISO 13485:2016 and the device regulations of your marketsISO 13485:2016²
2. Choose a bodyCheck its accreditation, and get its audit time and the justification for it in the contractISO;¹ TECH-3-004, 4.3 and 4.4³
3. Stage 1Documentation review and a check that you are ready for Stage 2; it should be on site for higher-risk devices such as GHTF classes C and DMDSAP audit approach;⁴ TECH-3-007, MD 9.3.1.2⁵
4. Stage 2Whether every applicable requirement is implemented and effectiveMDSAP audit approach⁴
5. Close findingsFor a major nonconformity, the body must accept your correction and corrective action and verify they workedTECH-3-007, MD 9.4.5⁵
6. CertificateIssued for a three-year cycleTECH-3-004, 5³
7. SurveillanceEach year, about a third of the initial audit time, including what you did about adverse event reports, advisory notices and recallsTECH-3-004, 5;³ TECH-3-007, MD 9.6.2.2⁵
8. RecertifyIn year three, about two thirds of the time a fresh initial audit would take thenTECH-3-004, 6;³ MDSAP audit approach⁴

2ISO 13485 certification readiness checklist.

What must exist before each stage, from NSAI’s published list: its Stage 1 and Stage 2 expectations for clauses 4.1.5, 5.6 and 8.2.4, and its common Stage 1 findings restated as what to have in place. Other bodies differ in detail, so ask yours for its list.

Table 2. Source: NSAI MTF-2031-B.⁶ Having no complaint handling or training system is the first example of a major nonconformity in Global ACI-TECH-3-007, MD 9.4.5(a).⁵
ClauseWhat must existBy
4.1.2, 4.1.3 ProcessesEach process in scope monitored, with an objective or metric, and on the internal audit scheduleStage 1
4.1.5 OutsourcingIdeally, providers identified, contracts drafted, evaluation under wayStage 1
4.1.5 OutsourcingContracts signed, providers on the approved supplier listStage 2
4.2.1 Change controlHow to identify a substantial change to the system and notify the bodyStage 1
4.2.2 Quality manualScope, exclusions correctly identified, how the processes interact, references to the required proceduresStage 1
4.2.4, 4.2.5 RecordsRetention periods defined, and aligned with the related documentsStage 1
5.3 Quality policyA commitment to comply with requirements and maintain effectiveness, and a framework for quality objectivesStage 1
5.5.2 RepresentativeThe management representative’s responsibilities and authority clearly statedStage 1
5.6 Management reviewCompleted, covering all applicable requirements, and concluding on “Continued suitability, adequacy and effectiveness”Stage 1; update after it if needed
6.2 TrainingA training system, with a method to judge training effectivenessStage 1
7.1 Risk managementA procedure saying what triggers a risk management file review or update, and how often periodic review happens; annual review may not be adequate for some technical filesStage 1
7.4 PurchasingSupplier re-evaluation documented, quality agreements with key suppliers signed, key suppliers on an approved list; for outsourced processes, the 4.1.5 rows applyStage 1
8.2.1 FeedbackA documented procedure for obtaining and using customer feedbackStage 1
8.2.2 ComplaintsA complaint handling system, with complaint defined as ISO 13485 defines itStage 1
8.2.4 Internal auditA schedule matching your processes, at least one audit done, and trained auditors who do not audit their own workStage 1
8.2.4 Internal auditA full round of internal audits completedStage 2
8.5.2 CAPAProcedures requiring corrective and preventive action without undue delayStage 1

3How long ISO 13485 certification takes.

No rule sets how long the whole project takes. Greenlight Guru says most device companies certify in six to 18 months, with at least three months running the system before Stage 1 and one to three months’ wait for the certification body.⁷ OpenRegulatory gives a typical four to nine months.⁸ After the build, the gates are fixed: records that meet the checklist, the body’s scheduling, and up to six months to clear a major nonconformity.

3.1Example: nine months from procedures to certificate.

Example. Corrowby Medical Ltd, an invented contract manufacturer, assembles and packs sterile single-use infusion sets for device companies whose supplier agreements require an ISO 13485 certificate. Months count from the day its procedures take effect; they are illustrative, the gates are not.

Table 3. Example. Gates from NSAI’s published expectations, which are one body’s, and from ISO/IEC 17021-1 as EA and Health Canada quote it.
MonthWhat happensGate
Before 0Gap analysis, two to four weeks; procedures written and staff trained, three to 12 monthsNone set by rule; durations are Greenlight Guru’s⁷
0Procedures take effect and records startNone: the clock starts
3First internal audit done; first management review complete; steriliser contract draftedNSAI’s Stage 1 expectations⁶
4Stage 1: documents reviewed, readiness for Stage 2 judgedThe body decides whether you are ready for Stage 2⁴
5Full round of internal audits complete; steriliser contract signed and the supplier approvedNSAI’s Stage 2 expectations⁶
6Stage 2 raises one major nonconformityThe body must accept the correction and corrective action⁵
8The body verifies they are implementedWithin 6 months of Stage 2’s last day, or Stage 2 is repeated⁹
9Certification decision; certificate takes effectNot before the decision¹⁰

4What ISO 13485 certification costs.

OpenRegulatory’s planning model puts initial certification for most small device companies at €15,000 to €100,000, counting consultants, internal labour, training and software as well as the body’s fees. Consultants are the biggest swing: €8,000 to €20,000 for targeted reviews, €25,000 to €60,000 for full implementation.⁸

4.1How audit days are set.

No accreditation rule sets fees; the rules set audit time, which each body prices at its own rates. An audit day is normally 8 hours, and a body may not cut the days by planning longer hours. Audit time includes time at your premises, physical or virtual, planning, document review and report writing; travel does not count as audit time on site.³

Days start from the effective number of personnel: everyone in the scope of certification, permanent, temporary or part-time, on every shift, and contractors when they are in scope. Depending on hours worked, part-timers may be converted to full-time equivalents: 30 working 4 hours a day count as 15.³

Table 4. Table D.1 of Global ACI-TECH-3-007, Stage 1 and Stage 2 together. It continues to larger organisations.⁵
Effective personnelInitial audit, days
1 to 53
6 to 104
11 to 154.5
16 to 255
26 to 456
46 to 657
66 to 858
86 to 12510
126 to 17511
176 to 27512
276 to 42513
426 to 62514

The table is a starting point. When more than one main technical area is audited, the time “shall be increased”, and sterilisation is a main technical area of its own, so a sterile device brings a second one. Time may also go up for complex devices, critical suppliers, installation at customers’ premises, poor regulatory compliance, and multiple shifts or lines. Reductions, for example for a scope without manufacturing, may not exceed 20% in total, or 50% for distribution or transport alone. Time to audit national or regional regulatory requirements comes on top.⁵ Days are rounded to the nearest half day, and the audit itself, from opening to closing meeting, should typically take at least 80% of the calculated time.³

4.2Example: checking a quote.

Corrowby has one site, no design in its scope, GHTF class B devices, and ethylene oxide sterilisation by a contractor.

Table 5. Example. Invented company. Rules from Global ACI-TECH-3-004 and -007;³⁵ the adjustment sizes are the body’s judgement, not set by the rules. Fees from NSAI’s published list.¹¹
LineWorking
Personnel31 full-time on two shifts, 20 part-time packers at 4 hours a day (10 full-time equivalents), 4 contract QC analysts in scope: 45
Starting pointTable D.1, 26 to 45 people: 6 days
AdjustmentsSterilisation is a second main technical area, so time must go up: the body adds 0.5 day. Two shifts: another 0.5 day. 7 days
SteriliserA critical supplier. If Corrowby cannot show enough evidence of its control, the body may add time to audit it
Initial audit7 days, Stage 1 and Stage 2 together
Surveillance7 ÷ 3 = 2.33, rounded to 2.5 days a year
First cycle’s auditsInitial and two surveillances: 7 + 2.5 + 2.5 = 12 audit days
Audit feesNSAI’s standalone ISO 13485 rate: 12 days × €1,900 = €22,800
Fixed fees€2,000 application, which includes year one’s annual fee, plus €1,500 in each of years two and three: €5,000
First cycle’s total€27,800, before travel
RecertificationIn year three, before expiry, if nothing changes: 7 × 2/3 = 4.67, rounded to 4.5 days; 4.5 × €1,900 = €8,550

NSAI’s list also prices pre- and post-audit activities at €1,900 a day and invoices handling of more than three nonconformities in half days, so ask what a quote includes.¹¹

If Corrowby sold sets under its own name, the certificate would not be enough:

  • EU. Sterile devices need an MDR notified body, whose audit can cover the ISO 13485 elements.¹²⁵ At NSAI’s MDR rate of €3,000 per auditor-day, the same 12 days cost €36,000, plus time to audit the MDR’s own requirements;⁵ technical documentation assessment starts at €19,500, and an unannounced audit, which the MDR requires at least once every five years, costs €6,000 a day with two auditors.¹¹¹²
  • Canada. An MDSAP certificate is required.¹³ MDSAP counts audit tasks, plus 25% for an initial audit, instead of using Table D.1,¹⁴ and NSAI charges €3,000 per auditor-day and €1,200 a year above the ISO 13485 annual fee.¹¹

4.3Comparing bodies’ published fees.

Day rates are close; minimums are not. NEOEMKI charges €234 an hour, or €1,872 for an 8-hour day, against NSAI’s €1,900, but sets minimums of €7,650 for initial certification, €4,000 for surveillance and €5,500 for renewal.¹⁵¹¹ Table D.1 gives a company of 1 to 5 people 3 audit days, worth €5,616 at €1,872, so the €7,650 minimum applies. Compare quotes on audit days and their written justification first, and on rates second.³

5Checking an ISO 13485 certificate.

Check the number on IAF CertSearch, which “aggregates data from accreditation bodies and certification bodies”, or with the body,¹ then read the certificate field by field:

Table 6. Example. The certificate is invented. Rules from Health Canada’s GD207 (2017), sections 2.1 to 2.6, written for registrars under its former CMDCAS scheme and citing ISO/IEC 17021-1.¹⁰
FieldExample: CorrowbyRule
NumberC-27-0417A unique identification code
Effective date9 September 2027, the decision dateNot before the certification decision
Expiry date8 September 2030No more than three years after the effective date
Name and addressCorrowby Medical Ltd, with its full street addressFor a Canadian licence, the same as on the licence and device label
SitesOne; the steriliser is not listedEvery covered site listed; subcontractors are not
ScopeManufacture of sterile single-use infusion setsProcesses and generic device groups, with no trade names, catalogue numbers or device classes

A contract manufacturer’s certificate does not stand in for its customers’: Health Canada will not accept a subcontractor’s certificate as evidence for a manufacturer’s licence.¹⁰ And a certificate shows the system was audited; it is not your supplier evaluation. See supplier qualification.

6What a certificate does in the US, EU, UK and Canada.

  • United States. No formal role. Part 820 incorporates ISO 13485:2016, but FDA “will not require certificates of conformance to ISO 13485”, and one “will not exempt a manufacturer from an FDA inspection”.¹⁶ FDA may accept MDSAP audit reports in place of routine inspections.¹⁷ See ISO 13485 vs 21 CFR 820.
  • European Union. It supports, but does not replace, the notified body’s audit for devices that need one. EN ISO 13485:2016 is a harmonised standard, and a notified body “shall assume” a system that meets it conforms to the requirements it covers, unless it substantiates otherwise.¹⁸¹²
  • United Kingdom. In Great Britain, EN ISO 13485:2016 is a designated standard under the UK MDR 2002,¹⁹ and a UK approved body does any third-party UKCA assessment. CE-marked devices remain accepted until 30 June 2030 under the EU MDR or IVDR, or until the sooner of certificate expiry and 30 June 2028 under the MDD or AIMDD (30 June 2030 under the IVDD). Northern Ireland applies the EU MDR.²⁰ On 21 October 2026 the government said it intends to legislate to recognise devices certified under the EU MDR indefinitely.²¹
  • Canada. Class II, III and IV licences need an ISO 13485 quality system certificate, and Health Canada requires it to be issued by an auditing organisation recognised under MDSAP.¹³

7Mistakes that add audit days or a major.

  • Narrowing the scope. A body may not exclude processes, products or services that affect safety and quality, “unless allowed by regulatory authorities”.⁵
  • Sampling sites. In a multi-site certification, sites that design, develop or manufacture devices cannot be sampled; each is audited.⁵
  • Repeat findings. Nonconformities repeated from earlier audits are a listed example of a major, as is failing to act when post-market data shows a pattern of defects.⁵
  • Surprises. New ownership, a new site or adding design to the scope can justify a short-notice audit, so tell the body first.⁵

8What changed in 2026.

  • 2 February 2026. FDA’s QMSR took effect, incorporating ISO 13485:2016 into Part 820.¹⁶ See the QMSR.
  • 28 August 2026. Global ACI-TECH-3-004 and -007 took effect, replacing IAF MD 5 and IAF MD 9, which several top-ranking guides still cite.²²³⁵

9How it runs in Seal.

Seal’s medical device blueprint records traceability as the work is done: protocols are created against requirements, results attach to executions, and risk controls, technical documentation and complaints link to the same records. The design and development file, the medical device file and the record for each device or batch are kept as connected records, whichever names your procedures use.

Audit management runs internal, supplier and system audits from one risk-based programme, and each gap becomes owned follow-up. For a certification audit or an inspection, inspection readiness logs requests, scope, packages, redactions and delivered evidence, and a mock inspection uses the same request log and package workflow as a real one. Neil assembles the evidence behind an auditor’s question and makes missing records explicit; your audit owner checks scope and completeness before anything is shared.

References

  1. 1Certification, ISO. https://www.iso.org/certification.html (read 9 November 2026)
  2. 2ISO 13485:2016, Medical devices, Quality management systems, Requirements for regulatory purposes, edition 3, ISO, March 2016, confirmed 2025. https://www.iso.org/standard/59752.html (read 9 November 2026)
  3. 3Global ACI-TECH-3-004, Determination of Audit Time of Quality, Environmental, and Occupational Health & Safety Management Systems, version 1.0, Global ACI, issued 10 July 2026, implemented 28 August 2026, clauses 1.8, 1.9, 2.1, 2.2, 2.3, 4.3, 4.4, 5 and 6. https://sys.global-aci.org/uploads/documents/Global_ACI_TECH-3-004_%28M%29_v1_0.pdf (read 9 November 2026)
  4. 4MDSAP AU P0002.011, MDSAP Audit Approach, Medical Device Single Audit Program, revision date 3 August 2026, MDSAP audit cycle and initial audit. https://www.mdsap.global/sites/default/files/2026-07/MDSAP%20AU%20P0002.011%20MDSAP%20Audit%20Approach.pdf (read 8 November 2026)
  5. 5Global ACI-TECH-3-007, Application of ISO/IEC 17021-1 in the Field of Medical Device Quality Management Systems (ISO 13485), version 1.0, Global ACI, issued 10 July 2026, implemented 28 August 2026, clauses MD 8.2.1, MD 9.1.4, MD 9.1.5, MD 9.2.2.1, MD 9.3.1, MD 9.4.5(a) to (f), MD 9.6.2.2 and MD 9.6.4.2, Table A.1.5 and Annex D. https://sys.global-aci.org/uploads/documents/Global_ACI_TECH-3-007_%28M%29_v1_0.pdf (read 9 November 2026)
  6. 6Common Findings: ISO 13485:2016 Stage 1 Audit, MTF-2031-B Rev 1.0, National Standards Authority of Ireland (NSAI), findings for clauses 4.1.2, 4.1.3, 4.2.1, 4.2.2, 4.2.4, 4.2.5, 5.3, 5.5.2, 5.6, 6.2, 7.1, 7.4, 8.2.1, 8.2.2, 8.2.4 and 8.5.2, and expectations at Stage 1 and Stage 2 for clauses 4.1.5, 5.6 and 8.2.4. https://www.nsai.ie/images/uploads/general/MTF-2031-B.pdf (read 9 November 2026)
  7. 7ISO 13485 certification: a realistic timeline and cost breakdown, Greenlight Guru, 18 August 2026, a vendor’s planning figures. https://www.greenlight.guru/blog/iso-13485-certification-timeline (read 9 November 2026)
  8. 8How much does ISO 13485 certification cost?, OpenRegulatory, reviewed 19 August 2026, a vendor’s planning model. https://openregulatory.com/iso-13485-certification-cost (read 9 November 2026)
  9. 9Question 36.14, Corrections and corrective actions, European co-operation for Accreditation (EA), September 2018, quoting ISO/IEC 17021-1 on initial certification. https://european-accreditation.org/sp_accordion_faqs/question-36-14-corrections-and-corrective-actions/ (read 9 November 2026)
  10. 10Guidance Document GD207: Guidance on the Content of ISO 13485 Quality Management System Certificates Issued by Health Canada Recognized Registrars, Health Canada, 14 March 2017, sections 2.1 to 2.6. https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/quality-systems-13485/guidance-document-gd207-guidance-content-13485-quality-management-system-certificates-issued.html (read 9 November 2026)
  11. 11NSAI MED EU Price List June 2026, MRF-1015 Rev 1.0, National Standards Authority of Ireland (NSAI): fees for MDR and IVDR conformity assessment, effective 1 June 2026, and for ISO 13485 and MDSAP certification, effective 11 November 2024. https://www.nsai.ie/images/uploads/medical-devices/MRF-1015.docx (read 9 November 2026)
  12. 12Regulation (EU) 2017/745 on medical devices, OJ L 117, 5 May 2017, Article 52 and Annex IX, sections 2.3 and 3.4, EUR-Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745 (read 9 November 2026)
  13. 13Guidance Document: Requirements in the Recognition Process for Medical Device Single Audit Program (MDSAP) Auditing Organizations, Health Canada, sections 1.2 to 1.4. https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/quality-systems-13485/requirements-recognition-process-mdsap-auditing-organizations-profile/guidance-document.html (read 9 November 2026)
  14. 14MDSAP AU P0008.008, Audit Time Determination Procedure, Medical Device Single Audit Program, effective 13 October 2023, section 5.2. https://www.mdsap.global/sites/default/files/2025-04/MDSAP-AU-P0008-008-Audit-Time-Determination-Procedure.pdf (read 8 November 2026)
  15. 15Applicable Fees and Rates / ISO 13485, valid from 1 April 2026, NEOEMKI. https://neoemki.hu/wp-content/uploads/2026/04/Szolgaltatasi-dijak-jegyzeke_EN_26-Apr.pdf (read 9 November 2026)
  16. 16Quality Management System Regulation: Frequently Asked Questions, update of 2 February 2026 and question 13, FDA. https://www.fda.gov/medical-devices/quality-management-system-regulation-qmsr/quality-management-system-regulation-frequently-asked-questions (read 9 November 2026)
  17. 17Medical Device Single Audit Program (MDSAP), FDA, update of 18 December 2025. https://www.fda.gov/medical-devices/cdrh-international-affairs/medical-device-single-audit-program-mdsap (read 9 November 2026)
  18. 18Commission Implementing Decision (EU) 2021/1182 on the harmonised standards for medical devices, consolidated text of 17 June 2026, Annex, entry 10, EUR-Lex. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:02021D1182-20260617 (read 9 November 2026)
  19. 19Designated standards: medical devices, notice of publication 0129/26 of 17 January 2026, Annex I, Part 2, row 126, Department of Health and Social Care, GOV.UK. https://assets.publishing.service.gov.uk/media/696900825fd54abe6531a80f/ds-0129-26-medical-devices-notice.pdf (read 9 November 2026)
  20. 20Regulating medical devices in the UK, MHRA, GOV.UK, updated 20 February 2026, including Northern Ireland. https://www.gov.uk/guidance/regulating-medical-devices-in-the-uk (read 9 November 2026)
  21. 21Government response to the targeted consultation on indefinite recognition of CE marked medical devices, MHRA, GOV.UK, 21 October 2026, proposals 1 and 2a and next steps. https://assets.publishing.service.gov.uk/media/6ad76f8be9713babfbe3f755/Government_response_to_the_targeted_consultation_on_indefinite_recognition_of_CE_marked_medical_devices.pdf (read 9 November 2026)
  22. 22Global ACI, IAF and ILAC Document Cross-Reference Table, version 15.0, Global ACI, 5 November 2026. https://sys.global-aci.org/uploads/documents/2026-11-05_Global_ACI_IAF_and_ILAC_Document_Cross-Reference_Table_v15_02.pdf (read 9 November 2026)

AQuestions and answers

Does ISO issue ISO 13485 certificates?

No. ISO writes the standards but does not certify organisations or issue certificates. Certification bodies do, and accreditation bodies check that those certification bodies work to the relevant international standard.

Can an individual get ISO 13485 certified?

Not to the standard itself. ISO 13485 certification applies to an organisation’s quality management system. People take ISO 13485 training or internal and lead auditor courses, and some personnel certification bodies certify auditors, but that is a certificate of the person’s competence, not of a quality system.

How long does ISO 13485 certification last?

Three years. The body runs a surveillance audit in each of the two years after the initial certification and a recertification audit in the third, before the certificate expires. If it does lapse, accreditation guidance lets the body restore it within six months once the outstanding recertification work is done; after that, at least a new Stage 2 audit is needed.

Does FDA require ISO 13485 certification?

No. Since February 2026 the QMSR has incorporated ISO 13485:2016 by reference, so its requirements apply, but FDA neither requires nor issues certificates, and holding one does not exempt a manufacturer from FDA inspection.

Can ISO 9001 and ISO 13485 be audited together?

Yes. The certification body adds at least 25% to the ISO 13485 audit days, whether it runs one integrated audit or a combined one, and may need more where the scopes or headcounts differ.

Can a Stage 1 audit be done remotely?

Often, at least in part: document review can be done away from your site. For higher-risk devices, such as GHTF classes C and D, the accreditation rules say Stage 1 should be on site. Ask the body how it plans to split remote and on-site time, because most of the calculated audit time should still be spent auditing your operation.